S TEST REAGENT CARTRIDGE ALP (ALKALINE PHOSPHATASE)
K130141 · Hitachi Chemical Diagnostics, Inc. · CJE · Mar 22, 2013 · Clinical Chemistry
Device Facts
Record ID
K130141
Device Name
S TEST REAGENT CARTRIDGE ALP (ALKALINE PHOSPHATASE)
Applicant
Hitachi Chemical Diagnostics, Inc.
Product Code
CJE · Clinical Chemistry
Decision Date
Mar 22, 2013
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 862.1050
Device Class
Class 2
Indications for Use
The S TEST Reagent Cartridge Alkaline Phosphatase (ALP) is intended for the quantitative measurement of alkaline phosphatase activity in serum, lithium heparinized plasma, or sodium citrate plasma using the HITACHI Clinical Analyzer. The S TEST Reagent Cartridge Alkaline Phosphatase (ALP) is intended for use in clinical laboratories or physician office laboratories. For in vitro diagnostic use only. Measurements of alkaline phosphatase are used in the diagnosis and treatment of liver, bone, parathyroid, and intestinal diseases.
Device Story
The Hitachi Clinical Analyzer is an automated, bench-top, wet chemistry system for clinical or physician office laboratories. It utilizes single-use plastic reagent cartridges containing two reservoirs (R1, R2) and a reaction cell/photometric cuvette. The system processes serum or plasma samples by pipetting sample and reagent into the cuvette, mixing, and incubating. A multi-wavelength photometer measures absorbance to calculate analyte concentration. The reagent cartridge includes a dot code label with chemistry parameters and calibration factors. Results are available in approximately 15 minutes. Healthcare providers use the quantitative ALP results to assist in the diagnosis and treatment of liver, bone, parathyroid, and intestinal diseases.
Clinical Evidence
Clinical performance was evaluated at three physician office laboratory (POL) sites. Precision studies (n=30 replicates per level) showed total precision %CVs ranging from 3.0% to 9.2%. Method comparison studies (n=72-77 per site) against a reference method yielded correlation coefficients (r) of 0.99, with slopes ranging from 0.942 to 0.980, demonstrating accuracy across the dynamic range of 11 to 745 U/L.
Indicated for quantitative measurement of alkaline phosphatase activity in serum, lithium heparinized plasma, or sodium citrate plasma for diagnosis and treatment of liver, bone, parathyroid, and intestinal diseases. For prescription and point-of-care use in clinical or physician office laboratories.
Regulatory Classification
Identification
An alkaline phosphatase or isoenzymes test system is a device intended to measure alkaline phosphatase or its isoenzymes (a group of enzymes with similar biological activity) in serum or plasma. Measurements of alkaline phosphatase or its isoenzymes are used in the diagnosis and treatment of liver, bone, parathyroid, and intestinal diseases.
Predicate Devices
Roche Cobas c systems ALP2 (k100853)
Related Devices
K131051 — HITACHI S TEST REAGENT CARTRIDGE TOTAL PROTEIN (TP) AND ALBUMIN (ALB) · Hitachi Chemical Diagnostics, Inc. · Jun 7, 2013
K130685 — HITACHI CLINICAL ANALYZER S TEST REAGENT CARTRIDGES FOR LD AND AMY · Hitachi Chemical Diagnostics, Inc. · Aug 9, 2013
K120945 — HITACHI CLINICAL ANALYZER S TEST REAGENT CARTRIDGE ALT, HITACHI CLINICAL ANALYZER S TEST REAGENT CARTRIDGE AST · Hitachi Chemical Diagnostics, Inc. · Jun 7, 2013
K130870 — HITACHI CLINICAL ANALYZER S TEST REAGENT CARTRIDGE CALCIUM (CA) · Hitachi Chemical Diagnostics, Inc. · May 2, 2013
K990024 — CARESIDE ALKALINE PHOSPHATASE · Careside, Inc. · Jan 13, 1999
Submission Summary (Full Text)
{0}
1
510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION
DECISION SUMMARY
ASSAY ONLY TEMPLATE
A. 510(k) Number:
k130141
B. Purpose for Submission:
New Device
C. Measurand:
Alkaline phosphatase (ALP)
D. Type of Test:
Quantitative, enzymatic activity
E. Applicant:
Hitachi Chemical Diagnostics
F. Proprietary and Established Names:
S TEST Reagent Cartridge Alkaline Phosphatase (ALP)
G. Regulatory Information:
1. Regulation section:
21 CFR § 862.1050 Alkaline Phosphatase (ALP) or isoenzymes test system
2. Classification:
Class II
3. Product code:
CJE
4. Panel:
Clinical Chemistry (75)
{1}
H. Intended Use:
1. Intended use(s):
See indications for use below.
2. Indication(s) for use:
The S TEST Reagent Cartridge Alkaline Phosphatase (ALP) is intended for the quantitative measurement of alkaline phosphatase activity in serum, lithium heparinized plasma, or sodium citrate plasma using the HITACHI Clinical Analyzer. The S TEST Reagent Cartridge Alkaline Phosphatase (ALP) is intended for use in clinical laboratories or physician office laboratories. For in vitro diagnostic use only.
Measurements of alkaline phosphatase are used in the diagnosis and treatment of liver, bone, parathyroid, and intestinal diseases.
3. Special conditions for use statement(s):
For prescription and point-of-care use.
4. Special instrument requirements:
Hitachi Clinical Analyzer E40
I. Device Description:
The S TEST reagent cartridges for the Hitachi Clinical Analyzer E40 are made of plastic and include two small reservoirs capable of holding two separate reagents R1 (2-Ethylaminoethanol Buffer and Magnesium Chloride) and R2 (p-Nitrophenyl phosphate disodium salt and 2-Amino-2-hydroxymethyl-1,3-propanediol Buffer), separated by a reaction cell/photometric cuvette. The cartridges also include a dot code label that contains all chemistry parameters, calibration factors, and other production-related information, e.g., expiration dating.
J. Substantial Equivalence Information:
1. Predicate device name(s):
Roche Cobas c systems ALP2
2. Predicate 510(k) number(s):
k100853
{2}
3. Comparison with predicate:
| | Hitachi S Test Reagent Cartridge ALP (Candidate Device) | Roche Cobas c systems ALP2 (Predicate Device, k100853) |
| --- | --- | --- |
| Similarities | | |
| Intended Use/Indications for Use | For the quantitative measurement of alkaline phosphatase activity in serum, lithium heparinized plasma, or sodium citrate plasma Measurements of alkaline phosphatase are used in the diagnosis and treatment of liver, bone, parathyroid, and intestinal diseases. | Same |
| Specimen Type | Human serum or (sodium citrate or lithium heparin) plasma | Same |
| Differences | | |
| Test Principle | In the presence of magnesium ions, ALP reacts with p-NPP to release p-nitrophenol. | In the presence of magnesium and zinc ions, p-NPP is cleaved by phosphatases into phosphate and p-nitrophenol |
| Testing Environment | Physician office or clinical lab | Clinical lab |
| Detection Wavelength | 405/508 nm | 480/450 nm |
| Claimed measuring range | 10 to 1,000 U/L | 5 to 1,200 U/L |
| Detection Limit | 1.8 U/L | 5 U/L |
K. Standard/Guidance Document Referenced (if applicable):
CLSI/NCCLS EP5-A2: Evaluation of Precision Performance of Quantitative Measurement Methods; 2004
CLSI/NCCLS EP6-A: Evaluation of Linearity of Quantitative Measurement Procedures, A Statistical Approach; 2003
{3}
CLSI EP7-A2: Interference Testing in Clinical Chemistry; Approved Guideline, 2005
CLSI EP17-A: Protocols for Determination of Limits of Detection and Limits of Quantitation; 2004
## L. Test Principle:
Alkaline phosphatase (ALP) in the sample reacts with its substrate, p-nitrophenyl phosphate (p-NPP), in ethylaminoethanol (EAE) buffer, to release p-nitrophenol (yellow). The ALP activity is determined by measuring the rate of p-nitrophenol production. This rate of change in absorbance is directly proportional to the ALP activity in the serum.
## M. Performance Characteristics (if/when applicable):
### 1. Analytical performance:
#### a. Precision/Reproducibility:
20-day In-house Precision: The studies followed CLSI EP5-A2, where three levels of archived (stored frozen) patient serum samples were each tested in two runs, twice a day, for 20 days on the Hitachi E40 Clinical Analyzer. The results were as follows:
ALP- Low, Level 1, Summary
| ALP | Within-Run | Total |
| --- | --- | --- |
| Mean (U/L) | 42.7 | 42.7 |
| SD (U/L) | 1.3 | 2.25 |
| %CV | 3.1% | 5.3% |
ALP- Middle, Level 2, Summary
| ALP | Within-Run | Total |
| --- | --- | --- |
| Mean (U/L) | 80.5 | 80.5 |
| SD (U/L) | 3.07 | 4.65 |
| %CV | 3.8% | 5.8% |
ALP- High, Level 3, Summary
| ALP | Within-Run | Total |
| --- | --- | --- |
| Mean (U/L) | 555.6 | 555.6 |
| SD (U/L) | 13.99 | 24.69 |
| %CV | 2.5% | 4.4% |
{4}
Point-of-Care Precision:
Three levels of samples A (low), B (middle), and C (high) were tested by three POL sites, six times a day for five days on the Hitachi E40 Clinical Analyzer. The samples were native (neat) serum specimens (stored frozen). The precision estimates are described below:
| Site # | Sample | Mean (U/L) | Within-run Precision | | Total Precision | |
| --- | --- | --- | --- | --- | --- | --- |
| | | | SD (U/L) | %CV | SD (U/L) | %CV |
| 1 | A | 13.2 | 0.80 | 6.1 | 0.96 | 7.3 |
| 2 | A | 12.8 | 0.89 | 6.9 | 1.00 | 7.8 |
| 3 | A | 13.4 | 1.18 | 8.8 | 1.23 | 9.2 |
| 1 | B | 76.2 | 1.22 | 1.6 | 2.26 | 3.0 |
| 2 | B | 68.8 | 4.39 | 6.4 | 5.24 | 7.6 |
| 3 | B | 70.0 | 4.56 | 6.5 | 6.28 | 9.0 |
| 1 | C | 408.6 | 4.22 | 1.0 | 26.01 | 6.4 |
| 2 | C | 390.1 | 23.21 | 6.0 | 23.21 | 6.0 |
| 3 | C | 402.9 | 13.21 | 3.3 | 12.33 | 3.1 |
b. Linearity/assay reportable range:
A linearity study was conducted based on the CLSI EP6-A guideline by comparing observed versus expected values for 10 samples. Ten (10) serial dilutions (4 to 1168 U/L were prepared and tested. The dilutions were prepared using a commercial linearity/calibration set. The calibration samples were assigned their reference values arithmetically from the labeled values and were tested in duplicate by the Hitachi E40 Clinical Analyzer. The mean Hitachi results (y-axis) were plotted against the expected values (x-axis). The parameters of linear regression are as follows:
$$
y = 0.9651x + 3.6328
$$
$$
R^2 = 0.999
$$
The ALP assay is linear between $4\mathrm{U/L}$ and $1106\mathrm{U/L}$. The results of the study support the sponsor's claim that the Hitachi S Test ALP Reagent Cartridge is linear across the measuring range of 10 to $1,000\mathrm{U/L}$.
{5}
c. Traceability, Stability, Expected values (controls, calibrators, or methods):
Each lot of S TEST ALP cartridges is calibrated by the manufacturer prior to shipment using material traceable to Japanese Enzyme Reference Material (JC-ERM). The barcode printed on each cartridge provides the analyzer with lot-specific calibration data. Alkaline Phosphatase concentration is directly determined by multiplying the change in absorbance of the unknown sample by the calibrator factor on the barcode. No user calibration is needed.
Commercially available controls are required but not provided. Users should follow federal, state, and local regulatory requirements regarding quality control practices.
d. Detection limit:
Detection limit studies were performed according to CLSI EP17-A Protocols for Determination of Limits of Detection and Limits of Quantitation. The blank sample for the reagent system was assayed on the Hitachi Clinical Analyzer E40, 20 times per day for three days for a total of 60 replicate results to determine LOB. Five low samples were assayed on the Hitachi Clinical Analyzer E40, 4 times per day for three days for a total of 60 replicate results with the specific reagent cartridges to determine LOD. Each of three samples targeted at 10 U/L were assayed 6 times per day for three days on three different analyzers for a total of 54 replicate results to determine LoQ. The LoQ was determined based on inter-assay precision of < 15% CV.
| Analyte | LoB (U/L) | LoD (U/L) | LoQ (U/L) |
| --- | --- | --- | --- |
| Alkaline Phosphatase | 1.02 | 1.8 | 10 |
The measuring range of the assay was 10 to 1000 U/L.
e. Analytical specificity:
Interference studies were performed according to CLSI EP7-A2, Interference Testing in Clinical Chemistry; Approved Guideline) were performed to determine the effects from potential interferents. Two levels of commercial control sera (ALP low and high, approximately 35 U/L and 100 U/L) were spiked to six levels with each interferent, and all seven samples (the 6 spiked samples and the neat, zero baseline sample) were tested in replicates of three on the Hitachi E40 Clinical Analyzer. In each case, the spiked sample result mean was compared to its neat control mean result, and recoveries were calculated. The sponsor claims non significant interference (≤10% difference) for the substances and concentrations listed in the table below.
{6}
Alkaline Phosphatase
| Interferent Compound | Highest Concentration Showing No Interference |
| --- | --- |
| Hemoglobin | 500 mg/dL |
| Unconjugated bilirubin | 50 mg/dL |
| Lipemia | 2,000 mg/dL |
| Ascorbic acid | 50 mg/dL |
f. Assay cut-off:
Not Applicable
2. Comparison studies:
a. Method comparison with predicate device:
In-house method comparison study was conducted using a total of 97 clinical specimens (4 diluted and 7 spiked) spanning the dynamic range (11 to 926 U/L), samples were assayed on the Hitachi Clinical Analyzer E40 in singleton on both the Hitachi S Test Reagent and the Roche Cobas c systems ALP2 (predicate device). The comparative data were analyzed by linear regression and are shown below.
Internal study summary – ALP (U/L)
| n | Hitachi Range | Regression Equation | “r” | 95% CI Slope | 95% CI Intercept |
| --- | --- | --- | --- | --- | --- |
| 97 | 11 to 926 U/L | y = 0.926x + 4.8075 | 0.996 | (0.909 to 0.943) | (-0.2 to 9.8) |
External site method comparison study: A series of approximately 70 serum specimens (4 diluted) with ALP values ranging from 11 to 745 U/L were assayed on the Hitachi Clinical Analyzer E40 at three sites using S TEST Reagent Cartridge ALP (y) and Roche Cobas c systems ALP2 (predicate device) as the reference method (x). The samples were native serum specimens. Linear regression analysis yielded the following results:
POL study summary- ALP (U/L)
| Site # | n | Sample Range | Regression Equation | “r” | 95% CI Slope | 95% CI Intercept |
| --- | --- | --- | --- | --- | --- | --- |
| 1 | 77 | 13 to 745 U/L | y=0.967x +2.9 | 0.99 | 0.949 to 0.984 | -0.6 to 6.4 |
| 2 | 72 | 11 to 688 U/L | y=0.942x -0.7 | 0.99 | 0.925 to 0.960 | -4.4 to 2.9 |
| 3 | 72 | 12 to 736 U/L | y=0.980x +0.8 | 0.99 | 0.961 to 0.999 | -3.2 to 4.7 |
{7}
# b. Matrix comparison
A study was performed to validate the use of two plasma types as an alternative to serum for the Hitachi Clinical Analyzer E40 with S TEST Reagent Cartridge ALP. The plasma types were sodium citrate and lithium heparin. Thirty-eight (38) matched serum/plasma samples that spanned the range of the assay (13 to 967 U/L) were assayed in singleton. The study set included five diluted samples and six spiked samples. The results were compared using linear regression (plasma = y-axis, each type). The performance characteristics were as follows.
$\mathrm{N} = 38$
Range (serum) = 13 to 967 U/L
| | Sodium Citrate Plasma | Heparinized Plasma |
| --- | --- | --- |
| Slope (95% CIs) | 1.03 (1.01 to 1.05) | 1.01 (1.00 to 1.02) |
| y-intercept (95% CIs) | -11.2 (-15.5 to -7.0) | -5.4 (-8.3 to -2.6) |
| r | 0.999 | 0.999 |
The sponsor claimed that lithium heparin and sodium citrate are acceptable anti-coagulants to be used with the ALP assay.
# 3. Clinical studies:
a. Clinical Sensitivity:
Not applicable
b. Clinical specificity:
Not applicable
c. Other clinical supportive data (when a. and b. are not applicable):
Not applicable
# 4. Clinical cut-off:
Not applicable. Clinical studies are not typically submitted for this device type.
# 5. Expected values/Reference range:
| | Males | Females |
| --- | --- | --- |
| ALP Reference Range1 | 53 to 128 U/L | 42 to 141 U/L |
{8}
9
¹ Tietz Fundamentals of Clinical Chemistry, 4th Edition, WB Saunders Company, (1996)
N. Proposed Labeling:
The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10.
O. Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.