← Product Code [JIT](/submissions/CH/subpart-b%E2%80%94clinical-chemistry-test-systems/JIT) · K032458

# ABBOTT ARCHITECT ESTRADIOL CALIBRATORS; ABBOTT IMX ESTRADIOL CALIBRATOR; ABBOTT IMX ESTRADIOL MODE 1 CALIBRATOR; ABBOTT (K032458)

_Abbott Laboratories · JIT · Oct 6, 2003 · Clinical Chemistry · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/CH/subpart-b%E2%80%94clinical-chemistry-test-systems/JIT/K032458

## Device Facts

- **Applicant:** Abbott Laboratories
- **Product Code:** [JIT](/submissions/CH/subpart-b%E2%80%94clinical-chemistry-test-systems/JIT.md)
- **Decision Date:** Oct 6, 2003
- **Decision:** SESE
- **Submission Type:** Abbreviated
- **Regulation:** 21 CFR 862.1150
- **Device Class:** Class 2
- **Review Panel:** Clinical Chemistry

## Indications for Use

Abbott ARCHITECT Estradiol Calibrators are devices intended for use in the ARCHITECT Estradiol assay test system to establish points of reference that are used in the quantitative determination of estradiol in human specimens. Estradiol measurements are used in the diagnosis and treatment of various hormonal sexual disorders and in assessing placental function in complicated pregnancy. Abbott AxSYM Estradiol Standard Calibrators and AxSYM Estradiol Master Calibrators are devices intended for use in the AxSYM Estradiol assay test system to establish points of reference that are used in the quantitative determination of estradiol in human specimens. Estradiol measurements are used in the diagnosis and treatment of various hormonal sexual disorders and in assessing placental function in complicated pregnancy. Abbott ARCHITECT FSH Calibrators are devices intended for use in the ARCHITECT FSH assay test system to establish points of reference that are used in the quantitative determination of follicle-stimulating hormone (FSH) in human specimens. FSH measurements are used in the diagnosis and treatment of pituitary gland and gonadal discorders ARCHITECT LH Calibrators are devices intended for use in the ARCHITECT LH assay test system to establish points of reference that are used in the quantitative determination of luteinizing hormone (LH) in human specimens. LH measurements are used in the diagnosis and treatment of gonadal function. ARCHITECT Progesterone Calibrators are devices intended for use in the ARCHITECT Progesterone assay test system to establish points of reference that are used in the quantitative determination of progesterone in human specimens. Progesterone measurements are used in the diagnosis and treatment of disorders of the ovaries or placenta. ARCHITECT Prolactin Calibrators are devices intended for use in the ARCHITECT Prolactin assay test system to establish points of reference that are used in the quantitative determination of prolactin in human specimens. Prolactin measurements are used in the diagnosis and treatment of disorders of the anterior pituitary gland or of the hypothalamus portion of the brain AxSYM Total T3 Standard Calibrators and AxSYM Total T3 Master Calibrators are devices intended for use in the AxSYM Total T3 assay test system to establish points of reference that are used in the quantitative determination of total T3 in human specimens. Total T3 measurements are used in the diagnosis and treatment of thyroid disease.

## Device Story

Calibrators function as reference standards for Abbott ARCHITECT and AxSYM immunoassay systems; establish calibration curves for quantitative measurement of hormones (estradiol, FSH, LH, progesterone, prolactin) and thyroid hormone (T3) in human specimens. Used in clinical laboratory settings by trained technicians. Input: calibrator material of known concentration; Output: reference points for assay system. Healthcare providers use resulting quantitative values to diagnose/treat endocrine, reproductive, and thyroid conditions. Benefits: ensures accuracy and consistency of patient test results.

## Clinical Evidence

No clinical data. Substantial equivalence demonstrated via bench testing and adherence to FDA Guidance for Industry 'Abbreviated 510(k) Submissions for In Vitro Diagnostic Calibrators'.

## Technological Characteristics

Calibrators consist of hormone-spiked matrices (Tris buffer, bovine serum, or human serum). Standardization achieved via gravimetric manufacturing and referencing to WHO International Standards or internal standards verified by HPLC/GCMS. Form factor: multi-level liquid calibrators. Connectivity: N/A (reagent component).

## Regulatory Identification

A calibrator is a device intended for medical purposes for use in a test system to establish points of reference that are used in the determination of values in the measurement of substances in human specimens. (See also § 862.2 in this part.)

## Special Controls

*Classification.* Class II (special controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9.

## Predicate Devices

- IMx® Estradiol Calibrators ([K951629](/device/K951629.md))
- Abbott FSH Calibrators ([K890135](/device/K890135.md))
- AxSYM® LH Standard Calibrators ([K935611](/device/K935611.md))
- AxSYM® Progesterone Standard Calibrators ([K955025](/device/K955025.md))
- Abbott Prolactin Calibrators ([K896162](/device/K896162.md))
- Ciba-Corning ACS:180 Prolactin Assay Calibrator ([K934517](/device/K934517.md))
- AxSYM® Total T3 Standard and Master Calibrators

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}

510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION
DECISION SUMMARY
DEVICE ONLY TEMPLATE

A. 510(k) Number:
K032458

B. Analyte:
Calibrator – estradiol, follicle stimulating hormone (FSH), leuteinizing hormone (LH), progesterone, prolactin, triiodothyronine (T3)

C. Type of Test:
Calibrator

D. Applicant:
Abbott Laboratories

E. Proprietary and Established Names:
Abbott ARCHITECT® Estradiol Calibrators
Abbott AxSYM® Estradiol Standard Calibrators
Abbott AxSYM® Estradiol Master Calibrators
Abbott ARCHITECT® FSH Calibrators
Abbott ARCHITECT® LH Calibrators
Abbott ARCHITECT® Progesterone Calibrators
Abbott ARCHITECT® Prolactin Calibrators
Abbott AxSYM® Total T3 Standard Calibrators
Abbott AxSYM® Total T3 Master Calibrators

F. Regulatory Information:
1. Regulation section:
21 CFR § 862.1150, Calibrator
2. Classification:
Class II
3. Product Code:
JIT, Secondary Calibrators
4. Panel:
Clinical Chemistry (75)

G. Intended Use:
1. Indication(s) for use:
Abbott ARCHITECT Estradiol Calibrators are devices intended for use in the ARCHITECT Estradiol assay test system to establish points of reference that are used in the quantitative determination of estradiol in human specimens.

Abbott AxSYM Estradiol Standard Calibrators and AxSYM Estradiol Master Calibrators are devices intended for use in the AxSYM Estradiol assay test system to establish points of reference that are used in the quantitative determination of estradiol in human specimens.

{1}

Page 2 of 7

Abbott ARCHITECT FSH Calibrators are devices intended for use in the ARCHITECT FSH assay test system to establish points of reference that are used in the quantitative determination of follicle-stimulating hormone (FSH) in human specimens.

ARCHITECT LH Calibrators are devices intended for use in the ARCHITECT LH assay test system to establish points of reference that are used in the quantitative determination of luteinizing hormone (LH) in human specimens.

ARCHITECT Progesterone Calibrators are devices intended for use in the ARCHITECT Progesterone assay test system to establish points of reference that are used in the quantitative determination of progesterone in human specimens.

ARCHITECT Prolactin Calibrators are devices intended for use in the ARCHITECT Prolactin assay test system to establish points of reference that are used in the quantitative determination of prolactin in human specimens.

AxSYM Total T3 Standard Calibrators and AxSYM Total T3 Master Calibrators are devices intended for use in the AxSYM Total T3 assay test system to establish points of reference that are used in the quantitative determination of total T3 in human specimens.

2. Special condition for use statement(s):

For professional use.

3. Special instrument Requirements:

The ARCHITECT test system or the AxSYM test system is required.

H. Device Description:

Calibrator components and validation procedures are described below.

I. Substantial Equivalence Information:

1. Predicate device name(s):

IMx® Estradiol Calibrators

Abbott FSH Calibrators

AxSYM® LH Standard Calibrators

AxSYM® Progesterone Standard Calibrators

Abbott Prolactin Calibrators

Ciba-Corning ACS:180 Prolactin Assay Calibrator

AxSYM® Total T3 Standard and Master Calibrators

2. Predicate K number(s):

K951629

K890135

{2}

K935611
K955025
K896162
K934517

3. Comparison with predicate:
All calibrators are identical to their predicate in intended use.

|  Similarities ARCHITECT® Estradiol Calibrators  |   |   |
| --- | --- | --- |
|  Item | Device | Predicate  |
|  matrix | Tris buffer based | Tris buffer based  |
|  traceability | Standardized to an internal reference standard – manufactured gravimetrically using a stock solution of estradiol (not less that 97% pure by HPLC) at each concentration level | Standardized to an internal reference standard – manufactured gravimetrically using a stock solution of estradiol (not less that 97% pure by HPLC) at each concentration level  |
|  Differences  |   |   |
|  Item | Device | Predicate  |
|  number | 2 levels (0 and 1600 pg/mL) | 6 levels (0, 50, 250, 750, 1500, 3000 pg/mL  |
|  Similarities AxSYM® Estradiol Standard Calibrators and AxSYM® Estradiol Master Calibrators  |   |   |
| --- | --- | --- |
|  Item | Device | Predicate  |
|  matrix | Tris buffer based | Tris buffer based  |
|  Differences  |   |   |
|  Item | Device | Predicate  |
|  traceability | Standardized to an internal reference standard for AxSYM Estradiol – standardized to correlate with GC/MS | Standardized to an internal reference standard – manufactured gravimetrically using a stock solution of estradiol (not less that 97% pure by HPLC) at each concentration level  |
|  concentration | 6 levels (0, 50, 100, 200, 500, and 1000 pg/mL) (Master contains two of those – 0 and 200 pg/mL – to establish points of reference) | 6 levels (0, 50, 250, 750, 1500, 3000 pg/mL  |

{3}

Page 4 of 7

|  Similarities ARCHITECT® FSH Calibrators  |   |   |
| --- | --- | --- |
|  Item | Device | Predicate  |
|  matrix | Bovine serum | Bovine serum  |
|  traceability | Manufactured by addition of FSH to a target concentration – referenced against WHO FSH 2nd International Reference Preparation (IRP) 78/549 | Manufactured by gravimetric addition of FSH to a target concentration – referenced against WHO FSH 2nd International Reference Preparation (IRP) 78/549  |
|  Differences  |   |   |
|  Item | Device | Predicate  |
|  number | 2 levels – 0, 100 mIU/mL | 6 levels – 0, 1, 10, 25, 100, 250 mIU/mL  |
|  Similarities ARCHITECT® LH Calibrators  |   |   |
| --- | --- | --- |
|  Item | Device | Predicate  |
|  matrix | calf serum | calf serum  |
|  traceability | Manufactured gravimetrically– referenced to WHO LH Human, Pituitary 2nd International Standard 80/552 at each concentration | Manufactured gravimetrically– referenced to WHO LH Human, Pituitary 2nd International Standard 80/552 at each concentration  |
|  Differences  |   |   |
|  Item | Device | Predicate  |
|  number | 2 levels – 2, 100 mIU/mL | 6 levels – 0, 2, 10, 25, 100, 250 mIU/mL  |
|  Similarities ARCHITECT® Progesterone Calibrators  |   |   |
| --- | --- | --- |
|  Item | Device | Predicate  |
|  Same intended use  |   |   |
|  Differences  |   |   |
|  Item | Device | Predicate  |
|  number | 2 levels – 0.7, 40 ng/mL | 6 levels – 0, 0.7, 2, 7, 20, 40 ng/mL  |
|  matrix | Human serum | Serum for level 0 and Tris buffer based for the remaining levels  |

{4}

Page 5 of 7

|  Similarities ARCHITECT® Prolactin Calibrators  |   |   |
| --- | --- | --- |
|  Item | Device | Predicate  |
|  traceability | Referenced to WHO 3rd International Standard 84/500 for Prolactin | Referenced to WHO 3rd International Standard 84/500 for Prolactin  |
|  matrix | Tris buffer based | Tris buffer based  |
|  Differences  |   |   |
|  Item | Device | Predicate  |
|  number | 2 levels – 5, 30 ng/mL | 6 levels – 0, 5, 10, 30, 80, 200 ng/mL  |
|  Similarities AxSYM® Total T3 Standard and Master Calibrators  |   |   |
| --- | --- | --- |
|  Item | Device | Predicate  |
|  matrix | Bovine serum | Bovine serum  |
|  number | 6 levels – 0, 0.5, 1, 2, 4, 8 ng/mL (only 0 and 1 ng/mL for Master) | 6 levels – 0, 0.5, 1, 2, 4, 8 ng/mL (only 0 and 1 ng/mL for Master)  |
|  Differences  |   |   |
|  Item | Device | Predicate  |
|  traceability | Manufactured using USP Grade L-triiodothyronine, sodium salt and signal matched to internal reference standards which are traceable to the USP Reference Standard L-triiodothyronine (free acid) at each concentration level. The stock solution concentration is determined by HPLC | Internal reference standards are manufactured using L-triiodothyronine Sodium (HPLC purity 95.0 – 101.0%). Primary and secondary standards are manufactured gravimetrically using this reference standard. All list material is tested against these primary and secondary calibrators.  |

J. Standard/Guidance Document Referenced (if applicable):
Abbreviated 510(k) Submissions for In Vitro Diagnostic Calibrators

K. Test Principle:
Not applicable as this submission is for calibrators (for two different test systems).

L. Performance Characteristics (if/when applicable):
1. Analytical performance:
a. Precision/Reproducibility:
Not applicable. See traceability below.
b. Linearity/assay reportable range:
Not applicable.

{5}

Page 6 of 7

c. Traceability (controls, calibrators, or method):

Calibrator validation and stability studies are summarized for each of the calibrators. The sponsor specifies the concentrations of materials evaluated in the studies, the frequency of testing, the method for testing the materials, environmental conditions of storage, and acceptance criteria for the study. Real time studies are being performed to support the stability claims in the labeling.

Abbott ARCHITECT® Estradiol Calibrators are standardized to an internal reference standard and manufactured gravimetrically using a stock solution of estradiol (not less than 97% pure by HPLC) at each concentration level.

Abbott AxSYM® Estradiol Standard Calibrators and Abbott AxSYM® Estradiol Master Calibrators are standardized to an internal reference standard for AxSYM Estradiol which correlates with GC/MS.

Abbott ARCHITECT® FSH Calibrators are manufactured by addition of follicle stimulating hormone to a target concentration and referenced against WHO FSH 2nd International Reference Preparation (IRP) 78/549.

Abbott ARCHITECT® LH Calibrators are manufactured gravimetrically and referenced to WHO Leuteinizing Hormone Human, Pituitary 2nd International Standard 80/552 at each concentration.

Abbott ARCHITECT® Progesterone Calibrators are manufactured by spiking stripped normal male human serum with calculated amounts of a progesterone stock solution. This stock solution is manufactured by adding Progesterone USP gravimetrically to a target concentration which is then assigned to the stock solution. The levels are tested by RLU matching to Progesterone Reference Calibrators. The concentration is adjusted, if necessary, by adding stripped normal male human serum or progesterone stock solution.

Abbott ARCHITECT® Prolactin Calibrators are referenced to WHO 3rd International Standard 84/500 for Prolactin.

Abbott AxSYM® Total T3 Standard Calibrators and Abbott AxSYM® Total T3 Master Calibrators are manufactured using USP Grade L-triiodothyronine, sodium salt and signal matched to internal reference standards which are traceable to the USP Reference Standard L-triiodothyronine (free acid) at each concentration level. The stock solution concentration is determined by HPLC.

d. Detection limit:
Not applicable.

e. Analytical specificity:
Not applicable.

{6}

Page 7 of 7

f. Assay cut-off:
Not applicable.

2. Comparison studies:
a. Method comparison with predicate device:
Not applicable.
b. Matrix comparison:
Not applicable. The sponsor summarizes the methods used for analysis and validation of the calibrator material and matrix.

3. Clinical studies:
a. Clinical sensitivity:
Not applicable.
b. Clinical specificity:
Not applicable.
c. Other clinical supportive data (when a and b are not applicable):
Not applicable.

4. Clinical cut-off:
Not applicable.

5. Expected values/Reference range:
Not applicable.

M. Conclusion:
I recommend that the Abbott ARCHITECT® Estradiol Calibrators, Abbott AxSYM® Estradiol Standard Calibrators, Abbott AxSYM® Estradiol Master Calibrators, Abbott ARCHITECT® FSH Calibrators, Abbott ARCHITECT® LH Calibrators, Abbott ARCHITECT® Progesterone Calibrators, Abbott ARCHITECT® Prolactin Calibrators, Abbott AxSYM® Total T3 Standard Calibrators, and Abbott AxSYM® Total T3 Master Calibrators are substantially equivalent to the legally marketed predicate devices.

---

**Source:** [https://fda.innolitics.com/submissions/CH/subpart-b%E2%80%94clinical-chemistry-test-systems/JIT/K032458](https://fda.innolitics.com/submissions/CH/subpart-b%E2%80%94clinical-chemistry-test-systems/JIT/K032458)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
