← Product Code [CEM](/submissions/CH/subpart-b%E2%80%94clinical-chemistry-test-systems/CEM) · K993083

# ALCYON 300I (WITH ISE MODULE) ANALYZER (K993083)

_Abbott Laboratories · CEM · Oct 5, 1999 · Clinical Chemistry · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/CH/subpart-b%E2%80%94clinical-chemistry-test-systems/CEM/K993083

## Device Facts

- **Applicant:** Abbott Laboratories
- **Product Code:** [CEM](/submissions/CH/subpart-b%E2%80%94clinical-chemistry-test-systems/CEM.md)
- **Decision Date:** Oct 5, 1999
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 862.1600
- **Device Class:** Class 2
- **Review Panel:** Clinical Chemistry

## Indications for Use

The ALCYON Analyzer is an automated chemistry analyzer for in vitro diagnostic use. The analyzer performs quantitative kinetic and endpoint determinations of specific analytes. The ALCYON 300i Analyzer with the ISE Module additionally measures the concentration of the electrolytes, sodium, potassium, and chloride, in samples, using indirect potentiometry. Per 21 CFR, 682.2160, the ALCYON Analyzer is a discrete photometric chemistry analyzer for clinical use. The device is intended to duplicate manual analytical procedures by performing various steps such as pipetting, preparing filtrates, heating, and measuring color intensity. This device is intended for use in conjunction with certain materials to measure a variety of analytes. The analyzer with the optional Ion-Selective Electrode (ISE) Module measures the concentration of the electrolytes, sodium, potassium, and chloride, in samples using indirect potentiometry.

## Device Story

ALCYON Analyzer is an automated, open-system clinical chemistry analyzer; performs quantitative kinetic and endpoint determinations of analytes. Optional Ion-Selective Electrode (ISE) module measures sodium, potassium, and chloride concentrations in serum or plasma samples using indirect potentiometry. Device automates manual laboratory procedures including pipetting, filtrate preparation, heating, and color intensity measurement. Used in clinical laboratory settings by trained personnel. Output consists of quantitative analyte concentrations; healthcare providers use these results for diagnostic assessment and clinical decision-making. Benefits include increased efficiency and standardization of clinical chemistry testing compared to manual methods.

## Clinical Evidence

Bench testing only. Anticoagulant studies compared recovery of sodium, potassium, and chloride in various plasma collection tubes (Lithium Heparin, Sodium Heparin, Ammonium Heparin, Sodium Citrate, Potassium-EDTA) against serum specimen baselines. Results demonstrated mean recovery within ±10% for most heparinized and plastic tubes, supporting the use of plasma specimens for the specified analytes.

## Technological Characteristics

Automated discrete photometric chemistry analyzer. ISE module utilizes indirect potentiometry for electrolyte measurement. System performs automated pipetting, heating, and colorimetric analysis. Open system architecture.

## Regulatory Identification

A potassium test system is a device intended to measure potassium in serum, plasma, and urine. Measurements obtained by this device are used to monitor electrolyte balance in the diagnosis and treatment of diseases conditions characterized by low or high blood potassium levels.

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

### 510(k) Summary

OCT 5 1999

| Submitter's name/address             | Contact Person                                            |
|--------------------------------------|-----------------------------------------------------------|
| Abbott Laboratories                  | Mark Littlefield                                          |
| 1920 Hurd Drive                      | Section Manager                                           |
| M.S. 1-8                             | Regulatory Affairs                                        |
| Irving, Texas 75038                  | (972) 518-6062                                            |
|                                      | Fax (972) 753-3367                                        |
| Date of Preparation of this Summary: | September 13, 1999                                        |
| Device Trade or Proprietary Name:    | Abbott ALCYON Analyzer                                    |
| Device Common Name:                  | Clinical Chemistry Analyzer (with<br>optional ISE Module) |
| Classification Numbers/Class:        | 75JJD, Class 1                                            |
|                                      | 75JGS, Class II                                           |
|                                      | 75CEM, Class II                                           |
|                                      | 75CGZ, Class II                                           |

## 510(k) Summary

This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.

The assigned 510(k) number is: k993083

#### Description:

The ALCYON Analyzer is an automated open system for quantitative analysis of clinical chemistries. The ALCYON Analyzer has an optional Ion-Selective Electrode (ISE) module which measures the concentration of the electrolytes, sodium, potassium, and chloride, in samples using indirect potentiometry.

## Substantial Equivalence:

Substantial equivalence has been demonstrated between the use of serum specimens and plasma specimens for the quantitative determination of sodium, and chloride using indirect potentiometry (ISE module) on the ALCYON 300i Analyzer.

{1}------------------------------------------------

## Intended Use:

The ALCYON Analyzer is an automated chemistry analyzer for in vitro diagnostic use. The analyzer performs quantitative kinetic and endpoint determinations of specific analytes. The ALCYON 300i Analyzer with the ISE Module additionally measures the concentration of the electrolytes, sodium, potassium, and chloride, in samples, using indirect potentiometry.

STATE AND ATTERNAL PROPERTY COLLECTION COLLECTION CONTROLLERS OF CONSULTION OF CONSULTION OF

### Performance Characteristics:

Anticoagulant studies were performed using Lithium Heparin, Sodium Heparin, Ammonia Heparin, Sodium Citrate, Potassium-EDTA, plastic tubes and baselined against serum specimen tubes.

Less than ± 10% difference from the serum specimen collection tube baseline was observed for sodium heparin, lithium heparin, ammonium heparin, plastic, and SST collection tubes for the quantitation of sodium, potassium, and chloride.

Sodium Anticoagulant Recovery

| Substance           | Mean of<br>% Recovery |
|---------------------|-----------------------|
| SST                 | 99.7                  |
| Na Hep (full draw)  | 100.7                 |
| Li Hep              | 99.6                  |
| NH4 Hep             | 99.9                  |
| KEDTA               | 95.0                  |
| Na Citrate          | 188.5                 |
| Plastic             | 99.8                  |
| Na Hep (1 / 2 draw) | 100.3                 |

{2}------------------------------------------------

## Potassium Anticoagulant Recovery

| S u b s t a n c e             | M e a n o f<br>% R e c o v e r y |
|-------------------------------|----------------------------------|
| S S T                         | 9 9 . 9                          |
| N a H e p ( f u l l d r a w ) | 9 1 . 0                          |
| L i H e p                     | 9 2 . 3                          |
| N H 4 H e p                   | 9 1 . 2                          |
| K E D T A                     | 7 5 9 . 5                        |
| N a C i t r a t e             | 7 6 . 7                          |
| P l a s t i c                 | 1 0 0 . 8                        |
| N a H e p ( 1 / 2 d r a w )   | 9 2 . 1                          |

# Chloride Anticoagulant Recovery

|  | S u b s t a n c e             |  | M e a n o f       |  |
|--|-------------------------------|--|-------------------|--|
|  |                               |  | % R e c o v e r y |  |
|  | S S T                         |  | 9 9 . 7           |  |
|  | N a H e p ( f u l l d r a w ) |  | 1 0 0 . 0         |  |
|  | L i H e p                     |  | 9 9 . 8           |  |
|  | N H 4 H e p                   |  | 9 9 . 6           |  |
|  | K E D T A                     |  | 9 9 . 5           |  |
|  | N a C i t r a t e             |  | 1 0 3 . 4         |  |
|  | P l a s t i c                 |  | 1 0 0 . 1         |  |
|  | N a H e p ( 1 / 2 d r a w )   |  | 9 9 . 9           |  |

- +

## Conclusion:

The data demonstrates that the use of plasma specimens is an acceptable specimen type for the quantitative determinations of sodium, potassium, and chloride using indirect potentiometry on the ALCYON 300i Analyzer.

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo consists of a stylized eagle with three lines representing its body and head. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular fashion around the eagle.

Food and Drug Administration 2098 Gaither Road Rockville MD 20850

OCT 5 1999

Mr. Mark Littlefield Section Manager, Regulatory Affairs Abbott Laboratories ADD Regulatory Affairs 1920 Hurd Drive Irving, Texas 75038

Re: K993083

Trade Name: ALCYON™ 300I (with ISE Module) Analyzer Regulatory Class: II Product Code: CEM, CGZ, JGS Dated: September 15, 1999 Received: September 15, 1999

Dear Mr. Littlefield:

We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.

{4}------------------------------------------------

#### Page 2

Under the Clinical Laboratory Improvement Amendments of 1988 (CLIA-88), this device may require a CLIA complexity categorization. To determine if it does, you should contact the Centers for Disease Control and Prevention (CDC) at (770) 488-7655.

This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4588. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification"(21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597, or at its internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html".

Sincerely yours,

Steven Putman

Steven I. Gutman, M.D, M.B.A. Director Division of Clinical Laboratory Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

510(k) Number (if known): 5993083

ALCYON™ Analyzer Device Name: _________________________________________________________________________________________________________________________________________________________________

Indications For Use:

જુદી -Per 21 CFR, 682.2160, the ALCYON Analyzer is a discrete photometric chemistry analyzer for clinical use. The device is intended to duplicate manual analytical procedures by performing various steps such as pipetting, preparing filtrates, heating, and measuring color intensity. This device is intended for use in conjunction with certain materials to measure a variety of analytes. The analyzer with the optional Ion-Selective Electrode (ISE) Module measures the concentration of the electrolytes, sodium, potassium, and chloride, in samples using indirect potentiometry.

Jancooper

(Division Sign-Off)
Division of Clinical Laboratory Devices
510(k) Number K993083

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Prescription Use_ (Per 21 CFR 801.109)

OR

Over-The-Counter Use_

(Optional Format 1-2-96)

---

**Source:** [https://fda.innolitics.com/submissions/CH/subpart-b%E2%80%94clinical-chemistry-test-systems/CEM/K993083](https://fda.innolitics.com/submissions/CH/subpart-b%E2%80%94clinical-chemistry-test-systems/CEM/K993083)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
