K970424 · Excel Scientific, Inc. · JHI · Mar 6, 1997 · Clinical Chemistry
Device Facts
Record ID
K970424
Device Name
EXCEL ONE STEP URINE HCG PREGNANCY DIPSTICK TEST
Applicant
Excel Scientific, Inc.
Product Code
JHI · Clinical Chemistry
Decision Date
Mar 6, 1997
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 862.1155
Device Class
Class 2
Indications for Use
Excel OneStep Urine hCG Pregnancy DipStick Test is Excel Scientific, Inc.'s name for professional use of solid phase migratory chromatography immunoassay in vitro diagnostic test kit for the qualitative detection of human Chorionic Gonadotropin (hCG) in urine.
Device Story
Device is a solid-phase migratory chromatography immunoassay dipstick for qualitative hCG detection in urine. Operated by healthcare professionals in clinical settings. Input is urine specimen; capillary action draws sample through gold sol-antibody conjugate pad. hCG binds to conjugate, forming complex that migrates to immobilized anti-hCG antibody zone, producing rose-pink band if hCG concentration ≥ 25 mIU/mL. Control zone confirms reagent/device function. Output is visual color band indicating presence or absence of hCG. Results assist clinicians in pregnancy diagnosis.
Clinical Evidence
No clinical data provided. Substantial equivalence is supported by comparative bench testing against the predicate device, demonstrating identical sensitivity (25 mIU/mL) and functional performance.
Technological Characteristics
Solid-phase migratory chromatography immunoassay. Employs gold sol-antibody conjugate and immobilized anti-hCG antibodies on a membrane. Form factor is a dipstick. No external energy source or software required.
Indications for Use
Indicated for professional use for the qualitative detection of human Chorionic Gonadotropin (hCG) in urine to aid in the early detection of pregnancy. Cutoff concentration is 25 mIU/mL.
Regulatory Classification
Identification
A human chorionic gonadotropin (HCG) test system is a device intended for the early detection of pregnancy is intended to measure HCG, a placental hormone, in plasma or urine. A human chorionic goadotropin (HCG) test system is a device intended for any uses other than early detection of pregnancy (such as an aid in the diagnosis, prognosis, and management of treatment of persons with certain tumors or carcinomas) is intended to measure HCG, a placental hormone, in plasma or urine.
Predicate Devices
Syntron Quikpac EIA
Reference Devices
QuikStrip (or E.Z. Strip) Pregnancy Home test
Related Devices
K962906 — SCRIPPS LABORATORIES HCG ONE-STEP · Scripps Laboratories, Inc. · Aug 23, 1996
K020968 — LIVESURE HCG URINARY PREGNANCY TEST STRIP AND TEST CARD · Pan Probe Biotech, Inc. · Jul 15, 2002
K963231 — EXCEL ONESTEP URINE HCG MODULE PREGNANCY TEST · Excel Scientific, Inc. · Dec 4, 1996
K051841 — WH ACCU TEST ONE-STEP URINE/SERUM COMBO PREGNANCY TEST · W.H.P.M., Inc. · Aug 22, 2005
K123050 — EGENS ONE STEP HCG URINE PREGNANCY TEST KIT(S) (STRIP, CASSETTE, MIDSTREAM I AND MIDSTREAM II) · Nantong Biotechnology Co., Ltd. · Feb 5, 2013
Submission Summary (Full Text)
{0}
EXCEL
"Commitment to Excellence"
Attachment #4
K970424
MAR - 6 1997
# Summary of Safety and Effectiveness
Excel OneStep Urine hCG Pregnancy DipStick Test
Excel OneStep Urine hCG Pregnancy DipStick Test is Excel Scientific, Inc.'s name for professional use of solid phase migratory chromatography immunoassay in vitro diagnostic test kit for the qualitative detection of human Chorionic Gonadotropin (hCG) in urine. Excel OneStep Urine hCG Pregnancy DipStick Test is based on comparative data with Syntron Quikpac EIA, an assay for human Chorionic Gonadotropin currently being marketed.
The Principle of Excel OneStep Urine hCG Pregnancy DipStick Test is as follows:
1. Both the Excel OneStep Urine hCG Pregnancy DipStick and the QuikStrip (or E.Z. Strip) Pregnancy Home tests are a sandwich solid phase dye-conjugated non-enzyme immunoassay. When the Excel OneStep Urine hCG Pregnancy DipStick is placed in a urine test cup containing the urine specimen, the urine will migrate by capillary action. The gold sol-antibody conjugate in the gold sol pad of the Excel OneStep Urine hCG Pregnancy DipStick binds to the hCG molecules in the specimen forming an antibody-antigen complex. This complex binds to the anti-hCG antibody immobilized in the positive reaction zone in the white membrane area and produces a rose-pink color band when the hCG concentration is at or greater than 25 mIU/mL. In the absence of detectable levels of hCG, there is no rose-pink band in the positive reaction zone. The reaction mixture continues to flow through the absorbent device, pass the positive reaction zone, producing a rose-pink color band in the control zone of the white membrane area, demonstrating that the reagents and the device are functioning correctly.
2. There is no difference in the testing procedure between the Excel OneStep Urine hCG Pregnancy DipStick Test and the Syntron's QuikStrip (or E.Z. Strip) Pregnancy Home test as shown below:
| | QuikStrip (or E.Z. Strip) Pregnancy Home test | Excel OneStep Urine hCG Pregnancy DipStick Test |
| --- | --- | --- |
| Specimen Volume | 6 drops | 6 drops of urine |
| Incubation | 5 minutes | 5 minutes |
| Reagent | No | No |
| Step | one step | one step |
| Cutoff | 25 mIU/mL | 25 mIU/mL |
| Urine Cup | Yes | Yes |
3. The differences between the Excel OneStep Urine hCG Pregnancy DipStick Test and the Syntron Quikpac EIA Pregnancy Test are as follows:
| | Syntron Quikpac EIA Pregnancy Test | Excel OneStep Urine hCG Pregnancy DipStick Test |
| --- | --- | --- |
| Specimen Volume | 3 drops of urine | 6 drops of urine |
| Incubation | 3 minutes with antibody and 3 minutes with substrate | 5 minutes |
| Reagent | 2 bottles | No |
| Step | 3 steps | one step |
| Cutoff | 25 mIU/mL | 25 mIU/mL |
Rev. C1, 9-96
23071 E. La Palma Ave., Yorba Linda, CA 92887, U.S.A. Tel: (714) 692-7151 Fax: (714) 692-7153
{1}
EXCEL
"Commitment to Excellence"
4. For questions concerning the safety and effectiveness of the Excel OneStep Urine hCG Pregnancy DipStick, please contact:
Min-Lee Cheng, Ph.D.
23071 E. La Palma Avenue
Yorba Linda, CA 92887
Tel: (714) 692-7151
Rev. C1, 9-06
23071 E. La Palma Ave., Yorba Linda, CA 92887, U.S.A. Tel: (714) 692-7151 Fax: (714) 692-7153
13
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.