THERMO ELECTRON ENZYMATIC CREATININE, MODEL TR35401
K061412 · Thermo Electron Corp. · JFY · Mar 23, 2007 · Clinical Chemistry
Device Facts
Record ID
K061412
Device Name
THERMO ELECTRON ENZYMATIC CREATININE, MODEL TR35401
Applicant
Thermo Electron Corp.
Product Code
JFY · Clinical Chemistry
Decision Date
Mar 23, 2007
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 862.1225
Device Class
Class 2
Indications for Use
The Thermo Electron Enzymatic Creatinine reagent is intended for the in vitro quantitative determination of Creatinine in human serum, plasma or urine. Creatinine determinations are used in the diagnosis and treatment of renal impairment, in monitoring renal dialysis, and as a calculation basis for measuring other urine analytes.
Device Story
Thermo Electron Enzymatic Creatinine reagent is an in vitro diagnostic chemical assay. It measures creatinine levels in human serum, plasma, or urine samples. The reagent facilitates quantitative determination of creatinine, which clinicians use to assess renal function, monitor dialysis patients, and normalize other urine analyte measurements. The device is intended for professional use in clinical laboratory settings. Results are provided to healthcare providers to inform clinical decision-making regarding kidney health and treatment efficacy.
Clinical Evidence
No clinical data provided; bench testing only.
Technological Characteristics
Enzymatic creatinine reagent for in vitro diagnostic use. Operates via enzymatic reaction principle for quantitative analysis. Formulated for use with clinical chemistry analyzers. No specific materials, software, or energy sources described.
Indications for Use
Indicated for in vitro quantitative determination of creatinine in human serum, plasma, or urine to diagnose and treat renal impairment, monitor renal dialysis, and calculate other urine analytes.
Regulatory Classification
Identification
A creatinine test system is a device intended to measure creatinine levels in plasma and urine. Creatinine measurements are used in the diagnosis and treatment of renal diseases, in monitoring renal dialysis, and as a calculation basis for measuring other urine analytes.
K983937 — DUAL VIAL CREATININE REAGENT · Intersect Systems, Inc. · Jan 6, 1999
K040976 — GENCHEM CREATININE REAGENT · Genchem, Inc. · Dec 27, 2004
K971908 — CREA OR CREATININE · Carolina Liquid Chemistries Corp. · Jun 17, 1997
K061944 — POLY-CHEM CREATININE · Polymedco, Inc. · Nov 3, 2006
Submission Summary (Full Text)
{0}
1
# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY AND INSTRUMENT COMBINATION TEMPLATE
A. 510(k) Number:
k061412
B. Purpose for Submission:
New device
C. Measurand:
Creatinine
D. Type of Test:
Quantitative, enzymatic colorimetric
E. Applicant:
Thermo Electron
F. Proprietary and Established Names:
Enzymatic Creatinine Reagent
G. Regulatory Information:
1. Regulation section:
21 CFR 862.1225
2. Classification:
II
3. Product code:
JFY, Enzymatic Method Creatinine
4. Panel:
Chemistry (75)
H. Intended Use:
1. Intended use(s):
See Indications for use.
2. Indication(s) for use:
The Thermo Electron Enzymatic Creatinine Reagent is intended for *in vitro* quantitative determinations of creatinine in human serum, plasma or urine. Creatinine determinations are used in the diagnosis and treatment of renal impairment, in monitoring renal dialysis, and as a calculation basis for measuring other urine analytes.
3. Special conditions for use statement(s):
Prescription use
4. Special instrument requirements:
Automated clinical chemistry analyzers. Performance characteristics were established on the Roche Hitachi 911 analyzer.
I. Device Description:
The Thermo Electron Enzymatic Creatinine test is a two reagent system for use on automated analyzers. Both reagents contain enzymes with stabilizers and/or buffers. The composition of the two ready-to-use liquid reagents is as follows:
Reagent 1: buffer, creatinine amidinohydrolase, sarcosine oxidase, catalase, ascorbate oxidase, TOPS (N-ethyl-sulfopropyl-m-toludine).
Reagent 2: buffer, creatinine amidohydrolase, peroxidase and 4-aminoantipyrine.
J. Substantial Equivalence Information:
1. Predicate device name(s):
Roche Diagnostics Creatinine Plus
2. Predicate 510(k) number(s):
k003261
{1}
3. Comparison with predicate:
| Similarities and Differences | | |
| --- | --- | --- |
| Item | New device | Predicate |
| Method | Enzymatic, colorimetric | Same |
| Storage | Store at 2-8°C | Same |
| Measuring Range | 0.2 – 30 mg/dL for serum & plasma 2.14 - 150 mg/dL for urine | 0.03-30 mg/dL for serum/plasma and 0.3 – 600 mg/dL for urine |
| Sample type | Serum, plasma, urine | Same |
# K. Standard/Guidance Document Referenced (if applicable):
CLSI - Evaluation of Precision Performance of Clinical Chemistry Devices - EP05-A2
CLSI - Method Comparison and Bias Estimation Using Patient Samples - EP09-A2
CLSI - Interference Testing in Clinical Chemistry - EP07-A2
# L. Test Principle:
The creatinine present in samples is acted upon by creatinine amidohydrolase, creatinine amidinohydrolase, sarcosine oxidase, and peroxidase. In the final reaction sequence, the formation of quinoneimine dye product results in an increase in absorbance at $550\mathrm{nm}$ which is directly proportional to the creatinine concentration.
# M. Performance Characteristics (if/when applicable):
1. Analytical performance:
a. Precision/Reproducibility:
Imprecision was evaluated using CLSI Document EP5-A2 as a guideline.
Samples: 3 serum controls and 2 urine controls
Analyzer: Roche Hitachi 911
Reagent lot: one
Calibrator lot: one
Operator: one
Number of runs per day: 2
Number of replicates: 2
Number of days: 20
The results are summarized below:
| Sample | Mean mg/dL | Number | Total | | Within-run | |
| --- | --- | --- | --- | --- | --- | --- |
| | | | SD | %CV | SD | %CV |
| Serum 1 | 0.9 | 80 | 0.03 | 3.7 | 0.2 | 2.5 |
| Serum 2 | 1.6 | 80 | 0.04 | 3.3 | 0.03 | 2.8 |
| Serum 3 | 5.3 | 80 | 0.07 | 1.3 | 0.04 | 3.3 |
| Urine 1 | 67.7 | 80 | 1.74 | 2.6 | 0.76 | 1.1 |
| Urine 2 | 145.3 | 80 | 3.3 | 2.7 | 0.94 | 0.6 |
{2}
b. Linearity/assay reportable range:
The claimed measuring range for serum/plasma is 0.2 to 30 mg/dL and for urine is 2.14 to 150 mg/dL.
Linearity studies were performed using CLSI EP6-A as a guideline. Samples were created by diluting a creatinine standard to concentrations within the claimed range. The concentrations tested in the study were 0, 20, 40, 60, 80, 100, 120, 140, 160, 180, and 200 mg/dL creatinine. The sample with a concentration of 200 mg/dL was excluded from the calculations. A slope close to 1 and correlation coefficient of 0.99 resulted. The results demonstrate that with this assay creatinine is linear across the claimed measuring ranges.
c. Traceability, Stability, Expected values (controls, calibrators, or methods):
The calibrator used to calibrate he device for the performance evaluations in this 510(k) was cleared under k022108.
The package insert recommends that an aqueous standard or serum based calibrator with an assigned value traceable to a primary standard (e.g., NIST or IRMM) be used with this reagent.
The reagent is stable until the expiration date when stored at 2 – 8°C.
d. Detection limit:
A blank sample of serasub, a serum matrix solution, and 10 serum samples containing low levels of creatinine were assayed on a Roche Hitachi 911 clinical chemistry analyzer. Ten (10) replicate assays were performed on each sample including the blank. This was repeated on two consecutive days. The lowest sample dilution which consistently met with a coefficient of variation percentage (%CV) below 20% for serum samples was 0.2 mg/dL. The lowest sample dilution which consistently met with a coefficient of variation percentage (%CV) below 20% for urine samples was 2.14 mg/dL. The results demonstrate that with this assay the lower end of the measuring range for creatinine is 0.2 mg/dL in serum or plasma and 2.14 mg/dL to 150 mg/dL in urine.
e. Analytical specificity:
Interference studies were performed using CLSI EP7-A as a guideline.
Interfering compounds were evaluated by spiking various concentrations of the interferents into a creatinine sample and comparing the results of the interferent spiked samples with an unspiked creatinine reference sample. Non-interference was defined as a recovery of 90% - 110% of the unspiked reference sample. The highest concentration of each compound that did not interfere is shown below.
| Interfering Compound | Maximum Concentration Tested |
| --- | --- |
| Hemoglobin | 1000 mg/dL |
| Bilirubin | 60 mg/dL |
| Conjugated Bilirubin | 33 mg/dL |
{3}
| Interfering Compound | Maximum Concentration Tested |
| --- | --- |
| Lipemia | 1490 mg/dL |
| Ascorbic Acid | 82 mg/dL |
| Creatine | 10 mg/dL |
| B-hydroxybuturate | 126 mg/dL |
| Cephatholin | 100 mg/dL |
| Cefotaxime | 100 mg/dL |
| Acetoacetate | 108 mg/dL |
| Proline | 70 mg/dL |
f. Assay cut-off:
Not Applicable
2. Comparison studies:
a. Method comparison with predicate device:
The method comparison was performed on the Roche Hitachi 911 using CLSI EP9-A2 as a guideline.
Serum
The sponsor compared 107 serum samples ranging from 0.5 to 28.5 mg/dL creatinine to the predicate device. Linear regression produced a slope of 00.951, a y-intercept of -0.1, and a correlation coefficient of 0.999.
Urine
The sponsor compared 107 urine samples ranging from 9.8 to 136.2 mg/dL creatinine to the predicate device. Linear regression produced a slope of 0.992, a y-intercept of -0.4, and a correlation coefficient of 0.999.
b. Matrix comparison:
A method comparison was performed using the Thermo Electron Enzymatic Creatinine method with 64 matched serum-EDTA plasma samples ranging in concentration from 0.49 to 28.48 mg/dL creatinine (in serum). Linear regression produced a slope of 0.953, a y-intercept of -0.022, and a correlation coefficient of 1.00
3. Clinical studies:
a. Clinical Sensitivity:
Not Applicable
b. Clinical specificity:
Not Applicable
c. Other clinical supportive data (when a. and b. are not applicable):
{4}
Not Applicable
4. Clinical cut-off:
Not Applicable
5. Expected values/Reference range:
The sponsor cites the following reference ranges for serum/plasma from Tietz Textbook of Clinical Chemistry and Molecular Diagnostics, Fourth Edition 2006 56:2251-2318:
Adult Male: 0.62 - 1.10 mg/dL
Adult Female: 0.45 - 0.75 mg/dL
The sponsor cites the following reference ranges for urine from Kaplan LA, Pesce AJ, Clinical Chemistry Theory, Analysis and Correlation, CV Mosby Company 1984, 59:1251-1252:
Adult Male: 40 - 278 mg/dL
Adult Female: 29 - 226 mg/dL
Q. Proposed Labeling:
The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10.
R. Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
5
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.