Lumipulse G Prolactin
Device Facts
| Record ID | K163544 |
|---|---|
| Device Name | Lumipulse G Prolactin |
| Applicant | Fujirebio Diagnostics,Inc. |
| Product Code | JIT · Clinical Chemistry |
| Decision Date | Jan 12, 2017 |
| Decision | SESE |
| Submission Type | Abbreviated |
| Regulation | 21 CFR 862.1150 |
| Device Class | Class 2 |
Indications for Use
Lumipulse G Prolactin Calibrators are for in vitro diagnostic use in the calibration of Lumipulse G Prolactin on the LUMIPULSE G System.
Device Story
Lumipulse G Prolactin Calibrators are liquid, ready-to-use reagents used to calibrate the Lumipulse G Prolactin assay on the LUMIPULSE G System. The kit consists of two levels: 0 ng/mL and 400 ng/mL. The calibrators are prepared gravimetrically using prolactin antigen in a Tris buffer with protein stabilizer and sodium azide preservative. They are used by laboratory professionals in a clinical setting to establish the calibration curve for the automated immunoassay analyzer. By ensuring accurate calibration, the device enables the analyzer to provide reliable quantitative prolactin measurements, which assist clinicians in diagnosing and monitoring conditions related to prolactin levels.
Clinical Evidence
Bench testing only. No clinical data provided. Performance was validated through analytical studies including real-time and accelerated stability testing, transport simulation, and value assignment verification. Stability studies confirmed a 12-month shelf life at 2-10°C.
Technological Characteristics
Liquid, ready-to-use calibrator kit. Contains prolactin in 0.15 M sodium chloride, Tris buffer, bovine protein stabilizer, and sodium azide preservative. Traceable to 3rd International Standard 84/500. Designed for use on the LUMIPULSE G System. Standards referenced: ISO 17511:2003, CLSI EP25-A.
Indications for Use
Indicated for in vitro diagnostic use in the calibration of the Lumipulse G Prolactin assay on the LUMIPULSE G System. No specific patient population or contraindications described.
Regulatory Classification
Identification
A calibrator is a device intended for medical purposes for use in a test system to establish points of reference that are used in the determination of values in the measurement of substances in human specimens. (See also § 862.2 in this part.)
Special Controls
*Classification.* Class II (special controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9.
Related Devices
- K163521 — Lumipulse G LH Calibrators · Fujirebio Diagnostics,Inc. · Jan 12, 2017
- K052982 — ROCHE DIAGNOSTICS COBAS ELECSYS PROLACTIN II CALSET · Roche Diagnostics · Oct 28, 2005
- K163546 — Lumipulse G Progesterone-N Calibrators · Fujirebio Diagnostics,Inc. · Jan 13, 2017
- K053059 — ROCHE DIAGNOSTICS ELECSYS PROLACTIN II CALCHECK · Roche Diagnostics · Nov 28, 2005
- K970147 — ELECSYS CALCHECK PROLACTIN · Boehringer Mannheim Corp. · Feb 10, 1997