For the quantitative measurement of intact parathyroid hormone (iPTH) in human serum and plasma (EDTA or heparin) using the VITROS ECi/ECiQ Immunodiagnostic Systems and the VITROS 5600 Integrated System. Intact PTH is indicated to aid in the differential diagnosis of hyperparathyroidism, hypoparathyroidism, or hypercalcemia of malignancy and can be used intraoperatively.
Device Story
Quantitative immunoassay for intact parathyroid hormone (iPTH) in human serum/plasma; utilizes VITROS ECi/ECiQ or 5600 systems. Principle: immunometric immunoassay; biotinylated goat polyclonal anti-PTH (39-84) and HRP-labeled goat polyclonal anti-PTH (1-34) antibodies bind PTH; complex captured on streptavidin-coated wells. HRP conjugate catalyzes luminogenic substrate oxidation; light signal measured is proportional to PTH concentration. Used in clinical laboratories by technicians; results aid clinicians in diagnosing parathyroid disorders and monitoring intraoperative tumor removal. Benefits include rapid, quantitative assessment of PTH levels to guide surgical and diagnostic decisions.
Clinical Evidence
Bench testing only. Precision evaluated per CLSI EP5-A2 (CV <10%). Linearity (3.4-5000 pg/mL) per CLSI EP6-A. LoB/LoD/LoQ determined per CLSI EP17-A (LoQ 3.4 pg/mL). Method comparison (N=456) against predicate showed high correlation (r=0.99). Intraoperative applicability evaluated in 39 patients; 31/39 showed >50% post-operative PTH drop, consistent with reference method.
Technological Characteristics
Solid-phase immunometric immunoassay. Reagents: biotinylated goat polyclonal anti-PTH, HRP-labeled goat polyclonal anti-PTH, streptavidin-coated wells. Energy source: luminescent reaction. Form factor: reagent packs for VITROS ECi/ECiQ and 5600 systems. Connectivity: integrated with VITROS systems. Standards: CLSI EP5-A2, EP6-A, EP7-A2, EP9-A2, EP17-A.
Indications for Use
Indicated for quantitative measurement of intact parathyroid hormone (iPTH) in human serum and plasma to aid in differential diagnosis of hyperparathyroidism, hypoparathyroidism, or hypercalcemia of malignancy; can be used intraoperatively.
Regulatory Classification
Identification
A parathyroid hormone test system is a device intended to measure the levels of parathyroid hormone in serum and plasma. Measurements of parathyroid hormone levels are used in the differential diagnosis of hypercalcemia (abnormally high levels of calcium in the blood) and hypocalcemia (abnormally low levels of calcium in the blood) resulting from disorders of calcium metabolism.
Predicate Devices
Roche Diagnostics Corporation Elecsys Parathyroid Hormone Test System (K070709)
Submission Summary (Full Text)
{0}
1
SUBSTANTIAL EQUIVALENCE DETERMINATION
DECISION SUMMARY
ASSAY ONLY TEMPLATE
A. 510(k) Number:
k101921
B. Purpose for Submission:
New device
C. Measurand:
Intact Parathyroid Hormone (iPTH)
D. Type of Test:
Quantitative, Enzyme-Linked Immunoassay (EIA)
E. Applicant:
Ortho-Clinical Diagnostics
F. Proprietary and Established Names:
1. VITROS Immunodiagnostics Products Intact PTH Reagent Pack
2. VITROS Immunodiagnostics Products Intact PTH Calibrators
3. VITROS Immunodiagnostics Products Range Verifiers
4. VITROS Immunodiagnostic Product Intact PTH Controls
G. Regulatory Information:
| Product Code | Classification | Regulation Section | Panel |
| --- | --- | --- | --- |
| CEW | Class II | 21 CFR 862.1545
Parathyroid Hormone
Test System | Clinical Chemistry (75) |
| JIT | Class II | 21 CFR 862.1150
Calibrator | Clinical Chemistry (75) |
| JJX | Class I, reserved | 21 CFR 862.1660
Quality Control Material | Clinical Chemistry (75) |
H. Intended Use:
1. Intended use(s):
{1}
Refer to indication for use below
2. Indication(s) for use:
For the quantitative measurement of intact parathyroid hormone (iPTH) in human serum and plasma (EDTA or heparin) using the VITROS ECi/ECiQ Immunodiagnostic Systems and the VITROS 5600 Integrated System. Intact PTH is indicated to aid in the differential diagnosis of hyperparathyroidism, hypoparathyroidism, or hypercalcemia of malignancy and can be used intraoperatively.
3. Special conditions for use statement(s):
For in vitro diagnostic use only
4. Special instrument requirements:
VITROS ECi/ECiQ Immunodiagnostic Systems and VITROS 5600 Integrated System were used to conduct performance studies below.
I. Device Description:
1. VITROS Intact PTH Reagent Pack Contents:
- 100 coated wells (streptavidin, binds >3 ng biotin/well)
- 6.2 mL biotinylated antibody reagent (biotin-goat polyclonal anti-PTH, binds >20,000 pg PTH/mL) in buffer with bovine gamma globulin, bovine serum albumin, and antimicrobial agent
- 8.4 mL conjugate reagent (HRP-goat polyclonal anti-PTH, binds >8,000 pg PTH/mL) in buffer with bovine serum albumin and antimicrobial agent
2. VITROS Intact PTH Calibrator Contents:
- 3 sets of VITROS Intact PTH Calibrators 1, 2 and 3 (freeze-dried, synthetic PTH in buffer with bovine serum albumin and antimicrobial agent, reconstitution volume 1.0 mL); nominal values 0, 100 and 1500 pg/mL
- Lot calibration card
- Protocol card
- 24 calibrator bar code labels (8 for each calibrator)
3. VITROS Intact PTH Range Verifier Contents:
2 sets of VITROS Intact PTH Range Verifiers, low and high (freeze-dried, synthetic PTH in buffer with bovine serum and antimicrobial agent, reconstitution volume 1.0 mL); nominal values 0 and 4750 pg/mL
4. Control Pack Contents:
{2}
3 sets of VITROS Intact PTH Controls 1, 2 and 3 (freeze-dried, synthetic PTH in buffer with bovine serum and antimicrobial agent, reconstitution volume 1.0 mL)
# J. Substantial Equivalence Information:
1. Predicate device name(s):
Roche Diagnostics Corporation Elecsys Parathyroid Hormone Test System
2. Predicate 510(k) number(s):
K070709
3. Comparison with predicate:
| Items | VITROS® Intact PTH assay(Candidate Device) | Roche Elecsys Intact PTH assay(Predicate Device) |
| --- | --- | --- |
| Similarity | | |
| Intended Use | Same | It is intended for the in vitro quantitative determination of intact parathyroid hormone in human serum and plasma for the differential diagnosis of hypercalcemia and hypocalcemia, and can be used intraoperatively. |
| Tracer | Same | Enzyme labeled |
| Precision | Same | Total imprecision <10% |
| Measuring Range | 3.4-5000 pg/mL | 1.20-5000 pg/mL |
| Time to First Result | Same | 18 minutes |
| Sample Type | Same | Serum and plasma (heparin and EDTA) |
| Difference | | |
| Test Principle | Solid phase immunoassay | Sandwich assay |
| Antibody | Goat polyclonal | Mouse monoclonal |
| Sample Volume | 80 μL | 50 μl |
| Hook Effect | None up to 1,218,400 pg/mL | None up to 17,000 pg/mL |
| Normal Range | 7.5-53.5 pg/mL | 16.0-65 pg/mL |
| Calibrator | | |
| Format | Same | Liquid |
| Levels | 3 | 2 |
| Matrix | Buffer with bovine serum and antimicrobial agent | Human serum with antimicrobial agent |
| Range Verifier | | |
{3}
| Format | Same | Lyophilized |
| --- | --- | --- |
| Levels | 2
Low and High (<3.4 & 4750 pg/mL) | 3
Low Medium and High |
| Matrix | Buffer with bovine serum antimicrobial agent | Human serum with antimicrobial agent |
| Control | | |
| Format | Same | Lyophilized |
| Levels | 3 | 3 |
| Values | Approximately 25, 75 and 500 pg/mL | Approximately 20, 60 and 700 pg/mL |
| Matrix | Buffer with bovine serum and antimicrobial agent | Human serum with antimicrobial agent |
# K. Standard/Guidance Document Referenced (if applicable):
CLSI Guideline EP5-A2
CLSI Guideline EP6-A
CLSI Guideline EP7-A2
CLSI Guideline EP9-A2
CLSI Guideline EP17-A
- Standards Report for C28-A3: Defining, Establishing, and Verifying Reference Intervals in the Clinical Laboratory; Approved Guideline - Third Edition.
- Guidance for Industry and FDA Staff - Use of Symbols on Labels and in Labeling of In Vitro Diagnostic Devices Intended for Professional Use
- Points to Consider for Collection of Data in Support of In-Vitro Device Submissions for 510(k) Clearance
- Points to Consider for Review of Calibration and Quality Control Labeling for In Vitro Diagnostic Devices/Cover Letter dated 3/14/1996
- Guidance for Industry and FDA Staff - Assayed and Unassayed Quality Control Material
# L. Test Principle:
An immunometric immunoassay technique is used, which involves the simultaneous reaction of PTH present in the sample with a biotinylated antibody (goat polyclonal anti-PTH39-84) and a horseradish peroxidase (HRP)-labeled antibody conjugate (goat polyclonal anti-PTH1-34). The antigen-antibody complex is captured by streptavidin on the wells. Unbound materials are removed by washing. The bound HRP conjugate is measured by a luminescent reaction. A reagent containing luminogenic substrates (a luminol derivative and a peracid salt) and an electron transfer agent is added to the wells. The HRP in the bound conjugate catalyzes the oxidation of the luminol derivative, producing light. The electron transfer agent (a substituted acetanilide) increases the level of light produced and prolongs its emission. The light signals are read by the system. The amount of HRP conjugate bound is directly proportional to the concentration of PTH present in the sample.
# M. Performance Characteristics (if/when applicable):
1. Analytical performance:
{4}
# a. Precision/Reproducibility:
# Study Protocol:
Precision was evaluated following CLSI guideline EP5-A2. Two replicates of each of 3 freeze-dried control samples and 6 patient serum sample pools were tested on 2 separate runs per day for 20 different days. When spiking synthetic PTH into a patient sample pool, the spike solution made up no more than $5\%$ of the total volume of the patient sample being spiked.
# Precision Summary:
Sample description
| Sample ID | Sample Description | iPTH Nominal Value (pg/mL) |
| --- | --- | --- |
| SC-1 | VITROS Control 1 | 30 |
| SC-2 | VITROS Control 2 | 80 |
| SC-3 | VITROS Control 3 | 500 |
| PP-1 | Endogenous Patient Serum Pool 1 | 15 |
| PP-2 | Endogenous Patient Serum Pool 2 | 40 |
| PP-3 | Spiked Patient Serum Pool 3 | 50 |
| PP-4 | Spiked Patient Serum Pool 4 | 100 |
| PP-H | Spiked Patient Serum Pool 5 | 3000 |
| PP-H2 | Spiked Patient Serum Pool 6 | 4000 |
Precision
| System | Units = pg/mL | | | | | | | No. Observ. Days |
| --- | --- | --- | --- | --- | --- | --- | --- | --- |
| | Mean PTH Conc. | Within-run* | | Within-calibration** | | Within-lab*** | | |
| | | SD | CV (%) | SD | CV (%) | SD | CV (%) | |
| ECi/ECiQ system 1 | 25.8 | 0.545 | 2.1 | 1.120 | 4.3 | 1.200 | 4.7 | 80 20 |
| | 75.4 | 1.25 | 1.7 | 3.72 | 4.9 | 3.73 | 5.0 | 80 20 |
| | 499.5 | 8.32 | 1.7 | 17.7 | 3.5 | 18.3 | 3.7 | 80 20 |
| | 12.8 | 0.32 | 2.5 | 0.561 | 4.4 | 0.662 | 5.2 | 80 20 |
| | 42.4 | 3.14 | 7.4 | 3.52 | 8.3 | 4.06 | 9.6 | 80 20 |
| | 45.5 | 0.704 | 1.5 | 1.62 | 3.6 | 2.24 | 4.9 | 80 20 |
| | 105.8 | 1.37 | 1.3 | 3.17 | 3.0 | 3.74 | 3.5 | 80 20 |
| | 2974 | 41.4 | 1.4 | 64.0 | 2.2 | 104 | 3.5 | 80 20 |
| | 3593 | 65.7 | 1.8 | 86.4 | 2.4 | 134 | 3.7 | 80 20 |
| ECi/ECiQ system 2 | 26.3 | 0.616 | 2.3 | 1.160 | 4.4 | 0.991 | 3.8 | 80 20 |
| | 76.7 | 1.39 | 1.8 | 2.42 | 3.2 | 2.31 | 3.0 | 80 20 |
| | 503.5 | 7.17 | 1.4 | 17.0 | 3.4 | 17.2 | 3.4 | 80 20 |
| | 13.6 | 0.357 | 2.6 | 0.611 | 4.5 | 0.553 | 4.1 | 80 20 |
| | 47.2 | 0.745 | 1.6 | 1.27 | 2.7 | 1.26 | 2.7 | 80 20 |
| | 49.4 | 0.773 | 1.6 | 1.68 | 3.4 | 1.71 | 3.5 | 80 20 |
| | 112 | 1.82 | 1.6 | 3.04 | 2.7 | 6.16 | 5.5 | 80 20 |
| | 3246 | 32.9 | 1.0 | 64.8 | 2.0 | 92.4 | 2.8 | 80 20 |
| | 3891 | 56.2 | 1.4 | 101 | 2.6 | 254 | 6.5 | 80 20 |
| 5600 | 27.6 | 0.588 | 2.1 | 1.93 | 7.0 | 2.09 | 7.6 | 80 20 |
| | 78.1 | 1.72 | 2.2 | 4.04 | 5.2 | 4.50 | 5.8 | 80 20 |
| | 505.3 | 8.13 | 1.6 | 18.3 | 3.6 | 20.0 | 4.0 | 80 20 |
{5}
6
| System | Units = pg/mL | | | | | | | No. Observ. | No. Days |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| | Mean PTH Conc. | Within-run* | | Within-calibration** | | Within-lab*** | | | |
| | | SD | CV (%) | SD | CV (%) | SD | CV (%) | | |
| | 14.2 | 0.257 | 1.8 | 0.863 | 6.1 | 0.928 | 6.5 | 80 | 20 |
| | 47.6 | 0.537 | 1.1 | 1.82 | 3.8 | 2.54 | 5.3 | 80 | 20 |
| | 51.7 | 0.745 | 1.4 | 1.55 | 3.0 | 2.57 | 5.0 | 80 | 20 |
| | 116.4 | 1.49 | 1.3 | 3.00 | 2.6 | 5.63 | 4.8 | 80 | 20 |
| | 3155 | 28.8 | 0.9 | 49.8 | 1.6 | 97.5 | 3.1 | 80 | 20 |
| | 3782 | 46.2 | 1.2 | 92.6 | 2.4 | 222 | 5.9 | 80 | 20 |
* Within-run (repeatability). Between Duplicate precision averaged over all runs
** Within-calibration. Total precision with weighted components of within-run, between-run and between-day variation.
*** Within-lab. A measure of the effect of recalibration on total precision, calculated within reagent lot, using data from at least 4 calibrations
## b. Linearity/assay reportable range:
### Study Protocol:
Linearity was evaluated following CLSI guideline EP6-A on 2 VITROS ECi/ECiQ Immunodiagnostic Systems and 1 VITROS 5600 Integrated System. Altered serum pool samples were used for the studies. The low pool was an endogenous serum sample stripped for PTH and had an estimated PTH concentration of 2.0 pg/mL. The high pool was either an endogenous patient pool (estimated PTH concentration 700 pg/mL) or a patient pool spiked with synthetic PTH (estimated PTH concentration 5500 pg/mL). For the whole range study, the spiked high serum pool and the low serum pool were mixed to give 11 additional pools of intermediate concentrations spanning the range of 2.0 - 5500 pg/mL. For the clinical relevant range study, the endogenous patient pool and the low serum pool were mixed to give 11 additional pools of intermediate concentrations spanning the range of 2.0 - 700 pg/mL.
### Result Summary:
Statistical evaluations for both studies using polynomial regression indicate that the data were best fitted by a 3rd order regression with less than 10% bias from the linear fitted results.
### Conclusion:
Based on the linearity results, the sponsor claimed that the assay is linear from 3.4 pg/mL to 5000 pg/mL.
## c. Traceability, Stability, Expected values (controls, calibrators, or methods):
### Traceability:
Calibration of the VITROS Intact PTH test is traceable to in-house reference calibrators, which value have been assigned to correlate to the predicate device, the Roche Elecsys PTH.
{6}
# Stability:
Real-time testing was conducted. The stability study protocol and the acceptance criteria have been reviewed and found to be acceptable. The study results support the stability claims summarized in the below table.
Close-Vial and Open-Vial Stability
| Item | Storage Conditions | | Claimed Stability |
| --- | --- | --- | --- |
| Reagent Packs | Close-Vial | 2-8°C | 26 weeks |
| | Open-Vial | On system | 8 weeks |
| | Open-Vial | 2-8°C | 8 weeks |
| Calibrators | Close-Vial | 2-8°C | 26 weeks |
| | Open-Reconstituted | 2-8°C | 1 day |
| | Open-Reconstituted | -20°C | 8 weeks |
| Range Verifiers | Close-Vial | 2-8°C | 26 weeks |
| | Open-Reconstituted | 2-8°C | 1 day |
| | Open-Reconstituted | -20°C | 5 days |
| Controls | Close-Vial | 2-8°C | 26 weeks |
| | Open-Reconstituted | 2-8°C | 1 day |
| | Open-Reconstituted | -20°C | 4 weeks |
# Calibration Interval:
Stability across a 28 day calibration interval was assessed by calculating the percentage bias of the precision samples on each day from the result obtained for that precision samples from the calibration on the first day. The results support the product claim of a 28-day calibration interval.
# Value Assignment:
- Calibrators: The in-house reference calibrators, which values have been assigned to correlate to the predicate device, were used to generate the master calibration curve. The lot specific calibrator value was determined as the mean of 20 assays per master lot, using 2 VITROS ECi/ECiQ Immunodiagnostic Systems and 1 VITROS 5600 Integrated System.
- Range Verifiers: The acceptance criterion for the Low Range Verifier is $< 3.4 \, \mathrm{pg/mL}$ , for the High Range Verifier is $4000 - 5500 \, \mathrm{pg/mL}$ .
- Controls: The target values are 25, 75 and $500~\mathrm{pg / mL}$ . The lot specific range was assigned as mean $\pm 3$ SD. The Mean and SD was calculated from 32 test results (2 tests per run, 5 or 6 runs per kit lot, using 3 kit lots and at least 2 VITROS systems).
{7}
d. Detection limit:
Study Protocol:
Limit of Blank (LoB), Limit of Detection (LoD) and Limit of Quantitation (LoQ) were determined following CLSI guideline EP17-A on 2 VITROS ECi/ECiQ Immunodiagnostic Systems and 1 VITROS 5600 Integrated System.
For LoB determination, 10 replicates of the zero calibrator buffer were run on 2 occasions per day for 5 days, assessed from three calibrations (days 1, 3 and 5) – giving 100 determinations in total.
For LoD determination, 10 replicates of each of five LoD pools were run on 2 occasions per day for 5 days across three calibrations (days 1, 3 and 5) – giving 100 determinations of each pool in total. The five LoD pools were made by diluting an endogenous patient pool in the zero calibrator buffer used for the Limit of Blank to give estimated iPTH concentrations of 2.5, 3.0, 3.5, 4.0 and 4.5pg/mL.
For LoQ determination, two additional pools with estimated iPTH concentrations of 5.0 and 5.5 pg/mL were included in addition to the 5 pools used in the LoD study. The level of imprecision used to accept the LoQ was within 20%.
Result Summary:
The following detection limit claims were made based on the worst case results:
| LoB | LoD | LoQ |
| --- | --- | --- |
| 1.2 pg/mL | 2.8 pg/mL | 3.4 pg/mL |
e. Analytical specificity:
- Interference
Study Protocol:
The sponsor evaluated the effect of the interfering substances on 2 VITROS ECi/ECiQ Immunodiagnostic Systems and 1 VITROS 5600 Integrated System using a patient serum pool with endogenous PTH between 20 and 80 pg/mL. For each substance, two aliquots of this pool were thawed, one spiked with the test substance up to the maximum level shown below, and the other spiked with the solvent (control). The PTH results (mean of 12 replicates) of the paired pools were compared and % interference was calculated using the following equation:
$$
\% \text{Interference} = \frac{(\text{Mean of test substance pool}) - (\text{Mean of "control" pool})}{\text{Mean of "control" pool}} \times 100
$$
Result Summary:
Based on the sponsor-defined interference limit of ± 10%, the following claims were
{8}
made:
- The below compounds at the indicated concentration do not cause significant interference with the assay.
| Compound | Concentration up to |
| --- | --- |
| Azide (sodium) | 20 mg/dL |
| Bilirubin | 20 mg/dL |
| Biotin | 500 ng/dL |
| Dipyrone | 100 mg/dL |
| Hemoglobin (hemolysate)* | 161 mg/dL |
| Intralipid | 850 mg/dL |
| Triolein | 2500 mg/dL |
- Interference by Hemoglobin at >=250 mg/dL is stated in “Limitations of the Procedure” in the labeling. Additionally, the sponsor added “Do not use hemolyzed specimens.” under section “Specimens Not Recommended”.
- No interference is seen with Rheumatoid Factor (RF concentration up to 4935 IU/mL) or HAMA samples (HAMA concentration up to 1825 ng/mL) or with HAMA spiked endogenous samples (HAMA concentration up to 2000 ng/mL).
- Cross-Reactivity
**Study Protocol:**
The sponsor evaluated cross-reactivity by using the zero calibrator buffer as the test matrix. This calibrator had a PTH concentration below the Limit of Blank. Three determinations of each cross-reactant and corresponding “control” were carried out with each of the two reagent lots using 3 VITROS Immunodiagnostic Systems (2 VITROS ECi/ECiQ Immunodiagnostic Systems and 1 VITROS 5600 Integrated System). % Cross-reactivity was calculated using the following the equation:
$$
\% \text{Cross-reactivity} = \frac{\text{Mean iPTH of the Cross-Reactant pool} \times 100}{\text{Concentration of Cross-Reactant}}
$$
**Result Summary:**
Based on the sponsor-defined % Cross-reactivity limit of 0.01%, none of the tested substances (except PTH 7-84) at the indicated concentration would cross react significantly with the proposed assay.
| Cross-reactant Tested | Concentration | Mean PTH Conc. of Cross-reactant Pool | % Cross-reactivity |
| --- | --- | --- | --- |
| | | Units pg/mL | |
| Bone Specific alkaline phosphatase | 7.5ng/mL | <3.4 | NA* |
| Calcitonin | 10,000 pg/mL | <3.4 | NA* |
| β-Cross laps | 1 ng/mL | <3.4 | NA* |
{9}
| Osteocalcin | 50 ng/mL | <3.4 | NA* |
| --- | --- | --- | --- |
| PTH 1–34 | 100,000 pg/mL | 3.9 | 0.00 |
| PTH 39–68 | 100,000 pg/mL | 4.8 | 0.00 |
| PTH 39–84 | 100,000 pg/mL | 14.8 | 0.01 |
| PTH 44–68 | 100,000 pg/mL | 4.5 | 0.00 |
| PTH 53–84 | 100,000 pg/mL | 4.8 | 0.00 |
| PTH 7-84 | 1,000 pg/mL | 925.2 | 92.5 |
*NA = Not Applicable. Concentration was below the measuring range of the test.
f. Assay cut-off:
Not applicable.
2. Comparison studies:
a. Method comparison with predicate device:
**Study Protocol:**
A total of 456 patient serum samples from a commercial vendor were used in the method comparison studies. These samples cover a variety of clinical categories (Hypoparathyroidism, Hyperparathyroidism, Chronic Renal Failures, Hypercalcemia of Malignancies and random donors). To cover the upper measuring range of the proposed iPTH assay, 21 patient samples were spiked with different concentrations of synthetic iPTH with target concentrations between 200 pg/ml and 4750 pg/ml. Regression analysis was performed between each of the VITROS systems (5600 and ECi/ECiQ) and the Predicate.
**Result Summary:**
The Result of Passing Bablok Regression analysis is summarized in the below Table.
| Instrument | N | Range (pg/mL) | Slope | Intercept | r |
| --- | --- | --- | --- | --- | --- |
| VITROS 5600 vs. Predicate | 412 | 3.5-4556 | 1.01 | +0.5 | 0.99 |
| VITROS ECi/ECiQ vs. Predicate | 204 | 4.0-4751 | 1.05 | -1.33 | 0.99 |
**Conclusion:**
Based on the regression analysis result, the sponsor claimed equivalency to the predicate assay.
b. Matrix comparison:
**Study Protocol:**
The matrix effect of various blood collection tubes on fresh and stored serum or plasma samples were evaluated. 40 apparently healthy volunteers were enrolled in the study. For each study participant, blood was collected into each of the following tube types: serum-plain glass, serum-plastic, lithium heparin plasma plastic, sodium heparin plasma plastic,
{10}
EDTA plasma plastic, SST plastic and lithium heparin PST plastic. To cover the entire measuring range of the assay, an additional 10 test samples were created by spiking different volumes of a synthetic PTH stock solution into 500 μl of serum or plasma samples from 10 patients. Aliquots from 40 endogenous samples and 10 spiked samples were made from the respective primary tubes and tested for various storage conditions.
Differences in PTH values (mean of three singlet determinations on VITROS 5600 and ECi/ECiQ system 1 and 2) between serum (glass) and each of the other six collection tubes on freshly collected samples were evaluated using Bias plots and correlation graphs.
To evaluate the effect of various storage conditions on PTH, % Difference was calculated using the following equation:
$$
\% \text{Difference} = \frac{\text{Test Condition} - \text{Baseline Condition}}{\text{Baseline Condition}} \times 100
$$
## Result Summary:
1. Matrix effect on freshly collected sample for PTH range 20-4876 pg/mL
| Tube Y | Tube X | Slope | Intercept | r | PTH range |
| --- | --- | --- | --- | --- | --- |
| Serum Plastic | Serum Glass | 1.05 | 1.38 | 0.999 | 20-4876 mg/dL |
| Lithium heparin plasma plastic | Serum Glass | 1.04 | 0.42 | 0.998 | |
| Lithium heparin PST plasma plastic | Serum Glass | 1.03 | 4.61 | 0.993 | |
| Sodium heparin plasma plastic | Serum Glass | 1.03 | -4.09 | 0.994 | |
| EDTA plasma plastic | Serum Glass | 0.97 | 10.01 | 0.999 | |
| SST plastic | Serum Glass | 1.02 | -9.45 | 0.998 | |
2. To demonstrate equivalence at the low dose iPTH range (3.4 - 20 pg/ml), the sponsor presented data from the method comparison study showing that the 95% confidence intervals for EDTA plasma and serum samples overlap, and there is no statistical difference between the two collection devices, see below:
| Sample Type | N | Slope (95% CI) | Intercept (95% CI) | Range of Samples (pg/mL) |
| --- | --- | --- | --- | --- |
| Serum | 134 | 0.97
(0.88 to 1.07) | -0.16
(-1.66 to 1.25) | 6.4 to 18.4 |
| Plasma | 34 | 1.04
(0.78 to 1.36) | 1.26
(-0.80 to 3.12) | 3.6 to 20.5 |
3. To address the concern regarding performance in serum separator tubes and other collection devices at the low end of the measuring range, the sponsor has conducted an additional study on twenty samples (see below table). This data suggests that the position of the intercept on the y-axis can be significantly increased by including sample values up to the top of the measuring range. The study also suggests that intercepts of +9.0 or +10.0 (seen in the original study) are not expected when assessing sample data around the normal range:
{11}
| Sample Types | n | Slope | Intercept | r | Sample Range (pg/mL) |
| --- | --- | --- | --- | --- | --- |
| Glass vs. EDTA | 20 | 1.01 | -0.40 | 0.997 | 28.10-103.13 |
| Glass vs. Heparin | 20 | 1.04 | -1.01 | 0.997 | 28.10-103.13 |
| Glass vs. Lithium | 20 | 1.05 | -0.29 | 0.994 | 28.10-103.13 |
| Glass vs. SST | 20 | 1.01 | -1.36 | 0.995 | 18.30-66.60 |
4. In addition to the data presented above, the sponsor has also evaluated the matrix effect on sample stability:
| | % Difference from fresh sample | | | | | |
| --- | --- | --- | --- | --- | --- | --- |
| Collection Tube | 2-8 °C | | | | -20 °C | |
| | 1 Day | 2 Days | 5 Days | 7 Days | 4 Weeks | 2X F/T* |
| Serum glass | -11% | -14% | -23% | -27% | -15% | -14% |
| Serum plastic | -11% | -14% | -25% | -33% | -5% | -4% |
| Lithium Heparin | -9% | -8% | -13% | -17% | -1% | -6% |
| Sodium Heparin | -9% | -9% | -14% | -16% | -2% | -2% |
| EDTA | -10% | -9% | -10% | -12% | -8% | 18% |
| SST Plastic | -12% | -16% | -27% | -34% | 0% | -10% |
| Lithium Heparin PST | -10% | -9% | -13% | -17% | -5% | 3% |
* F/T, Freeze-and-Throw
Conclusion: Based on the study results, the sponsor has included the following precautions in the Instructions for Use for collection devices:
# Special Precautions
IMPORTANT: Intact PTH is labile and susceptible to fragmentation. Correct handling of patient samples is necessary to ensure that the PTH molecule remains intact. The degree of fragmentation will depend on both time and temperature of storage.
# Specimen Collection, Preparation and Storage
Specimen types should not be used interchangeably during the serial monitoring of an individual patient as measured concentrations may vary slightly between sample types.
Certain collection devices have been reported to affect other analytes and tests. Owing to the variety of specimen collection devices available, Ortho Clinical Diagnostics is unable to provide a definitive statement on the performance of its products with these devices. Confirm that your collection devices are compatible with this test. Storage of samples collected and stored in serum separator tubes for more than one day may result in a decrease in concentration of up to $20\%$ .
{12}
13
3. Clinical studies:
a. Clinical Sensitivity:
Not applicable
b. Clinical specificity:
Not applicable
c. Other clinical supportive data (when a. and b. are not applicable):
The sponsor evaluated the proposed iPTH assay for its intraoperative applicability in 39 patients who underwent parathyroid surgery. Samples from 20 patients were sourced in the U.S.; samples from 19 patients were sourced in the E.U. The number of samples per patient was not pre-established but was decided during surgery and based on the observed decrease in PTH concentration. The study used the generally accepted guidelines that a >50% decrease in PTH value from baseline suggests complete tumor removal.
31 out of 39 sample series showed a post operative drop of >50% in the intact PTH values. For all 39 samples the VITROS Intact PTH assay results were similar to the reference method values.
4. Clinical cut-off:
Not applicable
5. Expected values/Reference range:
240 archived serum samples with normal Calcium, TSH, Creatinine and Vitamin D values were used to determine the reference range of this assay. These samples were obtained from commercial vendors or hospital laboratories sourced in the U.S. to represent a US population.
The samples were assayed across the instrument platforms using two reagent lots as shown below. The reference interval is the central 95% of the iPTH results.
| Instrument | Reagent Lot | Number of Samples | Reference Interval |
| --- | --- | --- | --- |
| VITROS ECi/ECiQ System 1 | 7 | 60 | 7.5 to 53.5 pg/mL |
| VITROS ECi/ECiQ System 2 | 8 | 60 | |
| VITROS 5600 System 1 | 7 | 60 | |
N. Proposed Labeling:
The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10.
{13}
O. Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
14
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.