VITAMIN D TEST SYSTEM NICHOLS ADVANTAGE 25-HYDROXYVITAMIN D
K012367 · Nichols Institute Diagnostics · JIT · Oct 9, 2001 · Clinical Chemistry
Device Facts
Record ID
K012367
Device Name
VITAMIN D TEST SYSTEM NICHOLS ADVANTAGE 25-HYDROXYVITAMIN D
Applicant
Nichols Institute Diagnostics
Product Code
JIT · Clinical Chemistry
Decision Date
Oct 9, 2001
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 862.1150
Device Class
Class 2
Indications for Use
The Nichols Advantage® 25-Hydroxyvitamin D assay is for use only on the Nichols Advantage® Specialty System. The Nichols Advantage 25-Hydroxyvitamin D assay is intended for the quantitative determination of 25-hydroxyvitamin D (25-OH-D) and other hydroxylated metabolites of vitamin D in serum or plasma to be used as an aid in the assessment of vitamin D sufficiency in an adult population. Assay results should be used in conjunction with other clinical data to assist the clinician in making individual patient management decisions.
Device Story
The Nichols Advantage 25-OH-D is an automated, hands-free immunoassay for use on the Nichols Advantage Specialty System. It utilizes 20 uL of serum or plasma as input. The device employs a competitive ligand binding technique using human vitamin D binding protein and acridinium-ester labeled anti-DBP, with 25-OH-D3 magnetic particles as the solid phase. The system performs automated incubation (two 21-minute steps at 37°C) without requiring manual extraction. The device outputs quantitative 25-OH-D concentrations (ng/mL). It is intended for clinical laboratory use to assist physicians in assessing vitamin D sufficiency and managing patient care. By providing automated, high-precision quantification of vitamin D levels, the device aids in the diagnosis and monitoring of vitamin D deficiency and insufficiency, particularly in patients with conditions like chronic renal insufficiency.
Clinical Evidence
Clinical evaluation included a method comparison study (n=111) against the DiaSorin RIA kit, showing an overall agreement of 93.9% using a 20 ng/mL cutoff. A separate clinical study (n=100) of patients with chronic renal insufficiency demonstrated the assay's ability to recognize vitamin D deficiency and insufficiency, with results consistent with biochemical markers of disordered calcium mineral ion metabolism. Performance metrics included Passing & Bablok slope of 1.1, Pearson correlation r=0.84, and within-run precision (CV) of 3.0-4.5%.
Technological Characteristics
Automated immunoassay system. Materials: 25-OH-D3 magnetic particles, human vitamin D binding protein, acridinium-ester labeled anti-DBP. Sensing principle: Competitive ligand binding. Sample volume: 20 uL serum/plasma. Incubation: 21 min @ 37°C (displacement) + 21 min @ 37°C (assay). Reportable range: 7-120 ng/mL. Connectivity: Integrated with Nichols Advantage Specialty System.
Indications for Use
Indicated for quantitative determination of 25-hydroxyvitamin D and metabolites in serum or plasma to aid in assessing vitamin D sufficiency in adult populations.
Regulatory Classification
Identification
A calibrator is a device intended for medical purposes for use in a test system to establish points of reference that are used in the determination of values in the measurement of substances in human specimens. (See also § 862.2 in this part.)
Special Controls
*Classification.* Class II (special controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9.
Predicate Devices
DiaSorin 25-Hydroxyvitamin D 125I RIA kit
Submission Summary (Full Text)
{0}------------------------------------------------
#### 510(k) Summary 12.0
This summary of safety and effectiveness is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.
# 1. Name of Manufacturer, Contact Person and Date Summary Prepared:
Nichols Institute Diagnostics 33051 Calle Aviador San Juan Capistrano, CA 92675 Phone: 949-240-5260 FAX: 949-240-5313 Contact Person: Jimmy Wong, Manager of Clinical and Technical Affairs Date Prepared: September 20, 2001
#### 2. Device Name:
| Trade/Proprietary Name: | Nichols Advantage® 25-Hydroxyvitamin D |
|-------------------------|----------------------------------------|
| Common Name: | 25-OH-D Immunoassay |
| Classification Name: | Vitamin D test system |
- Classification: Class II ട്.
Regulation Number: 862.1825 Product Code: MRG. Clinical Chemistry
- DiaSorin 25-Hydroxyvitamin D 1251 RIA kit 4. Predicate Device:
#### Device Description: 5.
The Nichols Advantage 25-OH-D (NA 25-OH-D) contains sufficient reagents for 100 tests. The NA 25-OH-D is an automated 25-OH-D assay for use only on the Nichols Advantage Immunoassay System.
#### Intended Use: 6.
The Nichols Advantage® 25-Hydroxyvitamin D assay is intended for the quantitative determination of 25-hydrxoyvitamin D (25-OH-D) and other hydroxylated metabolites of vitamin D in serum or plasma to be used as an aid in the assessment of vitamin D sufficiency. Assay results should be used in conjunction with other clinical data to assist the clinician in making individual patient management decisions.
### 7. Comparison to Predicate Device:
The Nichols Advantage 25-OH-D (Y) is substantially equivalent to the DiaSorin 25-Hydroxvyitamin D 125] RIA kit (X). N=99 samples were used in a split sample method comparison study following the NCCLS EP9-A quidelines. For n=80 samples within the reportable ranges of each method, the Passing & Bablok method comparison demonstrates: Y=1.1(X) - 0.6 (95% cf slope: 0.94 to 1.27; 95% cf intercept: -4.6 to 2.1), and a Pearson correlation r=0.84 (95%cf: 0.76-0.89). For the entire n=99 samples, relative agreement using a cut-off of 20 ng/mL or greater was 91%, and the relative agreement for values less than 20 ng/mL was 98%, with an overall agreement of 94%.
### 8. Clinical Study
Patients (n=100) with a diagnosis of chronic renal insufficiency were evaluated with the NA 25-OH-D. Using a cutoff of 20 ng/mL, the results demonstrate the assay recognizes vitamin D deficiency and vitamin D insufficiency, consistent with biochemical measures of disordered calcium mineral ion metabolism. As renal failure becomes progressive worse, the frequency of low 25-0H-D levels increases as do biochemical measures of worsening indices of calcium mineral ion metabolism.
{1}------------------------------------------------
### 9. Similarities:
- Intended use for each assay is identical. .
- Both assays use 25-OH-D3 in their standards, and both report values using the same units: ● ng/mL.
- Both assay use the competitive ligand binding technique. .
- Both assays are capable of using serum, EDTA and heparinized plasma as samples. .
## 10. Differences:
The following differences pertain to differences in immunoassay technology and do not affect the intended uses of the NA 25-OH-D assay.
| Feature | Nichols 25-OH-D | DiaSorin 25-OH-D |
|---------------------------------------|---------------------------------------------------------------------------------------|--------------------------------------------------------------------------------|
| Sample Size: | 20 uL serum or plasma | 50 ul serum or plasma |
| Solid Phase/Precipitating<br>Complex: | 25-OH-D3 magnetic particles | Donkey anti-goat + normal goat<br>serum + PEG precipitating<br>complex. |
| Label: | Acridinium-ester labeled Anti-DBP | $^{125}I$ analog o f 25-OH-D3 |
| Binder/Antibody: | Human vitamin D binding protein | Goat anti-25-OH-D |
| Technology | Automated, hands free set-up<br>No extraction is required. | Manual procedure<br>Separate extraction is required. |
| Incubation steps and temperature: | 21 minutes @ 37°C (25-OH-D<br>displacement)+ 21 minutes @<br>37°C (assay incubation). | 90 minutes @ 20-25°C + 20<br>minutes @ 20-25°C + 20 minutes<br>centrifugation. |
| Reportable Range | 7-120 ng/mL | 5.0-100 ng/mL |
### 11. Comparison of Performance Characteristics
| Feature | Nichols 25-OH-D | DiaSorin 25-OH-D |
|----------------------------|------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------|
| Sensitivity | Analytical: 4 ng/mL<br>Functional: 7 ng/mL | Analytical: ≤ 1.5 ng/mL<br>Functional: 2.5 ng/mL (optional std.) |
| Within-Run Precision (%CV) | 3.0-4.5% | 8.6-12.5% |
| Total Precision (%CV) | 6.4-14.5% | 8.2-11.0% |
| Recovery | 90-99% | 92-119% |
| Linearity | 80-109% | 96-110% |
| Specificity @ 50% B/Bo | Vitamin D2 = 1.6%<br>Vitamin D3 = 1.2%<br>25-OH-D2 = 100%<br>25-OH-D3 = 100%<br>24,25(OH)2D3 = 100%<br>25-26-OH2D3 = 39%<br>1,25(OH)2D3 = 1.1% | Vitamin D2 = 0.8%<br>Vitamin D3 = 0.8%<br>25-OH-D2 = 100%<br>25-OH-D3 = 100%<br>24,25(OH)2D3 = 100%<br>25-26-OH2D3 = 100%<br>1,25(OH)2D3 = 11.0% |
Conclusions: These data, which were provided to FDA, demonstrate safety and effectiveness of the Nichols Advantage 25-Hydroxyvitamin D for the intended in vitro diagnostic use. Furthermore, based on performance characteristics, the Nichols Advantage 25-OH-D assay is substantially equivalent to the predicate method.
{2}------------------------------------------------
### 10.5 Comparison to a Predicate Device
Method comparison was made against the DiaSorin 25-Hydroxyvitamin D 125RIA kit. Both the predicate device and the subject device were performed exactly as described by their respective manufacturer's directional inserts. The NCCLS Method Comparison (EP9A) procedures were followed. There were n=111 total samples assayed, each in duplicate. The samples were assayed over 3 days in 3 different runs, and guality control samples for each run were within their respective ranges. Of the 111 paired samples, 80 were within the reportable ranges of each method and demonstrated acceptable levels of imprecision. We used 4 standard deviations to delete outliers for reasons of imprecision. Twelve samples demonstrated unacceptable levels of imprecision (>4 sd) and were deleted. Nineteen samples were below the reportable range (<7 ng/mL) with the subject device, and of these, 4 were below the reportable range with the DiaSorin assay (<5 ng/mL). Tabulated data showing the above three groups are located in Appendix 6.
Method comparison statistical analysis was performed with the 80 samples that demonstrated acceptable levels of imprecision. The 80 samples with good precision and the 19 samples that were below the reportable ranges were combined to determine overall concordance using McNemars test. Below is a chart summarizing the levels of imprecision observed in the 80 samples with acceptable levels of imprecision. These data demonstrate that the Nichols Advantage 25-OH-D assay has better within-run precision.
| N=80 paired samples | Nichols Advantage<br>25-OH-D | DiaSorin 25-OH-D |
|---------------------|------------------------------|------------------|
| Ave. CV | 4.7% | 10.6% |
| Median CV | 3.3% | 9.1% |
| Minimum CV | 0.1% | 0% |
| Maximum CV | 25.4% | 35.0% |
NCCLS method comparison analysis was performed on the 80 paired samples. The range of results, the Deming method comparison analysis, the Passing and Bablok method comparison, the Pearson correlation, and the NCCLS bias
41
{3}------------------------------------------------
analysis are summarized in the chart below. These data, and a scatterplot of the entire method comparison are located in Appendix 6.
| | Nichols Advantage | DiaSorin 25-OH-D (X) |
|------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------|
| | 25-OH-D (Y) | |
| Range | 7.0-69.5 ng/mL | 5.9-88.8 ng/mL |
| Deming Method<br>Comparison: | Y=0.99(X) + 0.7<br>95% confidence slope: 0.85 to 1.14<br>95% confidence intercept: -3.8 to 5.2 | |
| Passing and Bablok Method<br>Comparison: | Y=1.1(X) – 0.6<br>95% confidence slope: 0.94 to 1.27<br>95% confidence intercept: -4.6 to 2.1 | |
| Pearson correlation | R=0.84<br>95% confidence: 0.76 to 0.89 | |
| Bias | Ave. bias: 0.6 ng/mL<br>95% confidence of ave. bias: -1.4 to 2.5 ng/mL<br>95% confidence of agreement:<br>lower limit: -16.5 ng/mL (-19.7 to -13.2)<br>upper limit: 17.6 ng/mL (14.4 to 20.9) | |
Either the Passing and Bablok analysis or the Deming analysis could have been used for describing the method comparison. Because the Passing and Bablok analysis is not dependent upon a constant level of imprecision for the two methods, we think this analysis accurately reflects the comparison between the two methods. The average bias was small, only +0.6 ng/mL, but the upper and lower limits of the bias was large (-16.5 ng/mL to 17.6 ng/mL). The large bias we observed could be due to several reasons. The bias could have been influenced by the larger imprecision we observed with the DiaSorin, and because the crossreactivity of each assay is slightly different for vitamin D sterols. The issue of crossreactivity differences and differences in technology were previously discussed in the Crossreactivity section of this 510k.
Pearson correlation was not ideal because the range of observed results could have been higher. Several high samples were deleted due to poor imprecision obtained using the DiaSorin assay. Here are two examples:
- 1. Nichols: 101.1, 98.1 vs. DiaSorin: 44.3, 70.8
- 2. Nichols: 103.8, 100.9 vs. DiaSorin: 89.9, 47.1
We think the Pearson's correlation coefficient would have been higher if those samples were included in the analysis, since the range of results would be
{4}------------------------------------------------
higher. Obtaining samples with high 25-OH-D levels is difficult because levels approaching 100 ng/mL or higher only come from individuals who are ingesting large quantities of vitamin D or are taking pharmacologic doses of vitamin D. Our method comparison study shows good agreement between methods across the reportable ranges, but potentially with large bias for values above approximately 20 ng/mL. The concordance chart below further illustrates this bias.
Concordance chart showing number of sample results falling within the specified ranges. The concordance chart include the 19 samples below the reportable range, and the 80 samples used in the method comparisons.
| Adv.<br>25-OH-D<br>(ng/mL) | 0-10 | 10.1-20 | 20.1-30 | 30.1-40 | 40.1-50 | 50.1-60 | 60.1-80 | ≥80.1 |
|----------------------------|------|---------|---------|---------|---------|---------|---------|-------|
| 0-10 | 18 | 7 | | | | | | |
| 10.1-20 | 4 | 16 | 4 | 1 | | | | |
| 20.1-30 | | | 13 | 1 | | | | |
| 30.1-40 | | | 11 | 5 | 5 | 1 | 1 | |
| 40.1-50 | | 1 | 2 | 3 | 1 | | | |
| 50.1-60 | | | | | | 1 | | 1 |
| 60.1-80 | | | | | | | 1 | 1 |
| >80.1 | | | | | | | | |
DiaSorin 25-OH-D (no/ml )
Using a cutoff of 20 ng/mL or higher, the relative agreement between methods were calculated.
Relative agreement at 20 ng/mL or less: [45/(45+1)] x 100 = 97.8% Relative agreement at 20 ng/mL and higher: [48/(5+48)] x 100 = 90.6% Overall agreement using 20 ng/mL as the cutoff: [45+48/99] x 100 = 93.9%
Using a cutoff of 20 ng/mL as the lower limit for vitamin D sufficiency, the subject device demonstrates a relative agreement of 90.6% with the DiaSorin assay at 20 ng/mL and higher. This demonstrates that the subject device is substantially equivalent to the DiaSorin kit in detecting absence of low 25-OH-D levels for determining vitamin D sufficiency. For values below 20 ng/mL, the subject device demonstrates a relative agreement of 97.8% with the DiaSorin kit in identifying low 25-OH-D levels. The overall agreement was 93.9%.
In conclusion, these results in addition to the previous clinical study that evaluates levels of 25-OH-D in patients who are at great risk for vitamin D insufficiency and vitamin D deficiency demonstrate that this device is safe and effective for its intended use.
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle or bird symbol, composed of three curved lines, representing the department's mission. The bird is encircled by the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA". The text is arranged in a circular fashion around the bird symbol.
Food and Drug Administration 2098 Gaither Road Rockville MD 20850
OCT - 92001
Mr. Jimmy Wong Manager. Clinical and Technical Affairs Nichols institute Diagnostics 33051 Calle Aviador San Juan Capistrano, CA 92675
Re: k012367
> Trade/Device Name: Nichols Advantage 25-Hydroxyvitamin D Regulation Number: 21 CFR 862.1825 Regulation Name: Vitamin D test system Regulatory Class: Class II Product Code: JIT, MRG Regulatory Class: Class I, reserved Product Code: JJX Dated: September 26, 2001 Received: September 27, 2001
Dear Mr. Wong:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{6}------------------------------------------------
Page 2 -
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed nouncation. The I Drice results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFF Part 801 and 1 IT you desire specific acrice for your dotic devices), please contact the Office of Compliance at additionally 607.10 for in Nue and successions on the promotion and advertising of your device, (301) 594-4568. Additionally, for quest (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Information on your responsionals and Consumer Assistance at its toll-free number (800) 638-2041 or Manufacturers International and Constant Constitution w.fda.gov/cdrh/dsma/dsmamain.html".
Sincerely yours,
Steven Sutman
Steven I. Gutman, M.D., M.B.A. Director Division of Clinical Laboratory-Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{7}------------------------------------------------
#### 2.0 Indications For Use Statement
| INDICATIONS FOR USE STATEMENT |
|-------------------------------|
|-------------------------------|
| 510(k) Number: | K012367 |
|----------------|---------|
|----------------|---------|
Nichols Advantage 25-Hydroxyvitamin D Device Name:
Indications for Use Statement: The Nichols Advantage® 25-Hydroxyvitamin D assay is for use only on the Nichols Advantage® Specialty System. The Nichols Advantage 25-Hydroxyvitamin D assay is intended for the quantitative determination of 25-hydroxyvitamin D (25-OH-D) and other hydroxylated metabolites of vitamin D in serum or plasma to be used as an aid in the assessment of vitamin D sufficiency in an adult population. Assay results should be used in conjunction with other clinical data to assist the clinician in making individual patient management decisions.
(Please Do Not Write Below This Line - Continue On Another Page If Needed)
# Concurrence of CDRH, Office of Device Evaluation (ODE)
**Prescription Use**
(Per 21 CFR 801.109)
Or 510(k) Number
Over -The-Counter Use (Optional Format 1-2-96)
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.