← Product Code [JIX](/submissions/CH/subpart-b%E2%80%94clinical-chemistry-test-systems/JIX) · K051452

# SENTINEL CLIN CHEM CAL (K051452)

_Sentinel Ch. Srl · JIX · Oct 25, 2005 · Clinical Chemistry · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/CH/subpart-b%E2%80%94clinical-chemistry-test-systems/JIX/K051452

## Device Facts

- **Applicant:** Sentinel Ch. Srl
- **Product Code:** [JIX](/submissions/CH/subpart-b%E2%80%94clinical-chemistry-test-systems/JIX.md)
- **Decision Date:** Oct 25, 2005
- **Decision:** SESE
- **Submission Type:** Abbreviated
- **Regulation:** 21 CFR 862.1150
- **Device Class:** Class 2
- **Review Panel:** Clinical Chemistry

## Indications for Use

The Sentinel Clin Chem Cal is a device intended for medical purposes for use in pancreatic amylase and cholinesterase assays to establish points of reference that are used in the determination of values in the measurement of pancreatic amylase and cholinesterase in human serum and plasma. For Prescription Use Only

## Device Story

Sentinel Clin Chem Cal is a multi-analyte calibrator consisting of pancreatic amylase and cholinesterase in a human serum matrix. It is used in clinical chemistry laboratories to establish reference points for the measurement of these analytes in human serum and plasma. The device is provided in vials that require reconstitution with distilled water. Value assignment is performed by assaying the calibrator on Hitachi 717 and Abbott ARCHITECT c8000 instruments, with the final assigned value being the mean of the averages from both systems. Healthcare providers use the calibrated assay results to determine patient analyte levels, aiding in clinical decision-making related to pancreatic and cholinesterase function. The calibrator ensures the accuracy and effectiveness of clinical chemistry measurements.

## Clinical Evidence

No clinical data. Bench testing only. Stability was validated through real-time studies confirming 24-month shelf life and 2-day stability at 2-8°C post-reconstitution. Value assignment was performed using commercially marketed analyzers with internal control monitoring.

## Technological Characteristics

Multi-analyte calibrator in human serum matrix. Analytes: pancreatic amylase and cholinesterase. Traceability: IFCC EPS/37°C for amylase (p-Nitrophenol standard) and DGKC Butyrylthiocholine/37°C for cholinesterase (Hexacyano-ferrate (III) standard). Form factor: lyophilized powder in vials, reconstituted with distilled water. Storage: -20°C for long-term, 2-8°C for short-term after reconstitution.

## Regulatory Identification

A calibrator is a device intended for medical purposes for use in a test system to establish points of reference that are used in the determination of values in the measurement of substances in human specimens. (See also § 862.2 in this part.)

## Special Controls

*Classification.* Class II (special controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9.

## Predicate Devices

- Roche Calibrator for Automated Systems (k033501)

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}

1

510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION
DECISION SUMMARY
ASSAY ONLY TEMPLATE

A. 510(k) Number:
k051452

B. Purpose for Submission:
Notification of intent to manufacture and market the device: Sentinel Clin Chem Cal.

C. Measurand:
Amylase and Cholinesterase Calibrator

D. Type of Test:
Calibrator

E. Applicant:
Sentinel CH. S.r.l.

F. Proprietary and Established Names:
Sentinel Clin Chem Cal

G. Regulatory Information:
1. Regulation section:
21 CFR 862.1150, Calibrator
2. Classification:
Class II
3. Product code:
JIX, Multi-Analyte Calibrator
4. Panel:
75, Chemistry

{1}

H. Intended Use:

1. Intended use(s):
See indications for use below

2. Indication(s) for use:
The Sentinel Clin Chem Cal is a device intended for medical purposes for use in pancreatic amylase and cholinesterase assays to establish points of reference that are used in the determination of values in the measurement of pancreatic amylase and cholinesterase in human serum and plasma.

3. Special conditions for use statement(s):
For Prescription Use Only

4. Special instrument requirements:
Abbott Aeroset and Architect Systems

I. Device Description:

The Sentinel Clin Cal is a lyophilized calibrator prepared with human serum. Porcine pancreatic amylase and human cholinesterase are added to the base serum. The product consists of 4 X 3 mL bottles of lyophilized material containing pancreatic amylase and cholinesterase in a human serum matrix. The values of the calibrator are lot-dependent.

All human source material was tested and found to be non-reactive for HBsAg, HCV, and HIV-1/2.

J. Substantial Equivalence Information:

1. Predicate device name(s):
Roche Calibrator for Automated Systems

2. Predicate 510(k) number(s):
k033501

3. Comparison with predicate:
|  Characteristics | New Device | Predicate  |
| --- | --- | --- |

{2}

3

|  Constituents | Pancreatic Amylase, Cholinesterase | Contains additional analytes  |
| --- | --- | --- |
|  Intended Use | For the calibration of clinical chemistry tests | Same  |
|  Storage | 2-8°C | 2-8°C  |
|  Reconstituted Stability | 2 days at 2-8°C
14 days at -20°C when stored in small volumes (300-600μL) and frozen only once. | 8 hours at 15-25°C
2 days at 2-8°C
4 weeks when stored at – (15°C) to - (25°C) and frozen only once.  |

K. Standard/Guidance Document Referenced (if applicable):

FDA’s Guidance for Industry, Abbreviated 510(k) Submissions for In Vitro Diagnostic Calibrators, February 22, 1999.

L. Test Principle:

Not applicable

M. Performance Characteristics (if/when applicable):

1. Analytical performance:

a. Precision/Reproducibility:

Not applicable

b. Linearity/assay reportable range:

Not applicable

c. Traceability, Stability, Expected values (controls, calibrators, or methods):

Sentinel Clin Chem Cal is traceable to the International Federation of Clinical Chemistry (IFCC) for Pancreatic Amylase and Deutsche Gesellschaft für Klinische Chemie (DGKC) for Cholinesterase.

|  Analyte | Method  |
| --- | --- |
|  Pancreatic Amylase | IFCC EPS/37°C  |
|  Cholinesterase | DGKC
Butyrylthiochholine37°C  |

{3}

Values were assigned in house on commercially marketed clinical chemistry analyzers. During each testing run, two levels of control material (low and high) were assayed to ensure the assays were meeting their expected performance.

On three consecutive days, three replicates of the product were assayed on a commercially marketed analyzer. The average of all replicates was calculated (X1).

Five vials of freshly reconstituted product were assayed in five replicates on a commercially marketed analyzer. The average of all replicates was calculated (X2).

The assigned value (Xm) was determined to be the mean of the two averages: Xm=(X1+X2)/2.

Shelf Life: The Sentinel Clin Chem Cal is stable for 24 months based on real time stability studies.

Reconstituted: Real time stability studies show the Sentinel Clin Chem Cal is stable for 2 days at 2-8°C and 14 days at -20°C when stored in small volumes (300-600μL) and frozen once.

d. Detection limit:
Not applicable

e. Analytical specificity:
Not applicable

f. Assay cut-off:
Not applicable

2. Comparison studies:
a. Method comparison with predicate device:
Not applicable
b. Matrix comparison:
Not applicable

3. Clinical studies:

{4}

a. Clinical Sensitivity:
Not applicable

b. Clinical specificity:
Not applicable

c. Other clinical supportive data (when a. and b. are not applicable):
Not applicable

4. Clinical cut-off:
Not applicable

5. Expected values/Reference range:
Not applicable

N. Proposed Labeling:
The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10.

O. Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.

---

**Source:** [https://fda.innolitics.com/submissions/CH/subpart-b%E2%80%94clinical-chemistry-test-systems/JIX/K051452](https://fda.innolitics.com/submissions/CH/subpart-b%E2%80%94clinical-chemistry-test-systems/JIX/K051452)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
