K964574 · Athena Medical Corp. · JHI · Jan 21, 1997 · Clinical Chemistry
Device Facts
Record ID
K964574
Device Name
AFFIRM ONE-STEP PROFESSIONAL PREGNANCY TEST
Applicant
Athena Medical Corp.
Product Code
JHI · Clinical Chemistry
Decision Date
Jan 21, 1997
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 862.1155
Device Class
Class 2
Indications for Use
Affirm One-Step Professional Pregnancy Test is a rapid one-step visual test for the qualitative detection of human Chorionic Gonadotropin in human urine to aid in the diagnosis of pregnancy. This test is intended for Professional use.
Device Story
Lateral flow immunoassay for qualitative hCG detection in human urine. Device consists of absorbent wick, conjugate pad with monoclonal anti-beta hCG antibodies, and nitrocellulose membrane with polyclonal goat anti-alpha hCG antibodies and goat anti-mouse IgG control line. Operation: urine applied via mid-stream or dip method; capillary action draws sample through reagent pad and membrane. Positive result indicated by color lines in both test and control windows; negative by control line only. Used in clinical settings by professionals. Provides rapid visual confirmation of pregnancy to aid clinical decision-making.
Clinical Evidence
Bench testing and clinical evaluation performed. Specificity testing showed no cross-reactivity with LH, FSH, or TSH. Sensitivity confirmed at 25 mIU/ml using spiked urine samples (n=20 per concentration). Interference testing with common substances (acetaminophen, caffeine, glucose, etc.) showed no interference. Clinical study (n=104) compared urine stream method to predicate dip method; demonstrated 99% user accuracy (53 true positives, 46 true negatives, 1 false negative) compared to 100% laboratory accuracy for the predicate.
Technological Characteristics
Lateral flow assay. Components: absorbent wick, conjugate pad, nitrocellulose membrane, wicking pad, plastic housing. Antibodies: monoclonal anti-beta hCG (conjugate), polyclonal goat anti-alpha hCG (test line), polyclonal goat anti-mouse IgG (control line). Sensitivity: 25 mIU/ml. Format: pen-like device with two result windows and moisture-resistant clear barriers. Visual readout.
Indications for Use
Indicated for professional use to detect human Chorionic Gonadotropin (hCG) in human urine to aid in pregnancy diagnosis. Applicable for use as early as the first day of a missed period.
Regulatory Classification
Identification
A human chorionic gonadotropin (HCG) test system is a device intended for the early detection of pregnancy is intended to measure HCG, a placental hormone, in plasma or urine. A human chorionic goadotropin (HCG) test system is a device intended for any uses other than early detection of pregnancy (such as an aid in the diagnosis, prognosis, and management of treatment of persons with certain tumors or carcinomas) is intended to measure HCG, a placental hormone, in plasma or urine.
Predicate Devices
SAS™ Pregnancy Strip
Related Devices
K250117 — FaStep Pregnancy Rapid Test Cassette; Fastep HCG Rapid Test Cassette · Assure Tech., LLC · Feb 13, 2025
K150063 — One Step hCG Pregnancy Test · Coretests, Inc. · Oct 2, 2015
K020968 — LIVESURE HCG URINARY PREGNANCY TEST STRIP AND TEST CARD · Pan Probe Biotech, Inc. · Jul 15, 2002
Submission Summary (Full Text)
{0}
K964574
Jan. 21, 1997
# 510 (k) SUMMARY
## AFFIRM ONE STEP PROFESSIONAL PREGNANCY TEST
86
{1}
# 510(k) Summary
**Submitted by:**
Athena Medical Corporation
10180 SW Nimbus Ave. Suite J-5
Portland, OR 97223
**Contact:**
Patricia Feetham
Phone: 503/968-8800 ext. 19
Fax: 503/639-3674
**Product Name:**
**Proprietary Name:**
Affirm One-Step Professional Pregnancy Test
**Classification Name:**
human Chorionic Gonadotropin test systems
**Intended Use:**
Affirm One-Step Professional Pregnancy Test is a rapid one-step visual test for the qualitative detection of human Chorionic Gonadotropin in human urine to aid in the diagnosis of pregnancy. This test is intended for Professional use.
**Description of Device:**
Affirm detects the presence of hCG in urine by using a combination of polyclonal and monoclonal antibodies on the reagent pad and the test strip. The assay begins functioning with the addition of urine onto the absorbent wick. Through capillary action the urine is pulled to the reagent pad where the hCG reacts with the conjugate, and then through the membrane containing bound capture antibodies. Positive results are determined by one color line in the test window and one color line in the control window. The absence of hCG produces only one color line in the control window.
**Predicate device for substantial equivalence comparison:**
SAS™ Pregnancy Strip
**Date Prepared:**
November 1, 1996
40
{2}
Substantial Equivalence:
Comparison Product: SAS™ Pregnancy Strip
Manufactured by: SA Scientific™
4919 Golden Quail
San Antonio TX 78240
Similarities to SAS™ Pregnancy Strip:
| Items to be compared | Affirm One-Step Professional Pregnancy Test | SAS™ Pregnancy Strip |
| --- | --- | --- |
| • Detection method: | hCG in urine specimens reacts with the colored-conjugate on the pad producing a solid color (red) line in the test region. Mouse conjugate binds to goat anti-mouse antibody in the control region on the strip. | hCG in urine specimens reacts with the colored-conjugate on the pad producing a solid color (red) line in the test region. Mouse conjugate binds to goat anti-mouse antibody in the control region on the strip. |
| • Method | Mid-stream or dip. | Dip only. |
| • Usage: | Anytime of the day. | Anytime of the day. |
| • Pregnancy determination: | As early as the first day of a women’s missed period. | As early as the first day of a women’s missed period. |
| • Antibodies: | Polyclonal goat anti-alpha hCG antibody is on the test strip, monoclonal anti-beta hCG antibody is on the conjugate pad, and polyclonal goat anti-mouse IgG antibody is on the control line. | Polyclonal goat anti-alpha hCG antibody is on the test strip, monoclonal anti-beta hCG antibody is on the conjugate pad, and polyclonal goat anti-mouse IgG antibody is on the control line. |
| • Results: | Positive = color line in test window and control window. Negative = color line in control window only. | Positive = color line in specimen zone and control zone. Negative = color line in control zone only. |
| • Format: | Lateral Flow assay. | Lateral Flow assay. |
| • Sensitivity: | Detects level of hCG as low as 25 mIU/ml. | Detects level of hCG as low as 25 mIU/ml. |
| • Device design: | Slender, pen-like shape, removable cap, absorbent tip partially enclosed, two window result view. | Test strip utilizing an immersion zone, a specimen zone and control zone. |
| • Visual detection: | A solid color line appears in control and test window in the presence of hCG. A solid color line appears in the control window only in the absence of hCG. | A solid color band appears in specimen and control zones in the presence of hCG. A solid color band appears in the control zone only in the absence of hCG. |
{3}
Differences between Affirm and SAS™ Pregnancy Strip:
| Items to be compared | Affirm One-Step Professional pregnancy test | SAS™ Pregnancy Strip |
| --- | --- | --- |
| • Device engineering: | Test device which contains an exposed absorbent wick, a conjugate pad, nitrocellulose membrane containing bound antibodies, and a wicking pad in a disposable plastic container. Result windows and two “tear drop” openings exposing the absorbent wick appear on the same side of the device. Area covering wick is a small cap. Result and control windows include a moisture resistant clear barrier. | Test strip only, which contains an exposed absorbent wick, a conjugate pad, nitrocellulose membrane containing bound antibodies, and a wicking pad. Test strips are stored in a desiccated canister until ready for use. |
| • Developing time: | Minimum of 4 minutes for negative results to appear; positive results as soon as 1 - 2 minutes (see Appendix A). Maximum of 20 minutes for any result. | Read results after 5 minutes. Do not interpret results after 30 minutes. |
| • Absorbent wick: | Partially enclosed with 2 tear drop openings in the plastic housing. | Wick is bottom 1/3 of test strip. |
| • Results windows: | Covered by a clear plastic shield which acts as splash guard to protect the results. | Colored bands are displayed in the specimen and/or control zones on the strip. |
| • Test window/Control window: | Test window shape is: “+”, control window shape is: “o”. | Test zones are bands in the middle to upper quadrant of the test strip. |
| • Color of holder: | Ivory | No holder |
The differences between the Affirm One-Step Professional Pregnancy Test and SAS™ Pregnancy Strip are not significant and do not effect the safety and effectiveness of the device.
42
89
{4}
# Performance characteristics:
## Specificity:
Affirm One-Step Professional Pregnancy Test was evaluated for specificity against LH (300 mIU/ml), FSH (1000 mIU/ml), and TSH (1000 μIU/ml). Affirm did not cross-react with these substances at these levels, nor were samples spiked with 25 mIU/ml concentration of hCG inhibited by these substances.
## Sensitivity:
The sensitivity of the Affirm One-Step Professional Pregnancy Test was determined using 20 negative urine samples spiked with hCG at varying levels. The results are described in the table below.
Sensitivity of Affirm One-Step Professional Pregnancy Test:
| Concentration Spiked at: | Results: Positive | Results: Negative | A 25 mIU/ml hCG urine reference control, SAS™ #087745P, Lot #5122101, Exp. 12/97 was ran during these studies. |
| --- | --- | --- | --- |
| 10 mIU/ml | 5 | 15 | |
| 25 mIU/ml | 20 | 0 | |
| 50 mIU/ml | 20 | 0 | |
| 100 mIU/ml | 20 | 0 | |
| 0 mIU/ml | 0 | 20 | |
Of the 20 samples that were tested at the 10 mIU/ml concentration, 5 of them generated a positive reading after about 15 minutes from the point of saturation. The 20 samples tested at the 25 mIU/ml concentration all generated a positive reading within 4 minutes. Therefore our claim is a 25 mIU/ml sensitivity with the Affirm One-Step Professional Pregnancy Test. Even though the test is sensitive enough to read concentrations lower than 25 mIU/ml this does not occur regularly for us to claim a lower sensitivity for the test.
43
90
{5}
# Interfering Substances:
The following potentially interfering substances were added to hCG negative and 25 mIU/ml hCG spiked urine samples. Interfering test samples were compared directly to the control result. The following results are described in the table below:
## Interference of the Affirm One-Step Professional Pregnancy Test
| Substance: | Concentration: | Interference: |
| --- | --- | --- |
| Acetaminophen | 20 mg/dI | NONE |
| Acetylsalicylic Acid | 20 mg/dI | NONE |
| Ascorbic Acid | 20 mg/dI | NONE |
| Atropine | 20 mg/dI | NONE |
| Caffeine | 20 mg/dI | NONE |
| Gentisic Acid | 20 mg/dI | NONE |
| Glucose | 2 g/dI | NONE |
| Hemoglobin | 200 mg/dI | NONE |
## Predicate Device Accuracy Correlation:
One hundred and four specimens were evaluated by the urine stream method and the dip method. The women participating in the study were solicited by a flyer appearing in clinics locally, and asked to participate at one of the following locations: Athena Medical Corporation, Planned Parenthood in either Southeast Portland or Beaverton, Oregon. Women who participated in the urine stream method answered a questionnaire about the product. The women participants identified 53 true positives and 46 true negatives, with 1 false negative, generating a 99% user accuracy rate for the Affirm One-Step Professional Pregnancy Test. The laboratory comparison of the same collected samples using the SAS™ Pregnancy Strip identified 54 true positives and 46 true negatives, generating a 100% laboratory accuracy rate. There were four invalid results due to insufficient urine placed on the device.
44
91
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.