ACS CKMB II IMMUNOASSAY
Device Facts
| Record ID | K962126 |
|---|---|
| Device Name | ACS CKMB II IMMUNOASSAY |
| Applicant | Ciba Corning Diagnostics Corp. |
| Product Code | JHS · Clinical Chemistry |
| Decision Date | Jun 21, 1996 |
| Decision | SESE |
| Submission Type | Traditional |
| Regulation | 21 CFR 862.1215 |
| Device Class | Class 2 |
Indications for Use
The intended use of ACS CKMB II is for the quantitative determination of CK-MB (CK-2) in serum or plasma using the Ciba Corning Automated Chemiluminescence Systems.
Device Story
ACS CKMB II is a two-site chemilumometric (sandwich) immunoassay; utilizes monoclonal mouse anti-CK-MB antibody labeled with acridinium ester and monoclonal mouse anti-CK-BB antibody coupled to paramagnetic particles. Input: serum or plasma samples. Operation: automated on Ciba Corning ACS:180 systems; detects relative light units (RLUs) proportional to CK-MB concentration. Output: quantitative CK-MB concentration (ng/mL). Used in clinical laboratory settings by trained technicians. Results assist clinicians in diagnosing myocardial infarction.
Clinical Evidence
Bench testing only. Accuracy evaluated using 1744 samples (0.0-280.0 ng/mL) comparing ACS CKMB II to FEIA reference method; correlation coefficient r=0.96; regression equation: y = 1.01x + 0.24. Sensitivity: minimum detectable concentration 0.18 ng/mL; range up to 300.0 ng/mL. Precision: total % CV 3.0-3.5%.
Technological Characteristics
Non-competitive chemiluminescence immunoassay. Reagents: monoclonal mouse anti-CK-MB (acridinium ester labeled) and monoclonal mouse anti-CK-BB (paramagnetic particle coupled). Automated system (ACS:180) detection of relative light units. Quantitative measurement range 0.18-300.0 ng/mL.
Indications for Use
Indicated for quantitative determination of CK-MB (CK-2) in human serum or plasma samples to aid in the diagnosis of myocardial infarction.
Regulatory Classification
Identification
A creatine phosphokinase/creatine kinase or isoenzymes test system is a device intended to measure the activity of the enzyme creatine phosphokinase or its isoenzymes (a group of enzymes with similar biological activity) in plasma and serum. Measurements of creatine phosphokinase and its isoenzymes are used in the diagnosis and treatment of myocardial infarction and muscle diseases such as progressive, Duchenne-type muscular dystrophy.
Predicate Devices
- Baxter Stratus CK-MB Fluorometric Enzyme Immunoassay (510(k) number not known)
Related Devices
- K993068 — VITROS IMMUNODIAGNOSTIC PRODUCTS CK-MB REAGENT PACK ; VITROS IMMUNODIAGNOSTIC PRODUCTS CK-MB CALIBRATORS · Ortho-Clinical Diagnostics · Oct 4, 1999
- K981495 — IMMULITE CK-MB, CATALOG # LKCP1 (100 TESTS), LKCP5 (500 TESTS) · Diagnostic Products Corp. · Jun 10, 1998
- K041596 — ARCHITECT STAT CK-MB IMMUNOASSAY · Fisher Diagnostics · Jul 21, 2004