DIMENSION VISTA CDT CAL, DIMENSION VISTA CDT CON L, DIMENSION VISTA CDT CON H

K110169 · Siemens Healthcare Diagnostics · JIT · Apr 7, 2011 · Clinical Chemistry

Device Facts

Record IDK110169
Device NameDIMENSION VISTA CDT CAL, DIMENSION VISTA CDT CON L, DIMENSION VISTA CDT CON H
ApplicantSiemens Healthcare Diagnostics
Product CodeJIT · Clinical Chemistry
Decision DateApr 7, 2011
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 862.1150
Device ClassClass 2

Indications for Use

Dimension Vista® CDT CAL CDT CAL is an in vitro diagnostic product for the calibration of the carbohydrate-deficient transferring (CDT) method on the Dimension Vista® System. Dimension Vista® CDT CON L CDT CON L is an assayed, low level, intra-laboratory quality control for the assessment of precision and analytical bias on the Dimension Vista® System in the quantitative determination of carbohydrate-deficient transferrin (CDT) Dimension Vista® CDT CON H CDT CON H is an assayed, high level, intra-laboratory quality control for the assessment of precision and analytical bias on the Dimension Vista® System in the quantitative determination of carbohydrate-deficient transferring (CDT).

Device Story

Dimension Vista® CDT CAL, CON L, and CON H are liquid, human serum-based products containing carbohydrate-deficient transferrin. Used as calibrators and assayed quality controls for the Dimension Vista® System. Laboratory personnel use these products to calibrate the CDT assay and monitor analytical precision and bias. The output provides reference values for system calibration and performance verification, ensuring accurate quantitative measurement of CDT in patient samples. These products support clinical decision-making by maintaining the accuracy of diagnostic testing for CDT, a biomarker often associated with chronic alcohol consumption.

Clinical Evidence

No clinical data. Bench testing only. Stability was validated via real-time studies confirming performance until expiration when stored at 2-8°C and 14-day open-vial stability on the system. Traceability established against an internally produced master calibrator calibrated against a highly purified in-house protein preparation. Value assignment performed via quadruplicate analysis across multiple runs on the Dimension Vista 1500 analyzer.

Technological Characteristics

Liquid human serum-based products. Calibrator and assayed quality control materials. Storage at 2-8 °C. Shelf life of 24 months. Designed for use on the Dimension Vista® System. No specific materials of construction or software algorithms described.

Indications for Use

Indicated for use as an in vitro diagnostic calibrator and assayed quality control material for the quantitative determination of carbohydrate-deficient transferrin (CDT) on the Dimension Vista® 1500 System. Intended for professional laboratory use.

Regulatory Classification

Identification

A calibrator is a device intended for medical purposes for use in a test system to establish points of reference that are used in the determination of values in the measurement of substances in human specimens. (See also § 862.2 in this part.)

Special Controls

*Classification.* Class II (special controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9.

Predicate Devices

Submission Summary (Full Text)

{0} 1 # 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY ONLY TEMPLATE A. 510(k) Number: k110169 B. Purpose for Submission: New device C. Measurand: Calibrator and Quality control material for Carbohydrate-Deficient Transferrin (CDT) D. Type of Test: Not applicable E. Applicant: Siemens Healthcare Diagnostics F. Proprietary and Established Names: Dimension Vista® CDT CAL Dimension Vista® CDT CON L Dimension Vista® CDT CON H G. Regulatory Information: | Product Code | Classification | Regulation Section | Panel | | --- | --- | --- | --- | | JIS | Class II | 21 CFR 862.1150 | Clinical Chemistry (75) | | JJX | Class I, reserved | 21 CFR 862.1660 | Clinical Chemistry (75) | H. Intended Use: 1. Intended use(s): See indications for use below. {1} 2. Indication(s) for use: Dimension Vista® CDT CAL CDT CAL is an *in vitro* diagnostic product for the calibration of the carbohydrate-deficient transferring (CDT) method on the Dimension Vista® System. Dimension Vista® CDT CON L CDT CON L is an assayed, low level, intra-laboratory quality control for the assessment of precision and analytical bias on the Dimension Vista® System in the quantitative determination of carbohydrate-deficient transferrin (CDT) Dimension Vista® CDT CON H CDT CON H is an assayed, high level, intra-laboratory quality control for the assessment of precision and analytical bias on the Dimension Vista® System in the quantitative determination of carbohydrate-deficient transferring (CDT). 3. Special conditions for use statement(s): For Prescription Use only 4. Special instrument requirements: Dimension Vista® 1500 System I. Device Description: Dimension Vista® CDT CAL CDT CAL is a liquid human serum based product containing carbohydrate-deficient transferring. CDT CAL is packed in 4 vials containing 2.0 ml calibrator material per vial. Dimension Vista® CDT CON L and CON H CDT CON L and CON H are liquid human serum based products containing human carbohydrate-deficient transferrin. CDT CON L and CDT CON H are packaged in 4 vials containing 2.0 ml quality control material per vial. All human source materials were tested and found to be negative for HIV ½, HBsAg, and HCV by FDA- approved methods. J. Substantial Equivalence Information: 1. Predicate device name(s) {2} N Latex CDT Kit with calibrator and control materials 2. Predicate 510(k) number(s): k060677 3. Comparison with predicate: | Similarities and Differences | | | | --- | --- | --- | | Item | Dimension Vista®CDT CAL Candidate Device | N Latex CDT Calibrator Predicate Device (k060677) | | Intended Use | For the calibration of the carbohydrate-deficient (CDT) method | Same | | Composition | Liquid human serum based product containing carbohydrate-deficient transferring. | Same | | Shelf-Life | 24 months | Same | | Storage Temperature | 2-8°C | Same | | Form | Liquid, ready to use | Same | | Instrumentation | Dimension Vista® 1500 System | Dimension BN™II and BN ProSpec System | | Similarities and Differences | | | | --- | --- | --- | | Item | Dimension Vista®CDT CON Candidate Device | N Latex CDT Control Predicate Device (k060677) | | Intended Use | For in vitro diagnostic use as an assayed quality control material to monitor the precision of laboratory testing procedures for the CDT method | Same | | Composition | Liquid human serum based product containing human carbohydrate-deficient transferring. | Same | | Shelf-life | 24 months | Same | | Storage Temperature | 2-8°C | Same | | Form | Liquid, ready to use | Same | | Instrumentation | Dimension Vista® 1500 System | Dimension BN™II and BN ProSpec System | {3} K. Standard/Guidance Document Referenced (if applicable): EP-5A2 Evaluation of Precision Performance of Quantitative Measurement Methods Guidance for Industry and FDA Staff-Assayed and Unassayed Quality Control Material L. Test Principle: Not applicable M. Performance Characteristics (if/when applicable): 1. Analytical performance: a. Precision/Reproducibility: Not applicable b. Linearity/assay reportable range: Not applicable c. Traceability, Stability, Expected values (controls, calibrators, or methods): Traceability The Dimension Vista CDT controls and calibrator are traceable to an internally produced master calibrator which is calibrated against a highly purified in-house protein preparation. Stability Stability testing protocols and acceptance criteria were reviewed and found to be acceptable. Stability characteristics of the Dimension CDT controls and calibrators were determined using real-time stability studies. The Dimension CDT controls and calibrators are stable until the expiration date printed on the vial when stored unopened at 2-8°C. Open vial stability is 14 days when stored on the Dimension Vista system. Storage recommendations are provided in the labeling. Value Assignment Three vials of each single level calibrator and control are analyzed in quadruplicate. The target value of the Dimension Vista CDT controls and calibrator is the median of the observed values which are obtained during multiple runs on the Dimension Vista 1500 analyzer. The expected ranges for the controls are ±20% of the target value. The expected values are provided in the labeling for each specific lot. 4 {4} d. Detection limit: Not applicable e. Analytical specificity: Not applicable f. Assay cut-off: Not applicable 2. Comparison studies: a. Method comparison with predicate device: Not applicable b. Matrix comparison: Not applicable 3. Clinical studies: a. Clinical Sensitivity: Not applicable b. Clinical specificity: Not applicable c. Other clinical supportive data (when a. and b. are not applicable): Not applicable 4. Clinical cut-off: Not applicable 5. Expected values/Reference range: The labeling for the control materials states that the values provided are intended only as a guideline. Each laboratory should determine its own individual values. A typical control value for the Vista CDT L control is 64 mg/L. A typical control value for the Vista CDT H control is 167 mg/L. {5} The lot specific value for the calibrator material is stated in the labeling. The target range is 299 mg/L – 365 mg/L. **N. Proposed Labeling:** The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10. **O. Conclusion:** The submitted information in this premarket notification is complete and supports a substantial equivalence decision. 6
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...