AEROSET SYSTEM

K980367 · Abbott Laboratories · CGZ · Apr 1, 1998 · Clinical Chemistry

Device Facts

Record IDK980367
Device NameAEROSET SYSTEM
ApplicantAbbott Laboratories
Product CodeCGZ · Clinical Chemistry
Decision DateApr 1, 1998
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 862.1170
Device ClassClass 2

Indications for Use

The AEROSET System is a fully automated, random access, clinical chemistry Analyzer which utilizes spectrophotometry (monochromatic and bichromatic modes of measurement) for photometric based determinations. The system also has a solid state Ion-Selective Electrode module, trademarked as Integrated Chip Technology™ (ICT), utilizing potentiometry, for electrolyte determinations (sodium, potassium, and chloride). Per 21 CFR, 682.2160, the AEROSET System is a discrete photometric chemistry Analyzer for clinical use. The device is intended to duplicate manual analytical procedures by performing various steps such as pipetting, heating, and measuring color intensity. This device is intended for use in conjunction with certain materials to measure a variety of analytes. The AEROSET System also has a solid state Ion-Selective Electrode module, trademarked as Integrated Chip Technology™ (ICT), utilizing potentiometry, for electrolyte determinations (sodium, potassium, and chloride).

Device Story

AEROSET System is a fully automated, random-access clinical chemistry analyzer for clinical laboratory use. It processes clinical samples to perform quantitative kinetic and endpoint chemistry assays. Input modalities include spectrophotometry (monochromatic and bichromatic modes) and potentiometry via a solid-state Ion-Selective Electrode (ISE) module (Integrated Chip Technology™). The device automates manual laboratory steps: pipetting, heating, and color intensity measurement. It supports up to 100 simultaneous assays with throughput up to 2000 tests/hour. Operated by laboratory personnel, the system provides quantitative results for analytes including total protein, alkaline phosphatase, creatinine, urea nitrogen, and electrolytes (Na, K, Cl). Output data assists clinicians in diagnostic decision-making. Benefits include increased laboratory efficiency and standardized analytical performance compared to manual methods.

Clinical Evidence

Bench testing only. Correlation analysis performed against the Hitachi 717 Chemistry Analyzer across multiple analytes (Total Protein, Alkaline Phosphatase, Creatinine, Urea Nitrogen, Sodium, Potassium, Chloride) in serum and urine. Correlation coefficients ranged from 0.933 to 0.998. Linearity testing confirmed performance across specified ranges (e.g., Sodium-Serum to 182.92 mmol/L, Creatinine-Serum to 23.82 mg/dL). Precision testing reported total %CVs for normal controls (1.0–5.3%) and abnormal controls (0.7–4.0%).

Technological Characteristics

Fully automated random-access clinical chemistry analyzer. Measurement principles: spectrophotometry (monochromatic/bichromatic) and potentiometry via solid-state Ion-Selective Electrode (Integrated Chip Technology™). Throughput: up to 2000 tests/hour. Capacity: up to 100 simultaneous assays. Connectivity: standalone clinical chemistry analyzer.

Indications for Use

Indicated for clinical use as a discrete photometric chemistry analyzer to perform automated analytical procedures including pipetting, heating, and measuring color intensity for various analytes, and potentiometric electrolyte determinations (sodium, potassium, chloride).

Regulatory Classification

Identification

A chloride test system is a device intended to measure the level of chloride in plasma, serum, sweat, and urine. Chloride measurements are used in the diagnosis and treatment of electrolyte and metabolic disorders such as cystic fibrosis and diabetic acidosis.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ KA80367 ﺮ ﺍﻟﻤﺮﺍﺟﻊ ﺍﻟﻤﺮﻛﺰ ﺍﻟﻤﺮﺍﺟﻊ APR - 1 1998 ## Section III # 510(k) Summary AEROSET™ System 510(k) January 27, 1998 1:25 PM Final {1}------------------------------------------------ ## 510(k) Summary #### Submitter's name/address Abbott Laboratories 1920 Hurd Drive M.S. 1-8 Irving, Texas 75038 Contact Person Andrew Johnson Section Manager Regulatory Affairs (972) 518-7861 Fax (972) 753-3367 Date of Preparation of this Summary: Device Trade or Proprietary Name: Device Common Name: Classification Number/Class: January 27, 1998 AEROSET™ System Clinical Chemistry Analyzer Classification Number 75JJD, Class I 75JGS, Class II 75CEM, Class II 75CGZ, Class II ### 510(k) Summary This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92. The assigned 510(k) number is: _______________________________________________________________________________________________________________________________________________ #### Description: The following is a brief description of the AEROSET System. The AEROSET System is a fully automated random access clinical chemistry Analyzer which utilizes spectrophotometry (mono and bichromatic modes of measurement). The AEROSET System has a solid state Ion-Selective Electrode module, trademarked as Integrated Chip Technology™ (ICT), utilizing potentiometry, which can perform electrolyte determinations for sodium, potassium, and chloride. The AEROSET System can run up to 100 different assays simultaneously with a maximum throughput approaching 2000 tests per hour (depending on configuration). AEROSET™ System 510(k) January 27, 1998 1:25 PM Final {2}------------------------------------------------ #### Substantial Equivalence: Substantial equivalence has been demonstrated between the AEROSET System and the Boehringer Mannheim Diagnostics' Hitachi® 717 Chemistry Analyzer (K872494). Both systems utilize monochromatic and bichromatic spectrophotometry to perform quantitative kinetic and endpoint clinical chemistries. Both Analyzers perform quantitative analysis of sodium, potassium, and chloride in clinical samples. Both Analyzers have an Ion-Selective Electrode module, the AEROSET is a solid state Ion-Selective Electrode module, trademarked as Integrated Chip Technology™ (ICT). #### Intended Use: The AEROSET System is a fully automated, random access, clinical chemistry Analyzer which utilizes spectrophotometry (monochromatic and bichromatic modes of measurement) for photometric based determinations. The system also has a solid state Ion-Selective Electrode module, trademarked as Integrated Chip Technology™ (ICT), utilizing potentiometry, for electrolyte determinations (sodium, potassium, and chloride). #### Performance Characteristics: A correlation analysis between the AEROSET System and the Boehringer Mannheim Diagnostics' Hitachi 717 Chemistry Analyzer yielded the following results: | Representative | Correlation | Slope | Y-axis intercept | |------------------|-------------|--------|------------------| | Method | Coefficient | | | | Total Protein | 0.977 | 0.931 | 0.105 g/dL | | Alkaline | 0.99 | 0.9 | 3.4 U/L | | Phosphatase | | | | | Creatinine-Serum | 0.995 | 0.924 | -0.198 mg/dL | | Creatinine-Urine | 0.933 | 0.9743 | 2.214 mg/dL | | Urea Nitrogen | 0.996 | 0.95 | -1.114 mg/dL | | Sodium-Serum | 0.94 | 0.86 | 16.14 mmol/L | | Sodium-Urine | 0.998 | 0.967 | 1.227 mmol/L | | Potassium-Serum | 0.99 | 1.08 | -0.49 mmol/L | | Potassium-Urine | 0.994 | 0.856 | 2.273 mmol/L | | Chloride-Serum | 0.96 | 0.93 | 3.9 mmol/L | | Chloride-Urine | 0.979 | 1.02 | 5.38 mmol/L | {3}------------------------------------------------ The linearity test yielded the following results: | Representative Method | Linearity | |-----------------------|------------------| | Total Protein | To 10.90 g/dL | | Alkaline Phosphatase | To 1484.16 U/L | | Creatinine-Serum | To 23.82 mg/dL | | Creatinine-Urine | To 191.65 mg/dL | | Urea Nitrogen | To 123.51 mg/dL | | Sodium-Serum | To 182.92 mmol/L | | Sodium-Urine | To 398.02 mmol/L | | Potassium-Serum | To 8.74 mmol/L | | Potassium-Urine | To 400.24 mmol/L | | Chloride-Serum | To 148.88 mmol/L | | Chloride-Urine | To 411.00 mmol/L | The precision of the representative assays was acceptable for both normal and abnormal controls. The total %CV's for both the normal and abnormal controls ranged from 1.0 to 5.3 and 0.7 to 4.0 respectively. 10 ### Conclusion: The data demonstrates that the AEROSET System and the Boehringer Mannheim Diagnostics' Hitachi 717 Chemistry Analyzer are substantially equivalent. AEROSET™ System 510(k) January 27, 1998 1:25 PM Final {4}------------------------------------------------ Image /page/4/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, stacked on top of each other. The profiles are simple and abstract, with only the basic outline of the head and face visible. ్లో ఒ Food and Drug Administration 2098 Gaither Road Rockville MD 20850 APR - 1 1998 Andrew Johnson Section Manager, Requlatory Affairs Abbott Laboratories 1920 Hurd Drive Irving, Texas 75038 K980367 Re: AEROSET™ System Regulatory Class: II Product Code: JJE, JGS, CEM, CGZ Dated: January 27, 1998 Received: January 29, 1998 Dear Mr. Johnson: We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions The general controls provisions of the Act of the Act. include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Requlations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical General requlation (21 CFR Part 820) and that, Devices: through periodic QS inspections, the Food and Druq Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory In addition, FDA may publish further announcements action. concerning your device in the Federal Reqister. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or requlations. {5}------------------------------------------------ Page 2 Under the Clinical Laboratory Improvement Amendments of 1988 (CLIA-88), this device may require a CLIA complexity categorization. To determine if it does, you should contact the Centers for Disease Control and Prevention (CDC) at (770) 488-7655. This letter will allow you to begin marketing your device as described in your 510 (k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4588. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html". sincerely yours, Steven Bitman Steven I. Gutman, M.D., M.B.A. Director Division of Clinical Laboratory Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {6}------------------------------------------------ 510(k) Number (if known): Device Name: Indications For Use: 86. Per 21 CFR, 682.2160, the AEROSET System is a discrete photometric chemistry Analyzer for clinical use. The device is intended to duplicate manual analytical procedures by performing various steps such as pipetting, heating, and measuring color intensity. This device is intended for use in conjunction with certain materials to measure a variety of analytes. The AEROSET System also has a solid state Ion-Selective Electrode module, trademarked as Integrated Chip Technology™ (ICT), utilizing potentiometry, for electrolyte determinations (sodium, potassium, and chloride). (Division Sign-Off) Division of Clinical Laboratory Devices 510(k) Number *k980367* (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Prescription Use 1 (Per 21 CFR 801.109) OR Over-The-Counter Use__________________________________________________________________________________________________________________________________________________________ (Optional Format 1-2-96) 500(x) Number saivs 1 101810068 I leading to noisivil (HO-ugis uoisiniQ) AEROSETTM System 510 (k) January 27, 1998 1:25 PM Final
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...