K052662 · Ind Diagnostic, Inc. · CGJ · Nov 30, 2005 · Clinical Chemistry
Device Facts
Record ID
K052662
Device Name
ONE STEP FSH MENOPAUSAL TEST
Applicant
Ind Diagnostic, Inc.
Product Code
CGJ · Clinical Chemistry
Decision Date
Nov 30, 2005
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 862.1300
Device Class
Class 1
Indications for Use
The One Step FSH Menopausal Test is a rapid chromatographic immunoassay for the qualitative detection of follicle stimulating hormone (FSH) in urine to aid in the detection of the onset of menopause. The test utilizes a combination of antibodies including mouse monoclonal anti-FSH antibodies and goat polyclonal anti-mouse antibodies to selectively detect elevated levels of FSH. This device is intended for both professional and lay person use.
Device Story
Rapid chromatographic immunoassay for qualitative FSH detection in urine; aids menopause onset detection. Device formats include dipstick, midstream, and cassette. Principle: two-site immunoassay using membrane strips coated with mouse monoclonal anti-FSH antibody and goat anti-mouse IgG polyclonal antibody; colloidal gold conjugate of monoclonal anti-FSH antibody used for visualization. Used by professionals or lay persons (home/clinic). User observes colorimetric results on test strip after 10 minutes. Elevated FSH levels indicate potential menopause onset; results assist clinical decision-making regarding menopausal status.
Clinical Evidence
No clinical data provided; device relies on analytical performance characteristics typical of rapid chromatographic immunoassays.
Indicated for qualitative detection of follicle stimulating hormone (FSH) in urine to aid in the detection of the onset of menopause. Intended for professional and lay person use.
Regulatory Classification
Identification
A follicle-stimulating hormone test system is a device intended to measure follicle-stimulating hormone (FSH) in plasma, serum, and urine. FSH measurements are used in the diagnosis and treatment of pituitary gland and gonadal disorders.
Related Devices
K030058 — EARLYDETECT MENOPAUSE TEST FOR FSH · Early Detect, Inc. · Jul 30, 2003
K043599 — MOMENTS MENOPAUSE CHECK, MODEL 9113 · Phamatech, Inc. · Feb 25, 2005
K040575 — PHARMATECH MENOPAUSE CHECK, MODELS 9111 AND 9112 · Phamatech, Inc. · May 20, 2004
K041165 — FSH MENOPAUSE PREDICTOR TEST · ACON Laboratories, Inc. · Jun 17, 2004
Submission Summary (Full Text)
{0}
1
510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION
DECISION SUMMARY
ASSAY ONLY TEMPLATE
A. 510(k) Number:
k052662
B. Purpose for Submission:
New device
C. Measurand:
Follicle stimulating hormone
D. Type of Test:
Qualitative
E. Applicant:
IND Diagnostic Inc.
F. Proprietary and Established Names:
One Step FSH Menopausal Test
G. Regulatory Information:
1. Regulation section:
21 CFR§862.1300, Follicle-stimulating hormone test system
2. Classification:
Class I, meets the limitations of exemptions 862.9 (c), (9)
3. Product code:
CGJ
4. Panel:
75, Chemistry
{1}
H. Intended Use:
1. Intended use(s):
See indications for use
2. Indication(s) for use:
The One Step FSH Menopausal Test is a rapid chromatographic immunoassay for the qualitative detection of follicle stimulating hormone (FSH) in urine to aid in the detection of the onset of menopause. The test utilizes a combination of antibodies including mouse monoclonal anti-FSH antibodies and goat polyclonal anti-mouse antibodies to selectively detect elevated levels of FSH. This device is intended for both professional and lay person use.
3. Special conditions for use statement(s):
This device is for professional and over-the-counter use.
4. Special instrument requirements:
Not applicable
I. Device Description:
The One Step FSH Menopausal Test is supplied in three formats and is a chromatographic immunoassay for the qualitative detection of follicle stimulating hormone (FSH) in urine. The devices contain membrane strips coated with mouse monoclonal anti-FSH antibody A in the test region, goat anti-mouse (IgG) polyclonal antibody in the control region and a non-woven cloth which contains colloidal gold conjugate of monoclonal anti-FSH antibody B. The devices are single-use and visually read. One is a dipstick device, one is a midstream device and the other is a cassette device.
J. Substantial Equivalence Information:
1. Predicate device name(s):
FSH Menopause Predictor Test - Midstream
2. Predicate 510(k) number(s):
k046115
{2}
3
3. Comparison with predicate:
| Similarities | | |
| --- | --- | --- |
| Item | Device | Predicate |
| Intended Use | Detection of FSH in urine | Detection of FSH in urine |
| Principle/Methodology | Uses colorimetric visual interpretation | Uses colorimetric visual interpretation |
| Sensitivity | 25 mIU/mL | 25 mIU/mL |
| Differences | | |
| --- | --- | --- |
| Item | Device | Predicate |
| Device type | Midstream, test strip and cassette | midstream |
| Test time | Read results after 10 minutes | Read results after 3 minutes |
K. Standard/Guidance Document Referenced (if applicable):
None referenced
L. Test Principle:
Two site immunoassay
{3}
# M. Performance Characteristics (if/when applicable):
# 1. Analytical performance:
# a. Precision/Reproducibility:
Five batches of normal human urine negative for FSH were spiked to levels of 0, 12.5, 25, 37.5 and $50\mathrm{mIU / mL}$ . Ten consumers blindly tested each of the 5 batched using all three test formats. The results are in the table below:
| FSH Added mIU/mL | | 0 | 12.5 | 25 | 37.5 | 50 |
| --- | --- | --- | --- | --- | --- | --- |
| Number of samples tested | | 30 | 30 | 30 | 30 | 30 |
| Negative | MS | 10 | 10 | 1 | 0 | 0 |
| | C | 10 | 10 | 1 | 0 | 0 |
| | S | 10 | 10 | 1 | 0 | 0 |
| Positive | MS | 0 | 0 | 9 | 10 | 10 |
| | C | 0 | 0 | 9 | 10 | 10 |
| | S | 0 | 0 | 9 | 10 | 10 |
| MS = Midstream, C = Cassette, S = Strip | | | | | | |
# b. Linearity/assay reportable range:
Not applicable
# c. Traceability, Stability, Expected values (controls, calibrators, or methods):
Not applicable
# d. Detection limit:
The sensitivity of the test is $25\mathrm{mIU / mL}$ . Seventy-five samples from known non-menopausal were pooled together and split. Each pool was then spiked to concentrations of 0, 12.5, 25, 37.5 and $50\mathrm{mIU / mL}$ . Two trained technicians performed 5 tests at each concentration for each test format over two days. The results are in the table below:
| FSH Added (mIU/mL) | 0 | 12.5 | 25 | 37.5 | 50 |
| --- | --- | --- | --- | --- | --- |
| N | 30 | 30 | 30 | 30 | 30 |
| Negative | 30 | 30 | 1 | 0 | 0 |
| Positive | 0 | 0 | 29 | 30 | 30 |
{4}
# e. Analytical specificity:
The specificity of the One Step FSH Menopausal Test was tested with similar hormones and compounds found in normal human urine. Luteinizing hormones (1,000 mIU/mL), human chorionic gonadotropin (100,000 mIU/mL) and thyroid stimulation hormone (1,000 mIU/mL) were added to urine containing different levels of FSH concentration and did not interfere with the results.
Interferne substances were added in urine containing the following FSH concentration 0, 25 and $50\mathrm{mIU / mL}$ . None of the substances in the table below at the concentrations tested interfered with the assay.
| Substance | Concentration |
| --- | --- |
| Acetaminophen | 20 mg/mL |
| Acetylicylic acid | 20 mg/mL |
| Albumin | 100 mg/mL |
| Ascorbic Acid | 20 mg/mL |
| Atropine | 20 mg/mL |
| Bilirubin | 2 mg/mL |
| Caffeine | 20 mg/mL |
| Gentestic Acid | 20 mg/mL |
| Glucose | 2g/dL |
| Hemoglobin | 1 mg/dL |
# f. Assay cut-off:
See Detection limit section above.
{5}
2. Comparison studies:
a. Method comparison with predicate device:
The One Step FSH Menopausal Test was compared to the FSH Menopause Predictor Test – Midstream (k041165) using 75 women. 25 test were run for each format.
75 women, ages 40-60 tested their own urine at home using one of the three formats of One Step Menopausal Test and FSH Menopause Predictor Test. The women were instructed to record their results and to keep their urine for testing by the trained technician. The technician tested each sample using the One Step FSH Menopausal Test, all three formats and the FSH Menopause Predictor Test.
There was an accuracy of 94.7% between the One Step Menopausal Test-all formats and the FSN Menopause Predictor Test when performed by the lay user. There was a 98.7% agreement between the One Step Menopausal Test-all formats and the FSN Menopause Predictor Test when performed by the trained technician. There was a 98.7% agreement between the results of the lay person and the trained technician when using the One Step FSH Menopausal Test. The midstream test device when used by the lay user was done with midstream test method and the trained technician used the dip method.
b. Matrix comparison:
Not applicable
3. Clinical studies:
a. Clinical Sensitivity:
Not applicable
b. Clinical specificity:
Not applicable
c. Other clinical supportive data (when a. and b. are not applicable):
4. Clinical cut-off:
Not applicable
{6}
5. Expected values/Reference range:
The expected values are based on literature and in previous sensitivity studies that demonstrated adequate performance at the cutoff of 25 mIU.mL.
N. Proposed Labeling:
The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10.
O. Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
7
{7}
P. Other Supportive Device and Instrument Information:
Q. Administrative Information:
1. Applicant contact information:
a. Name of applicant:
IND Diagnostic, Inc.
b. Mailing address:
1629 Fosters Way
Delta, British Columbia V3M 6S7
Canada
c. Phone #:
(604) 522-1619
d. Fax #:
(604) 522-6331
e. E-mail address (optional):
ra@ind.ca
f. Contact:
Jason Peng, Regulatory Affairs Manager
2. Review documentation:
R. Reviewer Name and Signature:
Ann Chappie
CDRH/OIVD/DCTD
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.