← Product Code [JIF](/submissions/CH/subpart-b%E2%80%94clinical-chemistry-test-systems/JIF) · K033921

# AMMONIA-L3K ASSAY, CATALOGUE NUMBER 233-10 (K033921)

_Diagnostic Chemicals , Ltd. · JIF · Mar 10, 2004 · Clinical Chemistry · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/CH/subpart-b%E2%80%94clinical-chemistry-test-systems/JIF/K033921

## Device Facts

- **Applicant:** Diagnostic Chemicals , Ltd.
- **Product Code:** [JIF](/submissions/CH/subpart-b%E2%80%94clinical-chemistry-test-systems/JIF.md)
- **Decision Date:** Mar 10, 2004
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 862.1065
- **Device Class:** Class 1
- **Review Panel:** Clinical Chemistry

## Indications for Use

The Ammonia-L3K® Assay is a testing device for the quantitative determination of Ammonia in plasma. Ammonia measurements are used in the diagnosis and treatment of sever liver disorders, such as cirrhosis, hepatitis, and Reye’s Syndrome. For In Vitro Diagnostic use only. Prescription Use

## Device Story

Ammonia-L3K Assay is an in vitro diagnostic reagent for quantitative ammonia measurement in plasma. Input: plasma sample; processed via Hitachi 911 analyzer. Principle: enzymatic colorimetric reaction; ammonia reacts with alpha-ketoglutarate and reduced cofactor (NADPH analog) catalyzed by glutamate dehydrogenase (GLDH); decrease in absorbance at 340 nm monitored; absorbance change proportional to ammonia concentration. Used in clinical laboratory settings by trained technicians. Output: quantitative ammonia concentration value. Assists clinicians in diagnosing and monitoring liver disorders.

## Clinical Evidence

No clinical data provided; bench testing only.

## Technological Characteristics

Reagent contains buffer (pH 8.0), 10 mmol/L alpha-ketoglutarate, >=6 KU/L microbial GLDH, 0.2 mmol/L NADPH analog, stabilizers, and preservative. Enzymatic colorimetric sensing principle. Designed for use on Hitachi 911 analyzer. No software/algorithm complexity beyond standard photometric absorbance monitoring.

## Regulatory Identification

An ammonia test system is a device intended to measure ammonia levels in blood, serum, and plasma, Ammonia measurements are used in the diagnosis and treatment of severe liver disorders, such as cirrhosis, hepatitis, and Reye's syndrome.

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}

510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION
DECISION SUMMARY
DEVICE ONLY TEMPLATE

A. 510(k) Number:
k033921

B. Analyte:
Ammonia

C. Type of Test:
Quantitative

D. Applicant:
Diagnostic Chemicals Limited

E. Proprietary and Established Names:
Ammonia-L3K® Assay

F. Regulatory Information:
1. Regulation section:
21 CFR § 862.1065 Ammonia Test System, Enzymatic method
2. Classification:
Class I
3. Product Code:
JIF,
4. Panel:
Clinical Chemistry (75)

G. Intended Use:
1. Indication(s) for use:
The Ammonia-L3K® Assay is a testing device for the quantitative determination of Ammonia in plasma. Ammonia measurements are used in the diagnosis and treatment of sever liver disorders, such as cirrhosis, hepatitis, and Reye’s Syndrome.
2. Special condition for use statement(s):
For In Vitro Diagnostic use only.
Prescription Use
3. Special instrument Requirements:
Hitachi® 911 analyzer

{1}

Page 2 of 4

H. Device Description:
Ammonia-L3K Reagent is a solution containing a buffer (pH 8.0 at 25°), 10 mmol/α-ketoglutarate, at least 6 KU/L GLDH (microbial), 0.2 mmol/L NADPH analog, stabilizers, and a preservative.

I. Substantial Equivalence Information:
1. Predicate device name(s):
Ammonia-Incorporating Dynamic Stabilization Technology (DST)
2. Predicate K number(s):
K974620
3. Comparison with predicate:
Both devices are for the quantitative determination of the same analyte in the same matrixes. Both devices employ enzymatic colorimetric reaction.

J. Standard/Guidance Document Referenced (if applicable):
NCCLS EP5-A - Evaluation of Precision Performance of Clinical Chemistry Devices
NCCLS EP6-P - Evaluation of the Linearity of Quantitative Analytical Methods
NCCLS EP9-A - Method Comparison and Bias Estimation Using Patient Samples

K. Test Principle:
Ammonia reacts with α-ketoglutarate and reduced cofactor to form L-glutamate and the cofactor. The reaction is catalyzed by glutamate dehydrogenase. The decrease in absorbance due to the oxidation of the reduced cofactor can be monitored at 340 nm and is proportional to the ammonia concentration.

L. Performance Characteristics (if/when applicable):
1. Analytical performance:
a. Precision/Reproducibility:
Data was collected on three aqueous standards in 40 runs conducted over 20 days.

|  Level | Level | Total SD | Total SD | Total CV | Within Run SD | Within Run SD | Within Run CV  |
| --- | --- | --- | --- | --- | --- | --- | --- |
|  umol/L | ug/mL | umol/L | (ug/mL) |  | Umol/L | (ug/mL) |   |
|  48.01 | 0.819 | 3.76 | 0.064 | 7.9 | 2.64 | 0.045 | 5.6  |
|  125.23 | 2.133 | 4.40 | 0.075 | 3.5 | 2.64 | 0.045 | 2.1  |
|  337.29 | 5.745 | 5.81 | 0.099 | 1.7 | 2.58 | 0.044 | 0.8  |

b. Linearity/assay reportable range:
Performed according to NCCLS Guideline EP6-P, the linearity of the procedure described is 1175.0 u mol/L (20.0 ug/mL). The lower limit of detection of the procedure described is 5.8 umol/L (0.1 ug/mL). This data results in a reportable range of 5.8-1175.0 umol/L.

c. Traceability (controls, calibrators, or method):
NA

{2}

Page 3 of 4

d. Sensitivity:

The lower limit of detection of the procedure is 5.8 umol/L (0.1 ug/mL)

e. Analytical specificity:

Interference from icterus, lipemia, hemolysis, and plasma pyruvate were evaluated on Hitachi 911 analyzer using a significance criterion of &gt;10% variance from control. No significant icterus interference was found at bilirubin levels from 0-684 umol/L (0-40 mg/dL) in a 167 umol/L (2.8 ug/mL) ammonia sample. Intralipid levels from 0-1000 mg/dL [0-33.87 mmol/L (0-3000 mg/dL) triglycerides] were studies with acceptable results to level of 250 mg/dL Intralipid [8.46 mmol/L (750 mg/dL) triglycerides] in a 154 umol/L (2.6 ug/mL) ammonia sample. No significant ascorbic acid interference was found at ascorbic acid levels from 0-3000 ug/dL in a 147 umol/L (2.5 ug/mL) ammonia sample. Plasma pyruvate levels up to 0.75 mmol/L (6.6 mg/dL) do not cause interference.

f. Assay cut-off:

NA

2. Comparison studies:

a. Method comparison with predicate device:

The performance of this method (y) was compared with the performance of similar ammonia method (x) on a Hitachi 911. Forty patient sample ranging from 11.1 - 811.4 umol/L (0.19 - 13.82 ug/mL) gave a correlation coefficient of 0.9990. Linear regression analysis gave the following equation:

$$Y = 0.95 X - 0.8 \text{ umol}(-0.01 \text{ ug/mL})$$

b. Matrix comparison:

NA

3. Clinical studies:

a. Clinical sensitivity:

Not applicable.

b. Clinical specificity:

Not applicable.

c. Other clinical supportive data (when a and b are not applicable):

Not applicable.

4. Clinical cut-off:

NA

{3}

Page 4 of 4

5. Expected values/Reference range:
12 – 47 umol/L (0.20 – 0.80 ug/mL). These values are suggested guidelines. It is recommended that each laboratory establish the normal range for the area in which it is located.

M. Conclusion:
I recommend that the Ammonia-L3K® Assay is substantially equivalent to the legally marketed predicate device.

---

**Source:** [https://fda.innolitics.com/submissions/CH/subpart-b%E2%80%94clinical-chemistry-test-systems/JIF/K033921](https://fda.innolitics.com/submissions/CH/subpart-b%E2%80%94clinical-chemistry-test-systems/JIF/K033921)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
