ONE TOUCH II HOSPITAL BLOOD GLUCOSE MONITORING SYSTEM DATA DOCK
K972473 · Lifescan, Inc. · CGA · Dec 16, 1997 · Clinical Chemistry
Device Facts
Record ID
K972473
Device Name
ONE TOUCH II HOSPITAL BLOOD GLUCOSE MONITORING SYSTEM DATA DOCK
Applicant
Lifescan, Inc.
Product Code
CGA · Clinical Chemistry
Decision Date
Dec 16, 1997
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 862.1345
Device Class
Class 2
Attributes
Pediatric
Indications for Use
The ONE TOUCH® II HOSPITAL Blood Glucose Monitoring System, consisting of the ONE TOUCH® II Hospital Meter, ONE TOUCH® Test Strips, control solutions, data management software, and the DataDock and other optional accessories is intended for quantitative measurement of glucose in whole blood in hospital, clinic, and home settings. It is intended for monitoring blood glucose levels in neonatal or non-neonatal whole blood samples. Non-neonatal blood glucose results can be reported in either whole blood equivalents or plasma/serum equivalents.
Device Story
DataDock is an optional accessory for the ONE TOUCH® II Hospital Blood Glucose Monitoring System; functions as a docking station for the meter. Improves ease-of-use and enhances data management capabilities for the existing system. Used in hospital, clinic, and home settings by healthcare providers or patients. Facilitates data transfer and management; output used by clinicians to monitor blood glucose levels and inform clinical decision-making. Benefits include streamlined data handling and improved workflow for glucose monitoring.
Clinical Evidence
No clinical data provided; device is an accessory to a previously cleared system.
Technological Characteristics
Docking station accessory for blood glucose meter; provides data management and ease-of-use features. No changes to the core glucose oxidase sensing technology of the primary meter.
Indications for Use
Indicated for quantitative measurement of glucose in whole blood for neonatal and non-neonatal patients in hospital, clinic, and home settings.
Regulatory Classification
Identification
A glucose test system is a device intended to measure glucose quantitatively in blood and other body fluids. Glucose measurements are used in the diagnosis and treatment of carbohydrate metabolism disorders including diabetes mellitus, neonatal hypoglycemia, and idiopathic hypoglycemia, and of pancreatic islet cell carcinoma.
Special Controls
*Classification.* Class II (special controls). The device, when it is solely intended for use as a drink to test glucose tolerance, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9.
Related Devices
K024234 — ASCENSIA GLUCOFACTS; ASCENSIA GLUCOFACTS PROFESSIONAL · Bayer Corp. · Jan 23, 2003
K062434 — ZERO-CLICK DATA MANAGEMENT SYSTEM, MODEL 8000-01528 · Agamatrix · Nov 13, 2006
K983047 — ACCU-CHEK HQ SYSTEM · Boehringer Mannheim Corp. · Feb 5, 1999
{0}------------------------------------------------
K972473
# 510(k) Summary
DEC I S 1997
| Submitter | LifeScan, Inc.<br>1000 Gibraltar Drive<br>Milpitas, CA 95037<br>John E. Hughes |
|---------------------|--------------------------------------------------------------------------------|
| Date | July 1, 1997 |
| Proprietary Name | ONE TOUCH® II Hospital Blood Glucose<br>Monitoring System<br>DataDock |
| Common Name | Blood Glucose Monitor |
| Classification Name | 75CGA Glucose oxidase, glucose |
## Device Description
The DataDock is an optional accessory for the ONE TOUCH® II Hospital Meter. When the ONE TOUCH® II Hospital Meter is inserted into the DataDock, it provides features which improve ease-of-use and enhance data management capability.
## Intended Use
The ONE TOUCH® II HOSPITAL Blood Glucose Monitoring System, consisting of the ONE TOUCH® II Hospital Meter, ONE TOUCH® Test Strips, control solutions, data management software, and the DataDock and other optional accessories is intended for quantitative measurement of glucose in whole blood in hospital, clinic, and home settings. It is intended for monitoring blood glucose levels in neonatal or non-neonatal whole blood samples. Non-neonatal blood glucose results can be reported in either whole blood equivalents or plasma/serum equivalents.
### Substantial Equivalence
The DataDock is an optional accessory modifying the ONE TOUCH® II Hospital Blood Glucose Monitoring System currently in commercial distribution. The modification neither significantly changes the safety and effectiveness nor makes any change to its intended use of the device.
{1}------------------------------------------------
Image /page/1/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" around the perimeter. Inside the circle is an abstract symbol resembling an eagle or bird in flight, composed of three curved lines.
Food and Drug Administration 2098 Gaither Road Rockville MD 20850
DEC 1 6 1997
John E. Hughes Manager, Regulatory Affairs LifeScan Inc. 1000 Gibraltar Drive Milpitas, California 95035-6312
K972473 Re : ONE TOUCH II DataDock Hospital Blood Glucose Monitoring System Regulatory Class: II Product Code: CGA October 1, 1997 Dated: October 3, 1997 Received:
Dear Mr. Hughes:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions The general controls provisions of the Act of the Act. include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Requlations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical General regulation (21 CFR Part 820) and that, Devices: through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory In addition, FDA may publish further announcements action. concerning your device in the Federal Register. Please note:
this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
{2}------------------------------------------------
Page 2
Under the Clinical Laboratory Improvement Amendments of 1988 (CLIA-88), this device may require a CLIA complexity categorization. To determine if it does, you should contact the Centers for Disease Control and Prevention (CDC) at (770) 488-7655.
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4588. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html".
Sincerely yours,
Steven Sitman
Steven I. Gutman, M.D., M.B.A. Director Division of Clinical Laboratory Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
K972473
# 510(k) Summary
DEC I S 1997
| Submitter | LifeScan, Inc.<br>1000 Gibraltar Drive<br>Milpitas, CA 95037<br>John E. Hughes |
|---------------------|--------------------------------------------------------------------------------|
| Date | July 1, 1997 |
| Proprietary Name | ONE TOUCH® II Hospital Blood Glucose<br>Monitoring System<br>DataDock |
| Common Name | Blood Glucose Monitor |
| Classification Name | 75CGA Glucose oxidase, glucose |
## Device Description
The DataDock is an optional accessory for the ONE TOUCH® II Hospital Meter. When the ONE TOUCH® II Hospital Meter is inserted into the DataDock, it provides features which improve ease-of-use and enhance data management capability.
## Intended Use
The ONE TOUCH® II HOSPITAL Blood Glucose Monitoring System, consisting of the ONE TOUCH® II Hospital Meter, ONE TOUCH® Test Strips, control solutions, data management software, and the DataDock and other optional accessories is intended for quantitative measurement of glucose in whole blood in hospital, clinic, and home settings. It is intended for monitoring blood glucose levels in neonatal or non-neonatal whole blood samples. Non-neonatal blood glucose results can be reported in either whole blood equivalents or plasma/serum equivalents.
### Substantial Equivalence
The DataDock is an optional accessory modifying the ONE TOUCH® II Hospital Blood Glucose Monitoring System currently in commercial distribution. The modification neither significantly changes the safety and effectiveness nor makes any change to its intended use of the device.
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" around the perimeter. Inside the circle is an abstract symbol resembling an eagle or bird in flight, composed of three curved lines.
Food and Drug Administration 2098 Gaither Road Rockville MD 20850
DEC 1 6 1997
John E. Hughes Manager, Regulatory Affairs LifeScan Inc. 1000 Gibraltar Drive Milpitas, California 95035-6312
K972473 Re : ONE TOUCH II DataDock Hospital Blood Glucose Monitoring System Regulatory Class: II Product Code: CGA October 1, 1997 Dated: October 3, 1997 Received:
Dear Mr. Hughes:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions The general controls provisions of the Act of the Act. include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Requlations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical General regulation (21 CFR Part 820) and that, Devices: through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory In addition, FDA may publish further announcements action. concerning your device in the Federal Register. Please note:
this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
{5}------------------------------------------------
Page 2
Under the Clinical Laboratory Improvement Amendments of 1988 (CLIA-88), this device may require a CLIA complexity categorization. To determine if it does, you should contact the Centers for Disease Control and Prevention (CDC) at (770) 488-7655.
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4588. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html".
Sincerely yours,
Steven Sitman
Steven I. Gutman, M.D., M.B.A. Director Division of Clinical Laboratory Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
### Indications for Use Form
510(k) Number (if known)
972473
Device Name:
ONE TOUCH® II Hospital Blood Glucose Monitoring System DataDock
Indications for Use
The ONE TOUCH® II HOSPITAL Blood Glucose Monitoring System, consisting of the ONE TOUCH® II Hospital Meter, ONE TOUCH® Test Strips, control solutions, data management software, the optional DataDock and other accessories is intended for quantitative measurement of glucose in whole blood in hospital, clinic, and home settings. It is intended for monitoring glucose levels in neonatal or non-neonatal whole blood samples. Non-neonatal glucose results can be reported in either whole blood equivalents or plasma/serum equivalents.
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use
(per 21 CFR 801.109) ✓
OR
Over-The-Counter Use (Optional Format 1-2-96)
-
(Division Sign-Off)
Division of Clincal Laboratory Devices,
510(k) Number. 972473 134
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.