← Product Code [CIC](/submissions/CH/subpart-b%E2%80%94clinical-chemistry-test-systems/CIC) · K023105

# ATAC CALCIUM REAGENT AND ATAC CALIBRATOR (K023105)

_Elan Diagnostics · CIC · Mar 13, 2003 · Clinical Chemistry · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/CH/subpart-b%E2%80%94clinical-chemistry-test-systems/CIC/K023105

## Device Facts

- **Applicant:** Elan Diagnostics
- **Product Code:** [CIC](/submissions/CH/subpart-b%E2%80%94clinical-chemistry-test-systems/CIC.md)
- **Decision Date:** Mar 13, 2003
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 862.1145
- **Device Class:** Class 2
- **Review Panel:** Clinical Chemistry

## Indications for Use

The ATAC Calcium Reagent Kit, the ATAC Calibrator and the ATAC 8000 Random Access Chemistry System are intended for use as a system for the quantitation of calcium in serum and plasma. Calcium results are for the diagnosis and treatment of parathyroid disease, a variety of bone disease and tetany (intermittent muscular contractions or spasms). This reagent is intended to be used by trained personnel in a professional setting and is not intended for home use.

## Device Story

ATAC Calcium Reagent Kit used with ATAC 8000 Random Access Chemistry System; performs quantitative calcium analysis on serum and plasma samples. Principle of operation: o-cresolphthalein complexone binding at alkaline pH; absorbance increase at 578 nm proportional to calcium concentration. Used in professional clinical laboratory settings by trained personnel. Output provides calcium concentration values to assist clinicians in diagnosing and managing parathyroid, bone, and neuromuscular disorders.

## Clinical Evidence

Bench testing only. Linearity established from 0.2 to 15 mg/dL (r=0.9984). Precision evaluated via 60-replicate assays across three serum levels (Total CV 1.8-2.0%). Method comparison against a competitive reagent (n=120) yielded r=0.942. Detection limit documented at 0.2 mg/dL. Reagent stability (14 days) and calibration stability (4 hours) verified.

## Technological Characteristics

Reagent-based colorimetric assay using o-cresolphthalein complexone. Operates on ATAC 8000 Random Access Chemistry System. Measures absorbance at 578 nm. Quantitative, in vitro diagnostic reagent kit.

## Regulatory Identification

A calcium test system is a device intended to measure the total calcium level in serum. Calcium measurements are used in the diagnosis and treatment of parathyroid disease, a variety of bone diseases, chronic renal disease and tetany (intermittent muscular contractions or spasms).

## Predicate Devices

- HiChem Calcium Reagent Kit (product no. 88815)

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

K023/05

MAR 1 3 2003

Image /page/0/Picture/2 description: The image shows the word "élan" in a stylized font. The "é" has a unique design, with the top half of the letter being a solid black circle and the bottom half being a white circle. Above the "é" is a black parallelogram with a textured pattern. The rest of the word "lan" is written in a cursive font.

1075 W. Lambert Road Building D Brea, California 92821 T (714) 672 3553 F (714) 672 3554

## SUMMARY OF 510(K) SAFETY AND EFFECTIVENESS INFORMATION

This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.

The ATAC Calcium Reagent Kit is intended for the quantitative determination of calcium in serum and plasma. Calcium results are for the diagnosis and treathyroid disease, a variety of bone diseases, chronic renal disease and tetany (intermittent muscular contractions or spasms). The ATAC Calcium Reagent determines calcium through binding by o-cresolphthalein complexone at alkaline pH. The resulting increase in absorbance at 578 nm is proportion al to the calcium concentration of the sample.

The ATAC Calcium Reagent Kit is substantially equivalent to the HiChem" Calcium Reagent Kit, product no. 88815, which is marketed by Elan Inc. of Brea California. The effectiveness of ATAC Calcium Reagent Kit on the ATAC 8000 Random Access Chemistry System is shown by the following studies.

The recovery of calcium using the ATAC Calcium Reagent is linear from 0.2 to 15 mg/dL, as shown by the recovery of linearity standards that span the usable range. Regression statistics, which are forced through the origin, compare standard recoveries to standard values. These statistics are shown below.

r = 0.9984, (ATAC Recoveries) = 1.069 x (Standard Value), sy.x = 0.32 mg/dL, n = 21

Precision is demonstrated by the replicate assay of commercially available control serum on two separate instruments, Precision statistics, calculated analogous to the method described in NCCLS Guideline EP3-T, are shown below.

| Sample  | n  | mean | Within Run |      | Total |      |
|---------|----|------|------------|------|-------|------|
|         |    |      | 1SD        | %CV  | 1SD   | %CV  |
| Serum 1 | 60 | 7.4  | 0.11       | 1.4% | 0.14  | 1.9% |
| Serum 2 | 60 | 10.4 | 0.20       | 1.9% | 0.21  | 2.0% |
| Serum 3 | 60 | 13.2 | 0.17       | 1.3% | 0.23  | 1.8% |

Mixed serum and plasma specimens, collected from adult patients, were assayed for calcium using the ATAC 8000 Random Access Chemistry System and another commercially available method. Results were compared by least squares linear regression and the following statistics were obtained.

ATAC 8000 = 0.7 mg/dL + 0.944 x Competitive Reagent r = 0.942 n = 120 range = 6.4 - 11.5 mg/dL

The detection limit of 0.2 mg/dL is documented through the repetitive assay of a diluted serum pool. The observed standard deviation of a 30 replicate within run precision study was 0.05 mg/dL. Consequently, the detection limit is reported as twice the round-off error of the assay.

The 14 day on board reagent stability and 4 hour calibration stability claims are documented through the assay of serum controls over the claimed periods. In all cases, the total imprecision of calcium recoveries over the test than 0.2 mg/dL or 2%.

Wynn Staley

Wynn Stocking Manager of Regulatory Affairs Elan

510(k) Notification, ATAC Calcium Reagent Kit, 17 September, 2002, p 49

{1}------------------------------------------------

Image /page/1/Picture/1 description: The image shows a logo for the Department of Health & Human Services - Visa. The logo features a stylized image of an eagle with three lines representing its wings. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - VISA" is arranged in a circular pattern around the eagle image.

Food and Drug Administration 2098 Gaither Road Rockville MD 20850

Mr. Wynn Stocking Regulatory Affairs Manager Elan Diagnostics 1075 W. Lambert Road Building D Brea, CA 92821

MAR 1 3 2003

Re: k023105

> Trade/Device Name: ATAC Calcium Reagent Regulation Number: 21 CFR 862.1145 Regulation Name: Calcium test system Regulatory Class: Class II Product Code: CIC Dated: January 21, 2003 Received: January 22, 2003

Dear Mr. Stocking:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820).

{2}------------------------------------------------

Page 2 -

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific information about the application of labeling requirements to your device, or questions on the promotion and advertising of your device, please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 594-3084. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html.

Sincerely yours,

Steven Sutman

Steven I. Gutman, M.D., M.B.A. Director Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

510(k) Number (if known):

K023105

Device Name:

ATAC Calcium Reagent

Indications for Use:

The ATAC Calcium Reagent Kit, the ATAC Calibrator and the ATAC 8000 Random Access Chemistry System are intended for use as a system for the quantitation of calcium in serum and plasma. Calcium results are for the diagnosis and treatment of parathyroid disease, a variety of bone disease and tetany (intermittent muscular contractions or spasms).

This reagent is intended to be used by trained personnel in a professional setting and is not intended for home use.

## (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Jean Cooper

(Division Sign-Off) Division of Clinical Laboratory De 510(k) Number .

Prescription Use L (Per 21 CFR 801.109)

OR

Over-The-Counter Use _________________________________________________________________________________________________________________________________________________________

(Optional Format 1-2-96)

---

**Source:** [https://fda.innolitics.com/submissions/CH/subpart-b%E2%80%94clinical-chemistry-test-systems/CIC/K023105](https://fda.innolitics.com/submissions/CH/subpart-b%E2%80%94clinical-chemistry-test-systems/CIC/K023105)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
