FORA TD-4251/U-RIGHT TD-4252/U-RIGHT TD-4254 BLOOD GLUCOSE MONITORING SYSTEM, MODEL TD-4251/TD-4252A/TD-4252B/TD-4254

K082482 · Taidoc Technology Corporation · CGA · Jan 30, 2009 · Clinical Chemistry

Device Facts

Record IDK082482
Device NameFORA TD-4251/U-RIGHT TD-4252/U-RIGHT TD-4254 BLOOD GLUCOSE MONITORING SYSTEM, MODEL TD-4251/TD-4252A/TD-4252B/TD-4254
ApplicantTaidoc Technology Corporation
Product CodeCGA · Clinical Chemistry
Decision DateJan 30, 2009
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 862.1345
Device ClassClass 2
AttributesPediatric

Indications for Use

The FORA TD-4251, URIGHT TD-4252, URIGHT TD-4254 Blood Glucose Monitoring System is intended for use in the quantitative measurement of glucose in fresh capillary whole blood from the finger, palm, forearm, upper arm, calf, or thigh. It is intended for use by healthcare professionals and people with diabetes mellitus at home as an aid in monitoring the effectiveness of diabetes control programs. It is not intended for the diagnosis of or screening for diabetes mellitus, and is not intended for use on neonates.

Device Story

Blood glucose monitoring system; measures glucose in capillary whole blood samples; utilizes electrochemical biosensor technology; provides quantitative glucose concentration output; used by patients at home or healthcare professionals in clinical settings; aids in diabetes management; modification of previously cleared TD-4230 Clever Check system; involves external casing redesign; fundamental scientific technology remains unchanged.

Clinical Evidence

No clinical data provided in the document. Substantial equivalence is based on bench testing and performance characteristics typical for glucose monitoring systems.

Technological Characteristics

Electrochemical blood glucose monitoring system; external casing redesigned from predicate; fundamental sensing technology unchanged; standalone device; intended for point-of-care or home use.

Indications for Use

Indicated for quantitative glucose measurement in fresh capillary whole blood (finger, palm, upper-arm, calf, thigh) for healthcare professionals and patients with diabetes mellitus to monitor glycemic control. Contraindicated for neonates and non-steady-state glucose conditions for alternative site testing.

Regulatory Classification

Identification

A glucose test system is a device intended to measure glucose quantitatively in blood and other body fluids. Glucose measurements are used in the diagnosis and treatment of carbohydrate metabolism disorders including diabetes mellitus, neonatal hypoglycemia, and idiopathic hypoglycemia, and of pancreatic islet cell carcinoma.

Special Controls

*Classification.* Class II (special controls). The device, when it is solely intended for use as a drink to test glucose tolerance, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9.

Related Devices

Submission Summary (Full Text)

{0} SPECIAL 510(k): Device Modification OIVD Review Memorandum (Decision Making Document is Attached) To: THE FILE RE: DOCUMENT NUMBER K082482 FORA TD-4251, URIGHT TD -4252, URIGHT TD-4254 Blood Glucose Monitoring System This 510(k) submission contains information/data on modifications made to the SUBMITTER'S own Class II, Class III or Class I devices requiring 510(k). The following items are present and acceptable (delete/add items as necessary): 1. The name and 510(k) number of the TaiDoc Technology Corporation's previously cleared device, TD-4230 Clever Check Blood Glucose Monitoring System k070472. (For a preamendments device, a statement to this effect has been provided.) 2. Submitter's statement that the INDICATION/INTENDED USE of the modified device as described in its labeling HAS NOT CHANGED along with the proposed labeling which includes instructions for use, package labeling, and, if available, advertisements or promotional materials (labeling changes are permitted as long as they do not affect the intended use). 3. A description of the device MODIFICATION(S), including clearly labeled diagrams, engineering drawings, photographs, user's and/or service manuals in sufficient detail to demonstrate that the FUNDAMENTAL SCIENTIFIC TECHNOLOGY of the modified device has not changed. This change was for external design modifications to the TD-4230 Clever Check Blood Glucose Meter. The outer casing of the TD-4230 meter (Clever Check TD-4230) has been redesigned to make additional meters: the FORA TD-4251, the U-Right TD-4252, and the U-Right TD-4254 meter. Comparison Information (similarities and differences) to applicant's legally marketed predicate device including, labeling, intended use, physical characteristics, and coding requirements. 4. A Design Control Activities Summary which includes: a) Identification of Risk Analysis method(s) used to assess the impact of the modification on the device and its components, and the results of the analysis. The sponsor used FMEA. b) Based on the Risk Analysis, an identification of the verification and/or validation activities required, including methods or tests used and acceptance criteria to be applied c) A declaration of conformity with design controls. The declaration of conformity should include: i) A statement signed by the individual responsible, that, as required by the risk analysis, all verification and validation activities were performed by the designated individual(s) and the results demonstrated that the predetermined acceptance criteria were met, and ii) A statement signed by the individual responsible, that the manufacturing facility is in conformance with design control procedure requirements as specified in 21 CFR 820.30 and the records are available for review. 5. A Truthful and Accurate Statement, a 510(k) Summary or Statement and the Indications for Use Enclosure (and Class III Summary for Class III devices). The labeling for this modified subject device has been reviewed to verify that the indication/intended use for the device is unaffected by the modification. In addition, the submitter's description of the particular modification(s) and the comparative information between the modified and unmodified devices demonstrate that the fundamental scientific technology has not changed. The submitter has provided the design control information as specified in The New 510(k) Paradigm and on this basis, I recommend the device be determined substantially equivalent to the previously cleared (or their preamendment) device.
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...