EASYSTAT PH, PCO2, PO2, HTC, NA, K, CL ANALYZER

K063376 · Medica Corp. · CHL · Jun 25, 2007 · Clinical Chemistry

Device Facts

Record IDK063376
Device NameEASYSTAT PH, PCO2, PO2, HTC, NA, K, CL ANALYZER
ApplicantMedica Corp.
Product CodeCHL · Clinical Chemistry
Decision DateJun 25, 2007
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 862.1120
Device ClassClass 2

Indications for Use

The EasyStat pH, PCO₂, PO₂, Htc, Na, K, Ca/Cl analyzer is designed for clinical laboratory use, making direct measurements of pH (hydrogen ion activity), PCO₂ (partial pressure of carbon dioxide), PO₂ (partial pressure of oxygen), Htc (Hematocrit), Na⁺ (sodium ions), K⁺ (potassium ions), Ca²⁺ (ionized calcium) and Cl⁻ (chloride ions) in whole blood samples from syringes or capillary tubes. pH measurements are used in the diagnosis and treatment of diseases involving imbalance in the acid-base equilibrium in blood. PCO₂ measurements are used in the diagnosis and treatment of diseases involving imbalance in the partial pressure of carbon dioxide in blood. PO₂ measurements are used in the diagnosis and treatment of diseases conditions characterized by low or high blood oxygen levels in blood. Hematotric (Htc) measurements are used in the diagnosis and treatment of diseases characterized by erythrocyte imbalances in whole blood. Sodium measurements are used in the diagnosis and treatment diseases involving electrolyte imbalance. Potassium measurements monitor electrolyte balance and in the diagnosis and treatment of diseases conditions characterized by low or high blood potassium levels. Calcium (ionized) measurements are used to determine the physiologically active form of calcium in blood and establish the patient’s calcium metabolism. Chloride measurements are used in the diagnosis and treatment of electrolyte and metabolic disorders.

Device Story

EasyStat analyzer performs direct electrochemical measurements of pH, PCO2, PO2, Htc, Na+, K+, Ca++, and Cl- in whole blood samples. Input samples are introduced via syringes or capillary tubes. Device utilizes electrochemical sensors to quantify ion activity and partial pressures; results are processed and displayed for clinical laboratory personnel. Output data assists clinicians in diagnosing and managing acid-base, electrolyte, and metabolic imbalances. Device is intended for professional use in clinical laboratory settings.

Clinical Evidence

No clinical data. Performance established via bench testing, including precision studies (CLSI EP-5) and method comparison studies against predicate devices using patient samples. Precision evaluated across three levels of aqueous QC material. Method comparison showed high correlation (R² > 0.96) for all analytes across syringe and capillary modes.

Technological Characteristics

Ion-selective electrode (ISE) technology for potentiometric measurement of pH, PCO₂, Na⁺, K⁺, Ca²⁺, Cl⁻; amperometric Pt electrode for PO₂; electrical conductivity for Htc. Dimensions: 14.25"W x 12.5"H x 7"D. Power: 100-115VAC or 220VAC. Connectivity: RS-232 port for external PC, optional barcode reader. Calibration: Automatic two-point.

Indications for Use

Indicated for clinical laboratory use to measure pH, PCO2, PO2, Hematocrit, Na+, K+, ionized calcium, and chloride in whole blood samples from syringes or capillary tubes for the diagnosis and treatment of acid-base, electrolyte, metabolic, and oxygenation disorders.

Regulatory Classification

Identification

A blood gases (PCO2 , PO2 ) and blood pH test system is a device intended to measure certain gases in blood, serum, plasma or pH of blood, serum, and plasma. Measurements of blood gases (PCO2 , PO2 ) and blood pH are used in the diagnosis and treatment of life-threatening acid-base disturbances.

Submission Summary (Full Text)

{1} 1 of 11 # 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY AND INSTRUMENT COMBINATION TEMPLATE A. 510(k) Number: k063376 B. Purpose for Submission: New submission for configuring a Cl⁻ option to the EasyStat pH, PCO₂, PO₂, Htc, Na⁺, K⁺, Ca²⁺ Analyzer (K021515). Medica's Cl⁻ option was originally cleared on the EasyElectroLyte Na⁺, K⁺, Cl⁻ Analyzer (K000926). C. Measurand: pH (hydrogen ion activity), PCO₂ (partial pressure of carbon dioxide), PO₂ (partial pressure of oxygen), Htc (Hematocrit), Na⁺ (sodium ions), K⁺ (potassium ions), Ca²⁺ (ionized calcium) and Cl⁻ (chloride ions) D. Type of Test: Quantitative Ion specific electrodes, Amperometric for PO₂ and conductivity for Hct E. Applicant: Medica Corp. F. Proprietary and Established Names: EasyStat pH, PCO₂, PO₂, Hct, NA⁺, K⁺, CA⁺⁺/Cl⁻ Analyzer G. Regulatory Information: 1. Regulation section: 21CFR Sec.-862.1120-Blood gases (PCO₂, PO₂) and blood pH test system. 21CFR Sec. 864.5600 Automated hematocrit instrument. 21CFR Sec. 862.1665 Sodium test system. 21CFR Sec. 862.1600 Potassium test system. 21CFR Sec. 862.1145 Calcium test system. 21CFR Sec. 862.1170 Chloride test system. 2. Classification: Class II 3. Product code: CHL - Electrode Measurement, Blood-Gases (PCO₂, PO₂) and Blood pH GKF - instrument, hematocrit, automated JGS - electrode, ion specific, sodium CEM - electrode, ion specific, potassium JFP - electrode, ion specific, calcium CGZ - electrode, ion-specific, chloride 4. Panel: Chemistry (75) H. Intended Use: 1. Intended use(s): See indications for use below 2. Indication(s) for use: The EasyStat pH, PCO₂, PO₂, Htc, Na, K, Ca/Cl analyzer is designed for clinical laboratory use, making direct measurements of pH (hydrogen ion activity), PCO₂ (partial pressure of carbon dioxide), PO₂ (partial pressure of oxygen), Htc {2} (Hematocrit), Na⁺ (sodium ions), K⁺ (potassium ions), Ca²⁺ (ionized calcium) and Cl⁻ (chloride ions) in whole blood samples from syringes or capillary tubes. pH measurements are used in the diagnosis and treatment of diseases involving imbalance in the acid-base equilibrium in blood. PCO₂ measurements are used in the diagnosis and treatment of diseases involving imbalance in the partial pressure of carbon dioxide in blood. PO₂ measurements are used in the diagnosis and treatment of diseases conditions characterized by low or high blood oxygen levels in blood. Hematotric (Htc) measurements are used in the diagnosis and treatment of diseases characterized by erythrocyte imbalances in whole blood. Sodium measurements are used in the diagnosis and treatment diseases involving electrolyte imbalance. Potassium measurements monitor electrolyte balance and in the diagnosis and treatment of diseases conditions characterized by low or high blood potassium levels. Calcium (ionized) measurements are used to determine the physiologically active form of calcium in blood and establish the patient’s calcium metabolism. Chloride measurements are used in the diagnosis and treatment of electrolyte and metabolic disorders. 3. Special conditions for use statement(s): For prescription use 4. Special instrument requirements: EasyStat pH, PCO₂, PO₂, Hct, Na⁺, K⁺, Ca²⁺/Cl⁻ Analyzer I. Device Description: The EasyStat (ES) pH, PCO₂, PO₂, Htc, Na, K, Ca/Cl Blood Gas Analyzer is an analyzer that has all the features of the Medica’s EasyStat (ES) pH, PCO₂, PO₂, Htc, Na, K, Ca analyzer, but the Ca ISE sensor may be replaced by a Cl sensor by either a distributor or an end-user. Depending on the operation mode chosen, the EasyStat Ca/Cl will analyze a whole-blood patient sample or a QC material for pH, PCO₂, PO₂, Htc, Na⁺, K⁺, Ca²⁺ or pH, PCO₂, PO₂, Htc, Na⁺, K⁺, Cl⁻. Both EasyStat analyzers (Ca²⁺ and Cl⁻ versions) utilize the same three aqueous calibrants enclosed in a sealed pack (Reagent Module) along with a waste container. These solutions are used in the calibration and sample analysis modes. The analyzer establishes the mathematical slope of the sensors by means of a two point calibration, and then determines the values of an unknown sample by direct comparison with one of the calibrants. One of the predicate devices for the Medica EasyStat chloride 2 of 11 {3} analyzer is its predecessor the Medica EasyStat ionized calcium analyzer (K021515). The calibrants for both analyzers have identical formulations. In addition, the packaging of the calibrants is identical. # J. Substantial Equivalence Information: 1. Predicate device name(s): EasyStat pH, PCO2, PO2, Htc, $\mathrm{Na^{+}}$ , $\mathrm{K^{+}}$ , $\mathrm{Ca^{++}}$ Analyzer EasyElectroLyte $\mathrm{Na^{+}}$ , $\mathrm{K^{+}}$ , Cl- Analyzer (for Cl- only) 2. Predicate 510(k) number(s): k021515, k000926 respectively 3. Comparison with predicate: | | New device | Predicate | Predicate | | --- | --- | --- | --- | | | EasyStat Ca/CL | EasyStat Ca | EasyElectroLytes Na/K/Cl | | 510(k) | k063376 | K021515 | k000926 | | Manufacturer | Medica Corp. | Medica Corp. | Medica Corp. | | Indications for use | See indications above | Within scope | Within scope | | pH | Ion Selective Electrode Technology | Ion Selective Electrode Technology | | | PC02 | Ion Selective Electrode Technology | Ion Selective Electrode Technology | | | P02 | Amperometric (Pt electrode) | Amperometric (Pt electrode) | | | Htc | Conductivity | Conductivity | | | Na+ | Ion Selective Electrode Technology | Ion Selective Electrode Technology | Ion Selective Electrode Technology | | K+ | Ion Selective Electrode Technology | Ion Selective Electrode Technology | Ion Selective Electrode Technology | | Ca++ | Ion Selective Electrode Technology | Ion Selective Electrode Technology | | | Cl- | Ion Selective Electrode Technology | | Ion Selective Electrode Technology | | Calibrator | Aqueous | Aqueous | Aqueous | | Measured Parameters | pH, PC02, P02, Htc, Na+, K+, Ca++, Cl- | pH, PC02, P02, Htc, Na+, K+, Ca++ | Na+, K+, Cl- | {4} | | New device | Predicate | Predicate | | --- | --- | --- | --- | | | EasyStat Ca/CL | EasyStat Ca | EasyElectroLytes Na/K/Cl | | Sample Volume | 120 μl Syringe / 95 μl Capillary | 120 μl Syringe / 95 μl Capillary | 55 μl Syringe / 50 μl Capillary | | Analysis time | 120 sec | 120 sec | 35 sec | | Measured range: | | | | | pH | 6.5 -8.0 pH units | 6.5 -8.0 pH units | | | PC02 | 5.0 - 150.0 mmHg | 5.0 - 150.0 mmHg | | | P02 | 5 - 700 mmHg | 5 - 700 mmHg | | | Htc | 10-70% | 10-70% | | | Na+ | 80 - 200 mmol/l | 80 - 200 mmol/l | | | K+ | 1.0 - 20.0 mmol/l | 1.0 - 20.0 mmol/l | | | Ca++ | 0.25 - 5.00 mmol/l | 0.25 - 5.00 mmol/l | | | Cl- | 50.0 - 150.0 mmol/l | | 50.0 - 150.0 mmol/l | | Dimensions | 14.25"W x 12.5"H x 7"D | 14.25"W x 12.5"H x 7"D | 14.25"W x 12.5"H x 7"D | | Voltage requirements | 1001115VAC,50-60Hz, 0.8A | 1001115VAC,50-60Hz, 0.8A | 1001115VAC,50-60Hz, 0.8A | | | 220VAC, 50-60Hz, 0.4A | 220VAC, 50-60Hz, 0.4A | 220VAC, 50-60Hz, 0.4A | K. Standard/Guidance Document Referenced (if applicable): CLSI - Evaluation of Precision Performance of Clinical Chemistry Devices - EP05-A2 CLSI - Method Comparison and Bias Estimation Using Patient Samples - EP09-A2 L. Test Principle: The EasyStat Ca/Cl in this submission utilizes the same basic technologies and software architecture as Medica's EasyElectroLytes Na, K, Cl (for chloride) and EasyStat pH, PCO2, PO2, Htc, Na, K, Ca line of analyzers. The EasyStat analyzer in this submission measures pH, PCO2, PO2, $\mathrm{Na^{+}}$ , $\mathrm{K^{+}}$ , and $\mathrm{Ca^{++}}$ or $\mathrm{Cl^-}$ using ion selective electrode (ISE) technology that measures changes in voltage (a potentiometric measurement). The PO2 electrode measures a change in current (an amperometric measurement). These changes are then compared with those occurring with on-board calibration standards to produce final results. Hct is measured using electrical conductivity, which is a technique used in other currently marketed devices. The hematocrit sensor is also calibrated with standard solutions. The analyzer then measures the impedance of the sample to obtain the Hct value. Next, the Hct value is corrected for the concentration of the sodium ions. {5} M. Performance Characteristics (if/when applicable): Because the modules used were all previously cleared, the sponsor demonstrated that the new configurations did not change performance through new precision and method comparison studies. All other performance parameters remain as cleared previously in k021515 and k000926. 1. Analytical performance: a. Precision/Reproducibility: The precision of the EasyStat Ca/Cl was evaluated following the guidelines of CLSI EP-5. This section provides data on the precision of the EasyStat Ca/Cl operating in the $\mathrm{Ca}^{++}$ and $\mathrm{Cl^-}$ modes for all sensors using both the syringe and capillary sampling modes. This study incorporated the use of three EasyStat Ca/Cl analyzers operating in the Cl mode (Chloride sensor and software) and three EasyStat Ca/Cl analyzers operating in the $\mathrm{Ca}^{++}$ mode (Calcium sensor and software). Three levels of ampuled aqueous based QC material were analyzed daily (AM & PM) for twenty days according to CLSI EP-5. Since the Hematocrit (Htc) QC material is available in two levels only, there are no precision data for level 3. The data summaries below are from one analyzer in both the Cl mode and the $\mathrm{Ca}^{++}$ mode. The other two analyzers had comparable precision in both modes. | Summary The within-run and Day-to-Day CV% and SD for Precision Testing for ES Ca/Cl (Ca++ Mode) Analyzers in Syringe Mode | | | | | | | | | --- | --- | --- | --- | --- | --- | --- | --- | | Level 1 | pH | PCO2 | PO2 | Hct | Na | K | Ca | | Total Average | 7.144 | 67.5 | 62.70 | 31.94 | 115.93 | 2.54 | 1.52 | | Swr (CLSI) | 0.01 | 0.32 | 1.77 | 0.22 | 0.47 | 0.02 | 0.01 | | CVwr (CLSI) | 0.08% | 0.47% | 2.83% | 0.70% | 0.41% | 0.89% | 0.57% | | Sdd (CLSI) | 0.00 | 0.48 | 0.65 | 0.27 | 0.00 | 0.01 | 0.01 | | CVdd | 0.00% | 0.71% | 1.03% | 0.85% | 0.00% | 0.59% | 0.52% | | Level 2 | pH | PCO2 | PO2 | Hct | Na | K | Ca | | Total Average | 7.398 | 42.6 | 103.3 | 51.19 | 136.7 | 4.17 | 1.11 | | Swr (CLSI) | 0.00 | 0.29 | 1.15 | 0.18 | 0.33 | 0.01 | 0.00 | | CVwr (CLSI) | 0.05% | 0.68% | 1.11% | 0.35% | 0.24% | 0.21% | 0.43% | | Sdd (CLSI) | 0.00 | 0.18 | 0.86 | 0.28 | 0.20 | 0.00 | 0.00 | | CVdd | 0.01% | 0.43% | 0.83% | 0.54% | 0.15% | 0.00% | 0.28% | | Level 3 | pH | PCO2 | PO2 | Hct | Na | K | Ca | | Total Average | 7.615 | 20.62 | 145.29 | | 159.40 | 5.89 | 0.64 | | Swr (CLSI) | 0.01 | 0.10 | 1.04 | | 0.54 | 0.02 | 0.01 | | CVwr (CLSI) | 0.12% | 0.48% | 0.72% | | 0.34% | 0.37% | 0.82% | | Sdd (CLSI) | 0.00 | 0.22 | 0.00 | | 0.22 | 0.02 | 0.00 | | CVdd | 0.06% | 1.06% | 0.00% | | 0.14% | 0.27% | 0.00% | | Summary The within-run and Day-to-Day CV% and SD for Precision Testing for ES Ca/Cl (Ca++ Mode) Analyzers in Capillary Mode | | | | | | | | | Level 1 | pH | PCO2 | PO2 | Hct | Na | K | Ca | {6} | Summary The within-run and Day-to-Day CV% and SD for Precision Testing for ES Ca/Cl (Cl Mode) Analyzers in Syringe Mode | | | | | | | | | --- | --- | --- | --- | --- | --- | --- | --- | | Level 1 | pH | PCO2 | PO2 | Hct | Na | K | Cl | | Total Average | 7.147 | 66.71 | 68.05 | 31.54 | 116.00 | 2.59 | 72.90 | | Swr (CLSI) | 0.003 | 0.36 | 1.78 | 0.17 | 0.42 | 0.03 | 0.36 | | CVwr (CLSI) | 0.04% | 0.55% | 2.62% | 0.55% | 0.36% | 1.03% | 0.49% | | Sdd (CLSI) | 0.003 | 0.16 | 0.27 | 0.15 | 0.00 | 0.01 | 1.09 | | CVdd | 0.05% | 0.24% | 0.40% | 0.48% | 0.00% | 0.54% | 1.50% | | Level 2 | pH | PCO2 | PO2 | Hct | Na | K | Cl | | Total Average | 7.400 | 41.92 | 106.66 | 50.69 | 136.38 | 4.16 | 97.96 | | Swr (CLSI) | 0.003 | 0.24 | 1.18 | 0.14 | 0.22 | 0.01 | 0.23 | | CVwr (CLSI) | 0.04% | 0.58% | 1.10% | 0.28% | 0.16% | 0.20% | 0.23% | | Sdd (CLSI) | 0.000 | 0.06 | 0.88 | 0.10 | 0.00 | 0.00 | 0.78 | | CVdd | 0.00% | 0.14% | 0.82% | 0.19% | 0.00% | 0.07% | 0.80% | | Level 3 | pH | PCO2 | PO2 | | Na | K | Cl | | Total Average | 7.637 | 20.37 | 144.03 | | 159.87 | 5.96 | 127.12 | | Swr (CLSI) | 0.003 | 0.08 | 1.17 | | 0.25 | 0.01 | 0.27 | | CVwr (CLSI) | 0.04% | 0.38% | 0.81% | | 0.15% | 0.19% | 0.21% | | Sdd (CLSI) | 0.000 | 0.08 | 0.00 | | 0.31 | 0.01 | 0.51 | | CVdd | 0.00% | 0.39% | 0.00% | | 0.20% | 0.13% | 0.40% | | Summary The within-run and Day-to-Day CV% and SD for Precision Testing for ES Ca/Cl (Cl Mode) Analyzers in Capillary Mode | | | | | | | | | Level 1 | pH | PCO2 | PO2 | Hct | Na | K | Cl | | Total Average | 7.159 | 65.57 | 73.25 | 30.79 | 117.42 | 2.67 | 76.62 | | Swr (CLSI) | 0.01 | 0.79 | 2.01 | 0.80 | 1.23 | 0.05 | 1.01 | | CVwr (CLSI) | 0.12% | 1.21% | 2.75% | 2.61% | 1.05% | 1.97% | 1.32% | | Sdd (CLSI) | 0.00 | 0.15 | 1.01 | 0.43 | 0.62 | 0.01 | 1.37 | | CVdd | 0.04% | 0.23% | 1.38% | 1.38% | 0.53% | 0.48% | 1.78% | | Level 2 | pH | PCO2 | PO2 | Hct | Na | K | Cl | {7} | Total Average | 7.403 | 42.40 | 110.78 | 49.96 | 136.54 | 4.12 | 101.24 | | --- | --- | --- | --- | --- | --- | --- | --- | | Swr (CLSI) | 0.01 | 0.35 | 1.34 | 0.34 | 0.97 | 0.04 | 0.83 | | CVwr (CLSI) | 0.09% | 0.83% | 1.21% | 0.68% | 0.71% | 0.92% | 0.82% | | Sdd (CLSI) | 0.00 | 1.53 | 0.90 | 0.33 | 0.52 | 0.01 | 1.00 | | CVdd | 0.02% | 3.62% | 0.81% | 0.66% | 0.38% | 0.28% | 0.99% | | Level 3 | pH | PCO2 | PO2 | | Na | K | Cl | | Total Average | 7.609 | 21.45 | 148.54 | | 157.48 | 5.66 | 127.78 | | Swr (CLSI) | 0.01 | 0.17 | 1.40 | | 0.69 | 0.07 | 0.55 | | CVwr (CLSI) | 0.10% | 0.81% | 0.94% | | 0.44% | 1.18% | 0.43% | | Sdd (CLSI) | 0.00 | 3.18 | 0.96 | | 0.30 | 0.03 | 0.33 | | CVdd | 0.03% | 14.82% | 0.65% | | 0.19% | 0.47% | 0.26% | b. Linearity/assay reportable range: | | Reportable Range | | --- | --- | | pH | 6.5 -8.0 pH units | | PC02 | 5.0 – 150.0 mmHg | | P02 | 5 700 mmHg | | Htc | 10-70% | | Na+ | 80 – 200 mmol/l | | K+ | 1.0 – 20.0 mmol/l | | Ca++ | 0.25 – 5.00 mmol/l | | Cl- | 50.0 – 150.0 mmol/l | Reportable Range was determined in predicate devices k021515 and k000926 for these instrument modules. c. Traceability, Stability, Expected values (controls, calibrators, or methods): There were no changes to the existing calibrator cleared under k021515. The Reagent Module value assignment method for pH, PCO2, PO2, Na, K, and Ca remains the same. The chloride (Cl-) value assignment for the EasyStat Calcium/ Chloride calibrants has been established via titration with a primary Silver Nitrate (AgNO3) solution. This standard volumetric method used a potentiometric end-point detection and it was applied to the Master Secondary Primary (MSP) calibrant A and calibrant C. Stability is based on accelerated shelf life study, and supported by a real time study. Controls were cleared under k021515, k955630 and k880447. d. Detection limit: see k021515 and k000926 e. Analytical specificity: {8} see k021515 and k000926 f. Assay cut-off: Not Applicable 2. Comparison studies: a. Method comparison with predicate device: A study was done to compare the performance of the pH, PCO2, PO2, Htc, Na, K, and Ca++ sensors on the EasyStat Ca/Cl analyzer operating in the Ca++ mode, against the same sensors on the predicate EasyStat analyzer cleared by k021515. This study was repeated to compare the performance of the pH, PCO2, PO2, Htc, Na, K and Cl- sensors on the EasyStat Ca/Cl analyzer operating in the chloride (Cl) mode, against the same sensors on the predicate EasyStat and EasyElectroLyte -Cl analyzers cleared by K021515 and K000926 respectively. Two predicate devices were employed using the "Syringe" option. The method comparison summary data below are from one representative analyzer configuration for each analyte and mode of measurement. Cl Linearity/Method Comparison for EasyStat Ca/Cl Syringe Values Vs. EasyElectrolyte-Cl syringe Values (N=102) range 68.45 - 133.1 mM/L y = 0.9771x + 2.6944 R2 = 0.9817 Cl Linearity/Method Comparison for EasyStat Ca/Cl Capillary Values Vs. EasyElectrolyte Cl- Values (N=102) range 93.4 - 140.5 mM/L y = 0.9726x + 4.5443 R2 = 0.9726 Ca Linearity/Method Comparison for EasyStat Ca/Cl Syringe Values Vs. EasyStat Ca++ syringe Values (N=100) range 0.765 - 1.64 mM/L y = 0.9797x + 0.0309 R2 = 0.9882 Ca Linearity/Method Comparison for EasyStat Cl Capillary Values Vs. EasyStat Ca++ Capillary Values (N=100) range 0.76 - 1.635 mM/L y = 0.9487x + 0.0632 R2 = 0.9783 pH Linearity/Method Comparison for EasyStat Ca/Cl Syringe Values Vs. EasyStat Ca++ Syringe Values (N=106) range 6.567 - 7.779 units Cl mode Ca mode y = 0.9669x + 0.2428 y = 0.9525x + 0.344 R2 = 0.9954 R2 = 0.996 8 of 11 {9} pH Linearity/Method Comparison for EasyStat Ca/Cl Capillary Values Vs. EasyStat Ca++ Capillary Values (N=106) range 6.571 - 7.731 units Cl mode Ca mode y = 0.9714x + 0.2113 y = 0.9541x + 0.3361 R2 = 0.9919 R2 = 0.9966 PCO2 Linearity/Method Comparison for EasyStat Ca/Cl Syringe Values Vs. EasyStat Ca++ Syringe Values (N=102) range 8.2 - 143.8 mmHg Cl mode Ca mode y = 0.9632x + 0.7453 Y = 0.9634x + 1.836 R2 = 0.996 R2 = 0.9946 PCO2 Linearity /Method Comparison for EasyStat Ca/Cl Capillary Values Vs. EasyStat Ca++ Capillary Values (N=102) range 10.3 - 141 mmHg Cl mode Ca mode y = 0.9494x + 1.5595 y = 0.956x + 2.431 R2 = 0.9968 R2 = 0.9963 PO2 Linearity/Method Comparison for EasyStat Ca/Cl Syringe Values Vs. EasyStat Ca++ Syringe Values (N=102) range 19 - 462 mmHg Cl mode Ca mode y = 0.9693x + 2.6519 y = 1.0032x + 0.4322 R2 = 0.9976 R2 = 0.9972 PO2 Linearity /Method Comparison for EasyStat Ca/Cl Capillary Values Vs. EasyStat Ca++ Capillary Values (N=102) range 17 - 423 mmHg Cl mode Ca mode y = 0.9805x + 2.4841 y = 1.0093x + 0.8746 R2 = 0.9988 R2 = 0.9994 Na Linearity/Method Comparison for EasyStat Ca/Cl Syringe Values Vs. EasyStat Ca++ Syringe Values (N=102) range 89.2 - 163.2 mM/L Cl mode Ca mode y = 0.9895x + 2.3029 y = 0.9903x + 1.8699 R2 = 0.992 R2 = 0.9937 Na Linearity /Method Comparison for EasyStat Ca/Cl Capillary Values Vs. EasyStat Ca++ Capillary Values (N=102) range 91.4 - 163.3 mM/L Cl mode Ca mode y = 0.9699x + 4.5856 y = 1.003x - 0.4345 R2 = 0.9869 R2 = 0.9918 K Linearity/Method Comparison for EasyStat Ca/Cl Syringe Values Vs. EasyStat Ca++ Syringe Values (N=102) range 2.84 - 8.06 mM/L Cl mode Ca mode y = 0.9451x + 0.2288 y = 0.9837x + 0.0773 R2 = 0.9977 R2 = 0.9981 K Linearity/Method Comparison for EasyStat Ca/Cl Capillary Values Vs. EasyStat Ca++ Capillary Values (N=102) range 2.90 - 7.84 mM/L Cl mode Ca mode y = 0.9471x + 0.2283 y = 0.9642x + 0.1592 R2 = 0.9851 R2 = 0.996 {10} Hct Linearity/Method Comparison for EasyStat Ca/Cl Syringe Values Vs. EasyStat $\mathrm{Ca^{+ + }}$ Syringe Values (N=102) range 21.5 - 66.7% Cl mode Ca mode $\mathrm{y} = 0.9955\mathrm{x} + 0.6535$ $\mathrm{y} = 1.0168\mathrm{x} - 0.5392$ R2 = 0.9795 R2 = 0.9877 Hct Linearity/Method Comparison for EasyStat Ca/Cl Capillary Values Vs. EasyStat $\mathrm{Ca^{+ + }}$ Capillary Values (N=102) range 21.0 - 68.1% Cl mode Ca mode $\mathrm{y} = 0.9861\mathrm{x} + 1.4148$ $\mathrm{y} = 1.0355\mathrm{x} - 1.2208$ R2 = 0.9649 R2 = 0.9857 b. Matrix comparison: Whole blood sample - see syringe verses capillary method comparisons above 3. Clinical studies: a. Clinical Sensitivity: Not Applicable b. Clinical specificity: Not Applicable c. Other clinical supportive data (when a. and b. are not applicable): Not Applicable 4. Clinical cut-off: Not Applicable 5. Expected values/Reference range: | | Arterial | Mixed Venous | Venous | | --- | --- | --- | --- | | pH | 7.350-7.450 | 7.340-7.360 | 7.340-7.360 | | PC02 | 35.0-45.0 mmHg | 44.0-46.0 mmHg | 44.0-46.0 mmHg | | P02 | 83-108 mmHg | 38-42 mmHg | 38-42 mmHg | | Htc | 35-50% | 35-50% | 35-50% | | Na+ | 136.0-146.0 mmol/l | 136.0-146.0 mmol/l | 136.0-146.0 mmol/l | | K+ | 3.5-5.1 mmol/l | 3.5-5.1 mmol/l | 3.5-5.1 mmol/l | | Ca++ | 1.05-1.2 mmol/l | 1.05-1.2 mmol/l | 1.05-1.2 mmol/l | | Cl- | 98.0-106.0 mmol/l | 98.0-106.0 mmol/l | 98.0-106.0 mmol/l | Cited from literature Tietz: Fundamentals of Clinical Chemistry, $4^{\mathrm{th}}$ edition, (1996) Fink: Clinical Practice in Respiratory Care N. Instrument Name: EasyStat pH, $\mathrm{PCO}_2$ , $\mathrm{PO}_2$ , Hct, $\mathrm{NA}^+$ , $\mathrm{K}^+$ , $\mathrm{CA}^{++}/\mathrm{CL}^-$ Analyzer {11} O. System Descriptions: 1. Modes of Operation: Single sample 2. Software: The analyzer can be interfaced to an external widows based computer via an RS-232 port. Connection for an optional Barcode Reader FDA has reviewed applicant’s Hazard Analysis and software development processes for this line of product types: Yes ☐ X or No ☐ The applicant provided software documentation that is typical for this device type. 3. Specimen Identification: Manual sample identification 4. Specimen Sampling and Handling: Manual - syringe or capillary mode 5. Calibration: Automatic Two-point calibration 6. Quality Control: Quality control results are stored P. Other Supportive Instrument Performance Characteristics Data Not Covered In The “Performance Characteristics” Section above: Q. Proposed Labeling: The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10. R. Conclusion: The submitted information in this premarket notification is complete and supports a substantial equivalence decision. 11 of 11
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...