K993211 · Abaxis, Inc. · JGS · Oct 25, 1999 · Clinical Chemistry
Device Facts
Record ID
K993211
Device Name
PICCOLO SODIUM TEST SYSTEM
Applicant
Abaxis, Inc.
Product Code
JGS · Clinical Chemistry
Decision Date
Oct 25, 1999
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 862.1665
Device Class
Class 2
Indications for Use
The Piccolo® MetLyte 7 Reagent Disc (contains Piccolo Sodium Test System) used with the Piccolo Point-of-Care Chemistry Analyzer is intended to be used for the in vitro quantitative determination of sodium in heparinized whole blood, heparinized plasma, or serum in a clinical laboratory setting or point-of-care location.
Device Story
Piccolo Sodium Test System uses heparinized whole blood, plasma, or serum samples; processed via Piccolo MetLyte 7 Reagent Disc. Device separates blood cells from plasma; meters plasma and diluent; mixes with dry reagent beads in reaction cuvettes. Chemical reactions monitored by Piccolo Point-of-Care Chemistry Analyzer at 37°C. Analyzer uses bar-coded disc-specific calibration data. Used in clinical labs or point-of-care settings by healthcare professionals. Output is quantitative sodium concentration; aids clinicians in diagnosing/treating fluid/electrolyte disorders like dehydration, hyponatremia, and hyperaldosteronism.
Clinical Evidence
Bench testing only. Linearity established at 99% significance level (F-ratio 0.99). Precision studies (n=80) showed CVs of 1.6-1.8%. Clinical correlation study (n=113) compared subject device to predicate, yielding correlation coefficient of 0.937, slope 0.782, and intercept 27.70.
Technological Characteristics
System consists of Piccolo Point-of-Care Chemistry Analyzer and MetLyte 7 Reagent Disc. Sensing principle: enzymatic activation. Reagents: dry, test-specific beads. Calibration: bar-coded disc-specific data. Operating temperature: 37°C. Assay range: 110-170 mmol/L.
Indications for Use
Indicated for quantitative determination of sodium in human heparinized whole blood, heparinized plasma, or serum. Used in diagnosis and treatment of dehydration, diabetes insipidus, loss of hypotonic gastrointestinal fluids, salt poisoning, thirst sense depression, skin losses, burns, sweating, hyperaldosteronism, CNS disorders, hyponatremia, and syndrome of inappropriate ADH secretion.
Regulatory Classification
Identification
A sodium test system is a device intended to measure sodium in serum, plasma, and urine. Measurements obtained by this device are used in the diagnosis and treatment of aldosteronism (excessive secretion of the hormone aldosterone), diabetes insipidus (chronic excretion of large amounts of dilute urine, accompanied by extreme thirst), adrenal hypertension, Addison's disease (caused by destruction of the adrenal glands), dehydration, inappropriate antidiuretic hormone secretion, or other diseases involving electrolyte imbalance.
{0}------------------------------------------------
Summary of Safety and Effectiveness, Page 4
## Sample Type Comparison:
A study was conducted to compare venous whole blood and serum run on the Piccolo® Point-of-Care Chemistry Analyzer.
Serum and whole blood comparability were established for sodium.
#### 8. Conclusions
The clinical and non-clinical tests performed using the Piccolo Sodium Test System (contained on the Piccolo MetLyte 7 Reagent Disc) demonstrate that the device is as safe, effective and performs as well as the legally marketed predicate device identified above.
{1}------------------------------------------------
OCT 25 599
1320 Chesapeake Terrace, Sunnyvale, CA 84089
Phone 408 • 734-0200 Fax 408 • 734-2874
Image /page/1/Picture/2 description: The image shows the word "ABAXIS" in a stylized font. The letters are in black and are bold. A black oval shape surrounds the letters "XIS" in the word. The font is sans-serif and appears to be a logo.
This 510(k) summary of safety and effectiveness information is being submitted in
tell and experience of the COMDA 4000 and 21 CER 807 92 This 510(K) summary of salety and Chocaveness .
accordance with the requirements of SMDA 1990 and 21 CFR 807.92.
The assigned 510(k) number is: K993211
### Applicant Information: 1.
| Date Prepared: | September 23, 1999 |
|----------------|----------------------------------------------|
| Name: | Abaxis, Inc. |
| Address: | 1320 Chesapeake Drive<br>Sunnyvale, CA 94089 |
| Contact Person: | Lisa G. McGrath |
|-----------------|-----------------|
| Phone Number: | (408) 745-6880 |
| Fax Number: | (408) 734-2874 |
### Device Information: 2.
| Classification | Class II |
|----------------|-----------------------------|
| Trade Name: | Piccolo® Sodium Test System |
Classification Name:
Sodium test system 862.1665
## ldentification of legally marketed device to which the submitter claims 3. equivalence:
The following table identifies the legally marketed device to which Abaxis claims equivalence:
| : 24 8 2 3 0 ( 2002 16) : 1 ( 1 ( 1 ) ( 1 ) ( 1 ) ( ) ( ) ( ) ( ) ( ) ( ) ( ) ( ) ( ) ( ) ( ) ( ) ( ) ( ) ( ) ( ) ( ) ( ) ( ) ( ) ( ) ( ) ( ) ( ) ( ) ( ) ( ) ( ) ( ) ( ) ( ) | 11:1011117 10:44 r : 5 - | 1131 - 101 - 14 | 1718: 0) - 295<br>. 1728 : } : 38 : 18 : 3 : 1 : 1 : 1 : 1 : 1 : 1 : 1 : 1 : 1 : 1 : 1 : 1 : 1 : 1 : 1 : 1 : 1 : 1 : 1 : 1 : 1 : 1 : 1 : 1 : 1 : 1 : 1 : 1 : 1 : 1 : 1 : 1 : 1 : 1 : 1 : 1 : 1 : |
|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------|-----------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| KNA™ 2 Sodium-<br>Potassium Analyzer | Radiometer<br>America, Inc. | K830805 | 4/8/83 |
### Description of the Device: 4.
The Piccolo Sodium Test System (contained on the Piccolo MetLyte 7 Reagent Disc) is designed to separate a heparinized whole blood sample into plasma and Disc) is designed to beparate a neguired quantity of plasma and diluent, mixes the discus blood cells. The diso meters the mixture to the reaction cuvettes along the disc
{2}------------------------------------------------
# Summary of Safety and Effectiveness, Page 2
perimeter. The diluted plasma mixes with the reagent beads, initiating the chemical reactions that are then monitored by the analyzer. Alternately, the disc may also be used with serum.
### 5. Statement of Intended Use:
The Piccolo® MetLyte 7 Reagent Disc (contains Piccolo Sodium Test System) used with the Piccolo Point-of-Care Chemistry Analyzer is intended to be used for the in vitro quantitative determination of sodium in heparinized whole blood, heparinized plasma, or serum in a clinical laboratory setting or point-of-care location.
### 6. Summary of the technological characteristics of the new device in comparison to those of the predicate device:
Table 1 outlines the technological characteristics of the Piccolo Sodium Test System in comparison to the legally marketed predicate device.
| | Piccolo Point-of-Care<br>Chemistry Analyzer | Radiometer KNA™ 2 Sodium<br>Potassium Analyzer |
|-------------------|--------------------------------------------------|--------------------------------------------------------------------------------------|
| Intended Use | quantitative analysis of<br>sodium | quantitative determination of<br>sodium |
| Methodology | enzymatic activation | ion-selective electrodes |
| Sample Type | whole blood, plasma,<br>serum | whole blood, plasma,<br>serum, urine |
| Sensitivity | 2.8 (A mA/min)/(mmol/L) | 3.1 (mmol/L)/(mV) |
| Reagents | dry, test specific reagent beads | N/A |
| Temperature | 37°C | 37° C |
| Calibration | bar code with disc-<br>specific calibration data | automatic 1 point calibration<br>every 2 hours; 2 point<br>calibration every 8 hours |
| Assay Range | 110 - 170 mmol/L | 7 - 350 mmol/L |
| Accuracy: | | |
| Sample size (n) | 113 | |
| Corr. Coefficient | 0.937 | |
| Slope | 0.782 | |
| Intercept | 27.70 | |
| SEE | 3.79 | |
| R-Square | 0.878 | |
Table 1 : Specification Comparison: Piccolo Sodium Test System
{3}------------------------------------------------
# Summary of Safety and Effectiveness, Page 3
### Brief discussion of the clinical and nonclinical tests relied on for a 7. determination of substantial equivalence.
Table 2 summarizes the results of clinical and non-clinical tests performed using the Piccolo® Sodium Test System.
# Linearity:
Data for sodium were found to be statistically linear at the 99% significance level by the Ftest.
| Table 2 | |
|----------------------|--------|
| Summary of Linearity | |
| | Sodium |
| F-Ratio | 0.99 |
| Slope | 0.99 |
| Intercept | -1.00 |
| Corr. Coefficient | 1.00 |
99% Critical F 2.99
# Precision:
Precision studies were designed to evaluate within-run and total precision of sodium included on the Piccolo MetLyte 7 Reagent Disc when run on the Piccolo Point-of-Care Chemistry Analyzer.
| Table 3:<br>Within-Run and Total Precision of Sodium Run<br>on the Piccolo Point-of-Care Chemistry AnalyzerA | | |
|--------------------------------------------------------------------------------------------------------------|------------------------|------------------|
| | Within-Run<br>(n = 80) | Total<br>(n= 80) |
| Moni-Trol 1 | | |
| Mean | 144 | 144 |
| SD | 2.3 | 2.3 |
| CV | 1.6 | 1.6 |
| Moni-Trol 2 | | |
| Mean | 120 | 120 |
| SD | 2.1 | 2.1 |
| CV | 1.8 | 1.8 |
A Results pooled from 4 instruments each running 20 discs.
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of an eagle or bird with three wing-like shapes, representing the department's mission to protect the health of all Americans and provide essential human services.
OCT 2 5 1999
Ms. Lisa G. McGrath Regulatory Affairs Specialist Abaxis 1320 Chesapeake Terrace Sunnyvale, California 94089
Re: K993211
> Trade Name: Piccolo® Sodium Test System Regulatory Class: II Product Code: JGS Dated: September 23, 1999 Received: September 24, 1999
Dear Ms. McGrath:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
Food and Drug Administration 2098 Gaither Road Rockville MD 20850
{5}------------------------------------------------
## Page 2
Under the Clinical Laboratory Improvement Amendments of 1988 (CLIA-88), this device may require a CLIA complexity categorization. To determine if it does, you should contact the Centers for Disease Control and Prevention (CDC) at (770) 488-7655.
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4588. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification"(21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597, or at its internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html".
Sincerely yours,
Steven Putman
Steven I. Gutman, M.D, M.B.A. Director Division of Clinical Laboratory Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
| | Indications for Use | |
|---------------------------|---------------------|--|
| 510(k) Number (if known): | K993211 | |
Piccolo® Sodium Test System Device Name:
Intended Use:
The Piccolo Sodium Test System (contained on the MetLyte 7 Reagent Disc), used with the Piccolo Point-of-Care Chemistry Analyzer is intended to be used for the in vitro quantitative determination of sodium in heparinized whole blood, heparinized plasma, or serum in a clinical laboratory setting or point-of-care location.
## Indications for Use:
The sodium assay is used for the quantitation of sodium in human heparinized whole blood, heparinized plasma, or serum. Sodium measurements are used in the diagnosis and treatment of dehydration, diabetes insipidus, loss of hypotonic gastrointestinal fluids, sait poisoning, selective depression of sense of thirst, skin losses, burns, sweating, hyperaldosteronism, CNS disorders, dilutional, depletional and delusional hyponatremia and syndrome of inappropriate ADH secretion.
| | and delusional hyponatr |
|---------------------|---------------------------------|
| | ean Cooper |
| (Division Sign-Off) | |
| | Division of Clinical Laboratory |
| | vice: |
| 510(k) Number | K993211 |
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
| Concurrence of CDRH, Office of Device Evaluation (ODE) |
|--------------------------------------------------------|
|--------------------------------------------------------|
ﺴﻠ Prescription Use (Per 21 CFR 801.109)
OR
Over- The Counter Use (Optional Format 1-2-96)
"
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.