NICHOLS ADVANTAGE ALDOSTERONE ASSAY

K032188 · Nichols Institute Diagnostics · CJM · Jul 31, 2003 · Clinical Chemistry

Device Facts

Record IDK032188
Device NameNICHOLS ADVANTAGE ALDOSTERONE ASSAY
ApplicantNichols Institute Diagnostics
Product CodeCJM · Clinical Chemistry
Decision DateJul 31, 2003
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 862.1045
Device ClassClass 2
Attributes3rd-Party Reviewed

Indications for Use

The Nichols Advantage® Aldosterone assay is intended for use with the Nichols Advantage® Specialty System to quantitatively measure aldosterone in human serum and EDTA plasma. Aldosterone measurements are used in the diagnosis and treatment of primary aldosteronism (a disorder caused by excessive secretion of aldosterone by the adrenal gland), hypertension caused by primary aldosteronism, selective hypoaldosteronism, edematous states, and other conditions of electrolyte imbalance.

Device Story

The Nichols Advantage Aldosterone assay is a competitive chemiluminescent immunoassay used with the Nichols Advantage Specialty System. It measures aldosterone levels in human serum or EDTA plasma samples. The assay utilizes acridinium ester-labeled aldosterone and mouse monoclonal antibodies to human aldosterone, with magnetic particles as the solid phase. The system processes samples to produce quantitative aldosterone concentrations. It is intended for professional use in clinical laboratory settings. Healthcare providers use the resulting aldosterone measurements to diagnose and manage conditions such as primary aldosteronism, hypertension, and electrolyte imbalances, aiding in clinical decision-making regarding patient treatment.

Clinical Evidence

Bench testing only. Precision evaluated via NCCLS EP5-A (21-day study). Linearity demonstrated across assay range. Analytical sensitivity (LOD) is ≤ 1.2 ng/dL. Specificity testing showed low cross-reactivity with other steroidal compounds. Method comparison with predicate (n=103) showed strong correlation (r=0.96) with Passing-Bablok slope of 1.04 and Deming slope of 1.09.

Technological Characteristics

Competitive immunoassay; avidin-coated magnetic particles; 3-step incubation at 37°C; 250 uL sample volume; analytical sensitivity 1.2 ng/dL. System is an automated clinical chemistry analyzer.

Indications for Use

Indicated for quantitative measurement of aldosterone in human serum and EDTA plasma to aid in the diagnosis and treatment of primary aldosteronism, associated hypertension, selective hypoaldosteronism, edematous states, and electrolyte imbalances.

Regulatory Classification

Identification

An aldosterone test system is a device intended to measure the hormone aldosterone in serum and urine. Aldosterone measurements are used in the diagnosis and treatment of primary aldosteronism (a disorder caused by the excessive secretion of aldosterone by the adrenal gland), hypertension caused by primary aldosteronism, selective hypoaldosteronism, edematous states, and other conditions of electrolyte imbalance.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0} 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY DEVICE ONLY TEMPLATE A. 510(k) Number: K032188 B. Analyte: Aldosterone C. Type of Test: Quantitative D. Applicant: Nichols Institute Diagnostics, Inc. E. Proprietary and Established Names: Nichols Advantage Chemilunescent Aldosterone Immunoassay F. Regulatory Information: 1. Regulation section: 862.1045; 862.1150; 862.1660 2. Classification: Class II; Class II; Class I 3. Product Code: CJM; JIS; JJX 4. Panel: Clinical Chemistry G. Intended Use: 1. Intended use(s): The Nichols Advantage® Aldosterone assay is intended for use with the Nichols Advantage® Specialty System to quantitatively measure aldosterone in human serum and EDTA plasma. 2. Indication(s) for use: Aldosterone measurements are used in the diagnosis and treatment of primary aldosteronism (a disorder caused by excessive secretion of aldosterone by the adrenal gland), hypertension caused by primary aldosteronism, selective hypoaldosteronism, edematous states, and other conditions of electrolyte imbalance. 3. Special condition for use statement(s): None 4. Special instrument Requirements: Nichols Advantage® Specialty System H. Device Description: The Nichols Advantage® Aldosterone assay contains sufficient reagents for 100 tests, is a competitive binding assay for aldosterone in serum or EDTA-plasma, and utilizes the Nichols Advantage® Specialty System, under FDA PMN device clearance (K961142). {1} Page 2 of 5 I. Substantial Equivalence Information: 1. Predicate device name(s): Diagnostic Product Corp. Coat-A-Count® Aldosterone 2. Predicate K number(s): K831178 3. Comparison with predicate: | Similarities | | | | --- | --- | --- | | Item | Device | Predicate | | specimen | serum or plasma | serum or plasma | | Differences | | | | Item | Device | Predicate | | Binding Technology | Magnetic particles - avidin coated | Antibody coated tubes | J. Standard/Guidance Document Referenced (if applicable): NCCLS EP5-A method (Evaluation of Precision Performance of Clinical Chemistry Devices; Approved Guideline) K. Test Principle: This Nichols Advantage® Chemiluminescent Aldosterone Immunoassay [i.e., Nichols Advantage® Aldosterone or NAA] is an in vitro diagnostic (IVD) chemiluminescence immunoassay designed for professional use with Nichols Advantage® Specialty System [NASS] for quantitatively assaying aldosterone in human serum and EDTA plasma. The NAA is based upon competitive binding between aldosterone in the patient sample and acridinium ester labeled aldosterone (labeled aldosterone for chemiluminescence) for a specific mouse monoclonal antibody to human aldosterone. L. Performance Characteristics (if/when applicable): 1. Analytical performance: a. Precision/Reproducibility: The within-run and total imprecision performance for the assay was estimated using the NCCLS EP5-A method (Evaluation of Precision Performance of Clinical Chemistry Devices; Approved Guideline). The data represent one run per day over 21 days with each serum pool or assay control run in duplicate. The study was performed on a single system. | Pool or Control | Mean (ng/dL) | Within-Run | | Total Imprecision | | | --- | --- | --- | --- | --- | --- | | | | SD | %CV | SD | %CV | | Pool 1 | 4.3 | 0.6 | 14.0% | 0.8 | 18.6% | | Pool 2 | 8.2 | 0.4 | 5.4% | 0.7 | 8.5% | | Pool 3 | 19.7 | 0.8 | 4.1% | 1.0 | 5.2% | | Pool 4 | 71.7 | 2.1 | 2.9% | 3.5 | 4.9% | | Control 1 | 6.3 | 0.5 | 7.9% | 1.0 | 16.0% | | Control 2 | 23.1 | 1.0 | 4.3% | 1.2 | 5.2% | | Control 3 | 76.5 | 3.6 | 4.7% | 5.8 | 7.6% | {2} # b. Linearity/assay reportable range: Samples with varying concentrations were manually diluted with Sample Diluent before placing onto the system. The observed and corrected results are presented. The results demonstrate linearity across the range of the assay. | Sample | Dilution | Observed (ng/dL) | Expected (ng/dL) | % Recovery | | --- | --- | --- | --- | --- | | A | Undiluted | 18.1 | | | | | 1:2 | 8.4 | 9.1 | 92% | | | 1:4 | 5.2 | 4.5 | 116% | | | 1:8 | 2.5 | 2.3 | 110% | | B | Undiluted | 22.2 | | | | | 1:2 | 11.8 | 11.1 | 106% | | | 1:4 | 5.2 | 5.6 | 93% | | | 1:8 | 2.8 | 2.8 | 100% | | C | Undiluted | 113.2 | | | | | 1:2 | 52.5 | 56.6 | 93% | | | 1:4 | 26.2 | 28.3 | 93% | | | 1:8 | 14.9 | 14.2 | 105% | | D | Undiluted | 53.8 | | | | | 1:2 | 27.7 | 26.9 | 103% | | | 1:4 | 12.8 | 13.5 | 91% | | | 1:8 | 6.5 | 6.7 | 97% | # c. Traceability (controls, calibrators, or method): Both the calibrators and controls contain aldosterone spiked into a liquid matrix containing processed human serum and $\leq 0.095\%$ sodium azide. The lyophilized calibrators and controls are assigned their concentrations against non-lyophilized materials. # d. Detection limit: The analytical sensitivity (Limit Of Detection, LOD) was determined by reading the -2SD response from $n = 20$ replicate measurements of the zero standard from the stored master curve from several runs and instruments. The sensitivity for this assay was estimated to be $\leq 1.2 \, \mathrm{ng/dL}$ . # e. Analytical specificity: The antisera for the assay is specific for aldosterone, and demonstrates low crossreactivity to other steroidal compounds in patient samples. Crossreactivity in the chart below were calculated on a weight-per-weight basis at approximately $50\%$ binding intercept or less. | Cross-reactant | Highest Amt. Tested (μg/dL) | Apparent Amt. Detected (ng/dL) | % Crossreactivity | | --- | --- | --- | --- | | Aldosterone | | | 100% | | Pregnenolone | 11,600 | <1.5 | none detected | | Progesterone | 50,000 | 25.6 | <0.0001% | | 11-Deoxycorticosterone | 50,000 | 18.9 | <0.0001% | | Corticosterone | 50,000 | 31.4 | <0.0001% | | 18-Hydroxycorticosterone | 5.0 | 20.9 | 0.42% | | 17-Hydroxyprogesterone | 50,000 | 20.1 | <0.0001% | {3} | 11-Deoxycortisol | 50,000 | 28.5 | <0.0001% | | --- | --- | --- | --- | | Cortisol | 50,000 | 26.1 | <0.0001% | | Testosterone | 100,000 | 25.1 | <0.0001% | | Androsterone | 100,000 | 11.4 | <0.0001% | | DHEA | 48,000 | 11.6 | <0.0001% | | Androstendione | 47,800 | 50.8 | 0.0001% | | Estradiol | 287,000 | <1.5 | none detected | | Estrone | 8,000 | 6.1 | <0.0001% | | Estriol | 8,400 | <1.5 | none detected | | Cortisone | 25,000 | 20.7 | <0.0001% | | Dexamethasone | 50,000 | 7.1 | <0.0001% | | Predinisone | 32,600 | 26.3 | <0.0001% | | Prednisolone | 50,000 | 27.3 | <0.0001% | | Spironolactone | 50,000 | 22 | <0.0001% | f. Assay cut-off: N/A # 2. Comparison studies: a. Method comparison with predicate device: Nichols Advantage® Aldosterone or NAA assay (Y) was compared to a current marketed predicate assay kit, the DPC Coat-A-Count Aldosterone RIA device (X), previously cleared by the FDA (K831178, 5/27/83). One hundred three (103) clinical undiagnosed remnant serum samples were assayed in duplicate by both methods following each manufacturers' directions. The aldosterone assay range observed with RIA method "X" was 2.7 to $125\mathrm{ng / dL}$ ; range for NAA method "Y" was 2.7 to $120\mathrm{ng / dL}$ . The Passing-Bablok regression analysis of these data yielded an equation of $\mathrm{Y} = 1.04\mathrm{X} + 0.1$ (95% confidence intervals for slope and intercept were 0.98 to 1.10, and -1.0 to +1.1, respectively). Deming regression analysis of these data yielded an equation of $\mathrm{Y} = 1.09\mathrm{X} - 0.6$ (95% confidence intervals for slope and intercept were 1.03 to 1.15, and -3.2 to +2.1 respectively). Pearson's correlation coefficient (r) of the paired data was 0.96 (95% confidence interval was 0.94 to 0.97), showing that essentially equivalent results were provided by both cleared predicate, the DPC Coat-a-Count® Aldosterone RIA device and the NAA immunoassay device. b. Matrix comparison: N/A # 3. Clinical studies: a. Clinical sensitivity: N/A b. Clinical specificity: N/A c. Other clinical supportive data (when $a$ and $b$ are not applicable): N/A # 4. Clinical cut-off: N/A # 5. Expected values/Reference range: To establish an expected reference range, AM (8-10 AM) and PM (4-6 PM) serum samples were obtained. For AM, serum from $n = 171$ healthy, prescription medication free fasting adults (102 females and 69 males, age: 18 to 79 years) were obtained in the upright/sitting posture and supine posture. The upright/sitting {4} blood sample was obtained while the fasting volunteer was ambulatory for at least 15 minutes prior to sitting down to have their blood drawn. Individuals who underwent further supine blood sampling laid down immediately after their upright sample was drawn. The supine blood sample was obtained after the volunteer was laying down for at least 30 minutes. All individuals had normal blood pressures (systolic $\leq 139\mathrm{mmHg}$ ; diastolic $\leq 89$ mm Hg), normal serum electrolytes, normal fasting glucose, BUN, and creatinine levels, and none were hyperlipidemic. None of the individuals were on restricted diets. Pregnancy and breast feeding mothers, individuals who were taking prescription medications, and anyone who were taking medications for weight loss were excluded. The supine cohort $(n = 62)$ was a subset of the larger AM population. The PM cohort $(n = 44)$ was a different subset of the AM population. Aldosterone levels were not age dependent. The $95\%$ confidence intervals observed from these studies are as follows. | Time and Posture | Serum: 95% Confidence Intervals: | | --- | --- | | 8-10 AM Upright/Sitting | 3-34 ng/dL | | 8-10 AM Supine | 2-19 ng/dL | | 4-6 PM Upright/Sitting | 2-23 ng/dL | To establish a suggested reference range using EDTA plasma samples, $n = 51$ adults (27 females, 24 males) ages 22-59 years of the above cohort was evaluated. | Time and Posture | EDTA: 95% Confidence Intervals: | | --- | --- | | 8-10 AM Upright/Sitting | 3-22 ng/dL | | 8-10 AM Supine | 2-14 ng/dL | # M. Conclusion: Based on Third Party Review Report, the Nichols Advantage Chemilunescent Adosterone Immunoassay is substantially equivalent (SE) to device regulated under: Aldosterone test system; 21 CFR §862.1045; CJM; Class II Calibrator; 21 CFR §862.1150; JIS; Class II Quality Control Material (Assayed and Unassayed); 21 CFR §862.1660; JJX Class I
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...