K251157 · Truvian Health · CIX · Jan 9, 2026 · Clinical Chemistry
Device Facts
Record ID
K251157
Device Name
Core Metabolic
Applicant
Truvian Health
Product Code
CIX · Clinical Chemistry
Decision Date
Jan 9, 2026
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 862.1035
Device Class
Class 2
Indications for Use
The Core Metabolic panel is part of the TruWellness Panel™ and is intended for use on the TruVerus™. The Core Metabolic panel (part of the TruWellness Panel™) is an in vitro diagnostic device and intended to be used for the quantitative determination of Glucose (GLU), Total Protein (TP), and Albumin (ALB) in lithium-heparinized venous whole blood in clinical laboratory or point-of-care settings. The Core Metabolic panel (part of the TruWellness Panel™) is an in vitro diagnostic test system that aids the physician in the diagnosis and treatment of the following disorders in adults 18 years of age or older: Glucose (GLU): Carbohydrate metabolism disorders, including diabetes mellitus, and idiopathic hypoglycemia, and of pancreatic islet cell carcinoma. Total Protein (TP): Variety of diseases involving the liver, kidney, or bone marrow, as well as other metabolic and nutritional disorders. Albumin (ALB): Numerous diseases involving primarily the liver and kidneys.
Device Story
The Core Metabolic panel is a component of the TruWellness Panel™ used on the TruVerus™ platform. It processes lithium-heparinized venous whole blood samples to provide quantitative measurements of Glucose, Total Protein, and Albumin. The system uses a Single-Use Consumable Kit (Disc and Support Pack) containing dried reagents and internal controls. Operation is automated, requiring no specialized skills or external calibration; the system performs photometric/colorimetric analysis. Results are displayed to the clinician to assist in diagnosing metabolic, liver, kidney, and bone marrow disorders. The device is intended for point-of-care or clinical laboratory settings. By providing rapid, integrated chemistry results from a single whole blood sample, it facilitates timely clinical decision-making and patient management.
Clinical Evidence
Bench testing only. Performance validated via precision studies (CLSI EP05-A3), linearity (CLSI EP06), interference (CLSI EP07/EP37), and method comparison (CLSI EP09c) against legally marketed comparators. Method comparison (N=341-345) showed high correlation (R=0.964-0.998) and acceptable bias at medical decision points. Analytical measuring ranges and LoQ established per CLSI EP17-A2.
Technological Characteristics
Single-use consumable kit containing dried reagents (HK, G-6-P-DH, Bromocresol Green, Copper Sulfate). Photometric/colorimetric sensing principle. Energy source: TruVerus™ analyzer. Dimensions/form factor: Single-use cartridge with integrated pipette tips and dilution wells. Connectivity: Integrated with TruVerus™ system. Sterilization: Not specified. Software: Embedded firmware for automated processing.
Indications for Use
Indicated for adults 18+ years for quantitative determination of Glucose, Total Protein, and Albumin in lithium-heparinized venous whole blood to aid diagnosis/treatment of carbohydrate metabolism disorders (diabetes, hypoglycemia, pancreatic islet cell carcinoma), liver/kidney/bone marrow diseases, and nutritional disorders. Contraindicated for use above 2,000 m altitude.
Regulatory Classification
Identification
An albumin test system is a device intended to measure the albumin concentration in serum and plasma. Albumin measurements are used in the diagnosis and treatment of numerous diseases involving primarily the liver or kidneys.
Abaxis Piccolo® Primary Health Panel Reagent Rotor (K950164)
Related Devices
K251058 — Tru Kidney Health Test Panel; Tru Analyzer · Truvian Health · Nov 26, 2025
K251074 — Tru Liver Health Test Panel · Truvian Health · Dec 19, 2025
Submission Summary (Full Text)
{0}
FDA
U.S. FOOD & DRUG
ADMINISTRATION
# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY
ASSAY ONLY
## I Background Information:
A 510(k) Number
K251157
B Applicant
Truvian Health
C Proprietary and Established Names
Core Metabolic
D Regulatory Information
| Product Code(s) | Classification | Regulation Section | Panel |
| --- | --- | --- | --- |
| CIX | Class II | 21 CFR 862.1035 - Albumin Test System | CH - Clinical Chemistry |
| CFR | Class II | 21 CFR 862.1345 - Glucose test system | CH - Clinical Chemistry |
| CEK | Class II, 510(k) exempt* | 21 CFR 862.1635 - Total Protein test system | CH - Clinical Chemistry |
* Total Protein meets the limitations of exemptions per 21 CFR § 862.9 (c)(9)
## II Submission/Device Overview:
A Purpose for Submission:
New Device
B Measurand:
Glucose (GLU), Total Protein (TP), Albumin (ALB)
Food and Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993-0002
www.fda.gov
{1}
C Type of Test:
Quantitative, photometric/colorimetric
III Intended Use/Indications for Use:
A Intended Use(s):
See Indications for Use below.
B Indication(s) for Use:
The Core Metabolic panel is part of the TruWellness Panel™ and is intended for use on the TruVerus™. The Core Metabolic panel (part of the TruWellness Panel™) is an in vitro diagnostic device and intended to be used for the quantitative determination of Glucose (GLU), Total Protein (TP), and Albumin (ALB) in lithium-heparinized venous whole blood in clinical laboratory or point-of-care settings.
The Core Metabolic panel (part of the TruWellness Panel™) is an in vitro diagnostic test system that aids the physician in the diagnosis and treatment of the following disorders in adults 18 years of age or older:
- Glucose (GLU): Carbohydrate metabolism disorders, including diabetes mellitus, and idiopathic hypoglycemia, and of pancreatic islet cell carcinoma.
- Total Protein (TP): Variety of diseases involving the liver, kidney, or bone marrow, as well as other metabolic and nutritional disorders.
- Albumin (ALB): Numerous diseases involving primarily the liver and kidneys.
C Special Conditions for Use Statement(s):
Rx - For Prescription Use Only
The instrument was validated up to an altitude of 2,000 m (6562 ft). The Core Metabolic panel should not be used above 2,000 m (6562 ft).
D Special Instrument Requirements:
TruVersus™
K251157 - Page 2 of 12
{2}
IV Device/System Characteristics:
# A Device Description:
Core Metabolic contains the following reagents:
| Core Metabolic | Quantity per single-use consumable kit |
| --- | --- |
| Adenosine-5'-Triphosphate | 17.5 μg |
| Bromocresol Green | 4.8 μg |
| Copper Sulfate | 38.4 μg |
| Glucose-6-Phosphate Dehydrogenase | 0.4 mU |
| Hexokinase | 0.4 mU |
| Magnesium Aspartate | 0.1 μg |
| Nicotinamide Adenine Dinucleotide | 0.3 μg |
| Potassium Iodide | 99.9 μg |
The Single-Use Consumable Kit (TruWellness Panel™) houses all the components needed to process as well as analyze samples on the TruVerus™, including dried reagents, internal process control solutions, barcodes that manage the identity of the kit lot (e.g., Disc and Support Pack ID), calibration information, dilution buffers, and single-use plastic pipette tips. It also serves as a waste container which the user discards at the end of the run. The Support Pack contains a feature to accept a standard 4 mL blood tube. The Support Pack also houses 22 pipette tips for transferring and mixing samples and reagents, 11 dilution wells to support reagent processing activities within the test system (e.g., sample dilution, reagent dilution, rehydration of dried reagents), and 6 x 2 ml tubes that contain additional wet and dry reagents.
# B Principle of Operation:
Glucose: This assay measures glucose by employing the hexokinase (HK) glucose-6-phophate dehydrogenase (G-6-P-DH) method. Glucose is phosphorylated by adenosine-5'-triphosphate (ATP) in a reaction catalyzed by HK. The glucose-6-phosphate (G-6-P) formed is oxidized by G-6-P-DH to 6-phosphogluconate (6-PG), producing an equimolar amount of NADH. The amount of NADH formed in the second reaction is directly proportional to the concentration of glucose in the sample and is measured via absorbance at $340~\mathrm{nm}$ .
$$
\mathrm {G l u c o s e} + \mathrm {A T P H K} \rightarrow \mathrm {G - 6 - P} + \mathrm {A D P G - 6 - P} + \mathrm {N A D G - 6 - P - D H} \rightarrow 6 - \mathrm {P G} + \mathrm {N A D H}
$$
Total Protein: This assay measures total protein via the biuret method. In alkaline solution, cupric ions react with all compounds with two amide or peptide bonds linked either directly or through an intermediate carbon atom to form a violet-colored complex. The absorbance of this colored complex at $550~\mathrm{nm}$ is directly proportional to protein concentration.
$$
\mathrm {C u} + \text {P r o t e i n} \rightarrow \text {P u r p l e C o m p l e x}
$$
Albumin: This assay measures albumin levels by monitoring the formation of a complex between albumin and bromocresol green (BCG). The blue-green color produced in the reaction is
K251157 - Page 3 of 12
{3}
measured at 628 nm, and its absorbance is directly proportional to the concentration of albumin in the sample.
BCG + Albumin → Green Complex
V Substantial Equivalence Information:
A Predicate Device Name(s):
Medicon Hellas Albumin, Medicon Hellas Glucose
B Predicate 510(k) Number(s):
K200898
C Comparison with Predicate(s):
Primary Device Predicate
| Device & Predicate Device(s): | K251157 | K200898 |
| --- | --- | --- |
| Device Trade Name | Core Metabolic | Medicon Hellas Albumin, Medicon Hellas Glucose |
| General Device Characteristic Similarities | | |
| Intended Use/Indications For Use | Quantitative determination of albumin and glucose | Same |
| General Device Characteristic Differences | | |
| Specimen Type | Lithium-heparinized venous whole blood | Serum and urine |
Secondary Device Predicate
| Device & Predicate Device(s): | K251157 | K950164 |
| --- | --- | --- |
| Device Trade Name | Core Metabolic | Abaxis Piccolo® Primary Health Panel Reagent Rotor |
| General Device Characteristic Similarities | | |
| Intended Use/Indications For Use | Quantitative determination of total protein | Same |
| General Device Characteristic Differences | | |
| Specimen Type | Lithium-heparinized venous whole blood | Heparinized whole blood, heparinized plasma, or serum |
| Analytical Measuring Range | 2.0-11.5 mg/dL | 2-14 mg/dL |
K251157 - Page 4 of 12
{4}
VI Standards/Guidance Documents Referenced:
- Clinical & Laboratory Standards Institute (CLSI) EP05-A3 Evaluation of Precision of Quantitative Measurement Procedures; Approved Guideline - Third Edition
- CLSI EP06 Evaluation of the Linearity of Quantitative Measurement Procedures - Second Edition
- CLSI EP07 Interference Testing in Clinical Chemistry - Third Edition
- CLSI EP09c Measurement Procedure Comparison and Bias Estimation Using Patient Samples -Third Edition
- CLSI EP17-A2 Evaluation of Detection Capability for Clinical Laboratory Measurement Procedures; Approved Guideline - Second Edition
- CLSI EP37 Supplemental Tables for Interference Testing in Clinical Chemistry - First Edition
VII Performance Characteristics (if/when applicable):
1. Analytical Performance:
1. Precision/Reproducibility:
A. Reproducibility with control material: This study was designed according to CLSI EP05-A3. The study was conducted at 3 external point-of-care (POC) sites, with 3 analyzers per site and at least 3 POC operators per site. Testing was conducted over 5 days using a single lot of consumables and consisted of at least 3 levels of control samples, with 6 replicates run per day, 3 run in the morning and 3 run in the afternoon (each replicate is run on an individual instrument). Overall, there were at least 90 replicates for each precision control sample level (3 sites x 1 sample x 3 replicates per run x 2 runs per day x 5 days = 90 replicates). The total precision as well as within run, between day, and between site precision were estimated. Results are displayed below:
| Analyte | N | Mean | Within-Run | | Between-Run | | Between-Day | | Between-Site | | Total Precision | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| | | | SD | CV% | SD | CV% | SD | CV% | SD | CV% | SD | CV% |
| GLU | 92 | 60 | 1 | 1.9 | 0 | 0.0 | 0 | 0 | 0.1 | 0.2 | 1 | 1.9 |
| GLU | 91 | 114 | 3 | 2.7 | 0 | 0.0 | 0 | 0 | 1 | 1.2 | 3 | 3.0 |
| GLU | 92 | 274 | 5 | 1.7 | 0 | 0.0 | 1 | 0.3 | 1 | 0.3 | 5 | 1.7 |
| GLU | 90 | 399 | 6 | 1.6 | 2 | 0.4 | 1 | 0.4 | 4 | 1.1 | 8 | 2.0 |
| TP | 92 | 4.8 | 0.1 | 2.3 | 0.0 | 0.0 | 0.04 | 0.8 | 0.1 | 1.9 | 0.1 | 3.1 |
| TP | 91 | 6.5 | 0.2 | 2.4 | 0.0 | 0.0 | 0.0 | 0.0 | 0.04 | 0.6 | 0.2 | 2.5 |
| TP | 92 | 7.8 | 0.1 | 1.2 | 0.0 | 0.0 | 0.04 | 0.5 | 0.03 | 0.4 | 0.1 | 1.4 |
| ALB | 92 | 3.1 | 0.1 | 1.8 | 0.0 | 0.0 | 0.0 | 0.0 | 0.1 | 1.9 | 0.1 | 2.6 |
| ALB | 91 | 4.3 | 0.1 | 2.3 | 0.0 | 0.5 | 0.0 | 0.0 | 0.1 | 1.3 | 0.1 | 2.7 |
| ALB | 92 | 5.0 | 0.1 | 1.2 | 0.0 | 0.0 | 0.0 | 0.3 | 0.1 | 1.1 | 0.1 | 1.7 |
K251157 - Page 5 of 12
{5}
B. Whole Blood Precision: This study was evaluated at five sites collecting samples from the intended use population with normal and abnormal analyte levels. At least 20 subjects were enrolled at each site and for each test subject, eight replicates were measured across four instruments and two operators. The SD and CV% precision estimates were calculated by pooling subject standard deviations and/or % CV for predefined Low, Medium or High sub-intervals selected to represent normal and abnormal regions of the analytical measuring range and encompassing medical decision levels. Pooled imprecision is considered a representative estimate of reproducibility that include variability of sites, instruments, operators, days and repeatability.
| Analyte | Range | Subject (N) | Replicate (N) | Mean | SD | CV% |
| --- | --- | --- | --- | --- | --- | --- |
| GLU (mg/dL) | 10–100 | 45 | 348 | 88.5 | 2.91 | 3.4% |
| GLU (mg/dL) | 100–150 | 28 | 221 | 118.4 | 3.61 | 3.0% |
| GLU (mg/dL) | 150–450 | 44 | 334 | 205.2 | 4.23 | 2.0% |
| TP (g/dL) | 2.0–6.0 | 27 | 210 | 5.27 | 0.165 | 3.2% |
| TP (g/dL) | 6.0–8.0 | 83 | 639 | 6.98 | 0.166 | 2.4% |
| TP (g/dL) | 8.0–11.5 | 3 | 24 | 8.48 | 0.269 | 3.3% |
| ALB (g/dL) | 1.0–3.5 | 53 | 409 | 2.92 | 0.108 | 3.7% |
| ALB (g/dL) | 3.5–5.0 | 63 | 487 | 4.22 | 0.108 | 2.6% |
| ALB (g/dL) | 5.0–6.5 | 1 | 8 | 5.15 | 0.076 | 1.5% |
C. Whole Blood Between Lot Precision: 22 subjects consisting of 6 healthy individuals and 16 individuals in various disease states were enrolled at a single point-of-care clinical site and tested across 3 lots of the candidate panel on 3 TruVersus™ analyzers by 2 operators. A total of 6 replicates per subject participant, 2 replicates per lot, were tested by 1 of the 2 operators. Results below present the number of subjects evaluated, the % difference from the median, and pooled %CV by lot and by analyte.
| Analyte | Subjects per Lot | Median relative difference % | | | Pooled % CV | | |
| --- | --- | --- | --- | --- | --- | --- | --- |
| | | Lot-A | Lot-B | Lot-C | Lot-A | Lot-B | Lot-C |
| GLU | 22 | 0.8% | -0.5% | -0.3% | 3.6% | 2.8% | 2.8% |
| TP | 22 | -0.2% | 0.9% | -0.7% | 2.5% | 2.4% | 1.9% |
| ALB | 22 | 0.6% | 0.5% | 0.2% | 2.4% | 1.5% | 1.6% |
D. Control Between Lot Reproducibility: To evaluate the lot-to-lot reproducibility, 3 control samples (low, medium and high) were tested on 3 TruVersus™ analyzers using 3 lots of the candidate test. Each sample level was tested with at least 5 replicates on one TruVersus™ analyzer each day with each single-use consumable kit lot over 3 days to achieve a minimum of 15 replicates per sample level and kit lot. Results are displayed below:
K251157 - Page 6 of 12
{6}
| Analyte | QC Level | N | Mean | Lot Mean | | | %CV | | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| | | | | Lot 1 | Lot 2 | Lot 3 | Lot 1 | Lot 2 | Lot 3 |
| GLU (mg/dL) | Low | 45 | 60.33 | 60.71 | 59.92 | 60.36 | 2.60% | 2.20% | 1.60% |
| | Medium | 45 | 117.88 | 118.34 | 117.45 | 117.86 | 2.60% | 2.10% | 2.40% |
| | High | 45 | 278.08 | 278.66 | 276.79 | 278.79 | 2.40% | 3.00% | 1.90% |
| TP (g/dL) | Low | 45 | 4.73 | 4.79 | 4.68 | 4.71 | 2.10% | 2.60% | 1.60% |
| | Medium | 45 | 6.53 | 6.53 | 6.49 | 6.56 | 1.70% | 2.70% | 1.50% |
| | High | 45 | 7.79 | 7.93 | 7.63 | 7.80 | 6.10%* | 3.70% | 1.70% |
| ALB (g/dL) | Low | 45 | 3.11 | 3.14 | 3.10 | 3.08 | 1.70% | 2.50% | 1.90% |
| | Medium | 45 | 4.27 | 4.30 | 4.26 | 4.26 | 1.20% | 1.20% | 1.40% |
| | High | 45 | 4.95 | 4.99 | 4.90 | 4.96 | 1.20% | 2.80% | 1.20% |
* A replicate for TP Lot-1 was identified as a statistical outlier and the table below provides the results without the outlier included.
| Analyte | QC Level | N | Mean | SD | %CV |
| --- | --- | --- | --- | --- | --- |
| TP (g/dL) | High | 14 | 7.81 | 0.10 | 1.31 |
2. Linearity:
A linearity study was conducted in accordance with CLSI EP06 - 2nd Edition. Linearity was assessed with multiple panels of samples. The linear range for the GLU, TP, and ALB assays were determined by testing 9 samples containing varying concentrations of analyte. Each sample was tested in quadruplicate, using 1 kit lot, across 4 TruVersus™ analyzers, on a single day. The results of the study supported the claimed linear interval for GLU, TP, and ALB. Within the claimed analytical measuring range, the maximum observed deviation for GLU, TP, and ALB are 5%, 4%, and 9%, respectively.
3. Analytical Specificity/Interference:
This study was designed according to CLSI EP07-A3 and CLSI EP37, to verify and validate the effect of potentially interfering exogenous and endogenous substances on the GLU, TP, and ALB assays. Testing was performed with contrived lithium heparin whole blood samples from a healthy donor population, with 2 analyte concentration levels targeted based on recommended levels in CLSI EP07. Potentially interfering substances were prepared in the appropriate diluent and then added to the low and high contrived whole blood samples to create test samples. For any substances identified as an interferent, dose response testing and analysis was conducted to assess the highest concentration limit below which no significant interference was observed. The results are displayed below:
K251157 - Page 7 of 12
{7}
Endogenous compounds:
| Substance (mg/dL) | Max Concentration without Interference | | |
| --- | --- | --- | --- |
| | GLU | TP | ALB |
| Triglycerides | 1,140 | 1,140 | 1,140 |
| Hemoglobin | 346 | 1,151 | 1,050 |
| Conjugated Bilirubin | 26.0 | 15.1 | 40 |
| Unconjugated Bilirubin | 37.9 | 17.5 | 40 |
Exogenous compounds:
| Substance | Max Concentration without Interference | | | |
| --- | --- | --- | --- | --- |
| | GLU | TP | ALB | Units |
| Acetaminophen | 15.6 | 15.6 | 15.6 | mg/dL |
| Acetylsalicylic acid | 3.0 | 3.0 | 3.0 | mg/dL |
| Ampicillin | 7.5 | 7.5 | 7.5 | mg/dL |
| Cefoxitin | 660 | 495 | 660 | mg/dL |
| Cyclosporine | 0.18 | 0.18 | 0.18 | mg/dL |
| Doxycycline | 1.8 | 1.8 | 1.8 | mg/dL |
| Heparin | 3,300 | 3,300 | 3,300 | U/L |
| Ibuprofen | 21.9 | 21.9 | 21.9 | mg/dL |
| Levodopa (L-Dopa) | 0.75 | 0.75 | 0.75 | mg/dL |
| Methyldopa | 2.25 | 2.25 | 2.25 | mg/dL |
| Metronidazole | 12.3 | 12.3 | 12.3 | mg/dL |
| Phenylbutazone | 24.08 | 32.1 | 32.1 | mg/dL |
| Rifampicin | 4.8 | 4.8 | 4.8 | mg/dL |
| Theophylline | 6.0 | 6.0 | 6.0 | mg/dL |
| Acetylcysteine | 15 | 15 | 15 | mg/dL |
| Ascorbic Acid | 5.25 | 5.25 | 5.25 | mg/dL |
| Calcium Dobesilate | 6.0 | 6.0 | 6.0 | mg/dL |
| α-ketoglutarate | N/A | N/A | 5 | mg/dL |
| α-1-acid Glycoprotein | N/A | N/A | 500 | mg/dL |
| Acetoacetate | 20 | N/A | 20 | mg/dL |
| Albumin | 15,000 | N/A | N/A | mg/dL |
| Caffeine | 10.8 | 10.8 | 10.8 | mg/dL |
| Calcium | N/A | N/A | 20 | mg/dL |
| Cefotaxime | N/A | N/A | 52.8 | mg/dL |
| Cephalothin | 180 | 180 | 180 | mg/dL |
| Chloramphenicol | 7.8 | N/A | N/A | mg/dL |
| Cimetidine | 3.0 | 3.0 | 3.0 | mg/dL |
| Desacetylcefotaxime | N/A | N/A | 6 | mg/dL |
| Dopamine | 0.0621 | N/A | N/A | mg/dL |
K251157 - Page 8 of 12
{8}
| Substance | Max Concentration without Interference | | | |
| --- | --- | --- | --- | --- |
| | GLU | TP | ALB | Units |
| Eltrombopag | 30 | N/A | N/A | mg/dL |
| Glutathione | 92 | N/A | N/A | mg/dL |
| Hydrochlorothiazide | 0.113 | N/A | N/A | mg/dL |
| IgG | 4,500 | N/A | N/A | mg/dL |
| Lidocaine | 1.5 | N/A | N/A | mg/dL |
| Lithium | 2.5 | N/A | N/A | mg/dL |
| Methotrexate | 136 | 68 | N/A | mg/dL |
| Nitrofurantoin | 0.213 | 0.213 | 0.213 | mg/dL |
| Oxaloacetate | 14.62 | N/A | N/A | mg/dL |
| p-Aminosalicylic Acid | N/A | N/A | 46.5 | mg/dL |
| Penicillin G | N/A | 900 | 1,800 | mg/dL |
| Phenytoin | 6 | N/A | N/A | mg/dL |
| Proline | N/A | N/A | 4 | mg/dL |
| Pyruvate | 5 | N/A | N/A | mg/dL |
| Salicylic Acid | 2.86 | 2.86 | 2.86 | mg/dL |
| Sulfadiazine | N/A | 25.5 | 25.5 | mg/dL |
| Sulfanilamide | N/A | 50 | 50 | mg/dL |
| Uric acid | 23.5 | N/A | N/A | mg/dL |
# 4. Detection Limit and Assay Reportable Range:
The Limit of Blank (LoB) was determined in accordance with the classical approach provided in section 5.3 of CLSI EP17-A2. LoB was determined by running four (4) blank samples on 4 TruVersus™ analyzers over 3 days of testing with 5 replicates per sample for a total of 60 tests for each single-use consumable kit lot.
The LoD (Limit of Detection) and LoQ (Limit of Quantitation) were determined based on recommendations from CLSI EP17-A2. The LoQ corresponds to the lowest concentration of analyte in a sample that had a $\mathrm{CV} \leq 20\%$ .
The results of the studies are summarized below and support the claimed range of analytes.
| Assay | Reportable Analytical Measuring Range | LoQ |
| --- | --- | --- |
| Glucose | 10 - 450 mg/dL | 6 mg/dL |
| Total Protein | 2.0 - 11.5 g/dL | 0.7 g/dL |
| Albumin | 1.0 - 6.5 g/dL | 0.2 g/dL |
# 5. Traceability, Stability, Expected Values (Controls, Calibrators, or Methods):
A. Traceability: Core Metabolic assay calibrators are traceable to the reference materials in the table below:
K251157 - Page 9 of 12
{9}
| Assay | Traceability |
| --- | --- |
| GLU | ID/MS→NIST SRM 1950 |
| TP | NIST SRM1950 |
| ALB | ERM-DA470k→Verichem 9466F |
B. Sample Stability: Sample Stability studies performed demonstrate that samples measured on the candidate tests are stable for 60-minutes post-collection.
C. Altitude Effects: An altitude study was conducted to evaluate performance of GLU, TP, and ALB at sea-level and the results are acceptable. This study supports the device performance at altitudes of up to 2000 m.
6. Assay Cut-Off:
Not applicable.
B. Comparison Studies:
1. Method Comparison with Predicate Device:
A method comparison study was conducted to establish the performance characteristics of the Core Metabolic on the TruVersus™ when used to analyze lithium heparin whole blood clinical specimens. Whole blood samples from apparently healthy patient and from patients with acute or chronic health conditions were collected from 5 point-of-care sites, tested by at least 3 point-of-care operators per site, on at least 4 TruVersus™ per site, using 5 lots of reagents. Results compared against a legally marketed comparator are displayed below. The data demonstrated that accuracy was consistent across sites and differences at the medical decision points were reviewed and found acceptable.
| Analyte | N | Range | Slope | Intercept | R |
| --- | --- | --- | --- | --- | --- |
| GLU (mg/dL) | 341 | 50-441 | 0.95
(0.94, 0.96) | 3.20
(1.73, 4.38) | 0.998 |
| TP (g/dL) | 344 | 2.5-11.5 | 1.00
(1.00, 1.05) | -0.20
(-0.53, -0.10) | 0.964 |
| ALB (g/dL) | 345 | 1.5-6.0 | 1.00
(1.00, 1.08) | 0.00
(-0.28, 0.10) | 0.969 |
2. Matrix Comparison:
Not applicable.
K251157 - Page 10 of 12
{10}
K251157 - Page 11 of 12
C. Clinical Studies:
1. Clinical Sensitivity:
Not applicable.
2. Clinical Specificity:
Not applicable.
3. Clinical Cut-Off:
Not applicable.
4. Other Clinical Supportive Data (When 1. and 2. Are Not Applicable):
Not applicable.
D. Expected Values/Reference Range:
The reference ranges are based on scientific literature.
| Analyte | Reference Interval | Units |
| --- | --- | --- |
| Glucose^{1} | Adult fasting plasma glucose: <100 mg/dL (5.55 mmol/L) | mg/dL (mmol/L) |
| | <140 mg/dL (7.77 mmol/L) 1-2 hours after meals. | |
| Total Protein^{2,3} | 6–7.8 | g/dL |
| Albumin^{2,3} | 3.5–5 | g/dL |
1 ElSayed NA, Aleppo G, Aroda VR, et al, American Diabetes Association. 2. Classification and diagnosis of diabetes: Standards of Care in Diabetes-2023. Diabetes Care 2023; 46 (Suppl. 1):S19-S40
2 Henry, J.B., Clinical Diagnosis and Management by Laboratory Methods. 20th Ed. W.B. Saunders Co., Philadelphia (1976).
3 Tietz, N.W., Textbook of Clinical Chemistry. 2nd Ed. W.B. Saunders Co., Philadelphia (1994).
{11}
VIII Proposed Labeling:
The labeling does support the finding of substantial equivalence for this device.
IX Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
K251157 - Page 12 of 12
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.