← Product Code [CDD](/submissions/CH/subpart-b%E2%80%94clinical-chemistry-test-systems/CDD) · K993571

# BAYER DIAGNOSTICS ACS:180 VB12, BAYER DIAGNOSTICS ADVIA CENTAUE VB12 (K993571)

_Bayer Corp. · CDD · Dec 16, 1999 · Clinical Chemistry · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/CH/subpart-b%E2%80%94clinical-chemistry-test-systems/CDD/K993571

## Device Facts

- **Applicant:** Bayer Corp.
- **Product Code:** [CDD](/submissions/CH/subpart-b%E2%80%94clinical-chemistry-test-systems/CDD.md)
- **Decision Date:** Dec 16, 1999
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 862.1810
- **Device Class:** Class 2
- **Review Panel:** Clinical Chemistry

## Indications for Use

In vitro diagnostic use in the quantitative determination of vitamin B12 in serum. heparinized plasma, or EDTA plasma using the ADVIA Centaur System using the ACS:180® Automated Chemiluminescence Systems. In vitro diagnostic use in the quantitative determination of vitamin B12 in serum, heparinized plasma, or EDTA plasma using the ADVIA® Centaur System.

## Device Story

Competitive immunoassay for vitamin B12 quantification in serum or plasma; utilizes acridinium ester-labeled vitamin B12 and purified intrinsic factor covalently coupled to paramagnetic particles; releasing agent (sodium hydroxide) and DTT liberate vitamin B12 from endogenous binding proteins; cobinamide prevents rebinding. Performed on ACS:180 or ADVIA Centaur automated chemiluminescence systems in clinical laboratory settings by trained personnel. System measures light emission proportional to vitamin B12 concentration; results assist clinicians in assessing vitamin B12 status and diagnosing deficiency.

## Clinical Evidence

Bench testing only. Performance evaluated using 298 serum samples (253 healthy, 45 deficient). Reference range established as 211–911 pg/mL. Sensitivity: minimum detectable concentration 30 pg/mL; reportable range up to 2000 pg/mL. Precision studies (n=231 for ACS:180; n=72 for ADVIA Centaur) showed total CVs ranging from 3.24% to 10.4%. Method comparison between serum and plasma yielded correlation coefficients (r) of 0.96 for both systems.

## Technological Characteristics

Competitive chemiluminescent immunoassay. Reagents: acridinium ester-labeled vitamin B12, intrinsic factor-coupled paramagnetic particles, sodium hydroxide, DTT, and cobinamide. Automated system platform. Measures light emission. No specific materials standards or software architecture details provided.

## Regulatory Identification

A vitamin B12 test system is a device intended to measure vitamin B12 in serum, plasma, and urine. Measurements obtained by this device are used in the diagnosis and treatment of anemias of gastrointestinal malabsorption.

## Predicate Devices

- ACS:180 and ADVIA Centaur VB12 ([K910343](/device/K910343.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows a handwritten alphanumeric string. The string is "K993571". The characters are written in a simple, slightly slanted style.

# Summary of Safety and Effectiveness

As required by 21 CFR 807.92, the following 510(k) Summary is provided:

## 1. Submitters Information

| Contact person:                 | Thomas F. Flynn<br>Manager of Regulatory Affairs                       |
|---------------------------------|------------------------------------------------------------------------|
| Address:                        | Bayer Diagnostics Corporation<br>63 North Street<br>Medfield, MA 02052 |
| Phone:                          | 508 359-3877                                                           |
| Date Summary Prepared:          | October 18, 1999                                                       |
| 2. Device Information           |                                                                        |
| Proprietary Name:               | Bayer Diagnostics ACS:180 VB12<br>Bayer Diagnostics ADVIA Centaur VB12 |
| Common Name:                    | Immunoassay for vitamin B12                                            |
| 3. Predicate Device Information |                                                                        |
| Name:                           | Current legally marketed device (ACS:180 and ADVIA<br>Centaur VB12)    |

Manufacturer:

Bayer Diagnostics Corporation

## 4. Device Description

The ACS:180 VB12 and ADVIA Centaur assays are competitive assays.

Vitamin B12, or cyanocobalamin, is a complex corrinoid compound containing four pyrrole rings that surround a single cobalt atom. Humans obtain vitamin B12 exclusively from animal dietary sources, such as meat, eggs, and milk. Vitamin B12 requires intrinsic factor, a protein secreted by the parietal cells in the gastric mucosa, for absorption. Vitamin B12 and intrinsic factor form a complex that attaches to receptors in the ileal mucosa, where proteins known as trans-cobalamins transport the vitamin B12 from the mucosal cells to the blood and tissues. Most vitamin B12 is stored in the liver as well as in the bone marrow and other tissues,

{1}------------------------------------------------

#### 5. Statement of Intended Use

### ACS180 VB12:

In vitro diagnostic use in the quantitative determination of vitamin B12 in serum. heparinized plasma, or EDTA plasma using the ADVIA Centaur System using the ACS:180® Automated Chemiluminescence Systems.

#### ADVIA Centaur:

In vitro diagnostic use in the quantitative determination of vitamin B12 in serum, heparinized plasma, or EDTA plasma using the ADVIA® Centaur System.

### 6. Summary of Technological Characteristics

The ACS:180 VB12 and ADVIA Centaur VB12 assays are competitive assays in which vitamin B12 from the patient sample competes with vitamin B12 labeled with acridinium ester in the Lite Reagent, for a limited amount of purified intrinsic factor, which is covalently coupled to paramagnetic particles in the Solid Phase. The assay uses Releasing Agent (sodium hydroxide) and DTT to release the vitamin B12 from the endogenous binding proteins in the sample and cobinamide to prevent rebinding after the Solid Phase is added to the sample.

### 7. Performance Characteristics

### Expected Results:

To determine the reference range for the ACS:180 and ADVIA Centaur VB12 assays, data was obtained on 298 serum samples, including 253 samples from apparently healthy individuals and 45 samples from physician-diagnosed vitamin B12 deficient patients. Ninety-five percent of the serumACS:180 VB12 values for the apparently healthy individuals fell in the range of 211 to 911 pg/mL (156 to 672 pmol/L).

| Category  | Median<br>(pg/mL) | Range<br>(pg/mL) | Median<br>(pmol/L) | Range<br>(pmol/L) |
|-----------|-------------------|------------------|--------------------|-------------------|
| Normal    | 382               | 211–911          | 282                | 156-672           |
| Deficient | 160               | 32–246           | 118                | 24-181            |

## Sensitivity and Assay Reportable Range:

The ACS:180 and ADVIA Centaur VB12 assays measures vitamin B12 concentrations up to 2000 pg/mL (1476 pmol/L) with a minimum detectable concentration of 30 pg/mL (22 pmol/L).

{2}------------------------------------------------

## Serum Versus Plasma Method Comparison:

### ACS 180:

A comparison of values for 213 pairs of serum and plasma specimens in the range of 160 to 1000 pg/mL (118 to 738 pmol/L) yielded the following regression equation: plasma VB12 = 1.04 (serum VB12) + 9 pg/mL, r = 0.96

## ADVIA Centaur:

A comparison of values for 207 pairs of serum and plasma specimens in the range of 147 to 1033 pg/mL (108 to 762 pmol/L) yielded the following regression equation: plasma VB12 = 1.03 (serum VB12) + 12 pg/mL, r = 0.96

### Precision:

## ACS180:

Four samples were assayed 3 times in 7 assays, on each of 11 systems (n=231 for each sample), over a period of 5 days. The following results were obtained:

| Mean    | Mean     | Within-run | Total |
|---------|----------|------------|-------|
| (pg/mL) | (pmol/L) | % CV       | % CV  |
| 1147    | 846      | 2.68       | 3.44  |
| 661     | 488      | 2.70       | 3.24  |
| 204     | 151      | 3.73       | 4.79  |
| 169     | 125      | 5.33       | 6.28  |

## ADVIA Centaur:

Four samples were assayed 6 times, in each of 12 runs, on 6 systems, (n = 72 for each sample), over a period of 3 days. The following results were obtained:

| Mean (pg/mL) | Mean<br>(pmol/L) | Within-run %<br>CV | Run-to-run %<br>CV | Total % CV |
|--------------|------------------|--------------------|--------------------|------------|
| 178.76       | 131.89           | 5.0                | 9.2                | 10.4       |
| 207.22       | 152.89           | 4.0                | 4.4                | 5.9        |
| 608.83       | 449.19           | 2.7                | 2.7                | 3.8        |
| 1343.87      | 991.51           | 2.4                | 3.0                | 3.9        |

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular border with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the top half of the circle. Inside the circle is a stylized image of an eagle with three wing-like shapes extending upwards and to the right.

Food and Drug Administration 2098 Gaither Road Rockville MD 20850

DEC I 6 1999

Mr. Thomas F. Flynn Manager of Regulatory Affairs Bayer Corporation Business Group Diagnostics 63 North Street Medfield, Massachusetts 02052-1688

> K993571 Trade Name: Chiron Diagnostics ACS:180 VB12 Assay Regulatory Class: II Product Code: CDD Dated: October 18, 1999 Received: October 21, 1999

Dear Mr. Flynn:

Re:

We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration,

If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.

{4}------------------------------------------------

#### Page 2

Under the Clinical Laboratory Improvement Amendments of 1988 (CLIA-88), this device may require a CLIA complexity categorization. To determine if it does, you should contact the Centers for Disease Control and Prevention (CDC) at (770) 488-7655.

This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4588. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification"(21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597, or at its internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html".

Sincerely yours,

Steven Putman

Steven I. Gutman, M.D, M.B.A. Director Division of Clinical Laboratory Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

Page ⊥ of ⊥

4993571 510(k) Number (if known): ____________________________________________________________________________________________________________________________________________________

Device Name: Chiron Diagnostics ACS:180 VB12 Assay

Indications for Use:

In vitro diagnostic use in the quantitative determination of vitamin B12 in serum, heparinized plasma, or EDTA plasma using the ADVIA Centaur System.

Qeen Cooper

**Division Sign-Off**

ivision of Clinical Laboratory Devices 0(k) Number _ K 99

(PLEASE DO NOT WRITE BELOW THIS LINE--CONTINUE ON ANOTHER PAGE, IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

| Prescription Use     | <div> <span style="text-decoration: overline;">✓</span> </div> |
|----------------------|----------------------------------------------------------------|
| (Per 21 CFR 801.109) |                                                                |

OR

| 510(k) Number (if known): | K910343 K993571 |
|---------------------------|-----------------|
|---------------------------|-----------------|

Over-The-Counter Use (Optional Format 1-2-96)

---

**Source:** [https://fda.innolitics.com/submissions/CH/subpart-b%E2%80%94clinical-chemistry-test-systems/CDD/K993571](https://fda.innolitics.com/submissions/CH/subpart-b%E2%80%94clinical-chemistry-test-systems/CDD/K993571)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
