← Product Code [CDZ](/submissions/CH/subpart-b%E2%80%94clinical-chemistry-test-systems/CDZ) · K152155

# ARCHITECT 2nd Generation Testosterone (K152155)

_Abbott Laboratories · CDZ · Dec 17, 2015 · Clinical Chemistry · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/CH/subpart-b%E2%80%94clinical-chemistry-test-systems/CDZ/K152155

## Device Facts

- **Applicant:** Abbott Laboratories
- **Product Code:** [CDZ](/submissions/CH/subpart-b%E2%80%94clinical-chemistry-test-systems/CDZ.md)
- **Decision Date:** Dec 17, 2015
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 862.1680
- **Device Class:** Class 1
- **Review Panel:** Clinical Chemistry

## Indications for Use

The ARCHITECT 2nd Generation Testosterone assay is a chemiluminescent microparticle immunoassay (CMIA) for the quantitative determination of testosterone in human serum and plasma. Measurements of testosterone are used in the diagnosis and treatment of disorders involving the male sex hormones (androgens), including primary and secondary hypogonadism, delayed or precocious puberty, impotence in males and, in females, hirsutism (excessive hair) and virilization (masculinization) due to tumors, polycystic ovaries, and adrenogenital syndromes.

## Device Story

ARCHITECT 2nd Generation Testosterone is a chemiluminescent microparticle immunoassay (CMIA) for quantitative testosterone measurement in human serum/plasma. Operation: sample, diluent, and anti-testosterone coated paramagnetic microparticles combine; testosterone binds to particles; acridinium-labeled conjugate added; pre-trigger/trigger solutions added. Chemiluminescent reaction measured as relative light units (RLUs) via ARCHITECT i System optics. Inverse relationship between testosterone concentration and RLUs; concentration interpolated from calibration curve. Used in clinical laboratories; operated by trained personnel. Output provides testosterone levels to assist clinicians in diagnosing/treating androgen-related disorders (hypogonadism, puberty disorders, impotence, hirsutism, virilization). Device also supports calculation of Free Testosterone Index (FTI/FAI) using SHBG results to assess free testosterone status.

## Clinical Evidence

No clinical data. Bench testing only. Reference range study conducted for FTI/FAI calculation using 307 healthy males and 349 healthy females to establish median and 2.5th/97.5th percentile values for total testosterone, SHBG, and FTI/FAI across age-stratified categories.

## Technological Characteristics

Chemiluminescent microparticle immunoassay (CMIA). Reagents: anti-testosterone (sheep, monoclonal) coated paramagnetic microparticles, acridinium-labeled conjugate, assay/specimen diluents. Buffer: BIS-TRIS. Preservative: ProClin 300. Platform: ARCHITECT i2000 SR System. Measurement: Relative Light Units (RLUs). Calculation: FTI/FAI = (Testosterone / SHBG) * 100.

## Regulatory Identification

A testosterone test system is a device intended to measure testosterone (a male sex hormone) in serum, plasma, and urine. Measurement of testosterone are used in the diagnosis and treatment of disorders involving the male sex hormones (androgens), including primary and secondary hypogonadism, delayed or precocious puberty, impotence in males and, in females hirsutism (excessive hair) and virilization (masculinization) due to tumors, polycystic ovaries, and adrenogenital syndromes.

## Predicate Devices

- Abbott ARCHITECT 2nd Generation Testosterone (k120009)

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}

1

# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY ONLY TEMPLATE

A. 510(k) Number:
k152155

B. Purpose for Submission:
Adding sex-hormone binding globulin (SHBG) and free testosterone index (FTI) reference ranges in the labeling

C. Measurand:
Testosterone

D. Type of Test:
Quantitative, Chemiluminescence assay

E. Applicant:
Abbott Laboratories

F. Proprietary and Established Names:
Abbott ARCHITECT 2nd Generation Testosterone

G. Regulatory Information:
1. Regulation section:
21 CFR § 862.1680
2. Classification:
Class I, reserved
3. Product code:
CDZ

{1}

4. Panel:
Clinical Chemistry (75)

H. Intended Use:

1. Intended use(s):
See Indications for use below.

2. Indication(s) for use:
The ARCHITECT 2nd Generation Testosterone assay is a chemiluminescent microparticle immunoassay (CMIA) for the quantitative determination of testosterone in human serum and plasma.

Measurements of testosterone are used in the diagnosis and treatment of disorders involving the male sex hormones (androgens), including primary and secondary hypogonadism, delayed or precocious puberty, impotence in males and in females, hirsutism (excessive hair) and virilization (masculinization) due to tumors, polycystic ovaries, and andrenogential syndromes.

3. Special conditions for use statement(s):
Prescription use only.

4. Special instrument requirements:
ARCHITECT I 2000 SR System

I. Device Description:

Each ARCHITECT 2nd Generation Testosterone Reagent Kit contains one bottle each of: Microparticles, Conjugate, Assay Specific Diluent, and Specimen Diluent.

- Microparticles (1 or 4 bottles) contain 6.6 mL Anti-Testosterone (sheep, monoclonal) coated microparticles in BIS-TRIS buffer [2,2-Bis(hydroxymethyl)-2,2',2''-nitrilotriethanol)] with protein (bovine) stabilizer and ProClin 300 preservative.
- Conjugate (1 or 4 bottles) contains 6.9 mL Testosterone acridinium-labeled conjugate in BIS-TRIS buffer [2, 2-Bis (hydroxymethyl)-2, 2', 2''-nitrilotriethanol)] with surfactant stabilizer and ProClin 300 preservative.

2

{2}

- Assay Specific Diluent (1 or 4 bottles) contains 25.0 mL Testosterone Assay Diluent consisting of phosphate and glycine in citrate buffer and ProClin 300 preservative.
- Specimen Diluent (1 or 4 bottles) contains 12.2 mL Testosterone Specimen Diluent consisting of phosphate-buffered saline (PBS) buffer and ProClin 300 preservative.

The testosterone reagents are identified to the reagents cleared in k120009; there are no changes to the reagents.

## J. Substantial Equivalence Information:

1. Predicate device name(s):

Abbott ARCHITECT 2nd Generation Testosterone

2. Predicate 510(k) number(s):

k120009

3. Comparison with predicate:

|  Similarities  |   |   |
| --- | --- | --- |
|  Item | Candidate Device
Abbott ARCHITECT 2nd Generation Testosterone
(k152155) | Predicate Device
Abbott ARCHITECT 2nd Generation Testosterone
(k120009)  |
|  Intended use | Immunoassay for the in vitro quantitative determination of testosterone in human serum and plasma. | Same  |
|  Platform | ARCHITECT i System
(immunoassay analyzer) | Same  |
|  Methodology | Chemiluminescence (CMIA) | Same  |
|  Specimen type | Serum and plasma | Same  |
|  Measuring range | 4.33 – 1500 mg/dL | Same  |
|  Calibrator levels | 6 Levels
A: 0 ng/dL
B: 2.88 ng/dL
C: 250 ng/dL
D: 500 ng/dL
E: 1,000 ng/dL
F: 2,000 ng/dL | Same  |

{3}

|  Difference  |   |   |
| --- | --- | --- |
|  Item | Candidate Device
Abbott ARCHITECT 2^{nd} Generation Testosterone
(k152155) | Predicate Device
Abbott ARCHITECT 2^{nd} Generation Testosterone
(k120009)  |
|  Expected values listed in the labeling | Testosterone, Sex Hormone Binding Globulin, and Free testosterone Index (FTI) or Free Androgen Index (FAI) | Testosterone  |

K. Standard/Guidance Document Referenced (if applicable):

CLSI EP28-A3c; Defining, Establishing, and Verifying Reference Intervals in the Clinical Laboratory; Approved Guideline-Third Edition

L. Test Principle:

The ARCHITECT 2nd Generation Testosterone immunoassay is based on a competitive test principle with anti-testosterone (sheep, monoclonal) coated paramagnetic microparticles and chemiluminescence detection. In the first step, sample, assay specific diluent and anti-testosterone (sheep, monoclonal) coated paramagnetic microparticles are combined. Testosterone present in the sample binds to the anti-testosterone coated microparticles. After incubation, testosterone acridinium-labeled conjugate is added to the reaction mixture. After further incubation and washing, Pre-Trigger and Trigger solutions are added to the reaction mixture. The resulting chemiluminescent reaction is measured as relative light units (RLUs). The concentration of testosterone is interpolated from a calibration curve established with calibrators of known testosterone concentration.

In order to obtain the free testosterone index (FTI) or Free Androgen index (FAI) result, the ARCHITECT 2nd Generation Testosterone assay and the Abbott ARCHITECT SHBG assays are utilized and are measured at the same time when the sample is tested on the analyzer. The free testosterone index (%FTI) or free androgen index (%FAI) was obtained using the following equation:

$$
\text{FTI or FAI (\%)} = \frac{\text{ARCHITECT 2}^{\text{nd}} \text{ Generation Testosterone Value (nmol/L)}}{\text{ARCHITECT SHBG (nmol/L)}} \times 100
$$

{4}

M. Performance Characteristics (if/when applicable):

1. Analytical performance:
a. Precision/Reproducibility:
Previously established in k120009
b. Linearity/assay reportable range:
Previously established in k120009
c. Traceability, Stability, Expected values (controls, calibrators, or methods):
Previously established in k120009
d. Detection limit:
Previously established in k120009
e. Analytical specificity:
Previously established in k120009
f. Assay cut-off:
Not applicable

2. Comparison studies:
a. Method comparison with predicate device:
Previously established in k120009
b. Matrix comparison:
Previously established in k120009

3. Clinical studies:
a. Clinical Sensitivity:
Not applicable

{5}

b. Clinical specificity:

Not applicable

c. Other clinical supportive data (when a. and b. are not applicable):

Not applicable

4. Clinical cut-off:

Not applicable

5. Expected values/Reference range:

Reference ranges are assay dependent for total testosterone, sex hormone-binding globulin (SHBG) and the free testosterone index (FTI). FTI is calculated from the total testosterone and SHBG results. Therefore, to determine the reference range for the FTI, new reference range studies were performed to obtain expected human SHBG and total testosterone values for apparently healthy males (n = 307) and females (n = 349) using the ARCHITECT 2nd Generation Testosterone assay and the Abbott ARCHITECT SHBG assays. The free testosterone index (%FTI) or free androgen index (%FAI) was obtained using the following equation:

$$
\text{FTI or FAI} (\%) = \frac{\text{ARCHITECT 2}^{\text{nd}} \text{Generation Testosterone Value (nmol/L)}}{\text{ARCHITECT SHBG (nmol/L)}} \times 100
$$

The expected values for total testosterone, SHBG, and FTI are presented in the tables below:

|  Total Testosterone (nmol/L and ng/dL)  |   |   |   |   |
| --- | --- | --- | --- | --- |
|  Category | n | median | 2.5^{th} Percentile | 97.5^{th} Percentile  |
|  Males
(21-49 years of age) | 163 | 15.33 nmol/L | 8.76 nmol/L | 27.85 nmol/L  |
|   |   |  442.07 ng/dL | 252.73 ng/dL | 803.24 ng/dL  |
|  Males
(≥50 years of age) | 144 | 14.42 nmol/L | 8.58 nmol/L | 23.37 nmol/L  |
|   |   |  415.85 ng/dL | 247.50 ng/dL | 674.13 ng/dL  |
|  Females
(Premenopausal, 21-49 years of age) | 174 | 1.05 nmol/L | 0.52 nmol/L | 1.72 nmol/L  |
|   |   |  30.43 ng/dL | 14.92 ng/dL | 49.56 ng/dL  |
|  Females
(Postmenopausal, ≥50 years of age) | 175 | 0.76 nmol/L | 0.46 nmol/L | 1.18 nmol/L  |
|   |   |  21.83 ng/dL | 13.34 ng/dL | 33.90 ng/dL  |

{6}

|  SHBG nmol/L  |   |   |   |   |
| --- | --- | --- | --- | --- |
|  Category | n | median | 2.5thPercentile | 97.5thPercentile  |
|  Males (21-49 years of age) | 163 | 31.1 | 16.2 | 68.5  |
|  Males (≥50 years of age) | 144 | 35.3 | 13.7 | 69.9  |
|  Females (Premenopausal, 21-49 years of age) | 174 | 48.6 | 14.7 | 122.5  |
|  Females (Postmenopausal, ≥50 years of age) | 175 | 49.9 | 16.7 | 124.4  |
|  Free Testosterone Index (FTI)/Free Androgen Index (FAI) (%)  |   |   |   |   |
| --- | --- | --- | --- | --- |
|  Category | n | median | 2.5th Percentile | 97.5th Percentile  |
|  Males (21-49 years of age) | 163 | 46.6 | 24.5 | 113.3  |
|  Males (≥50 years of age) | 144 | 40.7 | 19.3 | 118.4  |
|  Females (Premenopausal, 21-49 years of age) | 174 | 2.0 | 0.7 | 8.7  |
|  Females (Postmenopausal, ≥50 years of age) | 175 | 1.5 | 0.5 | 4.7  |

The sponsor recommended that each laboratory establish its own reference range that is appropriate for the laboratory patient population (i.e., a normal range that reflects the type of specimen and demographic variables such as age and sex, as applicable).

## N. Proposed Labeling:

The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10.

## O. Conclusion:

The submitted information in this premarket notification is complete and supports a substantial equivalence decision.

---

**Source:** [https://fda.innolitics.com/submissions/CH/subpart-b%E2%80%94clinical-chemistry-test-systems/CDZ/K152155](https://fda.innolitics.com/submissions/CH/subpart-b%E2%80%94clinical-chemistry-test-systems/CDZ/K152155)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
