UREA

K981918 · Abbott Laboratories · CDQ · Jul 28, 1998 · Clinical Chemistry

Device Facts

Record IDK981918
Device NameUREA
ApplicantAbbott Laboratories
Product CodeCDQ · Clinical Chemistry
Decision DateJul 28, 1998
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 862.1770
Device ClassClass 2

Indications for Use

The Urea Nitrogen assay is used for the quantitation of urea nitrogen in human serum, plasma, or urine. Measurements obtained by this device are used in diagnosis and treatment of certain renal and metabolic diseases.

Device Story

In vitro diagnostic assay for quantitative determination of urea nitrogen in human serum, plasma, or urine. Sample urea hydrolyzed by urease to ammonia and carbon dioxide; ammonia and a-ketoglutarate converted to glutamate and water via glutamate dehydrogenase (GLDH) with concurrent oxidation of NADH to NAD. Initial rate of decrease in absorbance at 340/380 nm proportional to urea concentration. Used in clinical laboratory settings; operated by laboratory technicians. Output provided as concentration values (mg/dL) to healthcare providers for assessment of renal and metabolic function.

Clinical Evidence

Method comparison studies performed on the AEROSET System against the predicate (Hitachi 717). Serum correlation: r=0.9974, slope=1.038, intercept=0.904 mg/dL. Urine correlation: r=0.9919, slope=0.983, intercept=9.993 mg/dL. Precision studies: Serum total %CV 1.8-2.0%; Urine total %CV 3.1-3.8%. Linearity: up to 234.4 mg/dL (serum) and 1,991.1 mg/dL (urine). Sensitivity (limit of quantitation): 0.9 mg/dL.

Technological Characteristics

In vitro diagnostic clinical chemistry assay. Reagents: Urease and glutamate dehydrogenase (GLDH). Detection: Spectrophotometric measurement of NADH oxidation at 340/380 nm. Linear range: 0.9 to 1,991.1 mg/dL depending on sample type. Designed for use on automated clinical chemistry analyzers.

Indications for Use

Indicated for the quantitative determination of urea nitrogen in human serum, plasma, or urine to aid in the diagnosis and treatment of renal and metabolic diseases.

Regulatory Classification

Identification

A urea nitrogen test system is a device intended to measure urea nitrogen (an end-product of nitrogen metabolism) in whole blood, serum, plasma, and urine. Measurements obtained by this device are used in the diagnosis and treatment of certain renal and metabolic diseases.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ # 510(k) Summary K981918 Submitter's Name/Address Abbott Laboratories 1920 Hurd Drive Irving, Texas 75038 Contact Person Mark Littlefield Section Manager MS 1-8 Regulatory Affairs (972) 518-6062 Fax (972) 753-3367 | Date of Preparation of this Summary: | May 29, 1998 | |--------------------------------------------------|----------------| | Device Trade or Proprietary Name: | Urea | | Device Common/Usual Name or Classification Name: | Urea Nitrogen | | Classification Number/Class: | 75CDQ/Class II | . • This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92. The assigned 510(k) number is: _______________________________________________________________________________________________________________________________________________ ## Test Description: Urea Nitrogen is an in vitro diagnostic assay for the quantitative determination of urea nitrogen in human serum, plasma, or urine. The Urea Nitrogen assay is a clinical chemistry assay in which the urea in the sample is hydrolyzed by urease to ammonia (NH3) and carbon dioxide (CO2). A second reaction, catalyzed by glutamate dehydrogenase (GLDH), converts ammonia and a-ketoglutarate to glutamate and water with the concurrent oxidation of reduced nicotinamide adenine dinucleotide (NADH) to nicotinamide adenine dinucleotide (NAD). Two moles of NADH are oxidized for each mole of urea present. The initial rate of decrease in absorbance at 340/380 nm is proportional to the urea concentration in the sample. Urea Nitrogen 510(k) May 29, 1998 BUN.IWD Section II Page 1 {1}------------------------------------------------ ### Substantial Equivalence: The Urea Nitrogen assay is substantially equivalent to the following device: - Boehringer Mannheim® Urea Nitrogen assay (K771923) on the Hitachi® 717 . Analyzer. Both assays yield similar Performance Characteristics. Similarities: - Both assays are in vitro clinical chemistry methods. . - Both assays can be used for the quantitative determination of urea nitrogen. . - Both assays vield similar clinical results. . #### Differences: - There is a minor difference between the assay range. . #### Intended Use: The Urea Nitrogen assay is used for the quantitation of urea nitrogen in human serum, plasma, or urine. #### Performance Characteristics: Comparative performance studies were conducted using the AEROSET™ System. The Urea Nitrogen assay method comparison vielded acceptable correlation with the the Boehringer Mannheim Urea Nitrogen assay on the Hitachi 717 Analyzer for the serum and urine applications. For the serum application, the correlation coefficient = 0.9974, slope = 1.038, and Y-intercept = 0.904 mg/dL. For the urine application, the correlation coefficient = 0.9919, slope = 0.983, and Y-intercept = 9.993 mg/dL. Precision studies were conducted using the Urea Nitrogen assay. Within-run, between-run, and between-day studies were performed using two levels of control material. For the serum application, the total %CV for Level 1/Panel 101 is 1.8% and Level 2/Panel 102 is 2.0%. For the urine application, the total %CV for Level 1/Panel 201 control is 3.8% and Level 2/Panel 202 is 3.1%. The Urea Nitrogen assay is linear up to 234.4 mg/dL for the serum application, and 1,991.1 mg/dL for the Section II Page 2 # 0000012 {2}------------------------------------------------ urine application. The limit of quantititation (sensitivity) of the Urea Nitrogen assay is 0.9 mg/dL. These data demonstrate that the performance of the Urea Nitrogen assay is substantially equivalent to the performance of the Boehringer Mannheim Urea Nitrogen assay on the Hitachi 717 Analyzer. # Conclusion: The Urea Nitrogen assay is substantially equivalent to the Boehringer Mannheim Urea Nitrogen assay on the Hitachi 717 Analyzer as demonstrated by results obtained in the studies. Urea Nitrogen 510(k) May 29, 1998 BUN JWD Section II Page 3 Image /page/2/Picture/5 description: The image shows a sequence of numbers, specifically '0000013'. The numbers are printed in a bold, sans-serif font. The numbers are black against a white background. {3}------------------------------------------------ Image /page/3/Picture/2 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features the department's emblem, which consists of a stylized human figure with three wavy lines extending from its head. The emblem is surrounded by the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" in a circular arrangement. JUL 28 1998 Food and Drug Administration 2098 Gaither Road Rockville MD 20850 Mark Littlefield Section Manager, Regulatory Affairs Abbott Laboratories 1920 Hurd Drive Irving, Texas 75038 K981918 Re : Urea Requlatory Class : II Product Code: CDO Dated: May 29, 1998 Received: June 1, 1998 Dear Mr. Littlefield: We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Druq, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions The general controls provisions of the Act of the Act. include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major requlations affecting your device can be found in the Code of Federal Requlations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical General regulation (21 CFR Part 820) and that, Devices: through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory In addition, FDA may publish further announcements action. concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations. {4}------------------------------------------------ Page 2 Under the Clinical Laboratory Improvement Amendments of 1988 (CLIA-88), this device may require a CLIA complexity categorization. To determine if it does, you should contact the Centers for Disease Control and Prevention (CDC) at (770) 488-7655. This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4588. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.qov/cdrh/dsmamain.html". Sincerely yours, Steven Litman Steven I. Gutman, M.D., M.B.A. Director Division of Clinical Laboratory Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {5}------------------------------------------------ 510(k) Number (if known): ____________________________________________________________________________________________________________________________________________________ Device Name: Urea Nitrogen ________________________________________________________________________________________________________________________________________________________________ Indications For Use: The Urea Nitrogen assay is used for the quantitation of urea nitrogen in human serum, plasma, or urine. Measurements obtained by this device are used in diagnosis and treatment of certain renal and metabolic diseases. (Division Sign-Off) Division of Clinical Laboratory Devices 510(k) Number 4981918 (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Congurrence of CDRH, Office of Device Evaluation (ODE) Prescription Use V Over-The-Counter Use _________________________________________________________________________________________________________________________________________________________ OR (Per 21 CFR 801.109) (Optional Format 1-2-96) 000000000
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...