← Product Code [JIT](/submissions/CH/subpart-b%E2%80%94clinical-chemistry-test-systems/JIT) · K103360

# DIMENSION VISTA DIGITOXIN CALIBRATOR - DGTX CAL (K103360)

_Siemens Corp. · JIT · Dec 17, 2010 · Clinical Chemistry · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/CH/subpart-b%E2%80%94clinical-chemistry-test-systems/JIT/K103360

## Device Facts

- **Applicant:** Siemens Corp.
- **Product Code:** [JIT](/submissions/CH/subpart-b%E2%80%94clinical-chemistry-test-systems/JIT.md)
- **Decision Date:** Dec 17, 2010
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 862.1150
- **Device Class:** Class 2
- **Review Panel:** Clinical Chemistry

## Indications for Use

The DGTX CAL is an in vitro diagnostic product for the calibration of digitoxin on the Dimension Vista® System.

## Device Story

DGTX CAL is a liquid, bovine serum-based calibrator containing digitoxin. It is used to calibrate digitoxin assays on the Dimension Vista® System. The device is intended for professional use in clinical laboratory environments. By providing a known concentration of digitoxin, it enables the system to establish a calibration curve, which is then used to quantify digitoxin levels in patient samples. This process ensures the accuracy of clinical measurements, assisting healthcare providers in monitoring therapeutic drug levels and making informed clinical decisions regarding patient treatment.

## Clinical Evidence

No clinical data provided; substantial equivalence is based on identical formulation and technological characteristics to the predicate device.

## Technological Characteristics

Liquid bovine serum-based calibrator; 5-level concentration set; stored at 2–8 °C. Designed for use with the Dimension Vista® System. No specific materials or software algorithms described beyond its function as a chemical calibrator.

## Regulatory Identification

A calibrator is a device intended for medical purposes for use in a test system to establish points of reference that are used in the determination of values in the measurement of substances in human specimens. (See also § 862.2 in this part.)

## Special Controls

*Classification.* Class II (special controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9.

## Predicate Devices

- Dimension® Drug Calibrator II ([K033809](/device/K033809.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

## 510(k) Summary of Safety and Effectiveness for the Dimension Vista® DGTX CAL

This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.

A. 510(k) Number: K103360

B. Date of Preparation:

C. Proprietary and Established Names:

Dimension Vista® DGTX CAL

D. Applicant:

Siemens Healthcare Diagnostics Inc.

P.O. Box 6101, Newark, DE 19714-6101

Rose T. Marinelli, Regulatory Technical Specialist

Office Number: (302) 631-8805 fax Number: (302) 631-6299

## E. Regulatory Information:

Drug Calibrator II:

1. Regulation section: 21 CFR § 862.1150 Calibrator

2. Classification: Class II

3. Product Code: JIT - Calibrator, Secondary

4. Panel: Immunology

#### F. Predicate Device:

The predicate device used to demonstrate substantial equivalence to the Dimension Vista® DGTX CAL is the Dimension® Drug Calibrator 11 cleared under k033809.

#### G. Device Description:

The DGTX CAL is a liquid bovine serum based product containing digitoxin.

### H. Intended Use:

The DGTX CAL is an in vitro diagnostic product for the calibration of digitoxin on the Dimension Vista® System.

{1}------------------------------------------------

## I. Substantial Equivalence Information:

The Dimension Vista® DGTX CAL was compared to the predicate, Dimension® Drug Calibrator II. The following table provides a comparison of the important similarities and differences:

| Feature      | Dimension Vista® DGTX CAL                                                                                                | Drug Calibrator II (DC49D)<br>(k033809) (predicate device)                                                                                                                                                                                                                                                                                                             |
|--------------|--------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Intended Use | The DGTX CAL is an in vitro<br>diagnostic product for the<br>calibration of digitoxin on the<br>Dimension Vista® System. | Drug Calibrator II is an in vitro<br>diagnostic product intended for the<br>calibration of the following methods<br>packaged in Flex® reagent<br>cartridges:<br>Acetaminophen (ACTM) Carbamazepine (CRBM) Digitoxin (DGTX) Gentamicin (GENT) Lidocaine (LIDO) N-acetylprocainamide (NAPA) Procainamide (PROC) Tobramycin (TOBR) Valproic Acid (VALP) Vancomycin (VANC) |
| Matrix       | Bovine Serum base                                                                                                        | Bovine Serum base                                                                                                                                                                                                                                                                                                                                                      |
| Levels       | 5 Levels                                                                                                                 | 5 Levels                                                                                                                                                                                                                                                                                                                                                               |
| Preparation  | Liquid                                                                                                                   | Liquid                                                                                                                                                                                                                                                                                                                                                                 |
| Storage      | 2 – 8 °C                                                                                                                 | 2 – 8 °C                                                                                                                                                                                                                                                                                                                                                               |

# J. Conclusion:

The Dimension Vista® DGTX CAL is substantially equivalent to the current Dimension® Drug Calibrator II cleared under k033809. The Dimension Vista® DGTX CAL product is the exact same formulation as Dimension® Drug Cal II.

{2}------------------------------------------------

Public Health Service

Siemens Healthcare Diagnostics Inc. c/o Ms. Rose Merinelli, Regulatory Technical Specialist PO Box 6101 MS 514 Newark, DE 19714-6101

Food & Drug Administration 10903 New Hampshire Avenue Building 66 Silver Spring, MD 20993

DEC 1 7 2010

# Re: k103360

Trade/Device Name: Dimension Vista® Digitoxin Calibrator – DGTX CAL Regulation Number: 21 CFR 862.1150 Regulation Name: Calibrator Regulatory Class: Class II Product Code: JIT Dated: November 12, 2010 Received: November 16, 2010

Dear Ms. Merinelli:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food., Orug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into class II (Special Controls), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21. Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further. announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820). This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

{3}------------------------------------------------

# Page 2

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 796-5450. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at (301) 796-5760. For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or ( 301 ) 796-5680 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

CA

Courtney C. Harper, Ph.D. Director Division of Chemistry and Toxicology Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

# Indications for Use Form

510(k) Number (if known): 103360

Device Name: _________________________________________________________________________________________________________________________________________________________________

Indications for Use: The DGTX CAL is an in vitro diagnostic product for the calibration of digitoxin on the Dimension Vista® System.

Prescription Use X (Part 21 CFR 801 Subpart D)

AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

Page 1 of

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)

Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)

Carol C. Benson

Division Sign-Off Office of In Vitro Diagnostic Device Evaluation and Safety

510(k) K103360

---

**Source:** [https://fda.innolitics.com/submissions/CH/subpart-b%E2%80%94clinical-chemistry-test-systems/JIT/K103360](https://fda.innolitics.com/submissions/CH/subpart-b%E2%80%94clinical-chemistry-test-systems/JIT/K103360)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
