ADDITIONAL IMS ASSAYS FOR THE BAYER ADVIA IMS SYSTEM
Applicant
Bayer Corp.
Product Code
CIJ · Clinical Chemistry
Decision Date
Jul 22, 1999
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 862.1070
Device Class
Class 2
Indications for Use
The Bayer ADVIA IMS Amylase assay is an in vitro diagnostic device intended to measure amylase activity in human serum, plasma or urine. Such measurements are used as an aid primarily in the diagnosis and treatment of pancreatitis (inflammation of the pancreas). The Bayer ADVIA IMS Cortisol assay is an in vitro diagnostic device intended to quantitatively measure cortisol in human serum. Measurements of cortisol are used as an aid in the diagnosis and treatment of disorders of the adrenal gland. The Bayer ADVIA IMS Iron assay is an in vitro diagnostic device intended to measure iron in human serum of plasma. Measurements of iron are used as an aid in the diagnosis, monitoring and treatment of a variety of diseases including iron deficiency anemias, hemochromatosis, hemosiderosis from excessive iron intake, and hemolytic anemias. The Baver ADVIA IMS Thyroxine assay is an in vitro diagnostic device intended to measure thyroxine (T4), both protein bound and free, in human serum and plasma. Measurements of T4 are used as an aid in the diagnosis and treatment of thyroid diseases. The Baver ADVIA IMS Free Thyroxine (FT4) assay is an in vitro diagnostic device intended to quantitatively measure free thyroxine in human serum. Measurements of free thyroxine in conjunction with other thyroid tests and clinical indicators are used as an aid in the diagnostic discrimination and assessment of thyroid diseases. The Bayer ADVIA IMS Triiodothyronine (T3) assay is an in vitro diagnostic device intended to quantitatively measure triiodothyronine (T3) in human serum. Measurements of triiodothyronine, in conjunction with other thyroid tests and clinical indicators, are used as an aid in the diagnostic discrimination and assessment of thyroid diseases. The Bayer ADVIA IMS Free Triiodothyronine (FT3) assay is an in vitro diagnostic device intended to quantitatively measure free triiodothyronine in human serum. Measurements of free triodothyronine, in conjunction with other first-line thyroid tests such as Thyroid Stimulating Hormone (TSH) and Free Thyroxine (Free T4), as well as other clinical indicators, are used as an aid in the diagnostic discrimination and assessment of thyroid diseases. The Bayer ADVIA IMS T Uptake (TUP) assay is an in vitro diagnostic device intended to quantitatively measure the total amount of available binding sites for thyroid hormone on the thyroxine-binding proteins, globulin, pre-albumin, and albumin in human serum. Measurements of T Uptake, in conjunction with other thyroid tests and clinical indicators, are used as an aid in the diagnostic discrimination and assessment of thyroid diseases. The Bayer ADVIA IMS Urea Nitrogen (BUN) method is an in vitro diagnostic device intended to measure urea nitrogen in human serum, plasma and urine. Such measurements are used as an aid in the diagnosis and treatment of certain renal and metabolic diseases.
Device Story
ADVIA IMS is an automated in vitro diagnostic system for clinical chemistry analysis. It processes human serum, plasma, or urine samples to measure specific analytes (Amylase, Cortisol, Iron, T4, Free T4, T3, Free T3, T-Uptake, Urea Nitrogen). The system utilizes reagent-based assays to produce quantitative results. Operated by laboratory technicians in clinical settings, the system automates the measurement process, providing data that clinicians use to diagnose and monitor conditions like pancreatitis, thyroid dysfunction, and renal disease. The output is a concentration value or ratio, which aids in clinical decision-making regarding patient treatment and disease management.
Clinical Evidence
Bench testing only. Performance validated through precision studies (Total CV%), analytical range determination, and correlation studies against predicate devices (n=50-72 per assay). Interference testing conducted for common substances (hemoglobin, bilirubin, lipids, etc.) to assess percentage change in analyte measurement.
Technological Characteristics
Automated in vitro diagnostic system; reagent-based colorimetric or immunoassay methods; serum/plasma/urine sample input; standalone system; quantitative measurement of biochemical analytes.
Indications for Use
Indicated for patients requiring diagnostic assessment of pancreatitis, adrenal gland disorders, iron-related conditions (anemia, hemochromatosis), thyroid disorders (hyperthyroidism, thyroid function assessment), and renal/metabolic diseases. Used for quantitative measurement of analytes in human serum, plasma, or urine.
Regulatory Classification
Identification
An amylase test system is a device intended to measure the activity of the enzyme amylase in serum and urine. Amylase measurements are used primarily for the diagnosis and treatment of pancreatitis (inflammation of the pancreas).
Predicate Devices
Bayer RA-XT Amylase method
Immuno 1 Cortisol assay
Technicon CHEM 1 Iron-II method
Technicon Immuno 1 T4 method
Bayer Immuno 1 FREE T4 Assay
Bayer Immuno 1 T3 Assay
Bayer Immuno 1 Free T3 Assay
Bayer Immuno 1 T-Uptake Assay
Technicon CHEM 1 Urea Nitrogen
Submission Summary (Full Text)
{0}------------------------------------------------
## Amylase (AMY) Method for the ADVIA IMS Systems
Listed below is a comparison of the performance of the Bayer ADVIA Amylase (AMY) method and a similar device that was granted clearance of substantial equivalence (Bayer RA-XT Amylase method). The information was extracted from the Bayer ADVIA AMY method and Bayer RA-XT Amylase method sheet.
### INTENDED USE
The Bayer ADVIA IMS Amylase (AMY) assay is an in-vitro diagnostic device intended to measure AMY in human serum, plasma or urine. Such measurements are used in the diagnosis and treatment of acute pancreatitis. This diagnostic method is not intended for -use on any other diagnostic system .- - -
| AMY METHOD: | ADVIA IMS | | RA-XT | | |
|------------------------------------------------|----------------------|------------------|----------------|------|--|
| Part Number: | Reagents B41-3721-23 | | T01-3508-01 | | |
| Analytical Range: | 0 to 1500 U/L | | 10 to 3400 U/L | | |
| Precision (Total): | mean<br>(U/L) | % CV | mean<br>(U/L) | % CV | |
| | 56 | 2.5 | 51 | 2.2 | |
| | 113 | 2.1 | 179 | 1.0 | |
| | 430 | 1.5 | 373 | 0.6 | |
| Regression Equation:<br>(serum) | y = 1.01x - 5.7 | | | | |
| | where: y | = ADVIA IMS | | | |
| | x | = RA-XT | | | |
| | n | = 72 | | | |
| | r | = 0.999 | | | |
| | Sy.x | = 11.9 | | | |
| | range | = 19 to 1318 U/L | | | |
| Regression Equation:<br>(plasma qualification) | y = 1.00x - 0.2 | | | | |
| | where: y | = plasma | | | |
| | x | = serum | | | |
| | n | = 60 | | | |
| | r | = 0.999 | | | |
| | Sy.x | = 1.0 | | | |
| | range | = 18 to 106 U/L | | | |
{1}------------------------------------------------
# Interference
| | Interfering Substance<br>Concentration | AMY<br>(U/L) | Effect<br>% Change |
|--------------------------|----------------------------------------|--------------|--------------------|
| Hemolysis (Hemoglobin) | 500 mg/dL | 112 | +2 |
| Bilirubin (conjugated) | 20 mg/dL | 113 | -4 |
| Bilirubin (unconjugated) | 25 mg/dL | 112 | -2 |
| Lipemia (Triglycerides) | 500 mg/dL | 113 | 3 |
Urine Samples:
1
| AMY METHOD: | ADVIA IMS | RA-XT | | |
|--------------------|----------------------|----------------|------------|------|
| Part Number: | Reagents B41-3721-23 | T01-3508-01 | | |
| Analytical Range: | 0 to 1500 U/L | 10 to 3400 U/L | | |
| Precision (Total): | mean (U/L) | % CV | mean (U/L) | % CV |
| | 50 | 1.3 | 57 | 3.4 |
| | 218 | 1.1 | 172 | 1.3 |
| | 493 | 1.6 | 510 | 1.3 |
(urine)
| where: | y | = ADVIA IMS |
|--------|-------|------------------|
| | x | = RA-XT |
| | n | = 72 |
| | r | = 0.999 |
| | Sy.x | = 9.7 |
| | range | = 17 to 1465 U/L |
Interference
:
| | Interfering Substance<br>Concentration | AMY<br>(U/L) | Effect<br>% Change |
|---------------|----------------------------------------|--------------|--------------------|
| Ascorbic Acid | 400 mg/dL | 115 | 1 |
| Acetaminophen | 50 mg/dL | 116 | -1 |
| Salicylate | 500 mg/dL | 116 | -2 |
Rqm. Jr. RA
5/21/99
{2}------------------------------------------------
# Cortisol Method for Bayer ADVIA® IMSTM
Listed below is a comparison of the performance between the ADVIA Cortisol method and a similar device that was granted clearance of substantial equivalence (Immuno 1 Cortisol assay). The information used in the Summary of Safety and Effectiveness was extracted from the ADVIA Cortisol method sheet and the Immuno 1 Cortisol method sheet.
#### INTENDED USED
This in viro method is intended to quantitatively measure cortisol, a hormone secreted by the human adrenal gland, in human serum using Bayer cortisol reagents on a Bayer ADVIA Modular System. Measurements of cortisol are used as a direct indicator of adrenal status or an indirect monitor of pituitary function.
| METHOD | | Immuno 1 Cortisol<br>(predicate device) | ADVIA Cortisol |
|--------------------------|--|---------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------|
| Part No. | | Reagents T01-2910-01<br>Calibrators T03-3252-01 | Reagents B42-3899-21<br>Calibrators B43-3931-01 |
| Minimum Detectable Conc. | | 0.2 µg/dL | 0.1 µg/dL |
| Precision (Total CV%) | | 3.2 µg/dL 7.9%<br>20.1 µg/dL 4.5%<br>33.3 µg/dL 4.7% | 4.4 µg/dL 6.0%<br>17.3 µg/dL 5.0%<br>37.5 µg/dL 3.8% |
| Correlation | | y = 1.013 x - 0.142<br><br>where<br>y = Bayer ADVIA Modular System<br>x = Bayer Immuno 1 System<br>n = 57<br>r = 0.996<br>Syx = 0.961 µg/dL | |
| Interfering<br>Substance | Interfering Substance<br>Concentration | | Analyte<br>Concentration | | Effect |
|--------------------------|----------------------------------------|---------|--------------------------|---------|--------|
| | SI Units | (mg/dL) | (nmol/L) | (µg/dL) | (%) |
| Hemoglobin | 10.0 g/L | 1000 | 466.4 | 16.9 | -4.8 |
| Lipids (Triglycerides) | 11.3 mmol/L | 1000 | 463.7 | 16.8 | -4.5 |
| Bilirubin | 171 µmol/L | 25 | 469.2 | 17.0 | -1.9 |
| Urea Nitrogen | 71.4 mmol/L | 200 | 554.8 | 20.1 | -4.4 |
Interforing Subotanco
Gabriel J. Muncey, Jr.
Gabriel J. Muraca, Jr. Manager Regulatory Affairs Bayer Corporation 511 Benedict Avenue Tarrytown, New York 10591-5097
5/21/99
Date
Date
{3}------------------------------------------------
## Iron Method for the Bayer ADVIA Integrated Modular System (IMS)
Listed below is a comparison of the performance between the Bayer ADVIA IMS Iron method and a similar device that was granted clearance of substantial equivalence (Technicon CHEM 1 Iron-II method). The information used in the Summary of Safety and Effectiveness was extracted from the Bayer ADVIA IMS Iron method sheet and the CHEM 1 Iron-II method sheet.
### INTENDED USE
This in vitro method is intended to quantitatively measure iron (Fe) in human serum and plasma on the Bayer ADVIA IMS. Measurements of iron are used in the diagnosis, monitoring and treatment of a variety of diseases including iron deficiency anemias, hemochromatosis, hemosiderosis from excessive iron intake, and hemolytic anemias.
| METHOD | ADVIA IMS | CHEM 1 |
|--------------------------|--------------------------------------------------------------------------------------------------|---------------------------------------------------------|
| Part No. | Reagents<br>B41-3735-43<br>Calibrators<br>T03-1291-62 | T01-3328-53<br>T03-1291-62 |
| Analytical Range | 0 to 800 ug/dL | 0 to 1200 ug/dL |
| Precision (Total) | 2.8% @ 50.7 ug/dL<br>1.1% @ 230.1 ug/dL<br>0.7% @ 438.3 ug/dL | 1.9% @ 96 ug/dL<br>1.3% @ 211 ug/dL<br>1.2% @ 383 ug/dL |
| Correlation | Y=0.93X+10.8 ug/dL<br>Where<br>Y=ADVIA IMS<br>X=CHEM 1<br>N=65<br>r=0.998<br>Sy.x=9.75 ug/dL | |
| Plasma/Serum Equivalence | Y=0.98X+0.46 ug/dL<br>Where<br>Y = plasma<br>X= Serum<br>N = 56<br>r = 0.99<br>Sy.x = 3.12 ug/dL | |
Gabriel J. Murea, Jr. 5/21/99
{4}------------------------------------------------
Interfering Substances
| Bilirubin (unconjugated) 25 mg/dL | 7.0% | effect change @ 215 ug/dL Fe |
|-----------------------------------|--------|------------------------------|
| Bilirubin (conjugated) 25 mg/dL | -1.0% | effect change @ 221 ug/dL Fe |
| Hemoglobin (hemolysate) 500 mg/dL | 44.0% | effect change @ 217 ug/dL Fe |
| Lipemia (Triglycerides) 500 mg/dL | -20.0% | effect change @ 207 ug/dL Fe |
Gabriel J. Murace, Jr.
Gabriel J. Muraca, Jr. Manager Regulatory Affairs Bayer Corporation 511 Benedict Avenue Tarrytown, New York 10591-5097
5/21/99
---
Date
{5}------------------------------------------------
## T4 Method for the Bayer ADVIA® IMS Systems
·
Listed below is a comparison of the performance between the Bayer ADVIA IMS T4 method and a similar device that was granted clearance of substantial equivalence (Technicon Immuno 1® method). The information used in the Summary of Safety and Effectiveness was extracted from the ADVIA IMS T4 method sheet and the Immuno 1 method sheet.
## INTENDED USE
This in vitro method is intended to quantitatively measure T4 in human serum and plasma on the Bayer ADVIA IMS systems. Measurements of T4 are used to aid in the diagnosis and treatment of thyroid diseases.
| METHOD | | ADVIA IMS | Immuno 1 |
|--------------------|-------------|-------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------|
| ` Part No. | Reagents | B41-3855-42<br>B41-3855-43 | T01-3260-51 |
| | Calibrators | B46-4092-01 | T03-3174-01 |
| Minimum Det. Conc. | | 0.25 µg/dL | 0.4 µg/dL |
| Precision (Total) | | 6.0% @ 3.5µg/dL<br>5.1% @ 7.9 µg/dL<br>4.7% @14.9 µg/dL | 3.6% @ 4.7 µg/dL<br>2.6% @ 8.2 µg/dL<br>2.5% @ 15.7 µg/dL |
| Correlation | SERUM | y = 1.06 x +0.11<br>where<br>y = ADVIA IMS<br>x = Immuno 1<br>n = 72<br>r = 0.994<br>Syx = 0.65 µg/dL | |
| | PLASMA | y = 0.98X + 0.06<br>Where<br>y = Serum (ADVIA IMS)<br>x = Plasma (ADVIA IMS)<br>n = 24<br>r = 0.977<br>Syx = 0.36 | |
{6}------------------------------------------------
| Interference<br>Substance | Interfering<br>Substance<br>Concentration | Thyroxine<br>Concentration,<br>ug/dL | Effect<br>% Change |
|-----------------------------|-------------------------------------------|--------------------------------------|--------------------|
| Bilirubin<br>(unconjugated) | 25 mg/dL | 15.3 | 0 |
| Bilirubin<br>(conjugated) | 20 mg/dL | 15.2 | -2 |
| Hemoglobin | 600 mg/dL | 15.2 | -3 |
| Lipemia<br>(Triglycerides) | 1000 mg/dL | 15.0 | +3 |
RgM, Jr.
5/24/95
{7}------------------------------------------------
#### Free T4 Assay for Bayer ADVIA® Integrated Modular System
Listed below is a comparison of the performance between the ADVIA FREE T4 (Free Thyroxine) method, and a similar device that was granted clearance of substantial equivalence (Bayer Immuno 1 FREE T4 Assay). The information used in the Summary of Safety and Effectiveness was extracted from the ADVIA FREE T4 insert and the Immuno 1 FREE T4 Assay method sheet.
#### INTENDED USED
This in vitro method is intended to quantitatively measure the unbound FREE T4 in human serum by using the ADVA FREE T4 Assay on the Bover ADVIA® Interrated Modular System. Less than 0.05% of the Total T4 is unbound to the serum proteins and it is this Free T4 fraction that directly regulates metabolic activity. Thus Free T4 measurements are generally used in the direct or first line diagnosis of thyroid disorders.
| METHOD | ADVIA FREE T4 Assay | | | Immuno 1 FREE T4 Assay<br>(predicate Device) | | |
|--------------------------|------------------------------------------------------------------------------------------------------------------------------------|-------------------------|--|----------------------------------------------|-------------------|--|
| Part No. | Reagents | B42-3905-41 (100 tests) | | Reagents | T01-3360-57 | |
| | Reagents | B42-3905-42 (250 tests) | | Calibrators | T03-3401-07 | |
| | Calibrators | B43-3936-01 | | | | |
| Minimum Detectable Conc. | 0.05 ng/dL | | | 0.10 ng/dL | | |
| Precision (Total CV) | 4.5% @ 0.85 ng/dL | 4.8% @ 1.46 ng/dL | | 4.9% @ 0.94 ng/dL | 3.5% @ 1.76 ng/dL | |
| | 3.0% @ 3.04 ng/dL | | | 2.2% @ 4.68 ng/dL | | |
| Correlation | y = 0.998 x + 0.2179<br>where<br>y = ADVIA FREE T4 Assay<br>x = Immuno 1 FREE T4 Assay<br>n = 50<br>r = 0.9928<br>S = 0.1368 ng/dL | | | | | |
#### Interfering Substances
| Interfering | Interfering Substance | | Analyte | | Effect |
|------------------------|-----------------------|-------------|-----------------------|----------|--------|
| Substance | Concentration | | Concentration, uIU/mL | | |
| | SI Units - | - (mg/dL) - | I - Expected | Observed | (%) |
| Hemoglobin | 10 g/L | 1000 | 1.83 | 1.76 | 3.8 |
| Lipids (Triglycerides) | 11.3 mmol/L | 1000 | 1.84 | 1.88 | 2.2 |
| Bilirubin | 428 umol/L | 25 | 1.78 | 1.79 | 0.6 |
| Urea Nitrogen | 71.4 mmol/L | 200 | 1.83 | 1.81 | 1.1 |
Gabriel J. Muraca, Jr.
Gabriel J. Muraca, Jr. Manager Regulatory Affairs Bayer Corporation 511 Benedict Avenue Tarrytown, New York 10591-5097
5/21/99
{8}------------------------------------------------
## Total T3 Assay for Bayer ADVIA® Integrated Modular System
Listed below is a companison of the performance between the ADVIA T3 (Trilodothyronine) method, and a similar device that was granted clearance of substantial equivalence (Bayer Immuno 10 T3 Assay). The Summary of Safety and Effectiveness was extracted from the ADVIA T3 insert and the Immuno 1 T3 Assay method sheet.
#### INTENDED USED
This in vitro method is intended to quantitatively measure T3 in human serum using ADVIA T3 Assay on a Bayer ADVIA Integrated Modular System. Measurements of T3 are used in the diagnosis and treatment of thyroid disorders such as hyperthyroidism.
| METHOD | ADVIA T3 Assay | | Immuno 1 T3 Assay<br>(predicate Device) | |
|--------------------------|--------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------|-----------------------------------------|--------------------------|
| Part No. | Reagents<br>Reagents<br>Calibrators | B42-3916-41 (100 tests)<br>B42-3916-42 (250 tests)<br>B43-3943-01 | Reagents<br>Calibrators | T01-2942-0<br>T03-2872-0 |
| Minimum Detectable Conc. | 0.13 ng/mL | 0.06 ng/mL | | |
| Precision (Total CV) | 10.3% @ 0.67 ng/mL<br>4.9% @ 1.74 ng/mL<br>3.6% @ 2.89 ng/mL | 13.3% @ 0.46 ng/mL<br>6.0% @ 1.34 ng/mL<br>3.9% @ 3.43 ng/mL | | |
| Correlation | y = 1.014x + 0.1368<br>where<br>y = ADVIA T3 Assay<br>x = Immuno 1 T3 Assay<br>n = 50<br>r = 0.996<br> | | | |
#### Interfering Substances
| Interfering<br>Substance | Interfering Substance<br>Concentration | | Analyte<br>Concentration, ng/mL | | Effect | |
|--------------------------|----------------------------------------|------|---------------------------------|----------|---------|--|
| | SI Units (mg/dL) | | Expected | Observed | · (%) | |
| Hemoglobin | 10 g/L | 1000 | 1.85 | 1.91 | 3.7 | |
| Lipids (Triglycerides) | 11.3 mmol/L | 1000 | 1.85 | ાં તેરે | 5.5 | |
| Bilirubin | 428 umol/L | 25 | I '85 | 1.90 | 3.0 | |
| Urea Nitrogen | 71.4 mmol/L | 200 | I જર | 1.95 | રેંદર્ભ | |
Gabriel J. Murray Jr.
Syx =
Gabriel J. Muraca, Jr. Manager Regulatory Affairs Bayer Corporation 511 Benedict Avenue Tarrytown, New York 10591-5097
5721/99
{9}------------------------------------------------
### Free T3 Assay for Bayer ADVIA® Integrated Modular System
Listed below is a companison of the performance between the ADVIA Free T3 Assay method and a similar device that was granted clearance of substantial equivalence (Bayer Immuno 1ª Free T3 Assay). The information used in the Summary of Safety and Effectiveness was extracted from the ADVIA Free T3 insert and the Immuno 1 " Free T3 Assay method sheet.
#### INTENDED USED
This in vitro method is intended to quantitatively measure Free T3, in human serven using ADVIA Free T3 Assay on a Bayer ADVIA® Integrated Modular System. Measurements of Free T3 are used in the diagnosis of thyroid or pituitary disorders.
| METHOD | ADVIA Free T3 Assay | | Immuno 1 Free T3 Assay<br>(predicate Device) | |
|--------------------------|--------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------|----------------------------------------------|-----------------------------------------------------------|
| Part No. | Reagents<br>Calibrators | B42-3904-42<br>B43-3935-01 | Reagents<br>Calibrators | T01-3662-57<br>T03-3663-02 |
| Minimum Detectable Conc. | | 0.4 pg/mL | | 0.3 pg/mL |
| Precision (Total CV) | | 10.0% @ 2.0 pg/mL<br>5.0% @ 4.88 pg/mL<br>4.1% @ 9.35 pg/mL | | 8.5% @ 1.8 pg/mL<br>3.8% @ 5.4 pg/mL<br>2.9% @ 10.8 pg/mL |
| Correlation | y = 0.97x + 0.41<br>where<br>y = ADVIA Free T3 Assay<br>x = Immuno 1 Free T3 Assay<br>n = 56<br>r = 0.998<br>Syx = 0.427 pg/mL | | | |
#### Interfering Substances
| Interfering<br>Substance | Interfering Substance<br>Concentration | | Analyte<br>Concentration, pg/mL | | Effect | |
|--------------------------|----------------------------------------|---------|---------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------|--|
| | SI Units | (mg/dL) | Expected | Observed | (%) | |
| Hemoglobin --- | 10 g/L | 1000 | 4.66 | રું રેણે રેણવાડી તેમ જ દૂધની ડેરી જેવી સવલતો પ્રાપ્ય થયેલી છે. આ ગામનાં પ્રાથમિક શાળા, આંગણવાડી તેમ જ દૂધની ડેરી જેવી સવલતો પ્રાપ્ય થયેલી છે. આ ગામનાં છે. આ ગામનાં પ્રાથમિક | 6:4 | |
| Lipids (Triglycerides) | l 1.3 mmol/L | 1000 | 4.80 | રું 30 | 9.4 | |
| Bilirubin | 428 umol/L | 25 | 4.66 | 4.50 | -3.4 | |
| Urea Nitrogen | 153.1mmol/L | 429 | 4.66 | 4.76 | 2.1 | |
Gabriel J. Munoz, Jr.
Gabriel J. Muraca, Jr. Manager Regulatory Affairs Bayer Corporation 511 Benedict Avenue Tarrytown, New York 10591-5097
5/2/99
{10}------------------------------------------------
## T-Uptake Assay for Bayer ADVIA® Integrated Modular System
Listed below is a comparison of the performance between the ADVIA T-Uptake method, and a similar device that was granted clearance of substantial equivalence (Bayer Immuno 19 T-Uptake Assay). The information used in the Summary of Safety and Effectiveness was extracted from the ADVIA T-Uptake insert and the Immuno 1 T-Uptake Assay method sheet.
#### INTENDED USED
This in vitro method is intended to quantiatively measure the total amount of binding thyroid hormone on the thyroxine-binding proteins, thyroxine-binding prealbuning prealburnin, and albumin in human serum using the ADVIA T-Untake Assay on the Boyer ADVIA® Integrated Modular System. Measurements of T-Uptake are used in the diagnosis and treatment of thyroid disorders.
| METHOD | ADVIA T-Uptake Assay | Immuno 1 T-Uptake Assay (predicate Device) | | |
|--------------------------|--------------------------------------------------------------------------------------------------------------|--------------------------------------------|-------------------------------------------|----------------------------|
| Part No. | Reagents<br>Calibrators | B42-3915-41 (100 tests)<br>B43-3994-01 | Reagents<br>Calibrators | T01-3036-51<br>T03-3076-01 |
| Minimum Detectable Conc. | N/A | | N/A | |
| Precision (Total CV) | 3.2% @ 0.96<br>2.6% @ 0.89<br>2.3% @ 1.14 | | 2.8% @ 0.71<br>2.6% @ 1.03<br>2.4% @ 1.41 | |
| Correlation | y = 0.96x + 0.0003<br>where<br>y = ADVIA T-UptakeAssay<br>x = Immuno 1 T-Uptake Assay<br>n = 51<br>r = 0.987 | | | |
#### Interfering Substances
| Interfering<br>Substance | Interfering Substance<br>Concentration | Analyte<br>Concentration, ratio | Effect | | |
|--------------------------|----------------------------------------|---------------------------------|----------|----------|------|
| | SI Units | (mg/dL) | Expected | Observed | (%) |
| Hemoglobin | 10 g/L | 1000 | 1.13 | 1.14 | 0.88 |
| Lipids (Triglycerides) | 11.3 mmol/L | 1000 | 1.11 | 1.11 | 0.90 |
| Bilirubin | 428 $\mu$ mol/L | 25 | 1.15 | 1.17 | 1.74 |
| Urea Nitrogen | 71.4 mmol/L | 200 | 1.10 | 1.11 | 0.91 |
Gabriel J. Munoz Jr.
Gabriel J. Muraca, Jr. Manager Regulatory Affairs Bayer Corporation 511 Benedict Avenue Tarrytown, New York 10591-5097
5/21/99
{11}------------------------------------------------
## 510(k) SUMMARY OF SAFETY AND EFFECTIVENESS Urea Nitrogen method for ADVIA® 400
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of the Safe Medical Device Act of 1990 and 21 CFR 807.92.
The assigned 510(k) number is: _______________________________________________________________________________________________________________________________________________
#### 1. Intended Use
This in vitro diagnostic method to measure urea nitrogen (an end product of nitrogen metabolism) in human serum, plasma or urine on the Bayer ADVIA 400 system
#### 2. Predicate Device
| Product Name | | Reagent Part # Calibrator Part # |
|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------|-----------------------------------|
| Technicon CHEM 1 " Urea | IT01-1452-53 T03-1291-62 | |
| Nitrogen · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · | | |
#### 3. Device / Method
| Product Name | Reagent Part # | Calibrator Part # |
|-------------------------|----------------|-------------------|
| ADVIA 400 Urea Nitrogen | B41-3745-26 | T03-1291-62 |
#### Imprecision(SERUM)
| ADVIA 400 | | CHEM 1 | |
|------------------|----------------|------------------|----------------|
| Level<br>(mg/dL) | Total<br>CV(%) | Level<br>(mg/dL) | Total<br>CV(%) |
| 7.3 | 4.3 | 21 | 3 |
| 17 | 2.5 | 54 | 3 |
| 52 | 1.6 | 97 | 3 |
#### B. Imprecision(URINE)
| ADVIA 400 | | CHEM |
|------------------|----------------|------------------|
| Level<br>(mg/dL) | Total<br>CV(%) | Level<br>(mg/dL) |
| 69 | 3.9 | 478 |
| 212 | 2.1 | 648 |
| 404 | 2.0 | |
| CHEM 1 | |
|------------------|----------------|
| Level<br>(mg/dL) | Total<br>CV(%) |
| 478 | 2.6 |
| 648 | 2.5 |
Total CV(%) 3.6 3.7 3.4
#### Correlation (Y=ADVIA 400, X=comparison system)
| Specimen type | Comparison System (X) | N | Regression Equation | Syx (mg/dL) | R | Sample Range (mg/dL) |
|---------------------|-----------------------|----|---------------------|-------------|-------|----------------------|
| Serum | CHEM 1 | 50 | Y=1.03X-0.6 | 2.8 | 0.997 | 5-124 |
| Plasma(y), Serum(x) | ADVIA 400 | 58 | Y=0.96X+0.6 | 1.1 | 0.975 | 6-32 |
| Urine | CHEM 1 | 53 | Y=1.08X-8.9 | 22.5 | 0.997 | 70-1010 |
Gabriel J. Murray Jr. 7/21/99
.
{12}------------------------------------------------
## Interfering Substances
| Interfering<br>Substance | Interfering Sub.<br>Conc. (mg/dL) | Urea Nitrogen<br>Conc. (mg/dL) | Effect<br>(% change) |
|--------------------------|-----------------------------------|--------------------------------|----------------------|
| Bilirubin | 33 | 17.6 | -4.0 |
| Hemoglobin | 500 | 36.1 | +4.4 |
| Lipids (Triglycerides) | 500 | 30.5 | +23.0 |
| Ascorbic Acid | 400 | 58.5 | +3.2 |
| Salicylate | 500 | 66.3 | -2.4 |
| Glucose | 500 | 61.9 | +9.5 |
| Acetominophen | 40 | 52.9 | +4.5 |
.
# Analytical Range
: 上
Serum/Plasma: 0 to 150 mg/dL Urine: 2 to 1030 mg/dL 1
Ядт. Jr.
SYMBA
{13}------------------------------------------------
Image /page/13/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol resembling a stylized caduceus or a bird-like figure with three parallel lines forming the body and a wavy line representing the tail or base.
JUL 22 1999
Food and Drug Administration 2098 Gaither Road Rockville MD 20850
Mr. Gabriel J. Muraca, Jr. Manager Regulatory Affairs Bayer Corporation Business Group Diagnostics 511 Benedict Avenue Tarrytown, New York 10591-5097
Re: K991817
> Trade Name: Additional IMS Assays for the Bayer ADVIA® IMS™System Regulatory Class: II Product Code: CIJ Dated: May 21, 1999 Received: May 27, 1999
Dear Mr. Muraca:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
{14}------------------------------------------------
#### Page 2
Under the Clinical Laboratory Improvement Amendments of 1988 (CLIA-88), this device may require a CLIA complexity categorization. To determine if it does, you should contact the Centers for Disease Control and Prevention (CDC) at (770) 488-7655.
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4588. Additionally, for questions on the promotion and advertising of your device. please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification"(21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597, or at its internet address "http://www.fda.gov/cdrl/dsma/dsmamain.html".
Sincerely yours,
Steven Sutman
Steven I. Gutman, M.D. M.B.A. Director Division of Clinical Laboratory Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{15}------------------------------------------------
Page 1 of 3
510(k) Number (if known): K991817
## Device Name: Bayer ADVIA® Integrated Modular System (IMS)
Indications For Use:
The Bayer ADVIA IMS Amylase assay is an in vitro diagnostic device intended to measure amylase activity in human serum, plasma or urine. Such measurements are used as an aid primarily in the diagnosis and treatment of pancreatitis (inflammation of the pancreas).
The Bayer ADVIA IMS Cortisol assay is an in vitro diagnostic device intended to quantitatively measure cortisol in human serum. Measurements of cortisol are used as an aid in the diagnosis and treatment of disorders of the adrenal gland.
The Bayer ADVIA IMS Iron assay is an in vitro diagnostic device intended to measure iron in human serum of plasma. Measurements of iron are used as an aid in the diagnosis, monitoring and treatment of a variety of diseases including iron deficiency anemias, hemochromatosis, hemosiderosis from excessive iron intake, and hemolytic anemias.
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) /
Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use ﮯ (Per 21 CFR 801.109) OR
Over-The-Counter Use
(Optional Format 1-2-96)
{16}------------------------------------------------
510(k) Number (if known):
## Device Name: Bayer ADVIA® Integrated Modular System (IMS)
Indications For Use:
The Baver ADVIA IMS Thyroxine assay is an in vitro diagnostic device intended to measure thyroxine (T4), both protein bound and free, in human serum and plasma. Measurements of T4 are used as an aid in the diagnosis and treatment of thyroid diseases.
The Baver ADVIA IMS Free Thyroxine (FT4) assay is an in vitro diagnostic device intended to quantitatively measure free thyroxine in human serum. Measurements of free thyroxine in conjunction with other thyroid tests and clinical indicators are used as an aid in the diagnostic discrimination and assessment of thyroid diseases.
The Bayer ADVIA IMS Triiodothyronine (T3) assay is an in vitro diagnostic device intended to quantitatively measure triiodothyronine (T3) in human serum. Measurements of triiodothyronine, in conjunction with other thyroid tests and clinical indicators, are used as an aid in the diagnostic discrimination and assessment of thyroid diseases.
## (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use L (Per 21 CFR 801.109) OR
Over-The-Counter Use ..
(Optional Format 1-2-96)
..............................................................................................................................................................................
{17}------------------------------------------------
### 510(k) Number (if known):
## Device Name: Bayer ADVIA® Integrated Modular System (IMS)
Indications For Use:
The Bayer ADVIA IMS Free Triiodothyronine (FT3) assay is an in vitro diagnostic device intended to quantitatively measure free triiodothyronine in human serum. Measurements of free triodothyronine, in conjunction with other first-line thyroid tests such as Thyroid Stimulating Hormone (TSH) and Free Thyroxine (Free T4), as well as other clinical indicators, are used as an aid in the diagnostic discrimination and assessment of thyroid diseases.
The Bayer ADVIA IMS T Uptake (TUP) assay is an in vitro diagnostic device intended to quantitatively measure the total amount of available binding sites for thyroid hormone on the thyroxine-binding proteins, globulin, pre-albumin, and albumin in human serum. Measurements of T Uptake, in conjunction with other thyroid tests and clinical indicators, are used as an aid in the diagnostic discrimination and assessment of thyroid diseases.
The Bayer ADVIA IMS Urea Nitrogen (BUN) method is an in vitro diagnostic device intended to measure urea nitrogen in human serum, plasma and urine. Such measurements are used as an aid in the diagnosis and treatment of certain renal and metabolic diseases.
Alan Cooper
(Division Sign-Off) Division of Clinical Laboratory Devices 1991811 510(k) Number _
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
1-Prescription Use (Per 21 CFR 801.109)
OR
Over-The-Counter Use
(Optional Format 1-2-96)
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.