K162399 · Siemens Healthcare Diagnostics, Inc. · JGJ · Jan 19, 2017 · Clinical Chemistry
Device Facts
Record ID
K162399
Device Name
Atellica CH Magnesium (Mg)
Applicant
Siemens Healthcare Diagnostics, Inc.
Product Code
JGJ · Clinical Chemistry
Decision Date
Jan 19, 2017
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 862.1495
Device Class
Class 1
Indications for Use
The Atellica™ CH Magnesium (Mg) assay is for in vitro diagnostic use in the quantitative determination of magnesium in human serum, plasma (lithium heparin), and urine using the Atellica™ CH Analyzer. Magnesium measurements are used in the diagnosis and treatment of hypomagnesemia (abnormally low levels of magnesium) and hypermagnesemia (abnormally high levels of magnesium).
Device Story
Atellica CH Magnesium (Mg) is a quantitative, photometric in vitro diagnostic assay for use on the Atellica CH Analyzer. The device processes human serum, lithium heparin plasma, or urine samples. It utilizes a modified xylidyl blue reaction in an alkaline medium where magnesium ions form a water-soluble purple-red complex; calcium is excluded via EGTA complexing. The increase in absorbance at 505/694 nm is measured, with the signal intensity proportional to magnesium concentration. The assay is intended for professional use in clinical laboratory settings. Healthcare providers use the resulting quantitative magnesium values to diagnose and manage patients with abnormal magnesium levels (hypomagnesemia or hypermagnesemia).
Clinical Evidence
No clinical data. Performance was established via bench testing, including precision (CLSI EP05-A3), linearity (CLSI EP06-A), detection limits (CLSI EP17-A2), and interference studies (CLSI EP07-A2). Method comparison against the predicate device (Dimension Magnesium Flex Reagent Cartridge) showed high correlation (r ≥ 0.996) across serum, plasma, and urine samples. Reference intervals were verified using 20 healthy adults per CLSI EP28-A3.
Technological Characteristics
Quantitative photometric assay using modified xylidyl blue reaction. Reagents are liquid-packaged in two-well packs. Sensing principle: absorbance measurement at 505/694 nm. Measuring range: 0.50–5.00 mg/dL (serum/plasma) and 1.00–14.00 mg/dL (urine). Connectivity: integrated with Atellica CH Analyzer. Storage: 2–8°C.
Indications for Use
Indicated for the quantitative determination of magnesium in human serum, plasma (lithium heparin), and urine to aid in the diagnosis and treatment of hypomagnesemia and hypermagnesemia in patients requiring clinical chemistry assessment.
Regulatory Classification
Identification
A magnesium test system is a device intended to measure magnesium levels in serum and plasma. Magnesium measurements are used in the diagnosis and treatment of hypomagnesemia (abnormally low plasma levels of magnesium) and hypermagnesemia (abnormally high plasma levels of magnesium).
{0}
1
510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION
DECISION SUMMARY
ASSAY ONLY TEMPLATE
A. 510(k) Number:
k162399
B. Purpose for Submission:
New Device
C. Measurand:
Magnesium
D. Type of Test:
Quantitative, photometric assay
E. Applicant:
Siemens Healthcare Diagnostics, Inc.
F. Proprietary and Established Names:
Atellica CH Magnesium (Mg)
G. Regulatory Information:
1. Regulation section:
CFR 862.1495, Magnesium test system
2. Classification:
Class I, reserved
3. Product code:
JGJ
{1}
4. Panel:
Clinical Chemistry
H. Intended Use:
1. Intended use(s):
See Indications for use below.
2. Indication(s) for use:
The Atellica CH Magnesium (Mg) assay is for in vitro diagnostic use in the quantitative determination of magnesium in human serum, plasma (lithium heparin), and urine using the Atellica CH Analyzer. Magnesium measurements are used in the diagnosis and treatment of hypomagnesemia (abnormally low levels of magnesium) and hypermagnesemia (abnormally high levels of magnesium).
3. Special conditions for use statement(s):
For in vitro diagnostic use.
For Professional use.
4. Special instrument requirements:
Atellica CH Analyzer
I. Device Description:
The Atellica CH Magnesium (Mg) reagents are liquid packaged in two separate reagent packs with two wells each. The saleable package contains 3 of each of the 2 packs described. These packs include a preserved Tris buffer and a preserved Xylidyl blue (0.84 mmol/L) reagent.
J. Substantial Equivalence Information:
1. Predicate device name(s):
Dimension Magnesium Flex Reagent Cartridge
2. Predicate 510(k) number(s):
k861700
{2}
3. Comparison with predicate:
| Similarities | | |
| --- | --- | --- |
| Item | Candidate Device Atellica CH Magnesium (Mg) | Predicate Device Dimension Magnesium Flex Reagent Cartridge (k861700) |
| Intended Use | The Atellica CH Magnesium (Mg) assay is for in vitro diagnostic use in the quantitative determination of magnesium | Same |
| Sample Type | Serum, Lithium Heparin plasma, and urine | Same |
| Expected Values | Urine : 24-255 mg/24 hr | Urine: Same |
| Traceability | NIST SRM 929 | Same |
| Differences | | |
| --- | --- | --- |
| Item | Candidate Device Atellica CH Magnesium (Mg) | Predicate Device Dimension Magnesium Flex Reagent Cartridge (k861700) |
| Device Technology | Xylidyl blue reaction | Methylthymol blue (MTB) complexometric procedure |
| Expected Values | Serum/plasma 1.60 to 2.60 mg/dL | Serum/Plasma 1.8 – 2.4 mg/dL |
| Calibration Frequency | 60 Days | 90 Days |
| Analytical Measuring Interval | Serum/Plasma: 0.50 to 5.00 mg/dL | Serum/Plasma: 0.0 – 20.0 mg/dL |
| | Urine: 1.00 to 14.00 mg/dL | Urine: 0.0 – 20.0 mg/dL |
K. Standard/Guidance Document Referenced (if applicable):
- CLSI Guideline, EP07-A2 (7-127): Interference Testing in Clinical Chemistry; Approved Guideline
- CLSI Guideline, EP09-A3 (7-245): Measurement Procedure Comparison And Bias Estimation Using Patient Samples; Approved Guideline - Third Edition
- CLSI Guideline, EP05-A3 (7-251): Evaluation of Precision Performance of Quantitative Measurement Methods: Approved Guideline
- CLSI Guideline, EP06-A (7- 93): Evaluation of the Linearity Of Quantitative Measurement Procedures: A Statistical Approach; Approved Guideline
- CLSI Guideline, EP17-A2 (7-233): Evaluation of Detection Capability for Clinical Laboratory Measurement Procedures
- CLSI Guideline, 25-A (7-235): Evaluation of Stability of In Vitro Diagnostic
{3}
Reagents; Approved Guideline
- CLSI Guideline, EP28-A3 (7-224): Defining, Establishing, And Verifying Reference Intervals In The Clinical Laboratory; Approved Guideline- Third Edition
# L. Test Principle:
The Atellica CH Mg assay is based on the modified xylidyl blue reaction. The reagent was modified to eliminate the use of organic solvents. Magnesium ions react with xylidyl blue in an alkaline medium to form a water-soluble purple-red complex. The increase in absorbance of xylidyl blue at $505 / 694\mathrm{nm}$ is proportional to the concentration of magnesium in the sample. Calcium is excluded from the reaction by complexing with EGTA.
# M. Performance Characteristics (if/when applicable):
# 1. Analytical performance:
# a. Precision/Reproducibility:
Precision testing was performed in accordance with CLSI EP05-A3. Precision was tested in 2 replicates per run, two runs per day for 20 days for a total of 80 results with controls, serum and lithium-heparin plasma pools on one instrument. The data are summarized in the following table.
| Sample Type | N | Mean mg/dL | Within Run | | Within-Lab Precision | |
| --- | --- | --- | --- | --- | --- | --- |
| | | | SD, mg/dL | %CV | SD, mg/dL | %CV |
| Serum | 80 | 0.78 | 0.23 | 3.0 | 0.31 | 3.9 |
| Plasma | 80 | 1.51 | 0.034 | 2.3 | 0.051 | 3.4 |
| Serum QC | 80 | 2.53 | 0.044 | 1.7 | 0.050 | 2.0 |
| Serum | 80 | 4.22 | 0.024 | 0.6 | 0.0477 | 1.1 |
| Urine QC1 | 80 | 4.61 | 0.047 | 0.8 | 0.097 | 2.1 |
| Urine QC2 | 80 | 11.19 | 0.108 | 1.0 | 0.141 | 1.3 |
# b. Linearity/assay reportable range:
Linearity was evaluated with 10 samples which spanned the assay measuring interval for serum specimens (0.42-5.61 mg/dL) and 10 samples which spanned the assay measuring interval for urine specimens (0.65-16.08 mg/dL). Each was prepared by mixing high and low concentration samples across the measurement interval as described in CLSI EP06-A. The high sample was prepared by spiking native serum or urine pools with magnesium acetate. Low pools were created by diluting serum and urine samples with saline solution. Four replicates were measured for each sample. The mean of these replicates was used for the regression analysis, as summarized below.
{4}
| Sample Type | Serum | Urine |
| --- | --- | --- |
| Linear regression equation | 0.999x - 0.022 | Y = 1.001x + 0.020 |
| Correlation Coefficient (r) | 0.999 | 1.000 |
| Mean values of samples tested: | 0.42, 0.75, 1.07, 1.72, 2.37, 3.02, 3.67, 4.32, 4.96, 5.61 | 0.65, 1.61, 2.58, 4.51, 6.43, 8.36, 10.29, 12.22, 14.15, 16.08 |
This study demonstrated linearity across the claimed measuring ranges: 0.50 to $5.00\mathrm{mg / dL}$ for serum and plasma specimens and 1.00 to $14.00\mathrm{mg / dL}$ for urine specimens.
c. Traceability, Stability, Expected values (controls, calibrators, or methods):
Traceability:
The Atellica CH Magnesium calibrators were previously cleared in k050374 and are traceable to the Atomic Absorption reference method, which is calibrated with NIST SRM 929 reference material.
Stability:
Real time stability protocols and acceptance criteria for the Atellica CH Magnesium reagents were reviewed and appear to support a closed vial storage claim of 12 months at $2 - 8\mathrm{C}^{\circ}$ and an onboard shelf life claim of 14 days at $2 - 8\mathrm{C}^{\circ}$ . Stability studies are ongoing.
d. Detection limit:
The Limit of Blank (LoB), Limit of Detection (LoD), and Limit of Quantitation (LoQ) were evaluated in accordance with CLSI EP17-A2. Studies were performed in the following manner.
To calculate the Limit of Blank (LoB), four samples with no analyte were measured in replicates of five for three days using three reagent lots. To estimate the Limit of Detection (LoD), four samples containing low levels of magnesium were measured in replicates of five for three days using three reagent lots. To estimate the Limit of Quantitation (LoQ), four samples containing low levels of magnesium were measured in replicates of five for three days using three reagent lots on one instrument.
The detection limit study results are summarized below:
| Sample Type | Serum, mg/dL | Urine, mg/dL |
| --- | --- | --- |
| LoB | 0.00 | 0.00 |
| LoD | 0.02 | 0.04 |
| LoQ | 0.46 | 0.57 |
{5}
The detection limit studies support the claimed measuring range of $0.50 - 5.0$ mg/dL for serum and plasma samples, and $1.00 - 14.0$ mg/dL for urine samples.
# e. Analytical specificity:
The Atellica CH Magnesium (Mg) reagent cartridge was evaluated for interferences according to CLSI EP7-A2. Potential interfering compounds were spiked into fresh serum or urine sample pools containing either low or high levels of magnesium. Bias was defined as the difference in the results between the control sample (without the interferent) and the test sample (contains the interferent) expressed in percent. The sponsor considered interference as bias exceeding $10\%$ .
The highest concentration of substances tested that did not cause significant interference is summarized in the tables below:
Interference Testing for Serum:
| Substance | Highest concentration tested that did not show significant interference |
| --- | --- |
| Hemoglobin | 500 mg/dL |
| Bilirubin, conjugated | 30 mg/dL |
| Bilirubin, unconjugated | 30 mg/dL |
| Lipemia (Intralipid®) | 500 mg/dL |
| EDTA | 12.5 mg/dL |
| Copper | 0.50 mg/dL |
| Calcium | 20 mg/dL |
| Iron | 0.50 mg/dL |
| Zinc | 0.25 mg/dL |
| Acetaminophen | 200 mg/dL |
| Ibuprofen | 500 mg/dL |
Interference Testing for Urine:
| Substance | Highest concentration tested that did not show significant interference |
| --- | --- |
| 6 N HCl | 0.01% HCl |
| Ascorbate | 50 mg/dL |
| Hemoglobin | 150 mg/dL |
| Calcium | 20 mg/dL |
| Conjugated Bilirubin | 30 mg/dL |
| Unconjugated Bilirubin | 15 mg/dL |
| Copper | 0.5 mg/dL |
| Iron | 0.5 mg/dL |
| Zinc | 0.25 mg/dL |
The sponsor has stated the following limitations in the labeling:
{6}
7
- Do not use hemolyzed samples. Because Mg concentrations are 3 times higher in red blood cells than in serum, hemolyzed samples may give spuriously elevated magnesium results. Bias from hemolysis may vary due to individual sample variations in intracellular magnesium.
- Hemoglobin concentrations above 500 mg/dL may result in a positive bias in serum/plasma specimens. Hemoglobin at 750 mg/dL increases the magnesium result in serum/plasma at 1.60 mg/dL by 13%.
- Unconjugated bilirubin concentrations above 15 mg/dL may result in a positive bias in urine specimens. Unconjugated bilirubin at 30 mg/dL increases the magnesium result in urine at 2.00 mg/dL by 11%.
- EDTA concentrations above 12.5 mg/dL may result in a negative bias in serum/plasma specimens. EDTA at 25 mg/dL decreases the magnesium result in serum/plasma at 1.60 mg/dL and 2.60 mg/dL by -15%.
- Zinc concentrations above 0.25 mg/dL may result in a positive bias in serum/plasma specimens. Zinc at 0.5 mg/dL increases the magnesium result in serum/plasma at 1.60 mg/dL by 19%. Zinc at 0.5 mg/dL increases the magnesium result in serum/plasma at 2.60 mg/dL by 14%.
2. Comparison studies:
a. Method comparison with predicate device:
The comparator device selected for the method comparison study was the Dimension Magnesium Flex Reagent Cartridge. Serum, lithium heparin plasma, and urine samples were tested. The study included native and diluted samples spanning the claimed assay intervals. For serum, 8 samples were diluted and the remainder were native. For lithium heparin plasma, 8 samples were diluted and the remainder were native. For urine, no samples were diluted; all were native. For all tests, native samples made up more than 90% of the sample pool.
The results across the full assay intervals were analyzed using Deming regression. One replicate of each sample was tested and used in the analysis.
| Specimen Type | N | R | Regression Equation | Sample Range Dimension Magnesium Flex (mg/dL) |
| --- | --- | --- | --- | --- |
| Serum | 108 | 0.996 | Y = 0.94x + 0.09 | 0.48-5.16 |
| Lithium Heparin Plasma | 109 | 0.998 | Y = 0.97x + 0.09 | 0.50-5.05 |
| Urine | 100 | 0.998 | Y = 0.96x - 0.06 | 1.10-13.22 |
b. Matrix comparison:
Not applicable. Please see sections M.1.a, M.1.b, M.1.d, M.1.e, and M.1.2.a, demonstrating the performance of the claimed sample types (serum, lithium
{7}
heparin plasma, and urine) for this device.
3. Clinical studies:
Not Applicable.
4. Clinical cut-off:
Not applicable
5. Expected values/Reference range:
Reference intervals are based on literature (see table below). In addition, the reference intervals were verified using 20 healthy adults on the Atellica CH Analyzer in accordance with CLSI Document EP28-A3c. As with all in vitro diagnostic assays, each laboratory should determine its own reference interval for the diagnostic evaluation of patient results. Consider these values as guidance only.
| Group | Specimen type | Reference Interval common unit (SI unit) |
| --- | --- | --- |
| Adults | Serum/plasma^{1} | 1.6 - 2.6 mg/dL (0.66 to 1.07 mmol/L) |
| Adults | Urine^{2} | 24 - 255 mg/24 hour (0.99 - 10.45 mmol/24 hour) |
1. Wu AHB. Tietz Clinical Guide to Laboratory Tests. 4th ed. Philadelphia, PA: WB Saunders Co; 2006:706.
2. Pesce, A.J. and Kaplan, L.A., Methods in Clinical Chemistry, C.V. Mosby Co., St. Louis, 1987.
## N. Proposed Labeling:
The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10.
## O. Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.