← Product Code [JIX](/submissions/CH/subpart-b%E2%80%94clinical-chemistry-test-systems/JIX) · K061838

# DIMENSION VISTA CHEMISTRY 1 CALIBRATOR, MODEL KC110 (K061838)

_Dade Behring, Inc. · JIX · Aug 21, 2006 · Clinical Chemistry · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/CH/subpart-b%E2%80%94clinical-chemistry-test-systems/JIX/K061838

## Device Facts

- **Applicant:** Dade Behring, Inc.
- **Product Code:** [JIX](/submissions/CH/subpart-b%E2%80%94clinical-chemistry-test-systems/JIX.md)
- **Decision Date:** Aug 21, 2006
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 862.1150
- **Device Class:** Class 2
- **Review Panel:** Clinical Chemistry

## Indications for Use

The CHEM 1 CAL is an in vitro diagnostic product for the calibration of Calcium (CA), Cholesterol (CHOL), Creatinine (CREA), Glucose (GLU), Lactic Acid (LA), Magnesium (MG), Thyroxine (T4), Thyronine Uptake (TU), Blood Urea Nitrogen (BUN), and Uric Acid (URCA) methods on the Dimension VistaTM System.

## Device Story

Dimension Vista Chemistry 1 Calibrator (CHEM 1 CAL) is a liquid, multi-analyte, bovine serum albumin-based product; used to calibrate clinical chemistry assays on the Dimension Vista System. Kit contains six vials (three of Calibrator A, three of Calibrator B). Device provides reference values for CA, CHOL, CREA, GLU, LA, MG, T4, TU, BUN, and URCA. Used by laboratory professionals in clinical settings to ensure accuracy of patient sample testing. Healthcare providers use the resulting calibrated instrument output to diagnose and monitor patient metabolic and endocrine health. Benefits include standardized, reliable diagnostic measurements.

## Clinical Evidence

Bench testing only. Stability testing performed per CEN 13640; shelf life established at 12 months at 4°C. Assigned values verified against Master Pools using N=45 replicates across multiple instruments. Traceability established to NIST SRMs (912, 915, 914, 917, 929A, 913), CDC Abell-Kendall method, USP, and A-Grade lithium salt.

## Technological Characteristics

Liquid, multi-analyte, bovine serum albumin-based calibrator. Two-level concentration. Traceable to NIST SRMs, USP, and reference methods. Stability: 12-month shelf life; 1-day on-board stability; 30-day open-vial refrigerated stability. Standards: CEN 13640, ISO 14971:2000.

## Regulatory Identification

A calibrator is a device intended for medical purposes for use in a test system to establish points of reference that are used in the determination of values in the measurement of substances in human specimens. (See also § 862.2 in this part.)

## Special Controls

*Classification.* Class II (special controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9.

## Predicate Devices

- CHEM I Calibrator (k860021)
- URCA (k862359)
- Chemistry II Calibrator (k861700)
- Cholesterol Calibrator (k861700)
- Thyroxine Calibrator (k862359)
- Thyronine Uptake Calibrator (k862359)

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}

1

# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY ONLY TEMPLATE

A. 510(k) Number:
k061838

B. Purpose for Submission:
New Device

C. Measurand:
Calibrator for Calcium (CA), Cholesterol (CHOL), Creatinine (CREA), Glucose (GLU), Lactic Acid (LA), Magnesium (MG), Thyroxine (T4), Thyronine Uptake (TU), Blood Urea Nitrogen (BUN) and Uric Acid (URCA).

D. Type of Test:
Not Applicable

E. Applicant:
Dade Behring, Inc.

F. Proprietary and Established Names:
Dimension Vista Chemistry 1Calibrator, (KC110)

G. Regulatory Information:

|  Product Code | Classification | Regulation Section | Panel  |
| --- | --- | --- | --- |
|  Calibrator, Multi-Analyte Mixture (JIX) | Class II | 21 CFR 862.1150, Calibrator. | 75 Clinical Chemistry (CH)  |

H. Intended Use:

1. Intended use(s):
See indications(s) for use below.

{1}

2. Indication(s) for use:

The CHEM 1 CAL is an in vitro diagnostic product for the calibration of Calcium (CA), Cholesterol (CHOL), Creatinine (CREA), Glucose (GLU), Lactic Acid (LA), Magnesium (MG), Thyroxine (T4), Thyronine Uptake (TU), Blood Urea Nitrogen (BUN) and Uric Acid (URCA) methods on the Dimension Vista™ System.

3. Special conditions for use statement(s):

For prescription use

4. Special instrument requirements:

Dimension Vista™ System

I. Device Description:

CHEM 1 CAL is a liquid, multi-analyte, bovine serum albumin based product containing calcium, cholesterol, creatinine, glucose, lactic acid, magnesium, thyroxine, urea nitrogen and uric acid. The kit consists of six vials, three vials of Calibrator A, and three vials of Calibrator B which are ready for use (no preparation is required). The volume per vial is 2.5 mL.

J. Substantial Equivalence Information:

|   | Device | Predicate Devices  |   |   |   |   |
| --- | --- | --- | --- | --- | --- | --- |
|  Item | Dimension Vista™ System Chemistry 1 Calibrator2 | CHEM I Calibrator k860021 (URCA-k862359) | Chemistry II Calibrator k861700 | Cholesterol Calibrator k861700 | Thyroxine Calibrator k862359 | Thyronine Uptake Calibrator k862359  |
|  Form | Liquid. | Lyophilized. | Liquid. | Lyophilized. | Lyophilized. | Lyophilized.  |
|  Traceability | BUN - NIST SRM 9122 CA - NIST SRM 915. CHOL - Abell-Kendall | BUN - NIST SRM 912 CA - NIST SRM 915. | MG - NIST SRM 929A | CHOL - NIST SRM 911. | T4 - Thyroxine Master Pool. | TU - Thyronine Uptake Master Pool.  |

{2}

3

|   | Device | Predicate Devices  |   |   |   |   |
| --- | --- | --- | --- | --- | --- | --- |
|   | (CDC-NCEP).
CREA – NIST SRM 914. GLU – NIST SRM 917. LA – Lactic acid – lithium salt A-Grade. MG – NIST SRM 929A. T4 – USP. TU – Calculated value. URCA – NIST SRM 913. | CREA - NIST SRM 914. GLU - NIST SRM 917. LA - Lactic acid - lithium salt A-Grade. URCA - NIST SRM 913. |  |  |  |   |
|  Matrix | Bovine serum albumin based product. | Bovine serum albumin based product. | Pure magnesium dissolved in a dilute Solution of HCL, reagent grade potassium dihydrogen phosphate and reagent grade glycerol. | Bovine serum albumin based product. | Human serum based product. | Human serum based product.  |
|  Number of Levels | Two levels. | Three levels. | Three levels. | Three levels. | Five levels. | Five levels.  |

K. Standard/Guidance Document Referenced (if applicable):

|  STANDARDS  |
| --- |
|  Title and Reference Number  |
|  Stability Testing of In Vitro Diagnostic Reagents (13640)  |
|  Medical devices - Application of risk management to medical devices (14971:2000)  |

{3}

4

|  GUIDANCE  |   |   |   |
| --- | --- | --- | --- |
|  Document Title | Office | Division | Web Page  |
|  Guidance for Industry - Abbreviated 510(k) Submissions for In Vitro Diagnostic Calibrators; Final | OIVD |  | http://www.fda.gov/cdrh/ode/calibrator.html  |
|  Guidance for Industry and FDA Staff - Use of Symbols on Labels and in Labeling of In Vitro Diagnostic Devices Intended for Professional Use |  |  | http://www.fda.gov/cdrh/ocd/guidance/4444.html  |

L. Test Principle:

Not Applicable

M. Performance Characteristics (if/when applicable):

1. Analytical performance:

a. Precision/Reproducibility:

Not Applicable

b. Linearity/assay reportable range:

Not Applicable

c. Traceability, Stability, Expected values (controls, calibrators, or methods):

Traceability: The assigned values of the Chemistry 1 Calibrator were verified on a Dimension® Vista™ System calibrated with an approved Master Pool. Master Pool values were assigned on multiple Dimension® clinical chemistry instruments. According to the sponsor the traceability of the assigned values of the Chemistry 1 Calibrator was standardized to the below table of assigned values.

|  Constituent | Traceability  |
| --- | --- |
|  BUN | NIST SRM_{®}912  |
|  AC | NIST SRM 915  |
|  CHOL | NIST SRM 911 (CDC_{®}) Abell-Kendall reference method  |
|  CREA | NIST SRM 914  |
|  GLU | NIST SRM 917  |
|  LA | Lactic Acid Lactic acid- lithium salt (A- Grade)  |
|  MG | NIST SRM 929A  |
|  T4 | USP_{®}  |

{4}

5

|  TU | Calculated value  |
| --- | --- |
|  URCA | NIST SRM 913  |

1 National Institute of Standards and Technology – Standard Reference Material.
b CDC: Centers for Disease Control
c United States Pharmacopeia.

Stability: The target shelf life for the Dimension Vista® Chemistry 1 Calibrator is 12 months. A vial punctured by the instrument and stored on board has a stability claim of one day. An open vial not stored on the instrument, but recapped and stored in a refrigerator has a stability claim of 30 days. Stability study protocols and acceptance criteria were described and found to be acceptable.

d. Detection limit:
Not applicable

e. Analytical specificity:
Not applicable

f. Assay cut-off:
Not applicable

2. Comparison studies:
a. Method comparison with predicate device:
Not applicable
b. Matrix comparison:
Not applicable

3. Clinical studies:
a. Clinical Sensitivity:
Not applicable
b. Clinical specificity:
Not applicable

{5}

c. Other clinical supportive data (when a. and b. are not applicable):

4. Clinical cut-off:
Not applicable

5. Expected values/Reference range:
Not applicable

N. Proposed Labeling:
The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10.

O. Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.

6

---

**Source:** [https://fda.innolitics.com/submissions/CH/subpart-b%E2%80%94clinical-chemistry-test-systems/JIX/K061838](https://fda.innolitics.com/submissions/CH/subpart-b%E2%80%94clinical-chemistry-test-systems/JIX/K061838)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
