FERRITIN CALIBRATOR

K963493 · Dade Chemistry Systems, Inc. · JIT · Dec 3, 1996 · Clinical Chemistry

Device Facts

Record IDK963493
Device NameFERRITIN CALIBRATOR
ApplicantDade Chemistry Systems, Inc.
Product CodeJIT · Clinical Chemistry
Decision DateDec 3, 1996
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 862.1150
Device ClassClass 2

Indications for Use

The Ferritin Calibrator is intended to be used to calibrate the ferritin (FERR) method on the Dimension® RxL clinical chemistry system.

Device Story

Ferritin Calibrator; lyophilized bovine serum-based product; kit contains ten vials across five levels. Level 1 contains no detectable ferritin; Levels 2-5 contain human liver ferritin. Used in clinical chemistry laboratory settings to calibrate ferritin assays on Dimension® RxL systems. Provides reference points for quantitative measurement of ferritin levels in patient samples; assists healthcare providers in diagnostic assessment of iron status.

Technological Characteristics

Lyophilized bovine serum-based matrix; five-level calibration kit; reconstituted volume 2.0 mL per vial. Intended for use with Dimension® RxL clinical chemistry system.

Indications for Use

Indicated for use as a calibrator for the ferritin (FERR) method on the Dimension® RxL clinical chemistry system. No specific patient population, age, or gender restrictions defined.

Regulatory Classification

Identification

A calibrator is a device intended for medical purposes for use in a test system to establish points of reference that are used in the determination of values in the measurement of substances in human specimens. (See also § 862.2 in this part.)

Special Controls

*Classification.* Class II (special controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0} DADE K963493 Dec. 3, 1996 DADE INTERNATIONAL Interoffice Memorandum # Summary of Safety and Effectiveness Information This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92. ## Submitter's Name: Cathy P. Craft Dade International Inc. P.O. Box 6101 Newark, DE 19714-6101 **Date of Preparation:** 8/29/96 **Name of Product:** Ferritin Calibrator **FDA Classification Name:** Calibrator **Predicate Device:** Abbott Laboratories*IMx® Ferritin Calibrator **Device Description:** The Ferritin Calibrator is a lyophilized bovine serum-based product. The Level 1 calibrator contains no detectable ferritin. Levels 2 through 5 contain human liver ferritin. The kit consists of ten vials; two at each of five levels. **Intended use:** The Ferritin Calibrator is intended to be used to calibrate the ferritin (FERR) method on the Dimension® RxL clinical chemistry system. ## Comparison to Predicate Device: | Item | Ferritin Calibrator | IMx® Ferritin Calibrator | | --- | --- | --- | | Intended Use | Calibrator | Calibrator | | Analytes | ferritin | ferritin | | Matrix | bovine serum base | buffered aqueous base | | Form | lyophilized | liquid | | Volume | 2.0 mL per vial, reconstituted | 4.0 mL per vial | | Levels | 5 levels | 6 levels | * IMx® is a trademark of Abbott Laboratories, Abbott Park, IL 60064 4 {1} # Comments on Substantial **Equivalence:** Both the Ferritin Calibrator and the Abbott IMx® Ferritin Calibrator are intended to be used as calibrators for ferritin methods. **Conclusion:** The Ferritin Calibrator is substantially equivalent to the IMx® Ferritin Calibrator based on the comparison discussed above. Cathy P. Craft Regulatory Affairs and Compliance Manager Date: 8/29/96 5 10
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...