K991532 · Metrika, Inc. · DHA · Sep 17, 1999 · Clinical Chemistry
Device Facts
Record ID
K991532
Device Name
DRX QT. HCG
Applicant
Metrika, Inc.
Product Code
DHA · Clinical Chemistry
Decision Date
Sep 17, 1999
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 862.1155
Device Class
Class 2
Indications for Use
The DRx® Ot. hCG is a rapid, single-use, point-of-care device for the quantitative measurement of human chorionic gonadotropin (total ßhCG) in capillary and venous whole blood. The device is for professional use as an aid in the assessment of pregnancy status.
Device Story
DRx® Qt. hCG is a single-use, disposable, four-channel reflectance photometer for point-of-care pregnancy assessment. Device integrates dry reagent chemistry strips within a sealed plastic case. Input: capillary or venous whole blood sample applied to a separation pad; red blood cells immobilized; plasma migrates via capillary action to assay strips. Principle: lateral flow immunoassay; anti-hCG antibodies conjugated to blue microparticles bind hCG; blue intensity measured by LEDs and silicon photodetectors before and after sample addition. Microprocessor calculates hCG concentration (mIU/mL) using stored calibration parameters; results displayed on LCD. Professional use only; results interpreted alongside medical history. Benefits: eliminates off-line dilutions and centrifugation required by predicate systems.
Clinical Evidence
Clinical accuracy evaluated by comparing DRx® Qt. hCG to Stratus® ßhCG predicate. Study included 96 venous whole blood samples and 63 matched capillary (fingerstick) samples. Results: venous vs. predicate (r=0.98, y=0.92x+20); fingerstick vs. predicate (r=0.96, y=1.05x+8); venous vs. fingerstick (r=0.99, y=1.06x-6). Precision evaluated at three sites using 3-level whole blood panel; CVs ranged from 8.3% to 15.1%.
Technological Characteristics
Four-channel reflectance photometer; includes internal reference channel, dry reagent chemistry strips, microprocessor, LEDs, silicon photodetectors, and LCD. Analyte: hCG (WHO 3rd IS 75/537). Reportable range: 5–10,000 mIU/mL. Battery-powered. Integrated whole blood separation chemistry pad. No external connectivity mentioned.
Indications for Use
Indicated for professional use as an aid in the assessment of pregnancy status by quantitatively measuring total ßhCG in capillary and venous whole blood samples.
Regulatory Classification
Identification
A human chorionic gonadotropin (HCG) test system is a device intended for the early detection of pregnancy is intended to measure HCG, a placental hormone, in plasma or urine. A human chorionic goadotropin (HCG) test system is a device intended for any uses other than early detection of pregnancy (such as an aid in the diagnosis, prognosis, and management of treatment of persons with certain tumors or carcinomas) is intended to measure HCG, a placental hormone, in plasma or urine.
{0}------------------------------------------------
# M = T R I K A
Metrika, Inc. 510 Oakmead Parkway Sunnyvale, CA 94086 main 408 524 2255 fax 408 524 2252
# 510(k) SUMMARY
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.
The assigned 510(k) number is K991532.
| 807.92 (a)(1): Name: | Metrika, Inc. |
|----------------------|---------------------|
| Address: | 510 Oakmead Parkway |
| | Sunnyvale, CA 94086 |
| Phone: | (408) 524-2255 |
| FAX: | (408) 524-2252 |
| Contact: | Stephen J. Hardt |
### 807.92 (a)(2): Device name- trade name and common name, and classification
| Trade name (Device): | DRx® Qt. hCG |
|--------------------------|------------------------------------|
| Common Name (Device): | whole blood hCG immunoassay system |
| Classification (Device): | 21 CFR 862.1155 |
| Trade name (Control): | DRx® Qt. hCG Control Set |
|------------------------|--------------------------|
| Common Name (Control): | assayed hCG control |
Classification (Control): 21 CFR 862.1660
### 807.92 (a)(3): Identification of the legally marketed predicate device
(Device)
DRx® Qt. hCG is substantially equivalent to the commercially available predicate device, Stratus® ßhCG (Dade Behring Inc., Miami, FL). DRx Ot. hCG reports a single result without additional off-line dilutions or whole blood specimen manipulations. Stratus® requires dilutions of samples greater than 1000 mIU/mL and sample centrifugation (for plasma) prior to evaluations in order to report the final result.
(Control)
Dade® Immunoassay Controls Comprehensive Tri-Level Name: Manufacturer: Dade International, Inc.
{1}------------------------------------------------
#### 807.92 (a)(4): Device Description
#### (Device)
The DRx® Ot. hCG device is a single-use, disposable four-channel reflectance photometer integrated with dry reagent chemistry strips and contained within a sealed plastic case. Each unit consists of:
. an optics subassembly that also supports the reagent strips
- a printed circuit board that includes a microprocessor, analog . electronics, battery clips, batteries, silicon diode photodetectors, and j light emitting diodes (LEDs)
- a plastic molded part ("spider") that supports the liquid crystal display ● (LCD), the sample pad and the auto start leads, and also provides optical shielding
- two reagent strips .
- whole blood separation chemistry pad ●
- a desiccant
- liquid crystal display (LCD)
Both test strips are lateral flow immunoassays for measuring hCG between zero and 10,000 mIU/mL. In both strips, a blue color is measured in discrete test zones. The test zones are the areas where the specific reactions occur and concentrations are measured.
The whole blood separation mechanism occurs when the whole blood ~ sample reaches the sample pad and immobilizes the red blood cells. As a result, the plasma freely migrates by capillary action onto the assay strips to complete the final reactions. The hCG reaction proceeds as follows: anti-hCG antibody, conjugated to blue microparticles, migrate across the strip upon the addition of whole blood sample. The amount of blue microparticles captured on the test zones is proportional to the amount of hCG in the sample.
The LEDs and silicon photodetectors compare the reflectances of the color intensities before and after the sample addition. Based on the calibration parameters stored in memory, the numerical concentrations of hCG is calculated. The final assay results are expressed in mIU/mL hCG.
### (Control)
The DRx® Qt. hCG Control Set is a bovine plasma-based, liquid, 2-level control set to be used in quality control procedures with the DRx® Qt. hCG test. The assayed control set is used for evaluating precision and systematic analytical deviations that may arise from device variations.
{2}------------------------------------------------
### 807.92 (a)(5): Intended use
(Device)
The DRx® Ot. hCG is a rapid, single-use, point-of-care device for the quantitative measurement of human chorionic gonadotropin (total ßhCG) in capillary and venous whole blood. The device is for professional use as an aid in the assessment of pregnancy status.
### (Control)
The DRx® Qt. hCG Control Set is a two level, assayed, liquid control containing human chorionic gonadotropin for the evaluation of precision and accuracy of the DRx® Qt. hCG device.
### 807.92 (a)(6): Technological Similarities and Differences to the Predicate
| CHARACTERISTIC | DRx® Qt. hCG | Stratus® βhCG<br>K850570 |
|-----------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Intended Use | DRx® Qt. hCG is a rapid, single-<br>use, point-of-care device for the<br>quantitative measurement of<br>human chorionic gonadotropin<br>(total βhCG) in capillary and<br>venous whole blood samples.<br>The device is for professional use<br>as an aid in the assessment of<br>pregnancy status. | The Stratus βhCG Fluorometric<br>Enzyme Immunoassay is for<br>quantitative determination of<br>total beta human chorionic<br>gonadotropin (βhCG) levels in<br>serum and plasma for the<br>detection of pregnancy. |
| Analyte(s) | hCG (mIU/mL)<br>WHO 3rd IS 75/537 | hCG (mIU/mL)<br>WHO 3rd IS 75/537 |
| Testing Matrix<br>(Specimen Type) | Plasma<br>(after integrated red blood cell<br>separation) | Serum and plasma |
| Testing Environment | professional use | professional use |
| Risk to Patient | not a sole discriminant; hCG<br>results are interpreted along with<br>medical histories and other<br>diagnostic procedures. | not a sole discriminant; hCG<br>results are interpreted along with<br>medical histories and other<br>diagnostic procedures. |
## Similarities Between DRx® Ot. hCG and Stratus® BhCG
{3}------------------------------------------------
807.92 (a)(6): Technological Similarities and Differences to the Predicate (continued)
| CHARACTERISTIC | DRx® Qt. hCG | Stratus® βhCG K850570 |
|----------------------------------|---------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------|
| Methodology/Testing<br>Platforms | DRx® Qt. hCG is a four channel, reflectance photometer (plus an internal reference channel) with integral dry chemistry strips. | Stratus® βhCG is a Fluorometric Enzyme Immunoassay. |
| Reportable Ranges | 5 to 10,000 mIU/mL | 2 to 1000 mIU/mL |
| Dilutions | Not required | Required for samples greater than 1000mIU/mL hCG |
| Specimen Processing | Not required<br>(straight addition of whole blood specimen into device for testing) | Required<br>(centrifugation of whole blood specimen to obtain serum or plasma before testing) |
Differences Between DRx® Qt. hCG and Stratus® ßhCG
:
{4}------------------------------------------------
807.92 (a)(6): Technological Similarities and Differences to the Predicate (continued)
| | and Dade® Immunoassay Controls Comprehensive Tri-Level | |
|---------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------|
| CHARACTERISTIC | DRx® Qt. hCG Control Set | Dade® Immunoassay<br>Controls Comprehensive<br>Tri-Level<br>K860906 |
| Intended Use | The DRx® Qt. hCG Control Set is<br>a two level, assayed, liquid control<br>containing human chorionic<br>gonadotropin for the evaluation of<br>precision and accuracy of the<br>DRx® Qt. hCG device. | Dade® Immunoassay Controls<br>Comprehensive Tri-Level are<br>intended to assist in<br>monitoring accuracy and<br>precision in clinical assays. |
| Testing Environment | professional use | professional use |
| Test Method | quantitative | quantitative |
# Similarities Between DRx® Qt. hCG Control Set d Dade® Immunoassay Controls Comprehensive Tri-Level
# Differences Between DRx® Qt. hCG Control Set
and Dade® Immunoassay Controls Comprehensive Tri-Level
| CHARACTERISTIC | DRx ® Qt. hCG Control Set | Dade® Immunoassay Controls Comprehensive<br>Tri-Level<br>K860906 |
|----------------|--------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------|
| Analyte(s) | hCG | (Various constituents), e.g.<br>Acetaminophen Lidocaine<br>Digoxin TSH<br>Vitamin B12 hGH<br>FSH PSA<br>hCG hLH, etc. |
| Matrix | Liquid, bovine plasma-based matrix with spiked human-based analyte | Lyophilized, human source materials |
The differences in the two testing platforms for either devices or control materials do not raise new issues of safety and effectiveness.
ﮐﮯ ﻣ
{5}------------------------------------------------
### 807.92 (b)(1): Brief Description of Nonclinical Data (Device)
Studies were conducted that evaluated hCG for analytical sensitivity. analytical specificity, and linearity.
Sensitivity studies demonstrated that the device is sensitive to 5 mIU/mL hCG.
DRx® Qt. hCG results were not affected by the addition of potential interferents such as biological compounds and therapeutic agents found routinely in blood. In addition, DRx® Qt. hCG results were not affected by potential cross/reactants similar to alpha hCG homology such as hLH, hFSH, and hTSH. The device is specific for total beta hCG subunit present as intact hCG or free beta subunit.
Linearity studies confirmed dynamic ranges of 5 to 10,000 mIU/mL for hCG.
### 807.92 (b)(2): Brief Description of Clinical Data (Device)
Accuracy was evaluated by comparative testing between DRx Ot. hCG and Stratus® BhCG as the predicate method. Ninety-six (96) venous whole blood samples were assayed by DRx® Ot. hCG among four "POL" clinical sites and in-house, and by the predicate method. In addition, 63 matched whole blood capillary (fingerstick) samples were collected at the These fingerstick samples were compared to venous clinical sites. samples tested in DRx® Qt. hCG assay to demonstrate equivalence of sample. The data are summarized below (predicate method = X-axis).
| | n samples | Linear Regression<br>Equation | Correlation<br>Coefficient |
|--------------------------------------------|-----------|-------------------------------|----------------------------|
| hCG<br>(venous compared to predicate) | 96 | $y = 0.92x + 20$ | 0.98 |
| hCG<br>(fingerstick compared to predicate) | 63 | $y = 1.05x + 8$ | 0.96 |
| hCG<br>(DRx® venous vs. DRx® fingerstick) | 63 | $y = 1.06x - 6$ | 0.99 |
CLINICAL LINEAR REGRESSION DATA
Precision was evaluated at three sites by repeat testing of a 3-level whole blood precision panel (low, middle, and high hCG levels) on a single day. The coefficients of variation ranged from 8.3% to 15.1% across the three sites.
{6}------------------------------------------------
## 807.92 (b)(3): Conclusions from Nonclinical and Clinical Testing (Device)
Studies demonstrated substantial equivalence between DRx® Qt. hCG and an existing product already being marketed in terms of accuracy and precision. Sensitivity and linearity studies confirmed the assay's reportable range, and interference substance studies confirmed the assay's specificity.
{7}------------------------------------------------
Image /page/7/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the top half of the circle. Inside the circle is a stylized image of three human profiles facing to the right, resembling a bird-like shape.
SEP 1 7 1999
Food and Drug Administration 2098 Gaither Road Rockville MD 20850
Mr. Ya-Chen Hsu Clinical Research Associate Metrika, Inc. 510 Oakmead Parkway Sunnyvale, California 94086
Re: K991532
> Trade Name: DRx® Qt. hCG Regulatory Class: II Product code: DHA Regulatory Class: I reserved Product Code: JJX Dated: July 20, 1999 Received: July 21, 1999
Dear Mr. Hsu:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (OS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions, Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
{8}------------------------------------------------
#### Page 2
Under the Clinical Laboratory Improvement Amendments of 1988 (CLIA-88), this device may require a CLIA complexity categorization. To determine if it does, you should contact the Centers for Disease Control and Prevention (CDC) at (770) 488-7655.
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4588. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification"(21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597, or at its internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html".
Sincerely yours,
Steven Autman
Steven I. Gutman, M.D, M.B.A. Director Division of Clinical Laboratory Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{9}------------------------------------------------
510(k) Number: K991532
Device Name: DRx® Qt. hCG
#### INDICATIONS FOR USE
DRx® Qt. hCG is a rapid, single-use, point-of-care device for the quantitative measurement of human chorionic gonadotropin (total ßhCG) in capillary and venous whole blood samples. The device is for professional use as an aid in the assessment of pregnancy status.
Device Name: DRx® Qt. hCG Control Set
### INDICATIONS FOR USE
The DRx® Qt. hCG Control Set is a two level, assayed, liquid control containing human chorionic gonadotropin for the evaluation of precision and accuracy of the DRx Ot. hCG device.
Jean Cooper
(Division Sign-off)
Division of Clinical Laboratory Devi
510(k) Number K991532
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use
(Per 21 CFR 801.109)
✓
OR
Over-The-Counter Use (Optional Format 1-2-96)
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.