← Product Code [CJY](/submissions/CH/subpart-b%E2%80%94clinical-chemistry-test-systems/CJY) · K981794

# HICHEM CALCIUM REAGENT KIT (K981794)

_Hichem Diagnostics · CJY · Jul 29, 1998 · Clinical Chemistry · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/CH/subpart-b%E2%80%94clinical-chemistry-test-systems/CJY/K981794

## Device Facts

- **Applicant:** Hichem Diagnostics
- **Product Code:** [CJY](/submissions/CH/subpart-b%E2%80%94clinical-chemistry-test-systems/CJY.md)
- **Decision Date:** Jul 29, 1998
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 862.1145
- **Device Class:** Class 2
- **Review Panel:** Clinical Chemistry

## Indications for Use

HiChem Calcium Reagent is intended for the quantitative determination of total calcium in serum, plasma and urine. Calcium measurements are used in the diagnosis and treatment of parathyroid disease, a variety of bone diseases, chronic renal disease and tetany. This reagent is intended to be used in a professional setting or by trained personnel and is not intended for home use.

## Device Story

HiChem Calcium Reagent is an in vitro diagnostic reagent for total calcium measurement in serum, plasma, and urine. It utilizes the arsenazo III method, where calcium reacts with arsenazo III to form a blue complex. The resulting increase in absorbance at 650 nm is measured via clinical analyzer or manual spectrophotometric procedure. The output is a quantitative calcium concentration value. Used in clinical laboratories by trained personnel to assist physicians in diagnosing and monitoring parathyroid, bone, and renal diseases. The device provides objective data to inform clinical decision-making regarding patient calcium levels.

## Clinical Evidence

Bench testing only. Performance validated through linearity studies (0.1–16 mg/dL manual; 2.0–15.0 mg/dL automated), precision studies (n=30 to 60 replicates), and method comparison against Beckman SYNCHRON Calcium Reagent. Serum/plasma comparison (n=92 manual; n=153 automated) showed high correlation (r²=0.887 to 0.924). Sensitivity documented at <0.15 mg/dL. Chemical additive interference (heparin, iodoacetate) showed bias <0.2 mg/dL. Stability validated for 14-day onboard calibration and 30-day onboard reagent use.

## Technological Characteristics

Colorimetric reagent based on arsenazo III chemistry. Measures absorbance at 650 nm. Compatible with manual procedures or automated clinical analyzers (e.g., SYNCHRON CX Systems). Chemical additives include sodium heparin, lithium heparin, and lithium iodoacetate. No specific materials of construction or connectivity standards provided.

## Regulatory Identification

A calcium test system is a device intended to measure the total calcium level in serum. Calcium measurements are used in the diagnosis and treatment of parathyroid disease, a variety of bone diseases, chronic renal disease and tetany (intermittent muscular contractions or spasms).

## Predicate Devices

- Beckman® Systems Calcium Reagent (product no. 442755)

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

JUL 20 1998

Image /page/0/Picture/1 description: The image shows the logo for HiChem Diagnostics. The logo features the text "HiChem" in a bold, stylized font, with the word "Diagnostics" underneath in a smaller font. Above and to the left of the text is a solid black circle. To the right of the circle are the words "Smart Solutions".

9817-94

## SUMMARY OF 510(K) SAFETY AND EFFECTIVENESS INFORMATION

This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.

HiChem Calcium Reagent is intended for the quantitative determination of total calcium in serum, plasma and urine. Calcium measurements are used in the diagnosis and treatment of parathyroid disease, a variety of bone diseases, chronic renal disease and tetany.

The HiChem Calcium Reagent determines calcium by its reaction with arsenazo III to form a blue complex. The resulting increase in absorbance at 650 nm is proportional to the calcium concentration in the sample.

The HiChern Calcium Reagent is an adaptation of the arsenazo III calcium method, first reported by Michaylova and Illikova and is intended to be used either as a manual procedure or on clinical analyzers which can automate the required manipulations.

The HiChern Calcium Reagent is substantially equivalent to the Beckman® Systems Calcium Reagent, product no. 442755, manufactured by Beckman® Instruments, Brea, CA. Both reagents support the same intended use and produce substantially equivalent results with the same clinical purpose. In addition, they are methodology which determines total calcium through the colorimetric measurement the calcium - arsenazo III complex.

The effectiveness of the manual procedure is shown by the recovery of linearity standards, the precision of control recoveries, the comparison of serum, plasma and urine recoveries to the Beckman® Calcium Reagent and the validation of the chemical additives and sensitivity claims.

The recovery of total calcium using HiChem Calcium Reagent as a manual procedure is linear between 0.1 to 16 mg/dL as shown by the recovery of linearity standards which span 0 to 18 mg/dL. Regression statistics are shown below.

sy.x = 0.14 mg/dL., (HiChem Recoveries) = - 0.1 mg/dL + 0.980 x (Standard Conc.), r2 = 1.000, df = 23

Precision, demonstrated by replicate assay of urine pools and commercially available control sera, is shown below.

| Specimen        | n  | mean        | within-run SD | total SD   |
|-----------------|----|-------------|---------------|------------|
| Serum control 1 | 30 | 6.69 mg/dL  | 0.21 mg/dL    | 0.24 mg/dL |
| Serum control 2 | 30 | 12.68 mg/dL | 0.29 mg/dL    | 0.30 mg/dL |
| Urine Pool 1    | 30 | 2.41 mg/dL  | 0.11 mg/dL    | 0.11 mg/dL |
| Urine Pool 2    | 30 | 11.10 mg/dL | 0.13 mg/dL    | 0.16 mg/dL |

Calcium recoveries of 92 mixed serum and plasma specimens are compared between the HiChem manual procedure and the Beckman® SYNCHRON® Calcium Reagent, used on the SYNCHRON CX® Systems. Least squares regression statistics are shown below.

| Serum/ Plasma Comparisons:                             |            |                     |
|--------------------------------------------------------|------------|---------------------|
| (HiChem Results) = -0.3 mg/dL + 1.020 × (BMD® Results) | r² = 0.887 | s(y.x) = 0.21 mg/dL |
| Urine Comparisons:                                     |            |                     |
| (HiChem Results) = 0.2 mg/dL + 1.022 × (BMD® Results)  | r² = 0.998 | s(y.x) = 0.15 mg/dL |

The use of sodium heparin, lithium heparin and lithium iodoacetate are shown to be acceptable chemical additives by comparison of spiked serum pools. In all cases, the biases produced by the additive were less than 0.1 mg/dL.

The sensitivity claim of 0.2 mg/dL is documented through the repettive assay of a diluted serum control. The observed sensitivity limit, calculated as three standard deviations of a 30 recision study, is 0.15 mg/dl. and is below the claimed limit of 0.2 mg/dL.

510(k) Notification, HiChem Calcium Reagent Kit, p 61

20 May, 1998

**2024 Native Student Orientation Schedule: August 18, 2024**

{1}------------------------------------------------

The effectiveness of the HiChem secondary reagent application for the Beckman® SYNCHRON CX® Systems is shown by the recovery of linearity standards, the precision of control recovery of serum controls over both the calibration stability and on-board stability claims, the validation of the chemical additives and sensitivity claims, and the comparison of patient specimen recoveries to the Beckman® SYNCHRON® Systems Calcium Reagent.

The recovery of total calcium using the HiChem CA Reagent on the SYNCHRON CX® Systems is linear from at least 2.0 mg/dL to 15.0 mg/dL as shown by the recovery of six linearity standards which span the claimed linear range. Regression statistics are shown below.

(HiChem Recoveries) = 0.0 mg/dL + 0.967 × (Standard Conc.), r2 = 1.000. sy.x = 0.08 mg/dL, df = 29

Precision, demonstrated by replicate assay of urine pools and commercially available control sera, is shown below.

| Specimen        | n  | mean        | within-run SD | total SD    |
|-----------------|----|-------------|---------------|-------------|
| Serum control 1 | 60 | 6.89 mg/dL  | 0.056 mg/dL   | 0.064 mg/dL |
| Serum control 2 | 60 | 10.03 mg/dL | 0.085 mg/dL   | 0.089 mg/dL |
| Serum control 3 | 60 | 13.21 mg/dL | 0.088 mg/dL   | 0.090 mg/dL |
| Urine Pool 1    | 60 | 3.24 mg/dL  | 0.048 mg/dL   | 0.059 mg/dL |
| Urine Pool 2    | 60 | 11.64 mg/dL | 0.075 mg/dL   | 0.095 mg/dL |

Calcium recoveries of 153 mixed serum and plasma specimens are compared between the HiChem and Beckman® CA Reagents on the SYNCHRON CX® Systems. Least squares regression statistics are shown below.

Serum/ Plasma Comparisons:

| (HiChem Results) = -0.3 mg/dL + 1.005 x (Beckman® Results) $r^2$ = 0.924 | s(y.x) = 0.20 mg/dL |
|--------------------------------------------------------------------------|---------------------|
| Urine Comparisons:                                                       |                     |
| (HiChem Results) = 0.1 mg/dL + 1.031 x (Beckman® Results) $r^2$ = 0.996  | s(y.x) = 0.16 mg/dL |

The use of sodium heparin, lithium heparin and lithium iodoacetate are shown to be acceptable chemical additives by comparison of spiked serum pools. In all cases, the observed biases were less than 0.2 mg/d..

The sensitivity claim of 2.0 mg/dL is documented through the repetitive assay of a diluted serum control. The observed sensitivity limit, calculated as three standard deviations of a 30 replicate within run precision study, is less than 0.15 mg/dL and is well below the claimed limit of 2.0 mg/dL.

The 14 day onboard calibration stability and the 30 day on board reagent stability claims are documented through the assay of serum controls and urine pools over the claimed periods. In all cases, estimates of imprecision of calcium recoveries over the claimed intervals are less than the greater of 0.26 mg/dL or 2.6%, which is the manufacturer's total precision claim for the SYNCHRON® Analyzer.

The HiChem Calcium Reagent is shown to be safe and effective and substantially equivalent to the Beckman® SYNCHRON® Systems Calcium Reagent, product no. 442755, manufactured by Beckman® Instruments, Brea, CA.

Wynne Stocking

Manager of Regulatory Affairs HiChem Diagnostics

510(k) Notification, HiChem Calcium Reagent Kit, p 62

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized depiction of an eagle or bird with outstretched wings. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the bird symbol.

Food and Drug Administration 2098 Gaither Road Rockville MD 20850

## JUL 2 9 1998

Wynn Stocking . Manager, Regulatory Affairs HiChem Diagnostics 231 North Puente Street Brea, California 92821

K981794 Re : HiChem Calcium Reagent Regulatory Class: II Product Code: CJY Dated: May 20, 1998 Received: May 21, 1998

Dear Mr. Stocking:

We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Druq, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major requlations affecting your device can be found in the Code of Federal Requlations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory In addition, FDA may publish further announcements action. concerning your device in the Federal Reqister. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.

{3}------------------------------------------------

Page 2

Under the Clinical Laboratory Improvement Amendments of 1988 (CLIA-88), this device may require a CLIA complexity categorization. To determine if it does, you should contact the Centers for Disease Control and Prevention (CDC) at (770) 488-7655.

This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4588. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to
premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html".

Sincerely yours,
Steven Autman

Steven I. Gutman, M.D., M.B.A. Director Division of Clinical Laboratory Devices ----------Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

Page 1 of 1

K981794 510(k) Number (if known):

Device Name:

HiChem Calcium Reagent Kit

Indications For Use:

HiChem Calcium Reagent is intended for the quantitative determination of total calcium in serum, Hichent Calcioin Reagent is mended for the qualifical disease, a variety of bone plasma and unne for the the tradinons and telany (intermittent muscular contractions or spasms).

This reagent is intended to be used in a professional setting or by trained personnel and Is not intended for home use.

Respectfully.

Wynn Stordins

Regulatory Affairs Manager HiChem Diagnostics

2 Juno, 1998

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Prescription Use
(Per 21 CFR 801.109)

OR

Over-The-Counter Use **__**
(Optional Format 1-2-96)

---

**Source:** [https://fda.innolitics.com/submissions/CH/subpart-b%E2%80%94clinical-chemistry-test-systems/CJY/K981794](https://fda.innolitics.com/submissions/CH/subpart-b%E2%80%94clinical-chemistry-test-systems/CJY/K981794)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
