K033061 · Diamond Diagnostics, Inc. · JIX · Nov 7, 2003 · Clinical Chemistry
Device Facts
Record ID
K033061
Device Name
CALIBRATING MATERIAL, ISE ELECTROLYTE REFERENCE
Applicant
Diamond Diagnostics, Inc.
Product Code
JIX · Clinical Chemistry
Decision Date
Nov 7, 2003
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 862.1150
Device Class
Class 2
Indications for Use
The three products encompassed by this request are intended for in-vitro diagnostics use. Mission Electrolyte Reference Reagent provides reference points for Na, K, CI, Ca (on Delta's only) and TCO2. When used in conjunction with CO2 Acid Reagent, CO2 Alkaline Buffer and Electrolyte Buffer, the ISE Electrolyte Reference is used for the calibration of the ISE electrodes; sodium, potassium chloride and calcium (on Delta's only) and CO2. The CO2 Acid reagent is used to release CO2 from samples. The CO2 Alkaline in used to provide a constant CO2 concentration as reference for the CO2 electrode. Mission reagents are intended to serve as direct replacements to the predicate devices manufactured by the OEM. The products encompassed are to be handled using normal laboratory precautions.
Device Story
The device consists of three liquid reagents (CO₂ Alkaline Buffer, CO₂ Acid Reagent, and ISE Electrolyte Reference) used as calibrators for Beckman Synchron CX® and CX® Delta clinical chemistry analyzers. These reagents provide reference points for Na, K, Cl, Ca, and TCO2 measurements. The CO₂ Acid reagent releases CO2 from samples; the Alkaline buffer provides a constant CO2 reference; the ISE Electrolyte Reference provides reference points for electrolytes. The device is used in clinical laboratory settings by trained laboratory personnel. It serves as a direct replacement for OEM reagents. By calibrating the analyzer, the device ensures the accuracy of patient sample measurements, which healthcare providers use to diagnose and monitor electrolyte and acid-base status.
Clinical Evidence
No clinical studies were performed. Evidence consists of analytical bench testing. Precision was evaluated on CX® Delta and CX®3 analyzers; total precision CVs ranged from 0.29% to 8.97% across analytes. Method comparison studies against the predicate device showed high correlation (R² 0.991–0.998) for CO2, Chloride, Potassium, and Sodium. Traceability was established via NIST standards and OEM calibrators.
Indicated for use as a multi-analyte calibrator for the Beckman Synchron CX® and CX® Delta clinical chemistry systems to provide reference points for sodium, potassium, chloride, calcium, and total CO2 measurements.
Regulatory Classification
Identification
A calibrator is a device intended for medical purposes for use in a test system to establish points of reference that are used in the determination of values in the measurement of substances in human specimens. (See also § 862.2 in this part.)
Special Controls
*Classification.* Class II (special controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9.
{1}
510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION
DECISION SUMMARY
DEVICE ONLY TEMPLATE
A. 510(k) Number:
K033061
B. Analyte:
Sodium, potassium, chloride, calcium, carbon dioxide (CO₂)
C. Type of Test:
Instrument calibrator
D. Applicant:
Diamond Diagnostics / Mission Diagnostics
E. Proprietary and Established Names:
CO₂ Alkaline Buffer for Beckman Synchron CX® & CX® Delta Systems
CO₂ Acid Reagent for Beckman Synchron CX® & CX® Delta Systems
ISE Electrolyte Reference for Beckman Synchron CX® & CX® Delta Systems
F. Regulatory Information:
1. Regulation section:
21 CFR § 862.1150, Calibrator
2. Classification:
Class II
3. Product Code:
JIX, Calibrator, multi-analyte mixture
4. Panel:
Clinical Chemistry (75)
G. Intended Use:
1. Indication(s) for use:
Mission CO₂ Acid Reagent is used to release CO2 from samples on the Beckman Synchron CX® & CX® Delta.
Mission CO₂ Alkaline Buffer provides a constant CO2 concentration as reference for the CO2 electrode on the Beckman Synchron CX® & CX® Delta.
Mission Electrolyte Reference Reagent provides reference points for Na, K, Cl, Ca (on Delta’s only) and TCO2 for Beckman Synchron CX® & CX® Delta.
2. Special condition for use statement(s):
{2}
Page 2 of 5
These reagents are intended to serve as direct replacements to like-named products of the original equipment manufacturer (OEM).
For use with the Synchron CX® & CX® Delta Systems.
For in vitro diagnostic use.
3. Special instrument Requirements:
The Synchron CX® & CX® Delta analyzers.
H. Device Description:
Three reagents make up the calibrating material: the CO₂ Alkaline buffer, the CO₂ Acid reagent, and the ISE Electrolyte Reference. These reagents are used to provide reference points for the listed analytes using the Beckman Synchron CX® & CX® Delta Systems.
The CO₂ Alkaline buffer (pH 8.8) contains 6 mmol/L potassium bicarbonate, 10 mmol/L potassium chloride, buffer, dye, surfactant, and preservative.
The CO₂ Acid reagent contains 6 mmol/L sulfuric acid and preservative.
The ISE Electrolyte Reference contains 140 mmol/L sodium, 4 mmol/L potassium, 100 mmol/L chloride, 10 mmol/L total CO₂, 2 mmol/L calcium, buffer, surfactant, and preservative.
I. Substantial Equivalence Information:
1. Predicate device name(s):
Beckman PN 443320, 443330, 450214
2. Predicate K number(s):
K942676, K864236
3. Comparison with predicate:
These reagents are intended to be able to be used in place of the OEM reagents. The device and its predicate have the same intended use, composition, packaging quantities and dimensions, storage conditions, and shelf life.
J. Standard/Guidance Document Referenced (if applicable):
K. Test Principle:
These reagents are for use in the calibration of the Synchron CX® & CX® Delta Systems. Please refer to the predicate submission for test principles.
L. Performance Characteristics (if/when applicable):
1. Analytical performance:
a. Precision/Reproducibility:
{3}
Total precision was measured on the CX® Delta analyzer. Samples were measured in triplicate 4 to 5 days in 2 to 5 runs per day. Results are summarized below (units = mmol/L):
| Analyte | Level | N | Mean | SD | % CV |
| --- | --- | --- | --- | --- | --- |
| Sodium (Na+) | DControl 1 | 43 | 142.0 | 6.4 | 4.5 |
| | DControl 2 | 38 | 155.7 | 6.9 | 4.4 |
| | CSF 1 | 13 | 127.4 | 6 | 3.8 |
| | CSF 2 | 12 | 159.9 | 5 | 4.9 |
| | Urine 1 | 48 | 65.6 | 4.9 | 7.5 |
| | Urine 2 | 44 | 208.2 | 7.9 | 3.8 |
| Potassium (K+) | DControl 1 | 43 | 4.22 | 0.22 | 5.10 |
| | DControl 2 | 45 | 8.10 | 0.37 | 4.59 |
| | Urine 1 | 48 | 33.7 | 2.2 | 6.4 |
| | Urine 2 | 45 | 108.4 | 6.1 | 5.6 |
| Chloride (Cl-) | DControl 1 | 43 | 107 | 5 | 4 |
| | DControl 2 | 38 | 118 | 5 | 4 |
| | CSF 1 | 13 | 89 | 4 | 4 |
| | CSF 2 | 12 | 105 | 3 | 3 |
| | Urine 1 | 47 | 100 | 6 | 6 |
| | Urine 2 | 47 | 256 | 10 | 4 |
| Calcium (Ca++) | DControl 1 | 31 | 2.51 | 0.08 | 3.06 |
| | DControl 2 | 27 | 3.35 | 0.18 | 5.23 |
| | Urine 1 | 29 | 1.57 | 0.14 | 8.97 |
| | Urine 2 | 32 | 3.50 | 0.14 | 3.99 |
| CO2 | DControl 1 | 40 | 14.8 | 0.7 | 4.4 |
| | DControl 2 | 43 | 25.4 | 1.1 | 4.2 |
Within run precision was tested by running serum controls (or samples for calcium) in triplicate six times. Sodium, potassium, chloride, and $\mathrm{CO}_{2}$ were tested on the CX®3 analyzer while calcium was tested on the CX® Delta analyzer. Results are summarized below (units = mmol/L):
| Analyte | Level | N | Mean | SD | % CV |
| --- | --- | --- | --- | --- | --- |
| Sodium (Na+) | DControl 1 | 18 | 139.7 | 0.08 | 0.6 |
| | DControl 2 | 18 | 154.1 | 1.03 | 0.7 |
| Potassium (K+) | DControl 1 | 14 | 4.07 | 0.01 | 0.29 |
| | DControl 2 | 12 | 7.36 | 0.03 | 0.42 |
| Chloride (Cl-) | DControl 1 | 18 | 104 | 0.5 | 0.5 |
| | DControl 2 | 18 | 118 | 0.8 | 0.7 |
| CO2 | DControl 1 | 18 | 14.8 | 0.2 | 1.4 |
| | DControl 2 | 18 | 26.5 | 0.3 | 1.3 |
| Calcium (Ca++) | Normal | 18 | 2.42 | 0.06 | 2.5 |
| | Low | 18 | 1.96 | 0.06 | 3.1 |
{4}
Page 4 of 5
| High | 18 | 3.65 | 0.13 | 3.6 |
| --- | --- | --- | --- | --- |
b. Linearity/assay reportable range:
Not applicable.
c. Traceability (controls, calibrators, or method):
Calibrators and reagents are prepared gravimetrically and traceable to the following standards (using the following methods):
| Analyte | Standard Used for Determination of Analyte Value | Instrument Used |
| --- | --- | --- |
| sodium & potassium | NIST | Flame Photometer |
| Calcium | NIST | Ion selective electrode (ISE) |
| chloride | OEM Calibrator | Ion selective electrode (ISE) |
| Total CO2 | OEM Calibrator | TCO2 module |
Reagents are tested prior to bottling, adjusted if necessary to meet specifications, tested during bottling process, and prior to release to stock for distribution.
Stability studies are summarized for the calibrators. Accelerated studies are being used by the sponsor to estimate the expiration date; however, on-going real time studies are being performed.
d. Detection limit:
Not applicable.
e. Analytical specificity:
Not applicable.
f. Assay cut-off:
Not applicable.
2. Comparison studies:
a. Method comparison with predicate device:
The analyzer was calibrated using the Mission reagents and the predicate reagents, and samples were measured and compared to yield the following correlations:
CO2:
- Range = 8 – 30.7 mmol/L, y = 0.93x – 1.58, R² = 0.991
Chloride:
- Range = 54 – 203 mmol/L, y = 0.97x – 1.49, R² = 0.998
Potassium:
- Range = 1.81 – 8.79 mmol/L, y = 0.99x – 0.03, R² = 0.998
Sodium:
- Range = 95.4 – 203.4 mmol/L, y = 0.95x – 3.59, R² = 0.992
{5}
Page 5 of 5
b. Matrix comparison:
Not applicable.
3. Clinical studies:
a. Clinical sensitivity:
Not applicable.
b. Clinical specificity:
Not applicable.
c. Other clinical supportive data (when a and b are not applicable):
4. Clinical cut-off:
Not applicable.
5. Expected values/Reference range:
Not applicable.
M. Conclusion:
I recommend that the Mission Diagnostic CO₂ Alkaline Buffer for Beckman
Synchron CX® & CX® Delta Systems, CO₂ Acid Reagent for Beckman Synchron
CX® & CX® Delta Systems, ISE Electrolyte Reference for Beckman Synchron CX®
& CX® Delta Systems are substantially equivalent to the legally marketed predicate
device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.