MODIFICATION TO: I-STAT PORTABLE CLINICAL ANALYZER
Applicant
I-Stat Corp.
Product Code
CHL · Clinical Chemistry
Decision Date
Aug 20, 2001
Decision
SESE
Submission Type
Special
Regulation
21 CFR 862.1120
Device Class
Class 2
Indications for Use
The i-STAT System is used by trained medical professionals for running clinical chemistry tests for sodium, potassium, chloride, urea nitrogen, glucose, hematocrit, calcium, blood-gases (PCO2, PO2) and blood pH, creatinine, lactic acid, and activated whole blood clotting time in a variety of test panels contained in the i-STAT test cartridges.
Device Story
i-STAT System is a portable clinical analyzer used by trained medical professionals in clinical settings. Device processes whole blood samples via disposable test cartridges to determine clinical chemistry parameters including sodium, potassium, chloride, urea nitrogen, glucose, hematocrit, calcium, blood gases (PCO2, PO2), pH, creatinine, lactic acid, and activated clotting time. System provides rapid diagnostic results to assist healthcare providers in clinical decision-making. This specific submission involves a software modification to reduce interference on the PCO2 test caused by thiopental sodium and propofol.
Clinical Evidence
No clinical data provided. The submission relies on bench testing to validate the software modification regarding PCO2 test interference.
Technological Characteristics
Portable clinical analyzer using disposable cartridges for electrochemical and enzymatic analysis of whole blood. System performs automated measurement of electrolytes, blood gases, metabolites, and clotting time. Software-controlled analysis.
Indications for Use
Indicated for use by trained medical professionals to perform clinical chemistry tests and test panels (including electrolytes, blood gases, pH, metabolites, and clotting time) on whole blood samples.
Regulatory Classification
Identification
A blood gases (PCO2 , PO2 ) and blood pH test system is a device intended to measure certain gases in blood, serum, plasma or pH of blood, serum, and plasma. Measurements of blood gases (PCO2 , PO2 ) and blood pH are used in the diagnosis and treatment of life-threatening acid-base disturbances.
Predicate Devices
Model 100 Portable Clinical Analyzer and Test Pack (K894914, K912387)
{0}------------------------------------------------
# AUG 2 0 2001
# 510(k) SUMMARY
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 C.F.R. § 807.92.
The assigned 510(k) number is K012478
Summary Prepared on: 31 July 2001
# Submitted by:
i-STAT Corporation
104 Windsor Center Drive
East Windsor, NJ 08520
Phone: 609-469-0242
Fax: 609-443-9310
# Contact:
Paul VanDerWerf, Ph.D., Vice President Regulatory Affairs and Quality Assurance
#### Establishment Registration Number: 2245578
#### Identification of the Device
Device Name: i-STAT System (see Table 1 below) Proprietary/Trade Name: i-STAT System Common Name: Portable clinical analyzer Classification Name: See Table 1 Device Classification: ll Regulation Number: See Table 1 Clinical Chemistry (75) Panel: Product Code: See Table 1
| i-STAT Corporation<br>104 Windsor Center Drive<br>East Windsor, NJ 08520 | |
|--------------------------------------------------------------------------|--|
|--------------------------------------------------------------------------|--|
{1}------------------------------------------------
| Device Name | 510(k)<br>Number | Classification | Regulation | Product<br>Code |
|----------------------------------------------------------|--------------------|-----------------------------------------------------------------------|----------------------|-----------------|
| Model 100 Portable<br>Clinical Analyzer and Test<br>Pack | K894914<br>K912387 | Electrode, ion specific,<br>sodium | 21 C.F.R. § 862.1665 | JGS |
| i-STAT Tests for Acid-Base<br>Parameters; | K936081 | Electrode<br>measurement, blood-<br>gases (pCO2, pO2) and<br>blood pH | 21 C.F.R. § 862.1120 | CHL |
| i-STAT 200 Portable<br>Clinical Analyzer | K940918 | Electrode<br>measurement, blood-<br>gases (pCO2, pO2) and<br>blood pH | 21 C.F.R. § 862.1120 | CHL |
| i-STAT Creatinine Test | K973292 | Electrode, ion based,<br>enzymatic, creatinine | 21 C.F.R. § 862.1225 | CGL |
| i-STAT Lactate Test | K982071 | Acid, lactic, enzymatic<br>method | 21 C.F.R. § 862.1450 | KHP |
| i-STAT Model 200 Portable<br>Clinical Analyzer | K001154 | Electrode, ion specific,<br>sodium | 21 C.F.R. § 862.1665 | JGS |
| i-STAT Model 300 Portable<br>Clinical Analyzer | K001387 | Glucose oxidase,<br>glucose | 21 C.F.R. § 862.1345 | CGA |
| i-STAT System | K011478 | Acid, Lactic, Enzymatic<br>Method | 21 C.F.R. § 862.1450 | KHP |
TABLE 1
# Identification of the Predicate Device.
The predicate device is the i-STAT System (see Table 1 above).
# Description of the Device.
The i-STAT System is used in conjunction with disposable i-STAT cartridges for determination of a variety of parameters in whole blood.
#### Intended Use of the Device.
The i-STAT PCA is used by trained medical professionals for running a variety of clinical chemistry tests and test panels contained in the i-STAT test cartridges.
#### Reason for Submission.
The software has been modified to diminish the interference on the PCO2 test caused by the drugs thiopental sodium and propofol.
# Conclusion.
These labeling changes do not affect prior substantial equivalence determinations.
| | i-STAT Corporation<br>104 Windsor Center Drive<br>East Windsor, NJ 08520 | 14 |
|--|--------------------------------------------------------------------------|----|
|--|--------------------------------------------------------------------------|----|
{2}------------------------------------------------
DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" arranged around the perimeter. Inside the circle is a stylized image of an eagle with its wings spread, with three wavy lines emanating from the eagle's body.
AUG 2 0 2001
Food and Drug Administration 2098 Gaither Road Rockville MD 20850
Paul VanDerWerf, Ph.D. Vice President, Regulatory Affairs and Quality Assurance i-STAT Corporation 104 Windsor Center Drive East Windsor, NJ 08520
Re: 510(k) Number: K012478 Trade/Device Name: i-STAT System Regulation Numbers: 862.1120, 862.1145, 862.1170, 862.1225, 862.1345, 862.1600, 862.1665, 862.1770, 864.6400, 864.7140, 862.1450 Regulatory Class: II Product Code: CHL, JGS, CDS, JPI, JBP, JFP, CGZ, CGL, CGA, CEM Regulatory Class: I Product Code: KHP Dated: July 31, 2001 Received: August 2, 2001
Dear Dr. VanDerWerf:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Good Manufacturing Practice for Medical Devices: General (GMP) regulation (21 CFR Part 820) and that, through periodic GMP inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
{3}------------------------------------------------
Page 2 -
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and 1 additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4588. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html".
Sincerely yours,
Steven Sutman
Steven I. Gutman, M.D., M.B.A. Director Division of Clinical Laboratory Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
#### INDICATIONS FOR USE
510(k) Number (if known): K012478
Device Name: i-STAT System
Indications for Use:
The i-STAT System is used by trained medical professionals for running clinical chemistry tests for sodium, potassium, chloride, urea nitrogen, glucose, hematocrit, calcium, blood-gases (PCO2, PO2) and blood pH, creatinine, lactic acid, and activated whole blood clotting time in a variety of test panels contained in the i-STAT test cartridges.
Kusia Alexander for JeanCrops
(Division Sign-Off)
Division of Clinical Laboratory Devices
a) Number K012478
#### (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
| i-STAT Corporation | |
|--------------------------|--|
| 104 Windsor Center Drive | |
| East Windsor, NJ 08520 | |
12
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.