← Product Code [CGZ](/submissions/CH/subpart-b%E2%80%94clinical-chemistry-test-systems/CGZ) · K973742

# SENDX 100 PH, BLOOD GAS AND ELECTROLYTE ANALYSIS SYSTEM (K973742)

_Sendx Medical, Inc. · CGZ · Dec 3, 1997 · Clinical Chemistry · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/CH/subpart-b%E2%80%94clinical-chemistry-test-systems/CGZ/K973742

## Device Facts

- **Applicant:** Sendx Medical, Inc.
- **Product Code:** [CGZ](/submissions/CH/subpart-b%E2%80%94clinical-chemistry-test-systems/CGZ.md)
- **Decision Date:** Dec 3, 1997
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 862.1170
- **Device Class:** Class 2
- **Review Panel:** Clinical Chemistry

## Indications for Use

The SenDx 100™ pH, Blood Gas and Electrolyte Analysis System (SenDx 100™) is a portable, automated analyzer that measures blood oxygen, carbon dioxide, pH, sodium, potassium, chloride, ionized calcium, and hematocrit in whole blood. The SenDx 100™ system is intended for use by trained technologists, nurses, physicians, and therapists. It can be used in laboratory, near patient or point-of-care settings.

## Device Story

Portable, automated blood gas and electrolyte analyzer; measures pH, pO2, pCO2, Na+, K+, Cl-, iCa++, and hematocrit in whole blood. System comprises analyzer, multi-use disposable sensor cassette, and calibration solution pack. Inputs: whole blood sample via aspiration stylus. Principle: miniaturized microelectrode technology (Clark cell for pO2; potentiometric ion-selective electrodes for pH, pCO2, Na+, K+, Cl-, iCa++; conductivity for hematocrit). Microprocessor-controlled electronics manage sensor output, touch screen display, and thermal printer. Used in labs, near-patient, or point-of-care settings by trained clinicians. Provides real-time measurement and derived values (e.g., HCO3, BEb, Hb, O2 Sat). Facilitates rapid clinical decision-making in critical care environments. Data output via touch screen, printer, or RS232 serial port for LIS/HIS integration.

## Clinical Evidence

Clinical studies conducted at four sites comparing SenDx 100 to Corning 865, Nova Stat Profile 5, and CO-oximeter/spun hematocrit. Evaluated >400 split-samples. Performance metrics (slope, r, intercept) reported for pH, pCO2, pO2, Na+, K+, iCa++, and Hct. Precision studies conducted over 5 days using QC materials across multiple sites; %CVs reported for within-run, within-day, day-to-day, and site-to-site variability. Results demonstrate performance equivalent to predicate devices.

## Technological Characteristics

Portable analyzer with microprocessor-controlled electronics. Sensors: miniaturized microelectrode 'chip' (1-3/8" x 3/8"). Measurement principles: polarographic (amperometric) for pO2; potentiometric for pH, pCO2, Na+, K+, Cl-, iCa++; conductance for hematocrit. Battery-powered with external charger. Connectivity: RS232 serial port for LIS/HIS. Calibration: automatic two-point using tonometered solution packs with bar-coded parameters. Thermal printer integrated. UL/TUV certified.

## Regulatory Identification

A chloride test system is a device intended to measure the level of chloride in plasma, serum, sweat, and urine. Chloride measurements are used in the diagnosis and treatment of electrolyte and metabolic disorders such as cystic fibrosis and diabetic acidosis.

## Predicate Devices

- StatPal II ([K922631](/device/K922631.md)/B)
- Mallinckrodt GEM-PREMIER ([K910305](/device/K910305.md))
- Ciba-Corning 865 Blood Gas and Electrolyte System ([K934907](/device/K934907.md))
- NOVA Biomedical NOVA STAT Profile 5 ([K874832](/device/K874832.md))

## Reference Devices

- StatPal ([K903965](/device/K903965.md)/B)
- StatPal ([K914560](/device/K914560.md))
- SenDx 100 ([K954482](/device/K954482.md))
- Corning 850 Blood Gas and Electrolyte Analyzer

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

15973742

SenDx Medical, Inc. 1945 Palomar Oaks Way Carlsbad, California 92009 Phone: 760.930.6300 Fax: 760.930.6320

DEC - 3 1997

SenDx Medical, Inc. Company Name:

1945 Palomar Oaks Way Address: Carlsbad, California 92009

Phone: 760.930.6300

760.930.6320 Fax:

Contact Person: Ruben Chairez

Date: September 15, 1997

#### 510(k) SUMMARY

This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR § 807.92.

The assigned 510 (k) number is: ___________________

ు గ్రామం
 నుండి

{1}------------------------------------------------

| A. Device Proprietary name:    | SenDx 100 Blood Gas and Electrolyte Analysis System      |
|--------------------------------|----------------------------------------------------------|
| B. Device common name:         | Blood Gas and Electrolyte Analyzer                       |
| C. Device classification name: |                                                          |
| • 21 CFR 862.1120:             | Blood gases (CO2, pO2) and blood pH test system Class II |
| • 21 CFR 862.1665              | Sodium test system Class II                              |
| • 21 CFR 862.1600              | Potassium test system Class II                           |
| • 21 CFR 862.1145              | Calcium test system Class II                             |
| • 21 CFR 862.1170              | Electrode, Ion-Specific, Chloride Class II               |
| • 21 CFR 864.6400              | Hematocrit measuring device Class II                     |

-

ﺳﺴﺴﺴ

ਸੂਨੀ ਦੇ ਸੰਗ੍ਰਹਿ ਹੈ।

ਹਵਾਲੇ 

..............................................................................................................................................................................

:

{2}------------------------------------------------

#### Statement of Substantial Equivalence

1

ﻟﺴﻴﺮ- ﺗﺮ

The SenDx 100 pH, Blood Gas and Electrolyte Analysis System is substantially equivalent to other blood gas, electrolyte and hematocrit measuring systems used with whole blood.

This 510(k) submission seeks clearance for modifications to the SenDx 100 line as embodied in the two models SD 100B and SD 100IB. These modifications include:

- 1. Addition of chloride to the electrolyte panel.
- 2. Modification to the Performance Characteristics section of the Operators Manual.

The SenDx 100 (Model SD100B and SD100IB) is substantially equivalent to other blood gas analyzers such as:

StatPal versions were previously submitted and cleared for marketing under 510(k) document Control numbers:

> K903965/B K914560 K922631/B.

These three StatPal versions were found by FDA to be substantially equivalent to the Mallinckrodt® GEM@-STAT, among other instruments.

The SenDx 100 system was originally submitted and cleared for marketing under Document Control number K954482. This SenDx 100 system was also found by FDA to be substantially equivalent to the Mallinckrodt® GEM®-PREMIER among other instruments.

The SenDx 100 is substantially equivalent to other blood gas and electrolyte systems such as:

- 1) The Ciba-Corning, Corning 865 Blood Gas and Electrolyte System 510(k) Document Control number K 934907.
- 2) The Mallinckrodt® GEM®-PREMIER Blood Gas and Electrolyte Monitor, 510(k) Document Control number K 910305.
- 3) The NOVA Biomedical, NOVA STAT Profile 5, 510(k) Document Control Number K874832.

{3}------------------------------------------------

#### Description

The SenDx 100™ system consists of the analyzer, a multi-use, disposable sensor cassette and a calibration solution pack. A bar code label containing lot and sensor-specific identifying parameters accompanies each new sensor cassette and calibration solution pack.

The SenDx 100™ line is available in two versions:

Model SD 100 B with modem Model SD 100 IB without modem for international use

The SD100B is sold in the United States and is UL certified. The unit is battery powered and requires an external battery charger that is provided by SenDx Medical, Inc. Use only Jerome Industries power supply model WSZ118M which requires 100 to 250 VAC, 50 to 60 Hertz with these units. It also has a modem.

The SD1001B is sold internationally and is certified by TUV Rheinland. This unit is also battery powered and requires an external battery charger that is provided by SenDx Medical, Inc. Use only Advance Power Solutions power supply model AD-740U-1180 which requires 100 to 240 VAC, 50 to 60 Hertz with these units. This model is not equipped with a modem.

The SenDx 100™ analyzer utilizes microprocessor controlled electronics. The microprocessor controls the touch screen display, the analog electronics for processing sensor output, and the integral thermal printer. The printer provides a hard copy of the measured and calculated values. Samples are introduced into the analyzer via an aspiration stylus.

The SenDx 100™ sensor incorporates microelectrode technology for the measurement of blood oxygen (p(), carbon dioxide (pCO2), pHx, sodium (Na*), potassium (K*), chloride (Cl1), ionized calcium (iCat'), and hematocrit (Hct) in a multi-use, disposable cassette assembly.

The sensors are analogous to traditional electrode methodologies for the measurement of blood gases and electrolytes. A Clark cell for DO, measurement and potentiometric ion-selective electrodes for the measurement of pH, pCO,, Na , K , Cl and iCa * have been miniaturized for placement on a 1-3/8" x 3/8" sensing "chip". Hematocrit is measured by using conductivity.

The sensors are contained in a low volume flow-through cell. The sensor cassette contains reference electrodes for the potentiometric sensors and an integral temperature sensor and heating element for precise temperature control. Calibration and analysis are carried out at 37.0°C. The SenDx 100™ analyzer allows for patient temperature correction over the range of 12-45°C.

All SenDx 100™ sensor cassettes are 100% tested at the factory with multiple precision reference solutions.

{4}------------------------------------------------

Calibration of the system is accomplished using a calibration solution pack which contains two levels of precision tonometered electrolyte solutions packaged in gas tight disposable cartridges. Precise pH, blood gas, and electrolyte calibration values for each pack are provided on an attached calibration bar code label.

The SenDx 100TM performs two point calibrations at automatic intervals. Throughout the calibration and sample analysis phases, sensor signals are continuously monitored. If any abnormal conditions are detected, an error message is generated.

Blood analysis and the subsequent system calibration and flush take approximately 50-75 seconds. After analysis, pH, pCO2, pO2, Na', K+, Cl; iCa*, and Hct measurement results are displayed on the analyzer touch screen. The measured values and derived values (HCO), BEb, BEecf, TCO3, SBC, Hb, and %02 Sat) are automatically printed.

The SenDx 100™ system can be interfaced via the RS232 serial data port with PC based or LIS/HIS data management systems.

{5}------------------------------------------------

### Intended Use

্রামের

The SenDx 100™ pH, Blood Gas and Electrolyte Analysis System (SenDx 100™) is a portable, automated analyzer that measures blood oxygen, carbon dioxide, pH, sodium, potassium, chloride, ionized calcium, and hematocrit in whole blood. The SenDx 100™ system is intended for use by trained technologists, nurses, physicians, and therapists. It can be used in laboratory, near patient or point-of-care settings.

{6}------------------------------------------------

510(k) Number (if known):

Device Name: SenDx 100™ pH, Blood Gas and Electrolyte Analysis System

Indications for Use:

The SenDx 100™ pH, Blood Gas and Electrolyte Analysis System (SenDx 100™) is intended to measure blood oxygen, carbon dioxide, pH, sodium, potassium, chloride, ionized calcium and hematocrit in whole blood.

The SenDx 100™ system can be used in the laboratory, near patient, or in point-of-care settings. It is intended for the measurement of blood gases, electrolytes and hematocrit in arterial or venous whole blood samples.

## (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUED ON ANOTHER PAGE IF NEEDED)

#### Concurrence of CDRH, Office of Device Evaluation (ODE)

Prescription Use (Per 21 CFR 801.109)

্রামের

OR

Over-the-Counter Use _________________________________________________________________________________________________________________________________________________________

(Optional Format 1-2-96)

{7}------------------------------------------------

#### Technological Characteristics

As described in a previous 510(k) (K922631/B) for StatPal II, the StatPal II and the Mallinckrodt® GEM® STAT are primarily intended for determination of arterial pH, pO2 and pCO2. Similarly, as described in the previous 510(k) (K954482) for the SenDx 100™ , this device intended to be used for determination of arterial or venous pH, pCO2, pO2, Na', K', ionized Ca* and hematocrit in whole blood. The present 510(k) submission adds chloride to the electrolyte panel of the SenDx 100™. The SenDx 100 is similar to the Mallinckrodte GEM®-PREMIER, Corning 850 & 865, and NOVA STAT Profile 5 in intended use, methods of application and principles of operation.

The methods of application involve trained medical personnel in a central lab, stat lab or critical care environment. The StatPal II, SenDx 100 and GEM®-PREMIER can also be used near the patient due to the portable nature of the devices. In each of these devices, the principles of operation are similar or the same:

| • polarographic<br>(amperometric) | pO₂ measurement                               |
|-----------------------------------|-----------------------------------------------|
| • potentiometric<br>measurement   | pH, Na+, K+, Cl-, iCa++, and pCO₂ measurement |
| • conductance                     | Hematocrit measurement                        |

The SenDx 100 poses the same type of questions about safety and effectiveness of the equivalent devices.

{8}------------------------------------------------

#### NON-CLINICAL TESTING

Performance data was generated on-site at SenDx Medical, Inc. Comparisons were done with the following reference instruments:

- Corning 850 and spun hematocrit (micro centrifuge) o
### Linearity of Recovery:

- SenDx Medical, Inc. a
Linearity of recovery for pCO, and pO, was assessed versus target recovery values employing blood tonometry. Linearity of recovery for pH, K', Na , Cl and iCa* was assessed by selecting from the pooled whole blood data from five ranges evenly spread throughout the total range of measurement. Each range consisted of approximately 50 - 100 blood samples. The values for these ranges were independently verified against a Corning 850 Blood Gas and Electrolyte Analyzer. The resulting data were analyzed using full linear regression analysis. Regression slope, intercept, and correlation coefficient are reported.

| Analyte | n   | Slope | r    | Intercept | Std Error<br>Sy.x |
|---------|-----|-------|------|-----------|-------------------|
| pH      | 333 | 1.21  | 0.99 | -1.56     | 0.017             |
| pCO2    | 384 | 1.00  | 0.99 | 0.92      | 1.83              |
| pO2     | 321 | 1.01  | 1.00 | 0.51      | 3.0               |
| Na+     | 465 | 1.09  | 0.99 | -11.17    | 2.15              |
| K+      | 475 | 1.06  | 1.00 | -0.18     | 0.13              |
| Cl-     | 414 | 1.03  | 0.99 | 1.10      | 1.66              |
| iCa++   | 600 | 1.01  | 1.00 | -0.04     | 0.04              |
| Hct     | 747 | 1.12  | 0.98 | -2.42     | 1.69              |

#### Correlations:

: 大 : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : :

- SenDx Medical, Inc. .
The study was performed at SenDx Medical, Inc. comparing the SenDx 100™ to a Corning 850 Blood Gas and Electrolyte Analyzer and a spun hematocrit (micro-centerfuge). Tonometered and spiked whole blood were tested in order to verify the accuracy of the SenDx 100™ system.

{9}------------------------------------------------

| Analyte | n    | Slope | r    | Intercept | Std Error<br>Sy.x |
|---------|------|-------|------|-----------|-------------------|
| pH      | 472  | 1.23  | 0.99 | -1.65     | 0.013             |
| pCO₂    | 413  | 1.00  | 0.99 | 0.49      | 1.62              |
| pO₂     | 1677 | 0.97  | 0.99 | 3.81      | 5.56              |
| Na⁺     | 619  | 1.17  | 0.99 | -24.22    | 2.11              |
| K⁺      | 601  | 1.02  | 1.00 | 0.07      | 0.14              |
| Cl⁻     | 619  | 1.01  | 0.99 | -0.10     | 1.54              |
| iCa⁺⁺   | 621  | 1.33  | 0.99 | -0.4      | 0.05              |
| Hct     | 1229 | 1.12  | 0.95 | -4.63     | 2.07              |

### Precision

- SenDx Medical, Inc. g
An in-house precision study using quality control materials was performed over a period of 15 days. This study utilized multiple lots of sensor cassettes and calibration packs.

RNA Medical Qualidata levels 1, 2, 3 were used for blood gases and electrolytes, and RNA Medical hematocrit levels low and high were used for hematocrit. Each level was run in duplicate twice a day.

| SenDx 100™ |                   | Mean  | S. D. | %CV   |                        | Mean  | S. D. | %CV  |
|------------|-------------------|-------|-------|-------|------------------------|-------|-------|------|
|            | pH                |       |       |       | $pCO_2$<br>(mmHg)      |       |       |      |
| Level 1    |                   | 7.177 | 0.006 | 0.089 |                        | 67.5  | 2.1   | 3.1  |
| Level 2    |                   | 7.416 | 0.003 | 0.047 |                        | 45.0  | 0.7   | 1.7  |
| Level 3    |                   | 7.612 | 0.006 | 0.073 |                        | 24.7  | 0.7   | 2.8  |
|            | $pO_2$<br>(mmHg)  |       |       |       | $Na^+$<br>(mmol/L)     |       |       |      |
| Level 1    |                   | 71.4  | 3.0   | 4.3   |                        | 107.6 | 0.7   | 0.6  |
| Level 2    |                   | 109.5 | 2.0   | 1.9   |                        | 130.1 | 0.5   | 0.4  |
| Level 3    |                   | 146.0 | 1.5   | 1.0   |                        | 157.4 | 0.6   | 0.4  |
|            | $K^+$<br>(mmol/L) |       |       |       | $iCa^{++}$<br>(mmol/L) |       |       |      |
| Level 1    |                   | 1.85  | 0.02  | 1.30  |                        | 1.55  | 0.02  | 1.32 |
| Level 2    |                   | 4.39  | 0.03  | 0.78  |                        | 1.19  | 0.01  | 0.97 |
| Level 3    |                   | 6.46  | 0.06  | 0.92  |                        | 0.54  | 0.01  | 2.03 |
|            | Hct<br>(%)        |       |       |       | Cl<br>(mmol/L)         |       |       |      |
| Low        |                   | 35.0  | 0.3   | 0.7   |                        |       |       |      |
| High       |                   | 58.2  | 0.5   | 0.8   | Level 1                | 74.0  | 1.1   | 1.5  |
|            |                   |       |       |       | Level 2                | 94.0  | 0.5   | 0.6  |
|            |                   |       |       |       | Level 3                | 127.0 | 1.0   | 0.8  |

{10}------------------------------------------------

## CLINICAL TESTING

Clinical studies were conducted of the SenDx 100™ system at four different geographical sites. This study involved four different SenDx 100™ analyzers, and multiple lots of sensor cassettes and calibrant solution packs. These clinical studies included split sample testing of over 50 samples per site and precision testing with QC control samples.

#### Hospital Stat Lab �

Clinical studies comparing the SenDx 100 system to the Corning 865, Nova Stat Profile 5, Corning Co-oximeter and spun hematocrit (micro centrifuge) methodologies have been conducted by evaluating over 400 split-samples at four clinical sites. Shown below is the performance of a representative site:

| Analyte | n   | Slope | r    | Intercept | Std Error Sy.x |
|---------|-----|-------|------|-----------|----------------|
| pH      | 120 | 0.89  | 0.99 | 0.841     | 0.013          |
| pCO2    | 114 | 0.84  | 0.99 | 5.93      | 1.03           |
| pO2     | 119 | 1.02  | 1.00 | -0.08     | 5.77           |
| Na+     | 120 | 0.98  | 0.89 | 3.04      | 2.48           |
| K+      | 120 | 1.01  | 0.99 | -0.15     | 0.07           |
| iCa++   | 118 | 0.85  | 0.94 | 0.17      | 0.03           |
| Hct     | 102 | 1.11  | 0.93 | -4.03     | 2.55           |

Reference Device: Corning 865 and CO-oximeter module

#### Hospital Operating Room Point-Of-Care .

The SenDx 100™ system was subjected to operating room point-of-care use and at two different hospitals, operated by typical OR personnel. Shown below is the performance of a representative site:

| Analvte | n   | Slope | r    | Intercept | Std Error<br>Sy.x |
|---------|-----|-------|------|-----------|-------------------|
| pH      | 112 | 0.78  | 0.99 | 1.66      | 0.01              |
| pCO2    | 106 | 0.80  | 0.96 | 4.96      | 1.56              |
| pO2     | 108 | 0.79  | 0.95 | 54        | 28.5              |
| Na+     | 112 | 0.85  | 0.90 | 17.1      | 1.98              |
| K+      | 112 | 1.06  | 0.98 | -0.4      | 0.20              |
| iCa++   | 97  | 0.88  | 0.95 | 0.12      | 0.06              |
| Hct     | 111 | 1.03  | 0.99 | -2.35     | 1.85              |

Reference Device: Mallinckrodt Gem-Premier

{11}------------------------------------------------

#### Hospital Stat Lab �

Precision studies of the SenDx 100™ were conducted at four clinical sites over a five-day period. Reference quality control materials for pH, blood gases, electrolytes and hematocrit were tested. Each level was run in duplicate twice a day. Shown below are summarized precision data collected from these clinical sites:

Reference Device: Corning 865 and Co-oximeter Module

|                      | Mean   | Level 1<br>S. D.                                                                                                                                                              | % CV                                                                                                                                                                           | Mean                                                                                                                                                                          | Level 2<br>S. D. | % CV       | Mean           | Level 3<br>S. D.                                                                                                                                                              | % CV                                                                                                                                                                          |
|----------------------|--------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------|------------|----------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
|                      |        |                                                                                                                                                                               |                                                                                                                                                                                |                                                                                                                                                                               | 6 3 3 3          |            | ,在此次           | الموالي الموالي                                                                                                                                                               | ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘ |
| DE EXP<br>Within Run | 7.188  | ,一起,<br>0.001                                                                                                                                                                 | 이 기자 10<br>0.020                                                                                                                                                               | 7.416                                                                                                                                                                         | 0.007            | 0.010      | 7.599          | 0.002                                                                                                                                                                         | 0.028                                                                                                                                                                         |
| Within Day           | 7.188  | 0.001                                                                                                                                                                         | 0.013                                                                                                                                                                          | 7.416                                                                                                                                                                         | 0.001            | 0.017      | 7.601          | 0.003                                                                                                                                                                         | 0.035                                                                                                                                                                         |
| Day to Day           | 7.193  | 0.003                                                                                                                                                                         | 0.046                                                                                                                                                                          | 7.423                                                                                                                                                                         | 0.004            | 0.029      | 7.607          | 0.005                                                                                                                                                                         | 0.072                                                                                                                                                                         |
| Site to Site*        | 7.192  | 0.007                                                                                                                                                                         | 0.100                                                                                                                                                                          | 7.422                                                                                                                                                                         | 0.008            | 0.102      | 7.602          | 0.006                                                                                                                                                                         | 0.075                                                                                                                                                                         |
| pCO2 .               |        |                                                                                                                                                                               |                                                                                                                                                                                | Shop The                                                                                                                                                                      |                  |            | ર્વારા તા      | 的一个人                                                                                                                                                                          |                                                                                                                                                                               |
| Within Run           | 67.7   | 0.354                                                                                                                                                                         | 0.523                                                                                                                                                                          | Production of the<br>46.8                                                                                                                                                     | 0.000            | 0.000      | 想的一个感情<br>24.1 | 0.354                                                                                                                                                                         | 1.470                                                                                                                                                                         |
| Within Day           | 67.9   | 0.374                                                                                                                                                                         | 0.551                                                                                                                                                                          | 47.0                                                                                                                                                                          | 0.238            | 0.507      | 23.9           | 0.263                                                                                                                                                                         | 1.099                                                                                                                                                                         |
| Day to Day           | 67.6   | 0.642                                                                                                                                                                         | 0.952                                                                                                                                                                          | 46.0                                                                                                                                                                          | 1.015            | 2.216      | 23.9           | 0.760                                                                                                                                                                         | 3.332                                                                                                                                                                         |
| Site to Site*        | 66.4   | 1.883                                                                                                                                                                         | 2.835                                                                                                                                                                          | 44.5                                                                                                                                                                          | 1.098            | 2.470      | 23.3           | 0.385                                                                                                                                                                         | 1.653                                                                                                                                                                         |
| po2 po2              | 사      | ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘ | 100000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000 |                                                                                                                                                                               |                  |            | ,              | ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘ |                                                                                                                                                                               |
| Within Run           | 68.5   | 0.636                                                                                                                                                                         | 0.930                                                                                                                                                                          | 106.8                                                                                                                                                                         | 1.061            | 0.994      | 144.5          | 0.071                                                                                                                                                                         | 0.049                                                                                                                                                                         |
| Within Day           | 68.2   | 1.049                                                                                                                                                                         | 1.538                                                                                                                                                                          | 106.5                                                                                                                                                                         | 0.707            | 0.664      | 144.4          | 0.419                                                                                                                                                                         | 0.290                                                                                                                                                                         |
| Day to Day           | 68.0   | 1.433                                                                                                                                                                         | 1.937                                                                                                                                                                          | 106.2                                                                                                                                                                         | 0.953            | 0.877      | 143.6          | 0.789                                                                                                                                                                         | 0.543                                                                                                                                                                         |
| Site to Site*        | 68.2   | 2.502                                                                                                                                                                         | 3.667                                                                                                                                                                          | 107.4                                                                                                                                                                         | 1.328            | 1.236      | 143.1          | 2.546                                                                                                                                                                         | 1.779                                                                                                                                                                         |
| 1 Naje J             |        | 1. Sales of the                                                                                                                                                               | ﺍﻟﻤﺴﺘﻘﻠﺔ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘ  | ,                                                                                                                                                                             | 儿的母婴儿            |            | 이러한 이름이 있      |                                                                                                                                                                               | a ka                                                                                                                                                                          |
| Within Run           | 108.1  | 0.141                                                                                                                                                                         | 0.131                                                                                                                                                                          | 129.8                                                                                                                                                                         | 0.071            | 0.054      | 157.4          | 0.424                                                                                                                                                                         | 0.270                                                                                                                                                                         |
| Within Day           | 108.0  | 0.263                                                                                                                                                                         | 0.244                                                                                                                                                                          | 129.8                                                                                                                                                                         | 0.141            | 0.109      | 157.1          | 0.432                                                                                                                                                                         | 0.275                                                                                                                                                                         |
| Day to Day           | 108.0  | 0.455                                                                                                                                                                         | 0.421                                                                                                                                                                          | 129.7                                                                                                                                                                         | 0.234            | 0.180      | 156.5          | 0.454                                                                                                                                                                         | 0.290                                                                                                                                                                         |
| Site to Site*        | 107.4  | 2.133                                                                                                                                                                         | 1.987                                                                                                                                                                          | 129.2                                                                                                                                                                         | 1.452            | 1.124      | 156.2          | 1.176                                                                                                                                                                         | 0.753                                                                                                                                                                         |
| AK K                 | રે રું | ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘ | ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘ  | الموقع الموقع الموقع الموقع الموقع الموقع الموقع الموقع الموقع الموقع الموقع الموقع الموقع الموقع الموقع الموقع الموقع الموقع الموقع الموقع الموقع الموقع الموقع الموقع الموق |                  | an de la p |                |                                                                                                                                                                               | ﺍﻟﻤﺴﺎﻋﺪ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘ |
| Within Run           | 1.80   | 0.028                                                                                                                                                                         | 1.571                                                                                                                                                                          | 4.37                                                                                                                                                                          | 0.028            | 0.647      | 6.50           | 0.007                                                                                                                                                                         | 0.109                                                                                                                                                                         |
| Within Day           | 1.80   | 0.017                                                                                                                                                                         | 0.950                                                                                                                                                                          | 4.38                                                                                                                                                                          | 0.030            | 0.681      | 6.48           | 0.021                                                                                                                                                                         | 0.318                                                                                                                                                                         |
| Day to Day           | 1.81   | 0.013                                                                                                                                                                         | 0.716                                                                                                                                                                          | 4.37                                                                                                                                                                          | 0.015            | 0.347      | 6.46           | 0.031                                                                                                                                                                         | 0.477                                                                                                                                                                         |
| Site to Site*        | 1.81   | 0.019                                                                                                                                                                         | 1.055                                                                                                                                                                          | 4.35                                                                                                                                                                          | 0.036            | 0.822      | 6.43           | 0.047                                                                                                                                                                         | 0.728                                                                                                                                                                         |
| PriCattle            |        | のお得な                                                                                                                                                                          | ﺍﻟﻤﺴﺘﻘﻠﺔ ﺍﻟﻤ                                                                                                                                                                   |                                                                                                                                                                               | 14.08            |            | ,但是            |                                                                                                                                                                               |                                                                                                                                                                               |
| Within Run           | 1.56   | 0.042                                                                                                                                                                         | 2.720                                                                                                                                                                          | 1.17                                                                                                                                                                          | 0.000            | 0.000      | 0.49           | 0.000                                                                                                                                                                         | 0.000                                                                                                                                                                         |
| Within Day           | 1.55   | 0.029                                                                                                                                                                         | 1.856                                                                                                                                                                          | 1.16                                                                                                                                                                          | 0.015            | 1.290      | 0.50           | 0.006                                                                                                                                                                         | 1.166                                                                                                                                                                         |
| Day to Day           | 1.55   | 0.019                                                                                                                                                                         | 1.236                                                                                                                                                                          | 1.16                                                                                                                                                                          | 0.009            | 0.763      | 0.50           | 0.004                                                                                                                                                                         | 0.714                                                                                                                                                                         |
| Site to Site*        | 1.52   | 0.045                                                                                                                                                                         | 2.959                                                                                                                                                                          | 1.15                                                                                                                                                                          | 0.021            | 1.841      | 0.49           | 0.008                                                                                                                                                                         | 1.681                                                                                                                                                                         |
| Hotel Hotel          | ::     |                                                                                                                                                                               |                                                                                                                                                                                | - 北海                                                                                                                                                                          |                  | ,          |                |                                                                                                                                                                               |                                                                                                                                                                               |
| Within Run           | 35.4   | 0.212                                                                                                                                                                         | 0.600                                                                                                                                                                          | 58.7                                                                                                                                                                          | 0.071            | 0.121      |                |                                                                                                                                                                               |                                                                                                                                                                               |
| Within Day           | 35.4   | 0.173                                                                                                                                                                         | 0.490                                                                                                                                                                          | 58.7                                                                                                                                                                          | 0.141            | 0.241      |                |                                                                                                                                                                               |                                                                                                                                                                               |
| Day to Day           | 34.8   | 0.425                                                                                                                                                                         | 1.220                                                                                                                                                                          | 57.8                                                                                                                                                                          | 0.722            | 1.249      |                |                                                                                                                                                                               |                                                                                                                                                                               |
| Site to Site*        | 34.2   | 0.546                                                                                                                                                                         | 1.594                                                                                                                                                                          | 57.3                                                                                                                                                                          | 0.797            | 1.391      |                |                                                                                                                                                                               |                                                                                                                                                                               |

### SUMMARY OF PRECISION STUDY

* For three clinical sites.

{12}------------------------------------------------

## CONCLUSION FROM NONCLINICAL AND CLINICAL TESTING

From the nonclinical data we conclude that the performance of the SenDx 100™ system is equivalent to that of the Mallinckrodt GEM-PREMIER, Corning 865 and spun hematocrit for measurement of pH, pO2, pCO2, Na*, K*, Cli, iCa** and hematocrit.

From the clinical study data we conclude that the performance of the SenDx 100™ system is substantially equivalent to that of the predicate devices routinely employed at the clinical sites, for measurement of pH, pO3, pCO2, Na K*, iCa* and hematocrit. From these nonclinical and clinical studies we conclude that the performance of the SenDx 100™ system is substantially equivalent to that of the predicate devices routinely employed at several clinical sites.

These data adequately show that the SenDx 100™ system is safe and effective for its intended use in the laboratory or point-of-care. The data shows the SenDx 100™ system can be used at clinical sites for equivalent performance to the predicate devices routinely in use.

R. Chans

Signature of applicant

Sept. 29, 1997
Date

Date

{13}------------------------------------------------

Image /page/13/Picture/1 description: The image shows the seal of the U.S. Department of Health & Human Services. The seal features the department's name encircling a symbol. The symbol consists of a stylized caduceus, which is a staff with two snakes entwined around it, representing medicine and health. The seal is in black and white.

Food and Drug Administration 2098 Gaither Road Rockville MD 20850

DEC - 3 1997

Ruben Chairez, Ph.D. . Vice President, Regulatory Affairs/Quality Assurance SenDx Medical, Inc. 1945 Palomar Oaks Way Carlsbad, California 92009

Re : K973742 SenDx 100 pH, Blood Gas and Electrolyte Analysis System Regulatory Class: II CHL, CGZ, CEM, JFP, JGS, GKF Product Code: Dated: September 23, 1997 Received: October 1, 1997

Dear Dr. Chairez:

We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions The general controls provisions of the Act of the Act. include requirements for annual registration, listing of devices, qood manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major requlations affecting your device can be found in the Code of Federal Requlations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP requlation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.

{14}------------------------------------------------

Page 2

Under the Clinical Laboratory Improvement Amendments of 1988 (CLIA-88), this device may require a CLIA complexity categorization. To determine if it does, you should contact the Centers for Disease Control and Prevention (CDC) at (770) 488-7655.

This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling requlation (21 CFR Part 801 and additionally 809.10 for in vitro diaqnostic devices), please contact the Office of Compliance at (301) 594-4588. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html".

Sincerely yours,
Steven Bitman

Steven I. Gutman, M.D., M.B.A. Director Division of Clinical Laboratory Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{15}------------------------------------------------

# 510(k) PRE-MARKET NOTIFICATION 510(k) PRE-MARKET NOTIFICATION
SenDx 100 pH, BLOOD GAS AND ELECTROLYTE ANALYSIS SYSTEM

Ka73742

510(k) Number (if known):

SenDx 100™ pH, Blood Gas and Electrolyte Analysis System Device Name:

Indications for Use:

The SenDx 100™ pH, Blood Gas and Electrolyte Analysis System (SenDx 100™) is The SellDX 100 - p11, D100 Gar ara Love Story Corporation, potassium, chloride, ionized calcium and hematocrit in whole blood.

The SenDx 100™ system can be used in the laboratory, near patient, or in point-of-care The SellDX 100" System can oo abou in the tas binn the tass of electrolytes and hematocrit in arterial or venous whole blood samples.

# (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUED ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

| (Division Sign-Off)                     |
|-----------------------------------------|
| Division of Clinical Laboratory Devices |

| 510(k) Number | k973742 |
|---------------|---------|
|---------------|---------|

| Prescription Use<br>(Per 21 CFR 801.109) | OR | Over-the-Counter Use |
|------------------------------------------|----|----------------------|
|------------------------------------------|----|----------------------|

(Optional Format 1-2-96)

---

**Source:** [https://fda.innolitics.com/submissions/CH/subpart-b%E2%80%94clinical-chemistry-test-systems/CGZ/K973742](https://fda.innolitics.com/submissions/CH/subpart-b%E2%80%94clinical-chemistry-test-systems/CGZ/K973742)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
