ST AIA-PACK ACTH, AND ST AIA-PACK ACTH CALIBRATOR SET MODEL 025221 AND 025321

K111335 · Tosoh Bioscience, Inc. · CKG · Dec 1, 2011 · Clinical Chemistry

Device Facts

Record IDK111335
Device NameST AIA-PACK ACTH, AND ST AIA-PACK ACTH CALIBRATOR SET MODEL 025221 AND 025321
ApplicantTosoh Bioscience, Inc.
Product CodeCKG · Clinical Chemistry
Decision DateDec 1, 2011
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 862.1025
Device ClassClass 2

Indications for Use

ST AIA-PACK ACTH is designed for IN VITRO DIAGNOSTIC USE ONLY for the quantitative measurement of adrenocorticotropic hormone (ACTH) in human EDTA plasma. Measurements of ACTH are useful in the differential diagnosis and treatment of certain disorders of the adrenal glands such as Cushing's syndrome, adrenocortical insufficiency, and the ectopic ACTH syndrome. ST AIA-PACK ACTH Calibrator Set is intended for IN VITRO DIAGNOSTIC USE ONLY for the calibration of the ST AIA-PACK ACTH assay on Tosoh AIA Systems Analyzers. The AIA-PACK ACTH Control Set is intended for IN VITRO DIAGNOSTIC USE ONLY for performing quality control procedures with the ST AIA-PACK ACTH Assay.

Device Story

ST AIA-PACK ACTH is a two-site immunoenzymometric assay performed in test cups on Tosoh AIA System Analyzers. Input: human EDTA plasma sample. Principle: ACTH binds to anti-ACTH goat polyclonal antibody on magnetic beads and enzyme-labeled anti-ACTH goat polyclonal antibody. After washing, beads are incubated with fluorogenic substrate 4-methylumbelliferyl phosphate (4MUP); alkaline phosphatase converts 4MUP to 4MU. Detector measures fluorescent energy at 448 nm (excited at 365 nm). Output: ACTH concentration calculated via standard curve. Used in clinical laboratories by technicians. Results assist physicians in diagnosing and managing adrenal/pituitary disorders. Benefits: provides quantitative diagnostic data for endocrine conditions.

Clinical Evidence

Bench testing only. Precision evaluated per CLSI EP5-A2 (n=80 determinants per specimen). Linearity evaluated per CLSI EP6 (range 2-2000 pg/mL). Method comparison (n=160) against predicate showed R=0.993, slope 1.10. Reference range study (n=121) established 95% interval of 7.4-64.3 pg/mL. No clinical diagnostic sensitivity/specificity studies performed.

Technological Characteristics

Two-site immunoenzymometric sandwich assay. Materials: magnetic beads, anti-ACTH goat polyclonal antibodies, alkaline phosphatase enzyme label, 4MUP fluorogenic substrate. Detection: fluorescence (365 nm excitation/448 nm emission). Form factor: test cups for automated AIA System Analyzers. Connectivity: standalone analyzer system. Standards: CLSI EP5-A2, EP6-A, EP9-A2, EP17-A, C28-A3.

Indications for Use

Indicated for quantitative measurement of ACTH in human EDTA plasma to assist in differential diagnosis and treatment of adrenal gland disorders, including Cushing's syndrome, adrenocortical insufficiency, and ectopic ACTH syndrome.

Regulatory Classification

Identification

An adrenocorticotropic hormone (ACTH) test system is a device intended to measure adrenocorticotropic hormone in plasma and serum. ACTH measurements are used in the differential diagnosis and treatment of certain disorders of the adrenal glands such as Cushing's syndrome, adrenocortical insufficiency, and the ectopic ACTH syndrome.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0} 1 # 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY ONLY TEMPLATE A. 510(k) Number: k111335 B. Purpose for Submission: New assay C. Measurand: Adrenocorticotropic hormone D. Type of Test: Quantitative, Automated Fluorescence Immunoenzymatic Assay E. Applicant: Tosoh Corporation F. Proprietary and Established Names: ST AIA-PACK ACTH; ST AIA-PACK ACTH Calibrator Set; ST AIA-PACK ACTH Control G. Regulatory Information: 1. Regulation section: 862.1025 – Adrenocorticotropic hormone (ACTH) test system 862.1150 – Calibrator 862.1660 – Quality control material 2. Classification: Class II, II, and I, reserved 3. Product code: CKG, JIT, JJX {1} 4. Panel: Chemistry (75) H. Intended Use: 1. Intended use(s): See indications for use. 2. Indication(s) for use: ST AIA-PACK ACTH is designed for IN VITRO DIAGNOSTIC USE ONLY for the quantitative measurement of Adrenocorticotropic hormone (ACTH) in human EDTA plasma on Tosoh AIA System Analyzers. Plasma ACTH measurements are useful in the differential diagnosis and treatment of certain disorders of the adrenal glands such as Cushing's syndrome, adrenocortical insufficiency, and ectopic ACTH syndrome. ST AIA-PACK ACTH Calibrator Set is intended for IN VITRO DIAGNOSTIC USE ONLY for the calibration of the ST AIA-PACK ACTH assay on Tosoh AIA Systems Analyzers. ST AIA-PACK ACTH Control Set is designed for IN VITRO DIAGNOSTIC USE ONLY for performing quality control procedures with the ST AIA-PACK ACTH assay on Tosoh AIA System Analyzers 3. Special conditions for use statement(s): For prescription use only. 4. Special instrument requirements: Tosoh AIA 2000 Analyzer. I. Device Description: Each kit contains 5 trays and 20 test cups (ST AIA-PACK ACTH Test Cup). (See principles of assay section for the specific components and their function it the test). The following materials are also required to perform Adrenocorticotropic hormone analysis using the ST AIA-PACK ACTH on the TOSOH AIA System Analyzer. They are sold separately from the test cup by Tosoh, and are listed under "materials required but not provided". All performance data shown below was obtained using all these materials. Package inserts were included in the 510(k) for all these separately sold materials. 2 {2} Substrate set II: 4-methylumbelliferyl phosphate, stabilizers and 0.01% sodium azide as a preservative (lyophilized). Wash concentrate set - Buffer solution with detergent and bacteriostatic agent. Diluent Concentrate Set - Buffer solution with detergent and 0.5% sodium azide as a preservative. Diluting solution - BSA with no detectable ACTH plus preservative. AIA-PACK ACTH Calibrator contains BSA with the assigned concentration of ACTH plus preservative. Target concentrations are: 0 pg/mL 15 pg/mL (approx.) 50 pg/mL (approx.) 300 pg/mL (approx.) 800 pg/mL (approx.) 2200 pg/mL (approx.) AIA PACK ACTH Control Set contains BSA with approximately 50 pg/mL ACTH (level 1) and approximately 300 pg/mL ACTH (level 2). J. Substantial Equivalence Information: 1. Predicate device name(s): Roche Diagnostics Elecsys ACTH 2. Predicate 510(k) number(s): k060585 3. Comparison with predicate: The intended uses are the same. The technology is similar. A table comparing the two is shown below: {3} | Item | Device | Predicate | | --- | --- | --- | | ST AIA-PACK ACTH | | Elecsys ACTH | | Intended Use/Indications for use | Device is designed for IN VITRO DIAGNOSTIC USE ONLY for the quantitative measurement of Adrenocorticotropic hormone (ACTH) in human EDTA plasma on Tosoh AIA System Analyzers. ACTH measurements are useful in the differential diagnosis and treatment of certain disorders of the adrenal glands such as Cushing’s syndrome, adrenocortical insufficiency, and ectopic ACTH syndrome. | Same | | Intended Use-Calibrators | ST AIA-PACK ACTH Calibrator Set is intended for the calibration of the ACTH assay. | Same | | Intended Use - Controls | The AIA-PACK ACTH CONTROL SET is intended for performing quality control procedures with the ACTH Assay. | Same | | Assay Protocol | Sandwich assay | Same | | Detection Method | Fluorescence | Electrochemiluminescent | | Sample Type | Human EDTA Plasma | Same | | Assay Low | 2.0 pg/mL | 1.0 pg/mL | | Assay High | 2000 pg/mL | 2000 pg/mL | | Reference Range observed in the sponsor’s study | 7.4-64.3 pg/mL | 7.2-63.3 pg/mL | | Number of control levels | Two levels of lyophilized control | Two levels of lyophilized control | {4} | Calibrator | ST AIA-PACK Calibrator Set 6-Point | ACTH CalSet 2-Point | | --- | --- | --- | | Traceability / Standardization | ACTH (Human, 1-39) Bachem AG: code. H-1160; gravimetric preparation | Standardized gravimetrically with synthetic ACTH produced by Roche | | Calibrator/Control Base Matrix | MOPSO Buffer with 5% Bovine Serum Albumin | Equine serum | # K. Standard/Guidance Document Referenced (if applicable): CLSI guidelines: C28-A3, Define and Determine Reference Intervals in the Clinical Laboratory; EP5-A2, Evaluation of Precision Performance of Quantitative Measurement Methods; EP6-A Evaluation of the Linearity of Quantitative Measurement Procedures; EP9-A2 Method Comparison and Bias Estimation Using Patient Samples; EP17-A Protocols for Determination of Limits of Detection and Limits of Quantitation. # L. Test Principle: The ST AIA-PACK ACTH is a two-site enzyme immunoassay which is performed in the ST AIA-PACK ACTH test cups. ACTH present in the test sample is bound with goat polyclonal antibody immobilized on magnetic beads and enzyme-labeled polyclonal antibody. The magnetic beads are washed to remove unbound enzyme-labeled polyclonal antibody and are then incubated with a fluorogenic substrate, 4-methylumbelliferyl phosphate (4MUP). The enzyme alkaline phosphatase causes oxidation of 4MUP to 4MU. 4MU is excited at $365~\mathrm{nm}$ and comes to ground state at $448~\mathrm{nm}$ releasing fluorescent energy. The amount of fluorescent energy is measured by the detector. The amount of enzyme-labeled polyclonal antibody that binds to the beads is directly proportional to the ACTH concentration in the test sample. A standard curve is constructed, and unknown sample concentrations are calculated using this curve. # M. Performance Characteristics (if/when applicable): Performance in this $510(\mathrm{k})$ was evaluated on the Tosoh AIA 2000 (Model 022100.) # 1. Analytical performance: # a. Precision/Reproducibility: The precision study followed the CLSI protocol, Evaluation of Precision Performance of Quantitative Measurement Methods (EP5-A2). The precision study for the ST AIA-PACK ACTH assay was evaluated utilizing three AIA-2000 analyzers and 3 different lots of reagents. Precision was assessed by assaying three levels of unaltered EDTA plasma specimens. Estimates of total and within {5} run precision were obtained from measurements of 2 replicates in a single run, 2 times a day for 20 non-consecutive days. This equaled to a total of 40 runs and 80 determinants. One calibration curve (obtained on the first day) was used throughout. The following results were obtained when tested with three unaltered plasma samples using three sets of reagents. A result of one reagent lot is shown below. Similar results were obtained with the two additional lots tested. Within-run precision | Specimen | Reagent Set # 1 | | | | --- | --- | --- | --- | | | Mean (pg/mL) | Pooled SD | CV % | | EDTA Plasma-A | 37.8 | 1.2 | 3.1 | | EDTA Plasma-B | 223.7 | 4.8 | 2.1 | | EDTA Plasma-C | 709.2 | 10.9 | 1.5 | Total precision: | Specimen | Reagent Set # 1 | | | | --- | --- | --- | --- | | | Mean (pg/mL) | Pooled SD | CV % | | EDTA Plasma-A | 37.8 | 1.2 | 3.3 | | EDTA Plasma-B | 223.7 | 5.7 | 2.5 | | EDTA Plasma-C | 709.2 | 15.6 | 2.2 | Also see Detection Limit section below regarding precision at lower levels. ## b. Linearity/assay reportable range: The sponsor's claimed reportable range is 2-2000 pg/mL. To evaluate linearity, high and low samples (EDTA plasma) were serially diluted and assayed on the new device. The high samples were prepared by spiking a plasma sample and the low samples were prepared by diluting a low concentration sample with the AIA diluting solution. For each level tested, the average of four replicates was determined. The recoveries relative to expected concentrations based on dilution of the high and low samples, and the mean and SD's of the replicates, are listed below. (In additions see the Detection {6} Limit Section for additional recovery determinations near the claimed assay lower limits.) | Expected Value (pg/mL) | Observed Mean (n=4) | Percent observed/expected | CV (%) (n=4) | | --- | --- | --- | --- | | 1.9 | 1.9 | 100.0 | 6.2 | | 38.1 | 35.9 | 94.2 | 2.7 | | 74.3 | 74.8 | 100.6 | 1.9 | | 146.7 | 152.0 | 103.6 | 2.5 | | 243.2 | 242.2 | 99.6 | 1.9 | | 485.6 | 481.6 | 99.2 | 1.0 | | 725.9 | 738.3 | 101.7 | 1.5 | | 967.3 | 983.3 | 101.7 | 2.7 | | 1208.6 | 1234.5 | 102.1 | 2.1 | | 1450.0 | 1440.2 | 99.3 | 0.7 | | 1691.3 | 1696.9 | 100.3 | 1.6 | | 1932.7 | 1946.7 | 100.7 | 2.6 | | 2174.0 | 2145.2 | 98.7 | 2.0 | The equation determined by linear regression was: $\mathrm{Y} = 0.998\mathrm{X} + 4.5$ , $\mathrm{R2} = 0.999$ . In addition the data was evaluated according to CLSI guideline EP6 and the sponsor's criteria for linearity. High dose hook was evaluated by spiking samples up to 1,000,000 pg/mL. The measured concentration read as $>$ high limit, (although the raw signal decreased at concentrations $>100,000~\mathrm{pg / mL}$ . Dilutions of high spiked samples, 2000 to 20,000 starting concentrations. No significant bias (relative to expected) was observed. c. Traceability, Stability, Expected values (controls, calibrators, or methods): ## Calibrators: Value assignment and traceability: The device is traceable to primary reference material containing commercially available purified ACTH (Human, 1-39), which is value assigned using another RIA test. Values were further validated by replicate measurements on multiple instruments for specific lots. Calibrator values are as follows: {7} | Target Values (pg/mL) | Estimated uncertainty in assigned values (pg/mL)^{1} | | --- | --- | | 0 | 0.1 | | 15 | 1.7 | | 50 | 5.4 | | 300 | 31 | | 800 | 81 | | 2700 | 215 | Real time storage: Expiration of ST AIA-PACK ACTH CALIBRATOR SET was determined by using three different lots of CALIBRATOR SET, one AIA-2000 analyzer and three reagent sets. Reagents were stored at 8 - 12 degree C. The system was calibrated using the stored calibrator set on the day of assay. Each calibrator was assayed in 5 replicates and the mean and %CV were calculated, at multiple time points including and past the expiration date. The protocol and acceptance criteria provided were adequate for the claim of 12 months at 2-8 degrees C. Calibrator in-use stability: Three sets of calibrator material were opened and reconstituted at time points before the measurement, and stored refrigerated. The system was calibrated using the stored (test) calibrators and recovery and precision of pooled samples stored at -70 degree C were evaluated as unknowns. The protocol and acceptance criteria provided were adequate to support the claim of in-use calibrator stability of one day when stored for one day tightly sealed and refrigerated at 2-8 degrees C. Control materials: Stability: Control materials opened and closed stability was evaluated as described for calibrators, except that open control stability testing extended to 7 days. The protocol and acceptance criteria were adequate to support shelf-life stability of months and open vial stability is 7 days when store at 2-8 degrees C. Value assignment: The AIA-PACK ACTH CONTROL SET contains assigned concentration ranges of ACTH. The assigned range is determined using two analyzers and two reagent lots with 5 replicates per sample using the manufacturer's working calibrators. The range is assigned as +/- 20% of the grand mean and is designed to provide target control levels of approximately 50 (target range approximately 40 to 60) and 300 (230 to 350) pg/mL of ACTH. 1 relative to the commercial material and methods to which calibrators are traceable {8} d. Detection limit: The lower limit of the assay range is based on the LoQ according to the CLSI EP17-A guideline. Fifteen samples were evaluated ranging from the LoB to 4 times the LoB. These samples were assayed 2 times a day for over 5 days. LoB is determined to be 0.5 pg/mL. LoD is determined by testing 60 blank and 60 low samples. LoD is determined to be 0.7 pg/mL. LoQ is determined to be 1.2 pg/mL. At that concentration CV was between 10-20% (and bias based on linearity was < 5%). The sponsor claimed that the measuring range of the assay is 2.0 to 2000 pg/mL e. Analytical specificity: ACTH fragments and related peptides were purchased from an outside vendor (Refer to the following table 1). ACTH fragments and related peptides were dissolved in ST AIA-PACK ACTH SAMPLE DILUTING SOLUTION (SDS) and adjusted to 10 ug/mL to prepare a stock solution. 30 uL of fragments were diluted with 2,970 uL of EDTA plasma to 100,000 pg/mL. Concentrations of 500 and 5,000 pg/mL were prepared by further dilution with EDTA plasma. The specimens were measured in duplicate using the ST AIA-PACK ACTH assay at one site using one AIA-2000 analyzer and a single lot of reagent each. The mean and percent recovery was calculated as (Apparent value / Control value) x 100. | ACTH Fragment | ACTH fragment conc. [pg/mL] | | | | | | --- | --- | --- | --- | --- | --- | | | | Average value | Difference (test- - control) | ACTH Recovery [%] | % Cross-reactivity | | (Control - without fragmets) | 0 | 54.3 | --- | --- | --- | | ACTH 1-10 | 500 | 54.3 | 0.0 | 100% | -0.01% | | | 5,000 | 54.7 | 0.4 | 101% | 0.01% | | | 100,000 | 55.7 | 1.3 | 102% | 0.00% | | ACTH 11-24 | 500 | 53.4 | -0.9 | 98% | -0.18% | | | 5,000 | 53.6 | -0.8 | 99% | -0.02% | | | 100,000 | 53.8 | -0.5 | 99% | 0.00% | | Beta-MSH | 500 | 53.8 | -0.5 | 99% | -0.10% | | | 5,000 | 54.5 | 0.2 | 100% | 0.00% | | | 100,000 | 51.6 | -2.8 | 95% | 0.00% | {9} | Beta-Endorphin | 500 | 53.0 | -1.3 | 98% | -0.26% | | --- | --- | --- | --- | --- | --- | | | 5,000 | 54.2 | -0.1 | 100% | 0.00% | | | 100,000 | 52.8 | -1.5 | 97% | 0.00% | | ACTH Fragment | ACTH fragment conc. [pg/mL] | Average Measured ACTH conc. [pg/mL] | Difference (test - control) | ACTH Recovery [%] | % Cross-reactivity | | --- | --- | --- | --- | --- | --- | | | | | | | | | (Control) | 0 | 54.3 | --- | --- | --- | | ACTH 1-17 | 500 | 50.1 | -4.2 | 92% | -0.85% | | | 5,000 | 25.6 | -28.8 | 47% | -0.58% | | | 100,000 | 2.0 | -52.3 | 4% | -0.05% | | ACTH 1-24 | 500 | 49.6 | -4.7 | 91% | -0.94% | | | 5,000 | 25.4 | -29.0 | 47% | -0.58% | | | 100,000 | 2.0 | -52.4 | 4% | -0.05% | | ACTH 18-39 | 500 | 53.3 | -1.1 | 98% | -0.21% | | | 5,000 | 39.0 | -15.3 | 72% | -0.31% | | | 100,000 | 3.5 | -50.8 | 6% | -0.05% | | ACTH 22-39 | 500 | 52.8 | -1.6 | 97% | -0.32% | | | 5,000 | 50.1 | -4.2 | 92% | -0.08% | | | 100,000 | 14.6 | -39.7 | 27% | -0.04% | | alpha-MSH | 500 | 49.7 | -4.7 | 91% | -0.93% | | | 5,000 | 25.6 | -28.7 | 47% | -0.57% | | | 100,000 | 10.3 | -44.1 | 19% | -0.04% | ACTH 1-10, 11-24, beta-MSH and beta-Endorphin do not interfere the assay up to $100,000~\mathrm{pg / mL}$ . ACTH recovery was less than $90\%$ in the presence of ACTH 1-17, 1-24, 18-39, alpha-MSH $(>5,000\mathrm{pg / mL})$ or ACTH 22-39 $(>100,000\mathrm{pg / mL})$ . These fragments may negatively affect the ACTH assay. It is likely that these fragments bind to either the antibodies on beads or the enzyme-labeled antibodies. Therefore, if these fragments were contained excessively in the test sample, lower values may be reported. {10} Evaluation of Common exogenous compounds: Specimen pools at three concentrations of ACTH were evaluated by adding interfering substance and assaying utilizing the ST AIA-PACK ACTH assay on the AIA-2000. The average of triplicate measurements for each level was calculated. The specimen value without adding interfering substance was used as the reference value. The sponsor defines non-significant interference as recovery within $\pm 10\%$ of the unspiked samples. No significant interference was observed when endogenous compounds were tested up to the following concentrations. Heparin (up to $50~\mathrm{U / mL}$ Acetominophen (up to $20\mathrm{mg / L}$ Acetylsalicylic acid (up to $300\mathrm{mg / L}$ Ampicillin (up to $200\mathrm{mg / L}$ Ibuprofen (up to $50~\mathrm{mg / L}$ Theophylline (up to $10\mathrm{mg / L}$ Hemoglobin (up to $440\mathrm{mg / dL}$ ) Conjugated bilirubin (up to $19\mathrm{mg / dL}$ ) Free bilirubin (up to $19\mathrm{mg / dL}$ ) Lipemia (up to $1667\mathrm{mg / dL}$ ) Rheumatoid factor (up to 500 IU/mL) # 2. Comparison studies: # a. Method comparison with predicate device: A total of 160 EDTA plasma specimens were assayed in singleton utilizing the ST AIA-PACK ACTH assay on the AIA-2000 analyzer and the Roche Elecsys-ACTH on the MODULAR ANALYTICS E170 (Elecsys module) analyzer. A combination of fresh and frozen specimens was utilized for this study. Sample selection criteria included only sample stability and integrity (e.g., no known hemolysis, lipemia or icteric). Of the 160 samples.154 were unaltered patient samples. Another 6 were prepared by mixing patient samples. Samples exposed to repeated freeze-thaw cycles were not used. Samples containing particulate matter were centrifuged prior to testing. Prior to assay, frozen samples were brought to room temperature slowly, and mixed gently. Testing was done at one site utilizing one AIA-2000 analyzers and one MODULAR ANALYTICS E170 (Elecsys module) analyzer. Results are shown below. | Regression Analysis | | | | --- | --- | --- | | Deming | | Linear regression | | Slope: | 1.10 (1.075 to 1.116) | 1.09 (1.067 to 1.107) | | Intercept: | -0.84 (-8.921 to 7.234) | 0.76 (-7.305 to 8.817) | | 95% Confidence Intervals are shown in parentheses | | | | Corr Coef (R): | 0.993 | | {11} 12 | Points (Plotted/Total): | 160/160 | | --- | --- | | Result Ranges: | 3.8 to 1986 pg/mL | ## b. Matrix comparison Not applicable; specimen collection and handling states EDTA plasma is required for the assay. 3. Clinical studies: a. Clinical Sensitivity: Not applicable; Clinical sensitivity and specificity is not typically provided in 510(k)s for this type of assay. b. Clinical specificity: See a, above. c. Other clinical supportive data (when a. and b. are not applicable): Not applicable 4. Clinical cut-off: Not applicable; this is a quantitative assay. 5. Expected values/Reference range: The package insert states that each laboratory should determine a reference interval corresponding to the characteristics of the population being tested. Clinical results must be interpreted with regard to concomitant medications administered to the patient. The reference range study was conducted with reference to the CLSI protocol entitled: How to Define and Determine Reference Intervals in the Clinical Laboratory (C28-A3). A total of 121 well-characterized unaltered EDTA plasma specimens from ambulatory patients with no known history of pituitary or adrenal disease were assayed in singleton utilizing the ST AIA-PACK ACTH assay on the AIA-2000 analyzer. Specimens from 110 males between the ages of 26 to 60 years and 11 females between the ages of 28 to 44 years were included in this study. Samples were collected between 8 AM and noon and frozen below -70 degree C until testing. Prior to assay frozen samples were brought to room temperature (18 - 25 °C) slowly, and mixed gently. Samples exposed to repeated freeze-thaw cycles were not used. Samples containing particulate matter were centrifuged prior to testing. Testing was done at one site utilizing one AIA-2000 analyzer. Instrument maintenance and QC records were reviewed to ensure the instrument was functioning within normal guidelines. {12} | Central 95% Interval | | | | --- | --- | --- | | N = 121 | | | | Lower Reference Limit 5% range | | Upper Reference Limit 95% range | | Nonparametric | 7.4 [6.0 to 8.9] pg/mL | 64.3 [48.4 to 72.4] pg/mL | N. Labeling: The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10. O. Conclusion: The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...