GAL-1F Blood Glucose Monitoring System, GAL-1F Pro Blood Glucose Monitoring System
Applicant
Apex BioTechnology Corp.
Product Code
CGA · Clinical Chemistry
Decision Date
Aug 6, 2015
Decision
SESE
Submission Type
Special
Regulation
21 CFR 862.1345
Device Class
Class 2
Attributes
Pediatric
Indications for Use
The GAL-1F Blood Glucose Monitoring System is intended for the quantitative measurement of glucose in fresh capillary whole blood samples drawn from the fingertips, forearm, or palm. Alternative site testing should be performed only during steady-state (when glucose is not changing rapidly). Testing is done outside the body (In Vitro diagnostic use). It is indicated for lay use by people with diabetes, as an aid to monitoring blood glucose levels in Diabetes Mellitus and should only be used by a single patient and it should not be shared. It is not indicated for the diagnosis or screening of diabetes or for neonatal use. The GAL-1F Pro Blood Glucose Monitoring System is intended for the quantitative measurement of glucose in fresh capillary whole blood samples drawn from the fingertips, forearm, or palm. Alternative site testing should be performed only during steady-state (when glucose is not changing rapidly). Testing is done outside the body (In Vitro diagnostic use). It is indicated to be used for multiple patients in a clinical setting by healthcare professionals, as an aid to monitoring blood glucose levels in Diabetes Mellitus. This system is only used with single-use, auto-disabling lancets. It is not indicated for the diagnosis or screening of diabetes or for neonatal use.
Device Story
GAL-1F/GAL-1F Pro systems measure glucose in fresh capillary whole blood (fingertip, forearm, palm). System comprises meter and test strips. GAL-1F intended for single-patient home use; GAL-1F Pro intended for multi-patient clinical use with single-use, auto-disabling lancets. Principle of operation involves electrochemical measurement of glucose; uses same test strip and algorithm as predicate. Meter design changes include relocated test strip holder and removal of strip ejection mechanism. Software supports data download. Healthcare providers/patients use output to monitor glucose levels in Diabetes Mellitus. Alternative site testing restricted to steady-state conditions.
Clinical Evidence
No clinical data. Evidence consists of bench testing, including disinfection efficacy studies demonstrating complete inactivation of HBV/HbsAg using specified EPA-registered wipes, and robustness studies simulating 5 years (single-patient) and 3 years (multiple-patient) of cleaning/disinfection cycles with no degradation in performance or materials.
Technological Characteristics
Electrochemical glucose monitoring system. Uses same test strip and algorithm as predicate. Features include data download capability. Meter design modified to relocate test strip holder and remove ejection mechanism. Tested for EMC, electrical safety, and disinfection robustness.
Indications for Use
Indicated for people with diabetes (lay use for GAL-1F; professional use for GAL-1F Pro) as an aid to monitor blood glucose levels. Not indicated for diagnosis, screening of diabetes, or neonatal use.
Regulatory Classification
Identification
A glucose test system is a device intended to measure glucose quantitatively in blood and other body fluids. Glucose measurements are used in the diagnosis and treatment of carbohydrate metabolism disorders including diabetes mellitus, neonatal hypoglycemia, and idiopathic hypoglycemia, and of pancreatic islet cell carcinoma.
Special Controls
*Classification.* Class II (special controls). The device, when it is solely intended for use as a drink to test glucose tolerance, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9.
Predicate Devices
GAL-1E and GAL-1E Multi (k113547) Blood Glucose Monitoring Systems
Related Devices
K113208 — GAL-1A BLOOD GLUCOSE MONITORING SYSTEM · Apex BioTechnology Corp. · Jan 26, 2012
K113670 — ELEMENT BLOOD GLUCOSE MONITORING SYSTEM · Infopia Co, Ltd. · May 8, 2012
K073648 — GLUCOSURE STAR BLOOD GLUCOSE MONITORING SYSTEM, MODEL AS90000E1 · Apex BioTechnology Corp. · Apr 10, 2008
K120989 — TRUE METRIX SELF-MONITORING BLOOD GLUCOSE SYSTEM, TRUE METRIX PROFESSIONAL MONITORING BLOOD GLUCOSE SYSTEM · Nipro Diagnostics, Inc. · Apr 24, 2013
K063068 — FERRARA BLOOD GLUCOSE MONITORING SYSTEM, MODEL 631100 · Arkray USA, Inc. · Dec 20, 2006
Submission Summary (Full Text)
{0}
SPECIAL 510(k): Device Modification OIR Decision Summary
To: THE FILE
RE: DOCUMENT NUMBER K142689
This 510(k) submission contains information/data on modifications made to the SUBMITTER'S own Class II, Class III or Class I devices requiring 510(k). The following items are present and acceptable (delete/add items as necessary):
1. The name and 510(k) number of the SUBMITTER'S previously cleared device. (For a preamendments device, a statement to this effect has been provided.) GAL-1E Blood Glucose Monitoring System and GAL-1E Multi Blood Glucose Monitoring System, k113547
2. Submitter's statement that the INDICATION/INTENDED USE of the modified device as described in its labeling HAS NOT CHANGED along with the proposed labeling which includes instructions for use, package labeling, and, if available, advertisements or promotional materials (labeling changes are permitted as long as they do not affect the intended use).
3. A description of the device MODIFICATION(S), including clearly labeled diagrams, engineering drawings, photographs, user's and/or service manuals in sufficient detail to demonstrate that the FUNDAMENTAL SCIENTIFIC TECHNOLOGY of the modified device has not changed.
This change was for the following items:
i. Re-design of the device appearance: moving the strip reader location from the top of the device to the bottom, and eliminating a strip eject button from the device design.
ii. Adding functionality that allows blood glucose data to be transferred to a personal computer
iii. Design changes to the device circuit board.
iv. Adding new disinfecting agents into the instructions for use.
4. Comparison Information (similarities and differences) to applicant's legally marketed predicate device including, labeling, intended use, physical characteristics, device performance and specifications.
5. A Design Control Activities Summary which includes:
a) Identification of Risk Analysis method(s) used to assess the impact of the modification on the device and its components, and the results of the analysis
b) Based on the Risk Analysis, an identification of the verification and/or validation activities required, including methods or tests used and acceptance criteria to be applied
The labeling for this modified subject device has been reviewed to verify that the indication/intended use for the device is unaffected by the modification. In addition, the submitter's description of the particular modification(s) and the comparative information between the modified and unmodified devices demonstrate that the fundamental scientific technology has not changed. The submitter has provided the design control information as specified in The New 510(k) Paradigm and on this basis, I recommend the device be determined substantially equivalent to the previously cleared (or their preamendment) device.
## Infection Control Studies:
## GAL-1F Blood Glucose Meter
The device system is intended for single-patient use only. Disinfection efficacy studies were performed on the materials comprising the meter by an outside commercial testing laboratory demonstrating complete inactivation of hepatitis B virus (HBV) or elimination of Hepatitis B surface antigen (HbsAg) with the chosen disinfecting agents, Dispatch Hospital Cleaner Disinfectant Towels with Bleach (EPA Registration # 56392-8), Clorox Bleach Germicidal Wipes (EPA Registration #67619-12), Medline Microkill Bleach Germicidal Bleach Wipes (EPA Registration # 69687-1-37549)
{1}
Page 2 of 2
and Medline Micro-Kill+ Disinfection, Deodorizing, Cleaning Wipes with Alcohol (EPA Registration # 59894-10-37549). Robustness studies were also performed by the sponsor demonstrating that there was no change in performance or external materials of the meter and lancing device after 1,825 cleanings and 1,825 disinfection steps with the Dispatch Hospital Cleaner Disinfectant Towels with Bleach, Clorox Bleach Germicidal Wipes, Medline Microkill Bleach Germicidal Bleach Wipes and Medline Micro-Kill+ Disinfection, Deodorizing, Cleaning Wipes with Alcohol. The robustness studies were designed to simulate 5 years of single-patient use. Labeling was reviewed for adequate instructions for the validated cleaning and disinfection procedures.
## GAL-1F Multi Blood Glucose Meter
The device system is intended for use in multiple patients with single use, auto-disabling lancing devices. Disinfection efficacy studies were performed on the materials comprising the meter by an outside commercial testing demonstrating complete inactivation of hepatitis B virus (HBV) or elimination of Hepatitis B surface antigen (HbsAg) with the chosen disinfecting agents, Dispatch Hospital Cleaner Disinfectant Towels with Bleach (EPA Registration # 56392-8), Clorox Bleach Germicidal Wipes (EPA Registration #67619-12), Medline Microkill Bleach Germicidal Bleach Wipes (EPA Registration # 69687-1-37549) and Medline Micro-Kill+ Disinfection, Deodorizing, Cleaning Wipes with Alcohol (EPA Registration # 59894-10-37549). Robustness studies were also performed by the sponsor demonstrating that there was no change in performance or external materials of the meter and lancing device after 10,950 cleanings and 10,950 disinfection steps with the Dispatch Hospital Cleaner Disinfectant Towels with Bleach, Clorox Bleach Germicidal Wipes, Medline Microkill Bleach Germicidal Bleach Wipes and Medline Micro-Kill+ Disinfection, Deodorizing, Cleaning Wipes with Alcohol. The robustness studies were designed to simulate 3 years of multiple-patient use Labeling was reviewed for adequate instructions for the validated cleaning and disinfection procedures.
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.