K963800 · Instrumentation Laboratory CO · CHL · Nov 13, 1996 · Clinical Chemistry
Device Facts
Record ID
K963800
Device Name
IL SYNTHESIS
Applicant
Instrumentation Laboratory CO
Product Code
CHL · Clinical Chemistry
Decision Date
Nov 13, 1996
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 862.1120
Device Class
Class 2
Indications for Use
The IL Synthesis (hardware, software, and reagents) is for the quantitative in vitro diagnostic determination of pH, pCO2 (partial pressure of carbon dioxide), pO2 (partial pressure of oxygen), sodium (Na+), potassium (K+), calcium (Ca++), chloride (Cl-), glucose (Glu), and Hematocrit (conductivity) in whole blood or pCO2 (partial pressure of carbon dioxide), pO2 (partial pressure of oxygen) in expired gases by direct ion selective electrode (ISE) potentiometry. Sodium and potassium results are used to monitor electrolyte balance. Note that the number of electrodes is limited and any given instrument will only contain two of the three functions: calcium (Ca++), chloride (Cl-) and Glucose (Glu). The IL Synthesis (hardware, software, and reagents) also permits the quantitative in vitro diagnostic determination of five forms of hemoglobin in whole blood samples, total hemoglobin (tHb), oxyhemoglobin (%O₂Hb), carboxyhemoglobin (%COHb), methemoglobin (MetHb) and reduced hemoglobin, also called deoxyhemoglobin (%RHb) and will also calculate the following parameters: oxygen content of hemoglobin (O₂ct), oxygen binding capacity (O₂ cap), and oxygen saturation (sO₂m). The IL Synthesis is designed for laboratory use to provide both measured and calculated results for blood gases, electrolytes, substrates, and co-oximeter parameters on blood and gas samples.
Device Story
IL Synthesis is a fully automatic, microprocessor-controlled analyzer family; integrates blood gas, electrolyte, glucose, hematocrit, and co-oximeter modules. Inputs: whole blood or expired gas samples. Operation: direct ion selective electrode (ISE) potentiometry for electrolytes/gases; co-oximetry for hemoglobin fractions. Output: measured concentrations of pH, pCO2, pO2, Na+, K+, Ca++, Cl-, glucose, hematocrit, and hemoglobin forms; calculated parameters include O₂ct, O₂ cap, and sO₂m. Used in clinical laboratories; operated by laboratory personnel. Healthcare providers use results to assess patient electrolyte balance and oxygenation status, informing clinical decisions regarding respiratory and metabolic management.
Clinical Evidence
Bench testing only. In-house studies evaluated between-day imprecision using three levels of ContrIL Plus and Hct CHECK QC materials, and dye-based controls for co-oximetry. Testing performed on three instruments (n=60 for blood gas/electrolytes; n=15 for co-oximetry). Results demonstrate consistent performance across parameters (pH, pCO2, pO2, Na, K, Ca, Cl, Glu, Hct, tHb, O2Hb, COHb, MetHb) with reported mean, SD, and %CV values.
Technological Characteristics
Fully automatic, microprocessor-controlled analyzer. Sensing: Ion Selective Electrode (ISE) potentiometry for electrolytes/gases; photometric/spectrophotometric analysis for co-oximetry. Measures pH, pCO2, pO2, Na+, K+, Ca++, Cl-, glucose, hematocrit (conductivity), and hemoglobin fractions. Modular configuration allows selection of two from Ca++, Cl-, and Glucose. Designed for laboratory use.
Indications for Use
Indicated for quantitative in vitro diagnostic determination of pH, blood gases (pCO2, pO2), electrolytes (Na+, K+, Ca++, Cl-), glucose, hematocrit, and hemoglobin fractions (tHb, O₂Hb, COHb, MetHb, RHb) in whole blood or expired gases for laboratory use to monitor electrolyte balance and oxygenation status.
Regulatory Classification
Identification
A blood gases (PCO2 , PO2 ) and blood pH test system is a device intended to measure certain gases in blood, serum, plasma or pH of blood, serum, and plasma. Measurements of blood gases (PCO2 , PO2 ) and blood pH are used in the diagnosis and treatment of life-threatening acid-base disturbances.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.