AQT90 FLEX CKMB TEST KIT, AQT90 FLEX LQC MULTI-CHECK, LEVELS 1-3, AQT90 FLEX CKMB CAL CARTRIDGE
K120326 · Radiometer Medical Aps · JHX · Nov 21, 2012 · Clinical Chemistry
Device Facts
Record ID
K120326
Device Name
AQT90 FLEX CKMB TEST KIT, AQT90 FLEX LQC MULTI-CHECK, LEVELS 1-3, AQT90 FLEX CKMB CAL CARTRIDGE
Applicant
Radiometer Medical Aps
Product Code
JHX · Clinical Chemistry
Decision Date
Nov 21, 2012
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 862.1215
Device Class
Class 2
Indications for Use
AQT90 FLEX CKMB Test is an in vitro diagnostic assay for the quantitative determination of creatine kinase isoform MB in EDTA or lithium-heparin whole blood or plasma specimens on the AQT90 FLEX analyzer in point of care and laboratory settings. It is intended for use as an aid in the diagnosis of myocardial infarction. AQT90 FLEX CKMB CAL cartridge is for in vitro diagnostic use for the calibration of the CKMB Test on the AQT90 FLEX analyzer by establishing points of reference to estimate CKMB values. AQT90 FLEX LQC Multi-CHECK, Levels 1-3, is for in vitro diagnostic use. For use with the AQT90 FLEX analyzer as a liquid quality control serum (LQC) to monitor the precision of laboratory testing procedures for the analytes listed on the specification insert.
Device Story
Cartridge-based immunoassay; uses time-resolved fluorescence with europium (Eu) chelate label. Input: EDTA or lithium-heparin whole blood or plasma. Process: sample and assay buffer dispensed into test cup; incubation (7-15 min at 37°C) allows immuno-complex formation; wash step removes unbound material; cup dried; excitation light applied; emitted light measured via single photon counting. Output: quantitative CKMB concentration compared to calibration curve. Used in point-of-care and laboratory settings; operated by healthcare professionals. Results aid clinicians in diagnosing myocardial infarction, enabling rapid cardiac assessment.
Clinical Evidence
Bench testing only. Precision evaluated per CLSI EP5-A2 (N=25 per site for whole blood; 58-96 per level for plasma). Linearity confirmed (r²=0.99). LoB 0.5 ng/mL, LoD 1.0 ng/mL, LoQ 1.0 ng/mL. Interference testing (52 drugs, 13 endogenous compounds) showed no significant interference. Method comparison against predicate (VIDAS) using Passing-Bablok regression showed high correlation (r² 0.95-0.99). Reference range established using 691 healthy individuals.
Technological Characteristics
Quantitative time-resolved fluorometric sandwich immunoassay. Uses europium (Eu) chelate label. Components: mouse monoclonal antibodies, bovine serum albumin, TRIS buffer. Form factor: test cartridges with test cups. Analyzer: AQT90 FLEX. Connectivity: barcode-based calibration. Standards: CLSI EP05-A2, EP6-A, EP07-A2, EP09-A2, EP17-A, C28-A3, ISO 14971, ISO 15223-1.
Indications for Use
Indicated for quantitative determination of creatine kinase isoform MB in EDTA or lithium-heparin whole blood or plasma specimens to aid in the diagnosis of myocardial infarction in point of care and laboratory settings.
Regulatory Classification
Identification
A creatine phosphokinase/creatine kinase or isoenzymes test system is a device intended to measure the activity of the enzyme creatine phosphokinase or its isoenzymes (a group of enzymes with similar biological activity) in plasma and serum. Measurements of creatine phosphokinase and its isoenzymes are used in the diagnosis and treatment of myocardial infarction and muscle diseases such as progressive, Duchenne-type muscular dystrophy.
Predicate Devices
Vidas CKMB assay (k962549)
Vidas CKMB calibrators (k962549)
Bio-Rad Laboratories Liquichek Cardiac Markers Plus (k050537)
Related Devices
K984067 — STRATUS (R) CS STAT FLUOROMETRIC ANALYZER, STRATUS (R) CS CKMB TEST PAK · Dade Behring, Inc. · Jan 13, 1999
K970343 — MMB METHOD · Dade Chemistry Systems, Inc. · Feb 14, 1997
K163462 — AQT90 FLEX Myo Test Kit, AQT90 FLEX CKMB Test Kit, AQT90 FLEX · Radiometer Medical Aps · Sep 21, 2017
Submission Summary (Full Text)
{0}
510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY ONLY TEMPLATE
A. 510(k) Number:
k120326
B. Purpose for Submission:
New device
C. Measurand:
Creatine kinase isoform MB (CKMB)
D. Type of Test:
Quantitative fluorometric immunoassay
E. Applicant:
Radiometer Medical ApS
F. Proprietary and Established Names:
AQT90 FLEX CKMB Test Kit
AQT90 FLEX CKMB CAL Cartridge
AQT90 FLEX LQC Multi-CHECK, Levels 1-3
G. Regulatory Information:
| Device Name | Product Code | Classification | Regulation | Panel |
| --- | --- | --- | --- | --- |
| AQT90 FLEX CKMB Test | JHX: Creatine phosphokinase/creatine kinase or isoenzymes test system | Class II | 21 CFR § 862.1215 | Clinical Chemistry (75) |
| AQT90 FLEX CKMB CAL cartridge | JIT: Calibrator | Class II | 21 CFR § 862.1150 | Clinical Chemistry (75) |
| AQT90 FLEX LQC Multi CHECK, Levels 1-3 | JJY: Quality control material | Class I, reserved | 21 CFR § 862.1660 | Clinical Chemistry (75) |
1
{1}
H. Intended Use:
1. Intended use(s):
See indications for use below.
2. Indication(s) for use:
AQT90 FLEX CKMB Test is an *in vitro* diagnostic assay for the quantitative determination of creatine kinase isoform MB in EDTA or lithium-heparin whole blood or plasma specimens on the AQT90 FLEX analyzer in point of care and laboratory settings. It is intended for use as an aid in the diagnosis of myocardial infarction.
AQT90 FLEX CKMB CAL cartridge is for *in vitro* diagnostic use for the calibration of the CKMB Test on the AQT90 FLEX analyzer by establishing points of reference to estimate CKMB values.
AQT90 FLEX LQC Multi-CHECK, Levels 1-3, is for *in vitro* diagnostic use. For use with the AQT90 FLEX analyzer as a liquid quality control serum (LQC) to monitor the precision of laboratory testing procedures for the analytes listed on the specification insert.
3. Special conditions for use statement(s):
For prescription use.
For point-of-care use.
4. Special instrument requirements:
AQT90FLEX analyzer.
I. Device Description:
CKMB assay and calibrator:
The AQT90 FLEX CKMB Test consists of ten Test cartridges and one Calibration cartridge. The Test cartridge contains test cups whereas the Calibration adjustment cartridge contains both test cups and calibration adjustments cups.
The components are listed below:
- Mouse monoclonal anti-CKMB capture and tracer antibodies, approximately 500 ng/cup.
- Bovine serum albumin. Bovine γ-globulin. Mouse IgG as blocker substance for heterophilic antibody interference.
- Tris-(hydroxymethyl)-aminomethane (TRIS) buffer
- Sodium azide < 0.1%.
- Calibrator Only: contains native purified CKMB (only in the eight CAL cups with added antigen), approximately 2 ng/cup.
{2}
Each human donor unit used to manufacture this calibrator was tested using FDA-accepted methods and found nonreactive for Hepatitis B Surface Antigen (HbsAg), antibody to Hepatitis C (HCV) and antibody to HIV-1/HIV-2.
Control Material: AQT90 FLEX LQC Multi-CHECK, Levels 1-3 is comprised of three levels of control material (7, 25 and 100 ng/mL) in human serum. Each human donor unit used to manufacture this control was tested using FDA-accepted methods and found nonreactive for Hepatitis B Surface Antigen (HbsAg), antibody to Hepatitis C (HCV) and antibody to HIV-1/HIV-2.
J. Substantial Equivalence Information:
| Predicate Device Name | Predicate 510(k) Number |
| --- | --- |
| Vidas CKMB assay | k962549 |
| Vidas CKMB calibrators | k962549 |
| Bio-Rad Laboratories Liquichek Cardiac Markers Plus | k050537 |
Comparison with predicate:
| CKMB Assay | | |
| --- | --- | --- |
| Item | Proposed Device AQT90 FLEX CKMB Test | Predicate Device VIDAS CKMB Immunoassay (k962549) |
| Intended Use | in vitro diagnostic assay for the quantitative determination of creatine kinase. It is intended for use as an aid in the diagnosis of myocardial infarction. | same |
| Matrix | Whole blood and plasma (EDTA, Li-heparin) | Serum and plasma (EDTA, heparin) |
| Principle | Quantitative time-resolved fluorimetric one-step sandwich immunoassay | Enzyme-linked fluorescent immunoassay. |
| Reportable Range | 1.5 to 300ng/mL | 0.8 to 300 ng/mL |
| Analyzer | AQT90 FLEX | Vidas Vitek |
{3}
| Calibrator | | |
| --- | --- | --- |
| Item | Proposed Device AQT90 FLEX CKMB CAL Cartridge | Predicate Device VIDAS CKMB Calibrators A-F (k962549) |
| Indications for Use | For calibration of the CKMB Test. | Same |
| Constituents | Sixteen calibration adjustment cups, which contain dried reagents including native purified CKMB. | Purified cardiac CKMB in Tris buffer with stabilizers |
| Calibration adjustment interval | Once per lot of AQT90 FLEX CKMB Test cartridges and as often as required. | Upon new assay reagent lot number |
| In-use stability | 24 hours on-board | 30 days at 2-8 °C |
| Storage temperature | 2-8 °C | -10 °C |
| Control | | |
| --- | --- | --- |
| Item | Proposed Device AQT90 FLEX LQC Multi-CHECK, Levels 1-3 | Predicate Device Liquichek Cardiac Markers Plus Control LT (k050537) |
| Indications for Use | For use as a liquid quality control serum (LQC) to monitor the precision of laboratory testing procedures for the analytes listed on the specification insert. | Same |
| Form | Liquid | Same |
| Matrix | Human serum based | Same |
| Levels | Three | Same |
{4}
K. Standard/Guidance Document Referenced (if applicable):
CLSI EP05-A2: Evaluation of Precision Performance of Quantitative Measurement Methods
CLSI EP6-A: Evaluation of Linearity of Quantitative Measurement Procedures
CLSI EP07-A2: Interference Testing in Clinical Chemistry
CLSI EP09-A2: Method Comparison and Bias Estimation Using Patient Samples
CLSI EP17-A: Protocols for Determination of Limits of Detection and Limits of Quantitation
CLSI C28-A3: Defining, Establishing, and Verifying Reference Intervals in the Clinical Laboratory
CLSI I/LA30-A: Immunoassay Interference by Endogenous Antibodies
EN 13640: Stability Testing of In Vitro Diagnostic Reagents
ISO 14971: Medical Devices - Application of risk management to medical devices
ISO 15223-1: Medical Devices - Symbols to be used with medical device labels, labeling and information to be supplied
L. Test Principle:
The test format for the AQT90 FLEX is an immunoassay based on time-resolved fluorescence using an europium (Eu) chelate as the fluorescent label. The test receptacles (300 µL test cups) for each assay contain the biotinylated antibodies used for the capture of the analyte, a separating layer of glucosides, and the europium-chelate labeled antibodies used to trace the captured proteins. The sample or diluted sample is added to the test cup together with assay buffer. The cup is then incubated to allow formation of the immuno-complex and washed to remove unbound antibodies and sample materials. Finally, the cup is exposed to excitation light and after a delay, the emitted light is measured by single photon counting; this measurement cycle is repeated up to 3,300 times. The total count (the response) is then compared to a calibration curve to obtain a quantitative measurement of the analyte's concentration in the sample.
M. Performance Characteristics (if/when applicable):
1. Analytical performance:
a. Precision/Reproducibility:
The precision of the AQT90 Flex CKMB Test on the AQT90 FLEX analyzer was evaluated using whole blood and plasma samples based upon the CLSI EP5-A2 guideline.
For plasma samples evaluations, five levels of frozen native or antigen spiked lithium-heparin plasma samples were tested in
{5}
duplicate in two separate runs for 20-25 days at three POC sites (one instrument, three test kit lots per site). The number of samples tested for each level ranged from 58 to 96. For whole blood samples evaluations, fresh native or antigen spiked lithium-heparin whole blood samples were tested in five replicates in five separate runs within three hours at four POC sites (one instrument, one test kit lot per site), total $N = 25$ .
The results are summarized in the following tables.
Li Heparin Plasma Precision Study Results
| Site | Mean ng/mL | Within run SD | Within run %CV | Total SD | Total %CV |
| --- | --- | --- | --- | --- | --- |
| 1 | L1=2.3 | 0.10 | 4.4 | 0.16 | 7.0 |
| | L2*=4.2 | 0.15 | 3.7 | 0.24 | 5.7 |
| | L3=12 | 0.4 | 3.0 | 0.7 | 5.5 |
| | L4=44 | 1.2 | 2.7 | 2.1 | 4.8 |
| | L5=288 | 7.9 | 2.7 | 10.7 | 3.7 |
| 2 | L1=2.3 | 0.12 | 5.2 | 0.18 | 7.7 |
| | L2*=4.1 | 0.20 | 5.0 | 0.27 | 6.6 |
| | L3=12 | 0.4 | 2.9 | 0.5 | 4.3 |
| | L4=44 | 1.1 | 2.4 | 1.8 | 4.1 |
| | L5=287 | 8.3 | 2.9 | 9.8 | 3.4 |
| 3 | L1=2.4 | 0.21 | 8.7 | 0.22 | 9.1 |
| | L2*=4.4 | 0.15 | 3.5 | 0.19 | 4.3 |
| | L3=12 | 0.3 | 2.5 | 0.4 | 3.3 |
| | L4=44 | 1.0 | 2.2 | 1.3 | 3.0 |
| | L5=290 | 4.5 | 1.6 | 6.9 | 2.4 |
\*native sample
{6}
Whole Blood Precision Study Results
| Site | Mean (ng/mL) | N | Within run SD | Within run %CV | Total SD | Total %CV |
| --- | --- | --- | --- | --- | --- | --- |
| 1 | L1= 1.9 | 25 | 10.11 | 6.0 | 0.11 | 6.0 |
| | L2*= 4.8 | 25 | 0.26 | 5.4 | 0.28 | 5.9 |
| | L3=18 | 25 | 0.4 | 2.1 | 0.4 | 2.4 |
| | L4=57 | 25 | 1.0 | 1.7 | 1.0 | 1.7 |
| 2 | L1=2.9 | 25 | 0.15 | 7.6 | 0.18 | 9.3 |
| | L2*=4.2 | 25 | 0.18 | 4.3 | 0.20 | 4.7 |
| | L3=16 | 25 | 0.6 | 3.7 | 0.6 | 3.7 |
| | L4=51 | 25 | 1.4 | 2.7 | 1.7 | 3.3 |
| 3 | L1=2.4 | 25 | 0.13 | 5.4 | 0.14 | 5.9 |
| | L2*=3.9 | 25 | 0.15 | 3.8 | 0.15 | 3.8 |
| | L3=51 | 25 | 0.5 | 3.1 | 0.5 | 3.1 |
| 4 | L4=217 | 25 | 2.2 | 4.4 | 2.2 | 4.4 |
| 5 | L5=237 | 25 | 8.4 | 3.5 | 8.4 | 3.5 |
*native sample
b. Linearity/assay reportable range:
A linearity study was performed using EDTA natural patient plasma spiked with CKMB antigen. The high concentration and low concentration samples were mixed together to create 10 intermediate dilution samples. All samples were tested in replicates of ten with samples ranging from 1.2 to 374-ng/mL. The linear regression correlation between the expected values and the measured values is:
$$
y = 1.01x + 0.68, r^2 = 0.9
$$
An additional linearity study was conducted using EDTA natural patient whole blood spiked with CKMB antigen. The high concentration and low concentration samples were mixed together to create 10 intermediate dilution samples. All samples were tested in replicates of ten with samples ranging from 1.3 ng/mL to 404 ng/mL. The linear regression correlation between the expected values and the measured values is: $y = 0.99x + 1.08, r^2 = 0.99$
The linearity data provided support the sponsor's claims that the reportable range of this assay is 1.5 to 300 ng/mL for both whole blood and plasma samples.
{7}
c. Traceability, Stability, Expected values (controls, calibrators, or methods):
## Traceability and Value Assignment
The antigen material, ERM-AD455/IFCC material, has been used to establish the metrological traceability chain of the AQT90 FLEX CKMB assay.
The calibrator value is internally assigned by the manufacturer and the factory-defined calibration data is provided in the form of a barcode. This data is used as a lot-specific reference curve, which is adjusted to the correct instrument specific signal level by using a CAL Cartridge with the same lot number. The protocols for value assignment and acceptance criteria were reviewed and are acceptable.
The control value assignment for the three CKMB levels of the AQT90 FLEX LQC Multi-CHECK control material is performed by Radiometer. The value assignment is traceable to master standards. The value assignment for each Multi-CHECK level includes multiple reagent lots, instruments, test days, and measurement replicates per control level, for a total of 60 measurements per control level. The assigned analyte value and range are lot-specific and printed on the specification label. The specification label has a lot-specific barcode, containing the same information that is printed (analyte level, assigned value, range levels, lot number and expiry date). This information is assessable through the AQT90 FLEX Analyzer. The protocols for value assignment and acceptance criteria were reviewed and are acceptable.
## Stability
AQT90 FLEX CKMB Test and Calibration Cartridges: The stability studies were performed on three product lots of test and calibration cartridges. Based on the results, the sponsor claims that test and calibration cartridges have a shelf stability of 8 months at 2 to 8 °C, while the on-board stability is 23 days at 32 °C for the test cartridge and 24 hours at 32 °C for the calibration cartridge. The protocols for stability and acceptance criteria were reviewed and are acceptable.
AQT90 FLEX LQC Multi-Check: Stability studies were performed on three lots. Based on the results, the sponsor claims a shelf stability of 12 months at -18 to -26 °C, while the on-board stability is 4 days unopened at 2 to 8 °C or 2 hours opened at room temperature. The protocols for stability and acceptance criteria were reviewed and are acceptable.
{8}
d. Detection limit:
The Limit of Blank (LoB) determination was based on 60 replicate measurements of AQT Assay Buffer on a single AQT90 FLEX instrument and one test kit lot number. The Limit of Blank as determined by the upper 95th percentile is 0.09 ng/mL. LoB is 0.5 ng/mL.
Limit of Detection (LoD) determination was based on 20 replicate measurements of 10 low samples on 2 AQT90 FLEX instruments and 2 test kit lot numbers. The LoD determination was calculated by using the determined LoB value and pooled SD values from LoD samples. The sponsor claimed that LoD is 1.0 ng/mL for both EDTA plasma and EDTA whole blood.
Limit of Quantitation (LoQ) determination was based on an interassay precision of >20% for 30 replicate measurements of 10 low samples on two AQT90 FLEX instruments and 2 test kit lot numbers. The Limit of Quantitation was determined to be 0.57 ng/mL and 0.62 ng/mL for EDTA plasma and EDTA whole blood samples, respectively. The sponsor claimed that LoQ is 1.0 ng/mL for both plasma and whole blood.
LoB, LoD and LoQ are summarized in the table below:
| CKMB | Value ng/ml |
| --- | --- |
| Limit of blank (LoB) | 0.5 |
| Limit of detection (LoD) | 1.0 |
| Limit of quantitation (LoQ) | 1.0 |
The sponsor's claimed measuring range is 1.5 to 300 ng/mL.
e. Analytical specificity:
Interference studies were designed according to the CLSI EP7-A guideline. Two levels of human plasma (CKMB approximately 13 and 300 ng/mL) were spiked with known interference substances (52 drugs and 13 endogenous compounds) and samples were analyzed in five replicates on the AQT90 FLEX analyzer. In addition, two levels of whole blood samples (13 and 300 ng/mL) were also tested with the 13 endogenous compounds. Non-significant interference, defined as
{9}
recovery within ±10% compared to reference without any interfering substance, was observed with the assay. Results are summarized in the table below for the endogenous interference substances (both plasma and whole blood yielded similar results). The tests for unconjugated bilirubin, conjugated bilirubin, biotin, cholesterol, creatinine, glucose, hemoglobin, human IgG, triglycerides and urea were performed with Lithium Heparin as the anticoagulant. All other tests were performed with K₂EDTA as anticoagulant. The interferent study results are summarized below:
| Interferents | Highest concentration tested showing non-significant interference |
| --- | --- |
| Bilirubin, conjugated | 27 mg/dL |
| Bilirubin, unconjugated | 27 mg/dL |
| Biotin | 3 μ/L |
| Cholesterol | 500 mg/dL |
| Creatinine | 5.0 mg/dL |
| Fibrinogen | 1000 mg/dL |
| Glucose | 1000 mg/dL |
| Hemoglobin | 1000 mg/dL |
| Human albumin | 10000 mg/dL |
| Human IgG | 5300 mg/dL |
| Triglycerides (Intralipid) | 1500 mg/dL |
| Urea | 200 mg/dL |
Results of the 52 drugs tested that have non-significant interference are listed in the labeling.
No hook effect was observed for concentrations up to 7600 ng/mL with either EDTA plasma or EDTA whole blood.
Cross reactivity with CKBB and CKMM were evaluated at two levels of CKMB concentrations (1 ng/mL and 15 ng/mL). Samples spiked with either 2500 ng/mL of CKBB or CKMM were used for the studies. Studies were carried out for both plasma and whole blood samples. All the cross-reactivity results were less than 0.1%.
Potential interference from human anti-mouse antibody (HAMA) and Rheumatoid factor (RF) was evaluated by comparing the results of seventeen human HAMA and/or RF
10
{10}
plasma samples spiked with CKMB to reference sample (HAMA/RF negative) containing the same amount of CKMB. No negative interference for the assay was observed, which was defined as recovery within $\pm 10\%$ of human HAMA/RF positive samples compared to a reference sample (HAMA/RF negative). The highest concentration test for HAMA was $1825~\mathrm{ng/mL}$ and for RF was $8530\mathrm{IU/mL}$. The sponsor claims no interference from human anti-mouse antibody (HAMA) and Rheumatoid factor (RF).
f. Assay cut-off: Not applicable
2. Comparison studies:
a. Method comparison with predicate device:
Paired human whole blood and lithium-heparin plasma samples were collected at three hospitals (POC sites) and measurements were obtained using the AQT90 FLEX CKMB Test run on the AQT90 FLEX analyzer. Frozen plasma samples were sent to a central laboratory and comparative measurements were obtained on the predicate device (VIDAS CKMB assay). Nine samples were spiked to supplement the samples in the hard-to-find, upper range. Singlet set of results from the candidate device was used to compare against the mean results of the predicate device. The results were analyzed by Passing-Bablok regression. The correlation data is summarized below:
| AQT Fresh Whole Blood vs. Vidas Frozen Plasma | POC 1 | POC 2 | POC 3 |
| --- | --- | --- | --- |
| Number of samples | 43 | 46 | 48 |
| Range of data (ng/mL) | 1.8-235 | 1.7-277 | 1.5-206 |
| Slope | 0.93 | 0.93 | 0.96 |
| Slope 95% C.I. | 0.87-0.99 | 0.85-1.00 | 0.89-1.05 |
| Intercept (ng/mL) | -0.2 | 0.0 | -0.5 |
| Intercept 95% C.I. | -0.5-0.1 | -0.4-0.3 | -1.0- -0.2 |
| r² | 0.99 | 0.98 | 0.95 |
{11}
| AQT Fresh Plasma vs. Vidas Frozen Plasma | POC 1 | POC 2 | POC 3 |
| --- | --- | --- | --- |
| Number of samples | 44 | 46 | 48 |
| Range of data (ng/mL) | 2.1-238 | 1.7-279 | 1.9-218 |
| Slope | 0.94 | 0.96 | 0.93 |
| Slope 95% C.I. | 0.89-0.98 | 0.88-1.01 | 0.89-1.05 |
| Intercept (ng/mL) | -0.3 | 0.0 | -0.4 |
| Intercept 95% C.I. | -0.5-0.1 | -0.2-0.3 | -1.0-0.0 |
| r² | 0.99 | 0.99 | 0.95 |
# b. matrix comparison:
Four matrix- anticoagulant combinations were evaluated in the study, which were Lithium-Heparin whole blood, Lithium-Heparin plasma, EDTA whole blood and EDTA plasma. The study followed CLSI guideline EP9-A2, Method Comparison and Bias Estimation Using Patient Samples. The results of plasma versus whole blood were analyzed by Passing-Bablok regression. Results are summarized below:
| Matrix | Regression | r2 | n | Data range |
| --- | --- | --- | --- | --- |
| Li-Hep Plasma vs. Li-Hep WB | y = 1.04x + 0.1 | 1.00 | 75 | 1.7-294 |
| EDTA Plasma vs. EDTA WB | y = 1.03x + 0.0 | 1.00 | 74 | 1.9-288 |
| Anticoagulant | Regression | r2 | n | Data range |
| --- | --- | --- | --- | --- |
| EDTA WB vs. Li-Hep WB | y = 1.03x + 0.0 | 1.00 | 65 | 1.7-198 |
| EDTA Plasma vs. Li-Hep Plasma | y = 1.02x + 0.0 | 1.00 | 71 | 1.9-197 |
{12}
The sponsor concluded that lithium heparin and EDTA are acceptable anticoagulant and the four anticoagulant-matrix combinations can be used.
3. Clinical studies:
a. Clinical Sensitivity: Not applicable
b. Clinical specificity: Not applicable
c. Other clinical supportive data (when a. and b. are not applicable): Not applicable
4. Clinical cut-off: None
5. Expected values/Reference range:
Whole blood (lithium-heparin and EDTA) and plasma (lithium-heparin and EDTA) were obtained from 691 apparently healthy individuals (ages: 21 to 91 years) from a geographically diverse U.S. adult population (388 women and 303 men) and analyzed using the AQT90 FLEX CKMB assay. The 97.5th Percentile was determined to be 6.9 ng/mL for women and 11 ng/mL for men. It is recommended that each laboratory should establish its own reference range.
N. Proposed Labeling:
The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10.
O. Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.