← Product Code [JIT](/submissions/CH/subpart-b%E2%80%94clinical-chemistry-test-systems/JIT) · K111952

# ACCESS FOLATE CALIBRATORS ON THE ACCESS IMMUNOASSAY SYSTEMS MODEL A98033 (K111952)

_Beckman Coulter, Inc. · JIT · Aug 19, 2011 · Clinical Chemistry · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/CH/subpart-b%E2%80%94clinical-chemistry-test-systems/JIT/K111952

## Device Facts

- **Applicant:** Beckman Coulter, Inc.
- **Product Code:** [JIT](/submissions/CH/subpart-b%E2%80%94clinical-chemistry-test-systems/JIT.md)
- **Decision Date:** Aug 19, 2011
- **Decision:** SESE
- **Submission Type:** Abbreviated
- **Regulation:** 21 CFR 862.1150
- **Device Class:** Class 2
- **Review Panel:** Clinical Chemistry

## Indications for Use

The Access Folate Calibrators are intended to calibrate the Access Folate assay for the quantitative determination of folic acid levels in human serum, plasma (heparin) and red blood cells using the Access Immunoassay Systems.

## Device Story

Access Folate Calibrators consist of a six-level set (0 to ~24.8 ng/mL) of buffered matrix containing folate (pteroylmonoglutamic acid), human serum albumin, sodium azide, and ProClin 300. Used to calibrate Access Immunoassay Systems for quantitative folate determination in serum, plasma, and red blood cells. Calibration cards with encrypted bar codes facilitate instrument setup. S0 calibrator also serves as a diluent for high-concentration samples. Stored at -20°C or colder; thawed vials stable at 2-10°C. Provides standardized reference points for clinical laboratory assays, ensuring accurate quantification of patient folate levels.

## Clinical Evidence

No clinical data. Bench testing only. Traceability established against USP Folic Acid Reference Standard and WHO International Standard 03/178. Value assignment performed via replicate analysis on Access 2 analyzers with 95% CI. Real-time stability studies confirmed shelf life and storage conditions.

## Technological Characteristics

Six-level liquid calibrator set; folate (pteroylmonoglutamic acid) in buffered matrix with human serum albumin, surfactant, <0.1% sodium azide, and 0.25% ProClin 300. Concentrations: 0, 1.2, 3.1, 6.2, 12.4, 24.8 ng/mL. Bar-coded calibration cards for automated system input. Storage: -20°C or colder. Compatible with Access Immunoassay Systems.

## Regulatory Identification

A calibrator is a device intended for medical purposes for use in a test system to establish points of reference that are used in the determination of values in the measurement of substances in human specimens. (See also § 862.2 in this part.)

## Special Controls

*Classification.* Class II (special controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9.

## Predicate Devices

- Access Folate Calibrators (k060774)

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}

1

# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY ONLY TEMPLATE

A. 510(k) Number:
k111952

B. Purpose for Submission:
New Device

C. Measurand:
Calibrators for Folic Acid.

D. Type of Test:
Not Applicable.

E. Applicant:
Beckman Coulter, Inc.

F. Proprietary and Established Names:
Access Folate Calibrators.

G. Regulatory Information:
|  Product Code | Classification | Regulation Section | Panel  |
| --- | --- | --- | --- |
|  JIT | Class II | 21 CFR § 862.1150, Calibrator | Clinical Chemistry (75)  |

H. Intended Use:

1. Intended use(s):
Please see intended use below.

2. Indication(s) for use:
The Access Folate Calibrators are intended to calibrate the Access Folate assay for the quantitative determination of folic acid levels in human serum, plasma (heparin) and red blood cells using the Access Immunoassay Systems.

{1}

3. Special conditions for use statement(s):

For prescription Use.

4. Special instrument requirements:

To be used on the Access Immunoassay Systems.

I. Device Description:

The Access Folate Calibrators are a six level calibrator set intended to calibrate the Access Folate assay for the quantitative determination of folic acid levels in human serum, plasma (heparin) and red blood cells using the Access Immunoassay System. The calibrator set provides calibrators at six levels – zero and approximately 1.2, 3.1, 6.2, 12.4 and 24.8 ng/mL (2.8, 7.0, 14.0, 28.1, and 56.2 nmol/L). The calibrators are contained in 4.0 mL vials. The calibrator vials are intended for storage at -20°C or colder.

Calibration cards are provided with each calibrator kit. Calibration cards contain bar codes that are encrypted with the individual calibrator concentrations for each calibrator level.

Folate Calibrator S0 is intended for use with Access Folate assay to dilute patient samples containing analyte concentrations greater than the analyte specific S5 calibrator. The Access Folate Calibrators are a buffered matrix with human serum albumin (HSA) surfactant, &lt; 0.1% sodium azide, and 0.25% ProClin 300.

J. Substantial Equivalence Information:

1. Predicate device name(s):

Access Folate Calibrators

2. Predicate 510(k) number(s):

k060774

3. Comparison with predicate:

|  Attribute | Access Folate Calibrator (Candidate Device) | Access Folate Calibrator (Predicate - k060774)  |
| --- | --- | --- |
|  Indication for use / Intended for Use | The Access Folate Calibrators are intended to calibrate the Access Folate assay for the quantitative determination of folic acid levels in human serum, plasma (heparin) and red blood cells using the Access Immunoassay Systems. | Same  |
|  Test System (Instrumentation / technology) | Access Immunoassay Systems | Same  |
|  Calibrators Antigen | Folate (pteroylmonoglutamic acid) in buffered matrix | Same  |
|  Calibrator Range | 0 – 25 ng/mL | 0 – 20 ng/mL  |

{2}

3

|  Storage Temperature after opening | 2 - 10°C | Same  |
| --- | --- | --- |
|  Shelf life | 6 months | 12 months  |

K. Standard/Guidance Document Referenced (if applicable):

ISO 14971:2007 Medical Devices - Application of risk management to medical devices.

WHO International Standard Vitamin B12 and Serum Folate; 03/178

L. Test Principle:

Not applicable.

M. Performance Characteristics (if/when applicable):

1. Analytical performance:

a. Precision/Reproducibility:

Not applicable.

b. Linearity/assay reportable range:

Not applicable.

c. Traceability, Stability, Expected values (controls, calibrators, or methods):

Traceability

The Access Folate Calibrator (A98033) is traceable to the USP Folic Acid Reference Standard. The reference standard is rehydrated and used to prepare six predetermined concentrations which are then compared to the WHO International Standard 03/178.

Calibrator Value Assignment

For each calibrator level, multiple replicates of the reference calibrators and multiple replicates of the commercial calibrators are analyzed on multiple Access 2 analyzers. These values are then used to determine the commercial calibrator values.

A reference curve is then established using multiple Access Analyzers with a 95% confidence interval (CI).

Stability

Real time stability study protocols and acceptance criteria were described and found to be acceptable.

The claimed calibrators' stability is 7 months at -20°C for unopened vials and at 2-10°C 90 days when thawed.

{3}

d. Detection limit:
Not applicable.

e. Analytical specificity:
Not applicable.

f. Assay cut-off:
Not applicable.

2. Comparison studies:
a. Method comparison with predicate device:
Not applicable.
b. Matrix comparison:
Not applicable.

3. Clinical studies:
a. Clinical Sensitivity:
Not applicable.
b. Clinical specificity:
Not applicable.
c. Other clinical supportive data (when a. and b. are not applicable):
Not applicable.

4. Clinical cut-off:
Not applicable.

{4}

5. Expected values/Reference range:

|  Level | Expected Values  |
| --- | --- |
|  Calibrator Level S0 | 0 ng/mL
0 nmol/L  |
|  Calibrator Level S1 | 1.2 ng/mL
2.8 nmol/L  |
|  Calibrator Level S2 | 3.1 ng/mL
7.0 nmol/L  |
|  Calibrator Level S3 | 6.2 ng/mL
14.0 nmol/L  |
|  Calibrator Level S4 | 12.4 ng/mL
28.1 nmol/L  |
|  Calibrator Level S5 | 24.8 ng/mL
56.2 nmol/L  |
|  Calibrator Range | 0 – 25 ng/mL  |

N. Proposed Labeling:

The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10.

O. Conclusion:

The submitted information in this premarket notification is complete and supports a substantial equivalence decision.

---

**Source:** [https://fda.innolitics.com/submissions/CH/subpart-b%E2%80%94clinical-chemistry-test-systems/JIT/K111952](https://fda.innolitics.com/submissions/CH/subpart-b%E2%80%94clinical-chemistry-test-systems/JIT/K111952)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
