ACCUCHEK INSTANT PLUS DUAL TESTING SYSTEM, ACCUCHEK INSTANT PLUS CHOLESTEROL, ACCUCHEK INSTANT PLUS GLUCOCOSE
Applicant
Roche Diagnostics
Product Code
CHH · Clinical Chemistry
Decision Date
Aug 23, 2005
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 862.1175
Device Class
Class 1
Indications for Use
The Roche AccuChek Instant Plus Dual testing System is intended to be used for the quantitative measurement of glucose and cholesterol in capillary whole blood.
Device Story
AccuChek Instant Plus Dual Testing System measures glucose and cholesterol in capillary whole blood; utilizes electrochemical biosensor technology. System comprises meter, glucose test strips, and cholesterol test strips. User applies blood to strip; reagents react to generate current proportional to analyte concentration. Amperometric technology measures current across electrodes. Used in home and professional settings; operated by patients or healthcare professionals. Submission focuses on usability study confirming user ability to distinguish between glucose and cholesterol test procedures.
Clinical Evidence
Usability study conducted with 30 subjects. Each subject performed two glucose and two cholesterol tests in alternating order. Results showed 100% correct identification and performance of tests (N=120 tests). Met acceptance criteria of >90% success rate within 95% confidence interval. Analytical performance (precision, linearity, specificity, detection limits) previously established in predicate submissions.
Technological Characteristics
Electrochemical biosensor; amperometric measurement. Glucose test strips use glucose oxidase; cholesterol test strips use esterase-oxidase. Quantitative measurement of capillary whole blood. Standalone meter system.
Indications for Use
Indicated for quantitative measurement of cholesterol and glucose in capillary whole blood. Glucose test strips for home and professional use. Cholesterol test strips for healthcare professional use and home use by patients with diabetes.
Regulatory Classification
Identification
A cholesterol (total) test system is a device intended to measure cholesterol in plasma and serum. Cholesterol measurements are used in the diagnosis and treatment of disorders involving excess cholesterol in the blood and lipid and lipoprotein metabolism disorders.
{0}
# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY ONLY TEMPLATE
A. 510(k) Number:
k051376
B. Purpose for Submission:
The purpose of this premarket notification is to provide results of a usability study performed to assess the ability of users to effectively run and distinguish between glucose and cholesterol tests offered on the AccuChek Instant Plus Dual Testing System, which is already being marketed. These tests were previously cleared individually under k944459 and k944458.
C. Measurand:
Glucose and Cholesterol
D. Type of Test:
Quantitative, electrochemical biosensor
E. Applicant:
Roche Diagnostics
F. Proprietary and Established Names:
AccuChek Instant Plus Dual Testing System
G. Regulatory Information:
1. Regulation section:
21 CFR 862.1345 – Glucose test system
21CFR 862.1175 – Cholesterol (total) test system
2. Classification:
glucose - Class II
cholesterol - Class I
{1}
3. Product code:
NBW, System, Test, Blood Glucose, Over The Counter
CGA, Glucose Oxidase, Glucose
CHH, Enzymatic Esterase-Oxidase, Cholesterol
4. Panel:
Chemistry 75
H. Intended Use:
1. Intended use(s):
The Roche AccuChek Instant Plus Dual testing System is intended to be used for the quantitative measurement of glucose and cholesterol in capillary whole blood.
2. Indication(s) for use:
The Roche AccuChek Instant Plus Dual Testing System is designed to quantitatively measure cholesterol and glucose in capillary whole blood. The AccuChek Instant Glucose Test Strips are for use in home and professional settings for testing glucose in whole blood. The AccuChek Instant Plus Cholesterol Test Strips are for use by healthcare professionals and for home use by people with diabetes for cholesterol screening.
3. Special conditions for use statement(s): N/A
4. Special instrument requirements: N/A
I. Device Description:
The AccuChek Instant Plus Dual Testing System consists of a blood glucose/cholesterol meter, glucose test strips, and cholesterol test strips.
J. Substantial Equivalence Information:
1. Predicate device name(s):
AccuChek Instant Glucose Test System
AccuChek Instant Plus Cholesterol Test System
2. Predicate 510(k) number(s):
{2}
k944459
k944458
3. Comparison with predicate:
The meter and test strips have not been modified. They are exactly the same as those previously cleared individually.
K. Standard/Guidance Document Referenced (if applicable):
None
L. Test Principle:
The test systems use electrochemical biosensor technology. When blood is applied to the test strip, reagents on the strip react with the blood and a current is generated. The monitors employ amperometric technology to measure the glucose concentrations in the blood sample by measuring the amount of current that is generated when glucose reacts with the glucose oxidase on the glucose test strip and with esterase oxidase on the cholesterol test strip.. The result is the generation of current across the electrodes, which is proportional to the concentration of analyte present in the whole blood sample.
M. Performance Characteristics (if/when applicable):
1. Analytical performance:
a. Precision/Reproducibility:
The precision/reproducibility was cleared with the original 510(k)s for the Roche AccuChek Instant Glucose Test System (k944459) and the Roche AccuChek Instant Plus Cholesterol Test System (k944458).
b. Linearity/assay reportable range:
The linear range was cleared with the original 510(k)s for the Roche AccuChek Instant Glucose Test System (k944459) and the Roche AccuChek Instant Plus Cholesterol Test System (k944458).
c. Traceability, Stability, Expected values (controls, calibrators, or methods):
Not applicable.
d. Detection limit:
Detection limits of glucose is 20 – 600 mg/dL
{3}
Detection limits of cholesterol is 150-300 mg/dL
The detection limits were cleared with the original 510(k)s for the Roche AccuChek Instant Glucose Test System (k944459) and the Roche AccuChek Instant Plus Cholesterol Test System (k944458).
e. Analytical specificity:
The analytical specificity was cleared with the original 510(k)s for the Roche AccuChek Instant Glucose Test System (k944459) and the Roche AccuChek Instant Plus Cholesterol Test System (k944458).
f. Assay cut-off:
Not applicable.
2. Comparison studies:
a. Method comparison with predicate device:
Clinical accuracy was demonstrated in the original 510(k)s for the Roche AccuChek Instant Glucose Test System (K944459) and the Roche AccuChek Instant Plus Cholesterol Test System (K944458).
b. Matrix comparison:
Not applicable.
3. Clinical studies:
a. Clinical Sensitivity:
Not applicable.
b. Clinical specificity:
Not applicable.
c. Other clinical supportive data (when a. and b. are not applicable):
Thirty subjects each performed two glucose and two cholesterol tests in alternating order using the AccuChek Instant Plus Dual Testing System. The study showed that users had no difficulty in differentiating between the two assays. Users identified and performed the tests correctly 100% of the time for N=30 users and N=120 tests. This met the study acceptance criteria of 90% or better within 95% confidence interval.
{4}
4. Clinical cut-off:
Not applicable.
5. Expected values/Reference range:
Not applicable.
N. Proposed Labeling:
The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10.
O. Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.