K042549 · Wako Chemicals USA, Inc. · JIX · Nov 8, 2004 · Clinical Chemistry
Device Facts
Record ID
K042549
Device Name
MULTI CHEM CAL B
Applicant
Wako Chemicals USA, Inc.
Product Code
JIX · Clinical Chemistry
Decision Date
Nov 8, 2004
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 862.1150
Device Class
Class 2
Indications for Use
The Multi-Chem Calibrator B is designed to be used with Wako's assays to determine the points of reference that are used in the determination of Glucose, Uric Acid, and Urea Nitrogen in human serum.
Device Story
Multi-Chem Calibrator B serves as a reference material for clinical chemistry assays. It is used in laboratory settings by trained personnel to calibrate Wako's diagnostic assays for Glucose, Uric Acid, and Urea Nitrogen. By establishing reference points, the calibrator ensures the accuracy and reliability of patient serum test results. This calibration process allows healthcare providers to make informed clinical decisions based on precise analyte measurements, ultimately benefiting patients through accurate diagnosis and monitoring of metabolic conditions.
Clinical Evidence
No clinical data provided; bench testing only.
Technological Characteristics
Liquid aqueous solution containing Glucose, Uric Acid, Urea Nitrogen, and Sodium Azide. Traceable to NIST SRM917b, SRM913a, and SRM912a. Storage 2-10°C. No electronic or software components.
Indications for Use
Indicated for use as a reference material for the determination of Glucose, Uric Acid, and Urea Nitrogen in human serum samples when used with Wako's clinical chemistry assays.
Regulatory Classification
Identification
A calibrator is a device intended for medical purposes for use in a test system to establish points of reference that are used in the determination of values in the measurement of substances in human specimens. (See also § 862.2 in this part.)
Special Controls
*Classification.* Class II (special controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9.
K042551 — MULTI-CHEM CALIBRATOR A · Wako Chemicals USA, Inc. · Nov 16, 2004
Submission Summary (Full Text)
{0}
Page 1 of 5
# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY DEVICE ONLY TEMPLATE
A. 510(k) Number:
k042549
B. Purpose For Submission:
Premarket Notification 510(k) of intention to manufacture and market the Wako Chemicals USA, Inc. Multi-Chem Calibrator B
C. Analyte:
Glucose, Uric Acid, Urea Nitrogen
D. Type of Test: Not Applicable
E. Applicant:
Wako Chemicals USA, Inc.
F. Proprietary and Established Names:
Multi-Chem Calibrator B
G. Regulatory Information:
Regulation section:
1. Regulation section: §862.1150, Calibrator
2. Classification: Class II
3. Product Code: JIX
4. Panel: 75 (Chemistry)
H. Intended use(s):
{1}
Page 2 of 5
1. Intended use(s):
The Multi-Chem Calibrator B is designed to be used with Wako's assays to determine the points of reference that are used in the determination of Glucose, Uric Acid, and Urea Nitrogen in human serum.
2. Indication(s) for use:
The Multi-Chem Calibrator B is designed to be used with Wako's assays to determine the points of reference that are used in the determination of Glucose, Uric Acid, and Urea Nitrogen in human serum.
3. Special condition for use statement(s): For Prescription Use.
4. Special instrument Requirements: Not Applicable
I. Device Description:
The Multi-Chem Calibrator B is liquid and ready to use. The calibrator consists of 4 x 5 mL bottles of aqueous solutions containing Glucose, Uric Acid, Urea Nitrogen and Sodium Azide. Calibrator traceability has been reference to NIST primary standards with assigned values verified by recognized enzymatic methods. This solution is stable until the date printed on the label when stored as directed.
J. Substantial Equivalence Information:
1. Predicate device name(s): Wako Glucose, Uric Acid and Urea Standard
2. Predicate K number(s): k871758, k871768, k871767
3. Comparison with Predicate:
| | Multi-Chem Calibrator B | Glucose | Uric Acid | Urea Nitrogen |
| --- | --- | --- | --- | --- |
| Format | Liquid | Liquid | Liquid | Liquid |
| Constituents | Glucose 300 mg/dL
Uric Acid 10 mg/dL
Urea Nitrogen 30 mg/dL | Glucose 300 mg/dL | Uric Acid 10 mg/dL | Urea Nitrogen 30 mg/dL |
| Storage | 2-10° C | 2-10° C | 2-10° C | 2-10° C |
| 510(K) # | K042549 | K871758 | K871768 | K871767 |
{2}
Page 3 of 5
K. Standard/Guidance Document Referenced (if applicable):
The Waco Multi-Chem Calibrator B is traceable to NIST referenced standards:
NIST, SRM917b
NIST, SRM913a
NIST, SRM912a
L. Test Principle:
Not Applicable
M. Performance Characteristics (if/when applicable):
1. Analytical performance:
a. Precision/Reproducibility: Not Applicable
b. Linearity/assay reportable range: Not Applicable
c. Traceability (controls, calibrators, or method):
The Waco Multi-Chem Calibrator B is traceable to NIST referenced standards
Primary Standard
| Item | Primary Standard |
| --- | --- |
| Glucose | NIST SRM917b |
| Uric Acid | NIST SRM913a |
| Urea Nitrogen | NIST SRM912a |
Multi Chem Calibrator B
| Item | Assigned Value | Method |
| --- | --- | --- |
| Glucose | 300 mg/dL | Assayed by enzymatic method |
| Uric Acid | 30 mg/dL | Assayed by enzymatic method |
| Urea Nitrogen | 10 mg/dL | Assayed by enzymatic method |
The stability of the product was tested over time and the results were compared to the start point of time 0. Percent recoveries were calculated. No recoveries exceeded $\pm 2.5\%$ from time 0. The product is stable at $2 - 10^{\circ}\mathrm{C}$ until the date printed on the label.
d. Detection limit: Not Applicable
e. Analytical specificity: Not Applicable
f. Assay cut-off: Not applicable
2. Comparison studies:
a. Method comparison with predicate device:
{3}
Page 4 of 5
The Multi-Chem Calibrator B is designed to be used with Wako's assays to determine the points of reference that are used in the determination of Glucose, Uric Acid and Urea Nitrogen in human serum.
The safety and effectiveness of the Multi-Chem Calibrator B is demonstrated by its substantial equivalency to Wako Glucose, Uric Acid and Urea Nitrogen Standard. The calibration materials are used to calibrate instruments to determine the points of reference. The following are results of the comparison studies against the predicates:
Glucose (L-Type Glucose 2)
| | n | 80 |
| --- | --- | --- |
| | Correlation coefficient | 0.9999 |
| | Slope | 1.0015 |
| | Intercept | -0.0592 |
| | | |
Uric Acid (L-Type UA F reagent)
| | n | 70 |
| --- | --- | --- |
| | Correlation coefficient | 0.9998 |
| | Slope | 0.9953 |
| | Intercept | -0.0096 |
| | | |
Urea Nitrogen (L-Type UN reagent)
| | n | 70 |
| --- | --- | --- |
| | Correlation coefficient | 0.9999 |
| | Slope | 0.9961 |
| | Intercept | 0.0743 |
| | | |
3. Clinical studies:
a. Clinical sensitivity:
Clinical studies are not typically submitted for this device type.
b. Clinical specificity:
Clinical studies are not typically submitted for this device type.
c. Other clinical supportive data (when a and b are not applicable):
Not applicable
{4}
Page 5 of 5
4. Clinical cut-off:
Not applicable
5. Expected values/Reference range:
Multi Chem Calibrator B
| Item | Assigned Value | Method |
| --- | --- | --- |
| Glucose | 300 mg/dL | Assayed by enzymatic method |
| Uric Acid | 30 mg/dL | Assayed by enzymatic method |
| Urea Nitrogen | 10 mg/dL | Assayed by enzymatic method |
N. Conclusion:
The submitted material in this premarket notification is complete and supports a substantial equivalence decision.
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.