← Product Code [JIS](/submissions/CH/subpart-b%E2%80%94clinical-chemistry-test-systems/JIS) · K032608

# GONOTEC CALIBRATION SOLUTION 300 MOSMOL/KG FOR OSMOMETER (K032608)

_Gonotec Gesellschaft Fuer Mess-Und Regeltechnik MB · JIS · Oct 21, 2003 · Clinical Chemistry · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/CH/subpart-b%E2%80%94clinical-chemistry-test-systems/JIS/K032608

## Device Facts

- **Applicant:** Gonotec Gesellschaft Fuer Mess-Und Regeltechnik MB
- **Product Code:** [JIS](/submissions/CH/subpart-b%E2%80%94clinical-chemistry-test-systems/JIS.md)
- **Decision Date:** Oct 21, 2003
- **Decision:** SESE
- **Submission Type:** Abbreviated
- **Regulation:** 21 CFR 862.1150
- **Device Class:** Class 2
- **Review Panel:** Clinical Chemistry

## Indications for Use

The primary calibrator “Calibration Solution for Osmomat 030/Auto” is used to calibrate the Osmometers “Osmomat 030” and “Osmomat Auto” which both are freezing point osmometers for use in clinical chemistry. The Osmolality of the calibrator is equal to the Osmolality of body fluids (isotonic solution), providing that the osmometer will be calibrated correctly for the fluids to be measured for medical purposes.

## Device Story

Calibration solution for freezing-point osmometers (Osmomat 010/030/auto); aqueous solution containing sodium chloride. Used in clinical chemistry laboratories to calibrate osmometers; ensures accurate measurement of osmolality in patient body fluids. Osmometers function by measuring freezing-point depression; dissolved substances in water depress freezing point; device calculates osmolality based on this physical effect. Calibrator provides known reference value for standard curve. Benefits patient by ensuring accurate diagnostic testing of body fluids (e.g., urine) to assess kidney function and physiological status.

## Clinical Evidence

No clinical data. Bench testing only; quality of ingredients (sodium chloride and distilled water) controlled via U.S. Pharmacopeia and European Pharmacopeia standards. Traceability established via certificate of analysis confirming quantity by weight.

## Technological Characteristics

Aqueous solution of sodium chloride; primary reference material. Designed for use with freezing-point osmometers. No active energy source or software; chemical calibrator only.

## Regulatory Identification

A calibrator is a device intended for medical purposes for use in a test system to establish points of reference that are used in the determination of values in the measurement of substances in human specimens. (See also § 862.2 in this part.)

## Special Controls

*Classification.* Class II (special controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9.

## Predicate Devices

- Sodium/Potassium standard modified ([K931834](/device/K931834.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}

510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION
DECISION SUMMARY
DEVICE TEMPLATE

510(k) Number: K032608

Analyte: Primary Calibrator for Osmolality

A. Type of Test: N/A
B. Applicant: Gonotec Gesellschaft Fuer Mess- UND Regeltechnik MBH
C. Proprietary and Established Names: Calibration Solution for Osmomat 010/030/Auto

Regulatory Information:
1. Regulation section: 21 CFR 862.1150
2. Classification: Class II
3. Product Code: JIS
4. Panel: 75

D. Intended Use:
1. Indication(s) for use: The primary calibrator “Calibration Solution for Osmomat 030/Auto” is used to calibrate the Osmometers “Osmomat 030” and “Osmomat Auto” which both are freezing point osmometers for use in clinical chemistry. The Osmolality of the calibrator is equal to the Osmolality of body fluids (isotonic solution), providing that the osmometer will be calibrated correctly for the fluids to be measured for medical purposes.
2. Special condition for use statement(s): none
3. Special instrument Requirements: none

Device Description: IVD device calibration is most commonly performed using calibrators (reference materials) specifically intended to be used as a standard curve or cut-off point for an assay.

A calibrator has an assigned value that is established by the manufacturer by a reference method. Calibrators exist in a variety of matrices such as simulated aqueous, serum, plasma or other types of specimens.

Primary reference calibrators are highly purified chemicals that can be directly weighed or measured to produce a solution of known concentration. Alternatively

{1}

Page 2 of 3

they may be more complex biological materials having received a value assignment using reference methodology. They are supplied with a certificate of analysis for each lot (For Example, Standard reference materials (SRMS) from the US National Institute of Standards. (NIST)).

The primary calibrator “Calibration Solution for Osmomat 030/Auto” is used to calibrate the Osmometers “Osmomat 030” and “Osmomat Auto” which both are freezing point osmometers for use in clinical chemistry. The Osmolality of the calibrator is equal to the Osmolality of body fluids (isotonic solution), providing that the osmometer will be calibrated correctly for the fluids to be measured for medical purposes.

## Substantial Equivalence Information:

4. Predicate device name(s): King Sodium/Potassium Standard modified.
5. Predicate K number(s): K931834

3. Comparison with predicate: The Predicate device is a sodium/potassium standard to calibrate ion-selective electrodes. The calibrator consists of an aqueous solution of salts in water. The difference in intended use is not relevant for the safety and effectiveness of the calibrator as the measuring principles in both calibrators are physical not chemical.

The respective calibrators do not react as chemicals in the measurement itself, but their physical properties are measured without changing their chemical composition.

In the case of the Osmomat 010/030/auto, the liquid calibrator is frozen under controlled conditions. The exact temperature needed to induce freezing of the sample is measured. This temperature is a measure for the exact number of molecules of any diluted substances in the aqueous sample.

## E. Standard/Guidance Document Referenced (if applicable)

The quality of sodium chloride and sevenfold distilled water as sole ingredients of the calibrator itself are checked and controlled under standards accepted by the U.S. Pharmacopeia. Both sodium chloride and water are regulated by European Pharmacopeia. The sodium chloride used is under control of the U.S. Pharmacopeia as well.

## F. Test Principle: N/A

Performance Characteristics (if/when applicable):

1. Analytical performance:
a. Precision/Reproducibility: N/A
b. Linearity/assay reportable range: N/A

{2}

Page 3 of 3

c. Traceability (controls, calibrators, or method): A certificate of analysis is commercially available confirming the quantity by weight of sodium and chloride in each calibrator lot.

d. Detection limit (functional sensitivity: N/A
e. Analytical specificity: N/A
f. Assay cut-off: N/A

2. Comparison studies:
a. Method comparison with predicate device: N/A
b. Matrix comparison: N/A

3. Clinical studies:
a. Clinical sensitivity: N/A
b. Clinical specificity: N/A

4. Clinical cut-off: N/A
5. Expected values/Reference range: N/A

G. Conclusion: Based upon the information provided, I recommend that the Gonotec Gesellschaft – Calibrator for the Osmomat 010/030 and Osmomat Auto be found substantially equivalent with the predicate devices as defined in 21 CFR 862.1150.

---

**Source:** [https://fda.innolitics.com/submissions/CH/subpart-b%E2%80%94clinical-chemistry-test-systems/JIS/K032608](https://fda.innolitics.com/submissions/CH/subpart-b%E2%80%94clinical-chemistry-test-systems/JIS/K032608)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
