AI/ML, Software as a Medical Device, Therapeutic, Pediatric
Indications for Use
EndoTool SubQ is a software application for use by trained healthcare professionals to calculate and recommend an individual patient's next dose of insulin to be administered subcutaneously to manage blood glucose levels in patients with Diabetes Mellitus in both adult and pediatric patients (age 2 years and above and 12 kg or more). The software is designed to recommend the insulin dose(s) (and when indicated a carbohydrate dose) based on the prescribing healthcare provider's nutritional regimen, insulin regimen, target glucose range, and patient specific characteristics. The EndoTool SubQ is not a substitute for clinical reasoning, but is an aid for trained healthcare professionals to manage patients. The System is based on obtained glucose readings and clinical data entered by the medical staff. Final dose decisions for a patient must be made only after consideration of the full clinical status of the patient. No medical decision should be made based solely on the recommendations provided by this software program.
Device Story
EndoTool SubQ is a hospital-based software application for glycemic management. It accepts inputs including physician-prescribed nutritional regimens, insulin regimens, target glucose ranges, patient characteristics, and bedside-entered blood glucose readings and food intake. The system utilizes proprietary feedback mathematics and a nonlinear dosing equation to calculate and recommend subcutaneous insulin doses and carbohydrate doses for hypoglycemia. It is operated by bedside caregivers (e.g., nurses) via an internal hospital website. The output is presented to the clinician as a dose recommendation; the clinician must review the patient's full clinical status before making final administration decisions. The device aims to achieve sustained, patient-specific glycemic control, assisting healthcare professionals in managing insulin therapy.
Clinical Evidence
Bench testing only. Evidence includes requirements-based testing for all functionality and risk-related requirements, integration testing for database data flow, and automated execution of algorithm test cases. Off-the-shelf (OTS) software components were verified for intended function.
Technological Characteristics
Software-based glucose management system deployed via internal hospital server/website. Uses feedback mathematics and nonlinear dosing equations for insulin/carbohydrate calculation. Features include configurable therapy goals, audit trails, and visual/audible alerts. Operates on patient-specific data inputs to optimize dosing models over time.
Indications for Use
Indicated for adult and pediatric patients (age 2+ years, weight 12kg+) with Diabetes Mellitus to manage blood glucose levels via subcutaneous insulin dose recommendations.
Regulatory Classification
Identification
A predictive pulmonary-function value calculator is a device used to calculate normal pulmonary-function values based on empirical equations.
Predicate Devices
EndoTool IV Glucose Management System (EGMS8) (K132547)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the seal of the U.S. Department of Health and Human Services. The seal features a stylized design of three human profiles facing right, stacked on top of each other. The profiles are rendered in a dark color, creating a sense of depth and unity. The seal is encircled by the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" in a circular arrangement.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
April 24, 2015
Monarch Medical Technologies Mr. Al Pacheco Certified Compliance Solutions 11665 Avena Place, Suite #203 San Diego, CA 92128
Re: K142918
Trade/Device Name: EndoTool SubQTM Regulation Number: 21 CFR 868.1890 Regulation Name: Predictive pulmonary-function value calculator Regulatory Class: II Product Code: NDC Dated: March 23, 2015 Received: March 24, 2015
### Dear Mr. Pacheco:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Page 2 - Mr. Al Pacheco
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Tina
Kiang -S
for Erin I. Keith, M.S. Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K142918
Device Name EndoTool SubQ
#### Indications for Use (Describe)
EndoTool SubQ is a software application for use by trained healthcare professionals to calculate and recommend an individual patient's next dose of insulin to be administered subcutaneously to manage blood glucose levels in patients with Diabetes Mellitus in both adult and pediatric patients (age 2 years and above and 12 kg or more). The software is designed to recommend the insulin dose(s) (and when indicated a carbohydrate dose) based on the prescribing healthcare provider's nutritional regimen, insulin regimen, target glucose range, and patient specific characteristics.
The EndoTool SubQ is not a substitute for clinical reasoning, but is an aid for trained healthcare professionals to manage patients. The System is based on obtained glucose readings and clinical data entered by the medical staff. Final dose decisions for a patient must be made only after consideration of the full clinical status of the patient. No medical decision should be made based solely on the recommendations provided by this software program.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
| <span style="font-size:12px">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) |
|------------------------------------------------------------------------------------|
| <span style="font-size:12px">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) |
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
## 510(k) Summary
| Submitter: | Monarch Medical Technologies |
|------------------------|--------------------------------------------------------------------------------------|
| Address: | 2137 South Blvd, Suite 300<br>Charlotte, NC 28217<br>Website: www.monarchmedtech.com |
| Phone number: | (855) 363-7475 |
| Fax number: | (704) 746-3913 |
| Contact person: | Al Pacheco |
| Phone number: | (858) 675-8200 |
| Fax number: | (858) 675-8201 |
| Date prepared: | March 23, 2015 |
| Trade name: | EndoTool SubQ™ |
| Common name: | EndoTool SubQ |
| Product Code, Primary: | NDC, Calculator, Drug Dose |
| Regulation: | 21 CFR 868.1890, Predictive pulmonary-function value calculator |
Substantial equivalence claimed to: EndoTool IV Glucose Management System (EGMS8) - K132547 Glytec LLC, Glucommander G+ Enterprise System - K113853
#### Description:
Monarch Medical Technology's EndoTool SubQ Glucose Management System is a glucose management software solution for healthcare institutions. This system uses proprietary control technology to provide patient-specific glycemic control across a broad population of adult and pediatric patients using basal/bolus insulin therapy. This system is designed to correctly dose subcutaneous insulin (and amount of carbohydrates if and when indicated to treat hypoglycemia) to achieve patient-specific, sustained control with different nutritional and insulin regimens, selected by the patient's physician responsible for glycemic control.
EndoTool SubQ Glucose Management System is designed to be used following a physician order with physician set optional diet, insulin regimen, basal/bolus distribution, initial total daily dose of insulin, and glucose target range. The primary user of the EndoTool SubQ Glucose Management System is the bedside caregiver (e.g. nurse) who will use the system to enter clinical data (e.g. food intake and scheduled blood glucose readings). With confirmation of previous data entered, the system makes the next dose calculation of subcutaneous insulin and the next time for a scheduled glucose determination.
{4}------------------------------------------------
#### Indications for Use:
EndoTool SubQ is a software application for use by trained healthcare professionals to calculate and recommend an individual patient's next dose of insulin to be administered subcutaneously to manage blood glucose levels in patients with Diabetes Mellitus in both and pediatric patients (age 2 years and above and 12 kg or more). The software is designed to recommend the insulin dose(s) (and when indicated a carbohydrate dose) based on the prescribing healthcare provider's nutritional regimen, insulin regimen, target glucose range, and patient specific characteristics.
EndoTool SubQ is not a substitute for clinical reasoning, but is an aid for trained healthcare professionals to manage patients. The System is based on obtained glucose readings and clinical data entered by the medical staff. Final dose decisions for a patient must be made only after consideration of the full clinical status of the patient. No medical decision should be made based solely on the recommendations provided by this software program.
### Technological Characteristics
The EndoTool SubQ Glucose Management System uses feedback mathematics to individualize insulin dosing by calculating limited proportional incremental changes in the insulin dosing model (total daily dose and the physician set basal/bolus distribution) or carbohydrate recommendations based on a patient's previous blood glucose readings in response to prior insulin doses. These calculations are repeated by the software when new data is entered into the system, constantly seeking the optimal, patient specific insulin dose for the targets set by the physician.
The software provides three main functions: set-up or initial mathematical model of a patient, physician confirmed or altered daily update of patient's mathematical model through time as new blood glucose values and carbohydrate intake is entered and calculated doses of insulin or carbohydrates based on the physician confirmed total daily dose, basal/bolus distribution, and insulin regimen.
The system is installed on a server and deployed via an internal website on the hospital's infrastructure. Insulin and carbohydrate therapy are managed using blood glucose measurements, available patient information and a set of algorithms that includes a nonlinear dosing equation that is individualized and optimized through time based on blood glucose response to previous doses administered. Therapy goals and limits are configurable by the patient's physician.
{5}------------------------------------------------
#### Substantial Equivalence
The EndoTool SubQ Glucose Management System is substantially equivalent to other FDA cleared, marketed drug dosage calculators. Specifically, the EndoTool SubQ Glucose Management System is substantially equivalent to the Glytec Glucommander G+ Enterprise System (K113853) and EndoTool IV Glucose Management System (K132547). Minor differences in the technological characteristics between these devices do not raise new questions of safety or efficacy.
| Functionality | This<br>device | EndoTool IV<br>(Predicate 1) | Glucommander<br>(Predicate 2) |
|-------------------------------------------------------------------------------------|----------------|------------------------------|-------------------------------|
| Patient selection and initiation | Yes | Yes | Yes |
| Patient setup (demographics, clinical, protocol) -<br>directed via physician orders | Yes | Yes | Yes |
| Carbohydrate intake (meals and continuous nutrition) | Yes | Yes | Yes |
| Blood glucose entry | Yes | Yes | Yes |
| Display of advisories/guardrails | Yes | Yes | Yes |
| Display of calculated insulin dose or carbohydrate dose | Yes | Yes | Yes |
| Reports | Yes | Yes | Yes |
| User Security | Yes | Yes | Yes |
| Visual and audible alerts | Yes | Yes | Yes |
| Audit trails | Yes | Yes | Unknown |
| Data storage | Yes | Yes | Yes |
#### Functionality Comparison Between the Subject Device and Predicates
#### Substantial Equivalence Conclusion
The EndoTool SubQ Glucose Management System has the same intended use and similar indications, technological characteristics, principles of operation as the previously cleared Glytec Glucommander G+ Enterprise System and EndoTool IV Glucose Management System, thus establishing substantial equivalence.
#### Test Summary
Testing included the following:
- Requirements-based testing for all functionality. 1)
- 2) Requirements-based testing for all risk-related requirements.
- 3) Integration testing to ensure that data flows correctly into and out of the database.
- Automated algorithm test case execution. 4)
- 5) Off The Shelf (OTS) software embedded in the application was included in the technical verification protocols. Each OTS component was tested to ensure that it functioned as intended.
#### Test Summary Conclusion:
The performance of the EndoTool SubQ Glucose Management System is substantially equivalent to that of the Glytec Glucommander G+ Enterprise System (K113853) and EndoTool IV Glucose Management System (K132547) and raises no new safety or effectiveness issues and performs as well as the predicate.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.