← Product Code [JHX](/submissions/CH/subpart-b%E2%80%94clinical-chemistry-test-systems/JHX) · K962549

# VIDAS CREATINE KINASE MB ASSAY 30-421 (K962549)

_Biomerieux Vitek, Inc. · JHX · Aug 30, 1996 · Clinical Chemistry · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/CH/subpart-b%E2%80%94clinical-chemistry-test-systems/JHX/K962549

## Device Facts

- **Applicant:** Biomerieux Vitek, Inc.
- **Product Code:** [JHX](/submissions/CH/subpart-b%E2%80%94clinical-chemistry-test-systems/JHX.md)
- **Decision Date:** Aug 30, 1996
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 862.1215
- **Device Class:** Class 2
- **Review Panel:** Clinical Chemistry

## Indications for Use

The VIDAS CKMB assay determines the concentration of creatine kinase MB isoenzyme in serum or plasma (heparin or EDTA).

## Device Story

VIDAS CKMB assay utilizes ELFA (Enzyme Linked Fluorescent Assay) technology to measure creatine kinase MB isoenzyme concentrations in serum or plasma samples. Performed on VIDAS instrument; system uses specific antibodies to detect CKMB isoform while excluding BB and MM isoforms. Calibrator ensures validity of stored master curve. Provides quantitative results (0.8 ng/ml to 300 ng/ml) to clinicians for diagnostic assessment of myocardial infarction. Operates in clinical laboratory settings; requires professional operation. Results assist in clinical decision-making regarding cardiac status.

## Clinical Evidence

Bench testing only. Comparison study against predicate yielded y = 0.83x - 0.656 (r=0.97). Sensitivity established at 0.8 ng/ml. Specificity confirmed via cross-reactivity testing with BB and MM isoforms. Interference testing performed for hemoglobin, lipids, bilirubin, and various collection tube types. Precision/reproducibility: intra-assay CV 3.2-6.2%; inter-assay CV < 7.2%; inter-instrument CV < 16.7%.

## Technological Characteristics

ELFA (Enzyme Linked Fluorescent Assay) technology. Quantitative immunochemistry. Measures CKMB in serum/plasma. Reportable range 0.8-300 ng/ml. Instrument-based automated analysis using stored master curves.

## Regulatory Identification

A creatine phosphokinase/creatine kinase or isoenzymes test system is a device intended to measure the activity of the enzyme creatine phosphokinase or its isoenzymes (a group of enzymes with similar biological activity) in plasma and serum. Measurements of creatine phosphokinase and its isoenzymes are used in the diagnosis and treatment of myocardial infarction and muscle diseases such as progressive, Duchenne-type muscular dystrophy.

## Predicate Devices

- Ciba-Corning Magic Lite CK-MB assay

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}

AUG 30 1996
k962549

# SUMMARY: Safety and Effectiveness Information for VIDAS CKMB Assay

The VIDAS CKMB assay determines the concentration of creatine kinase MB isoenzyme in serum or plasma (heparin or EDTA). It is substantially equivalent to the Ciba-Corning Magic Lite CK-MB assay.

Safety and effectiveness issues relevant to ELFA methodology for quantitative immunochemistry assays such as those for creatine kinase MB may include poor correlations, poor sensitivity, cross-reactivity, interference, or imprecision.

## 1. Correlations:

a. Comparison of the VIDAS CKMB assay with the Ciba-Corning Magic Lite CK-MB assay yields a line with the equation $y = 0.83x + -0.656$ and a correlation coefficient of 0.97.

b. The calibrator in the kit ensures that the master curve stored by the VIDAS instrument is valid for the shelf life of that kit. The body of data supports the use of a single calibrator run in duplicate for this purpose.

## 2. Sensitivity:

The VIDAS CKMB assay is designed to measure creatine kinase MB levels between 0.8 ng/ml and 300 ng/ml. The lowest measurable level of creatine kinase MB (assay sensitivity) that can be distinguished from zero with 95 % probability using the VIDAS assay is 0.8 ng/ml.

## 3. Cross-reactivity:

Specificity studies demonstrate that the antibodies used in the VIDAS CKMB assay is specific for Creatine Kinase MB isoform. No cross-reactivity is seen with isoforms BB and MM.

## 4. Interfering Substances

a. No interference in VIDAS CKMB assay performance was seen with serum collected in dry glass tubes with glycerol cap, tubes containing a separating gel, siliconized tubes, and non-siliconized tubes, tubes containing EDTA and tubes containing heparin.

b. No interference in the VIDAS CKMB assay performance was seen with the range of concentrations of hemoglobin, lipids, or bilirubin tested.

## 5. Precision/Reproducibility

a. Intra-assay precision studies showed coefficients of variation ranging from 3.2 % to 6.2 % over the reportable range of the assay.

b. Inter-assay reproducibility over an eight-week time period yields coefficients of variation that do not exceed 7.2 %.

c. Inter-assay, inter-instrument reproducibility for five different serum samples in nine assays on different instruments yields coefficients of variation that do not exceed 16.7 %.

When the VIDAS CKMB assay is used as instructed in the package insert, the above statements are true. The package insert should always be consulted along with the Operator's Manual to ensure that the assay is being performed properly. For additional information, references are listed in the package insert.

---

**Source:** [https://fda.innolitics.com/submissions/CH/subpart-b%E2%80%94clinical-chemistry-test-systems/JHX/K962549](https://fda.innolitics.com/submissions/CH/subpart-b%E2%80%94clinical-chemistry-test-systems/JHX/K962549)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
