ELITECH CLINICAL SYSTEMS PHOSPHORUS, URIC ACID MONO SL AND UREA UV SL REAGENTS
K100263 · Seppim S.A.S. · CEO · May 6, 2011 · Clinical Chemistry
Device Facts
Record ID
K100263
Device Name
ELITECH CLINICAL SYSTEMS PHOSPHORUS, URIC ACID MONO SL AND UREA UV SL REAGENTS
Applicant
Seppim S.A.S.
Product Code
CEO · Clinical Chemistry
Decision Date
May 6, 2011
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 862.1580
Device Class
Class 1
Indications for Use
ELITech Clinical Systems PHOSPHORUS reagent is for the quantitative in vitro diagnostic determination of inorganic phosphorus in human serum and plasma on the Vital Scientific Selectra/Flexor analyzers. It is not intended for use in Point of Care settings. Measurements of phosphorus (inorganic) are used in the diagnosis and treatment of various disorders, including parathyroid gland and kidney diseases, and vitamin D imbalance. ELITech Clinical Systems URIC ACID MONO SL reagent is for the quantitative in vitro diagnostic determination of uric acid in human serum and plasma on the Vital Scientific Selectra/Flexor analyzers. It is not intended for use in Point of Care settings. Measurements obtained by this device are used in the diagnosis and treatment of numerous renal and metabolic disorders, including renal failure, gout, leukemia, psoriasis, starvation or other wasting conditions, and of patients receiving cytotoxic drugs. ELITech Clinical Systems UREA UV SL reagent is for the quantitative in vitro diagnostic determination of urea nitrogen in human serum and plasma on the Vital Scientific Selectra/Flexor analyzers. It is not intended for use in Point of Care settings. Measurements obtained by this device are used in the diagnosis and treatment of certain renal and metabolic diseases. ELITech Clinical Systems ELICAL 2 is a multi-parametric calibrator for in vitro diagnostic use in the calibration of quantitative ELITech Clinical Systems methods on the Vital Scientific Selectra Junior Analyzer and the Vital Scientific Flexor Junior Analyzer. ELITech Clinical Systems ELITROL I is a multi-parametric control serum for in vitro diagnostic use in accuracy control of quantitative ELITech Clinical Systems methods on the Vital Scientific Selectra Junior Analyzer and the Vital Scientific Flexor Junior Analyzer. ELITech Clinical Systems ELITROL II is a multi-parametric control serum for in vitro diagnostic use in accuracy control of quantitative ELITech Clinical Systems methods on the Vital Scientific Selectra Junior analyzer and the Vital Scientific Flexor Junior analyzers.
Device Story
System includes three clinical chemistry reagents (Phosphorus, Uric Acid, Urea) plus calibrator (ELICAL 2) and controls (ELITROL I & II). Reagents are liquid, ready-to-use. Phosphorus assay uses phosphomolybdate UV method; Uric Acid uses enzymatic Trinder method (uricase/peroxidase); Urea uses enzymatic UV method (urease/glutamate dehydrogenase). Device operates on Vital Scientific Selectra/Flexor Junior analyzers in clinical laboratory settings. Input: human serum or lithium heparin plasma. Output: quantitative concentration values displayed to laboratory personnel. Results assist clinicians in diagnosing/monitoring renal, metabolic, and parathyroid conditions. System requires professional operation; not for point-of-care use.
Clinical Evidence
Bench testing only. Performance validated via precision (CLSI EP5-A2), linearity (CLSI EP6-A), and method comparison (CLSI EP9-A2) against predicate devices using 100 patient samples per analyte. Results showed high correlation (r > 0.99) and acceptable precision (CVs < 3%). Interference studies defined limitations for turbidity, hemolysis, and specific drugs.
Indicated for: 1) ELICAL 2: Calibration of quantitative ELITech Clinical Systems methods. 2) UREA UV SL: Measurement of urea nitrogen in human serum/plasma for diagnosis/treatment of renal and metabolic diseases. 3) PHOSPHORUS: Measurement of inorganic phosphorus in human serum/plasma for diagnosis/treatment of parathyroid/kidney diseases and vitamin D imbalance. 4) URIC ACID MONO SL: Measurement of uric acid in human serum/plasma for diagnosis/treatment of renal/metabolic disorders (e.g., gout, leukemia, renal failure). 5) ELITROL I/II: Accuracy control for quantitative ELITech Clinical Systems methods. All devices are for prescription use only; not for Point of Care.
Regulatory Classification
Identification
A phosphorus (inorganic) test system is a device intended to measure inorganic phosphorus in serum, plasma, and urine. Measurements of phosphorus (inorganic) are used in the diagnosis and treatment of various disorders, including parathyroid gland and kidney diseases, and vitamin D imbalance.
{0}
1
# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY ONLY TEMPLATE
A. 510(k) Number:
k100263
B. Purpose for Submission:
New devices and the addition of phosphorus, uric acid and urea assays with calibrator and control materials to a previously cleared device in k093883
C. Measurand:
Phosphorus, Uric Acid and Urea in human serum or lithium heparin plasma
D. Type of Test:
Quantitative spectrophotometry
E. Applicant:
SEPPIM S.A.S.
F. Proprietary and Established Names:
ELITech Clinical Systems Phosphorus, ELITech Clinical Systems Uric Acid Mono SL, ELITech Clinical Systems Urea UV SL, ELITech Clinical Systems ELICAL 2, ELITech Clinical Systems ELITROL 1 & 2
G. Regulatory Information:
1. Regulation section:
| Device | Regulation | Classification | Pro code |
| --- | --- | --- | --- |
| ELITech Clinical Systems Urea UV SL | 21 CFR § 862.1770, Urea nitrogen test system | II | CDQ |
| ELITech Clinical Systems ELICAL 2 | 21 CFR § 862.1150, Calibrator | II | JIX |
| ELITech Clinical Systems Phosphorus | 21 CFR § 862.1580, Phosphorus (inorganic) test system | I (reserved) | CEO |
| ELITech Clinical Systems Uric Acid | 21 CFR § 862.1775, Uric acid test system | I (reserved) | KNK |
{1}
| Mono SL | | | |
| --- | --- | --- | --- |
| ELITech Clinical Systems ELITROL 1 & 2 | 21 CFR § 862.1660, Quality Control material | I (reserved) | JJY |
2. Panel:
75, Clinical Chemistry
## H. Intended Use:
1. Intended use(s):
See Indication(s) for use
2. Indication(s) for use:
ELITech Clinical Systems PHOSPHORUS reagent is for the quantitative *in vitro* diagnostic determination of inorganic phosphorus in human serum and plasma on the Vital Scientific Selectra/Flexor analyzers. It is not intended for use in Point of Care settings. Measurements of phosphorus (inorganic) are used in the diagnosis and treatment of various disorders, including parathyroid gland and kidney diseases, and vitamin D imbalance.
ELITech Clinical Systems URIC ACID MONO SL reagent is for the quantitative *in vitro* diagnostic determination of uric acid in human serum and plasma on the Vital Scientific Selectra/Flexor analyzers. It is not intended for use in Point of Care settings. Measurements obtained by this device are used in the diagnosis and treatment of numerous renal and metabolic disorders, including renal failure, gout, leukemia, psoriasis, starvation or other wasting conditions, and of patients receiving cytotoxic drugs.
ELITech Clinical Systems UREA UV SL reagent is for the quantitative *in vitro* diagnostic determination of urea nitrogen in human serum and plasma on the Vital Scientific Selectra/Flexor analyzers. It is not intended for use in Point of Care settings. Measurements obtained by this device are used in the diagnosis and treatment of certain renal and metabolic diseases.
ELITech Clinical Systems ELICAL 2 is a multi-parametric calibrator for *in vitro* diagnostic use in the calibration of quantitative ELITech Clinical Systems methods on the Vital Scientific Selectra Junior Analyzer and the Vital Scientific Flexor Junior Analyzer.
ELITech Clinical Systems ELITROL I is a multi-parametric control serum for *in vitro* diagnostic use in accuracy control of quantitative ELITech Clinical Systems methods on the Vital Scientific Selectra Junior Analyzer and the Vital Scientific Flexor Junior Analyzer.
{2}
ELITech Clinical Systems ELITROL II is a multi-parametric control serum for in vitro diagnostic use in accuracy control of quantitative ELITech Clinical Systems methods on the Vital Scientific Selectra Junior analyzer and the Vital Scientific Flexor Junior analyzers.
3. Special conditions for use statement(s):
Serum and lithium heparin plasma only
Prescription use
4. Special instrument requirements:
For use on the Vital Scientific Selectra Junior (also trademarked as the Vital Scientific Flexor Junior) Analyzer
All studies were performed on the Vital Scientific Selectra Junior.
I. Device Description:
The ELITech Clinical Systems PHSOPHORUS kit reagent is a one (1) reagent system with reagent R. Reagent R is supplied in liquid ready-to-use form and contains sulfuric acid and ammonium molybdate.
The ELITech Clinical Systems URIC ACID MONO SL kit reagent is a one (1) reagent system. Reagent R is supplied in liquid ready-to-use form and contains Phosphate buffer, N-Ethyl-N-(2-Hydroxy-3-Sulfopropyl) m-Toluidine (EHSPT), Ferrocyanide, Amino-4-antipyrine (4-AAP), Uricase (microorganisms), Peroxidase (horseradish) and sodium azide. The measuring range can be extended by making manual dilutions with 0.9% saline.
The ELITech Clinical Systems UREA UV SL kit reagent is a two (2) reagents system with reagent R1 and reagent R2. Reagents R1 and R2 are supplied in liquid ready-to-use form. Reagent R1 contains Tris buffer (pH 7.60), Adenosine diphosphate potassium salt (ADP), α-Ketoglutarate, Urease (jack bean), Glutamate dehydrogenase (GIDH) (bovine liver) and sodium azide. Reagent R2 contains NADH and sodium azide. The measuring range can be extended by making manual dilutions with 0.9% saline.
ELITech Clinical Systems Elical 2 is a lyophilized calibrator based on human serum containing added constituents.
ELITech Clinical Systems Elitrol I and II is a two level quality control product consisting of lyophilized human serum containing constituents at desired levels.
All human source materials were tested with FDA-approved methods and found to be negative for HbsAG and to antibodies to HCV and HIV 1/2.
3
{3}
J. Substantial Equivalence Information:
1. Predicate device name(s) and 2. Predicate 510(k) number(s):
| Device | Predicate Device Name | Predicate 510(k) number |
| --- | --- | --- |
| The ELITech Clinical Systems PHOSPHORUS | ABX Pentra Phosphorus CP | k060205 |
| The ELITech Clinical Systems URIC ACID MONO SL | ABX Pentra Uric Acid CP | k060205 |
| The ELITech Clinical Systems UREA UV SL | ABX Pentra Urea CP | k060205 |
| ELITech Clinical Systems Elical 2 | Roche Diagnostics Calibrator for Automated Systems (C.f.a.s) | k033501 |
| ELITech Clinical Systems Elitrol I and II | Roche Diagnostics Precinorm U and Precipath U | k041227 |
3. Comparison with predicate:
| Similarities | | |
| --- | --- | --- |
| Item | The ELITech Clinical Systems PHOSPHORUS | ABX Pentra Phosphorus CP k060205 |
| Intended use | For in vitro diagnostic use in the quantitative determination of inorganic phosphorus. | Same |
| Assay Principle | UV method using phosphomolybdate | Same |
| Appearance of reagents | Liquid form, ready to use | Same |
| Composition | Reagent R: Sulfuric acid, Ammonium molybdate | Same |
| Expected values | Serum, plasma : 2.7 – 4.5 mg/dL | Same |
| Differences | | |
| --- | --- | --- |
| Item | The ELITech Clinical Systems PHOSPHORUS | ABX Pentra Phosphorus CP k060205 |
| Sample type | Serum
Plasma in lithium heparin | Serum
Plasma in lithium heparin
Urine |
| Reagent storage | Store at 2-25 °C and protect from light. | Store at 2-8 °C. |
{4}
| Differences | | |
| --- | --- | --- |
| Item | The ELITech Clinical Systems PHOSPHORUS | ABX Pentra Phosphorus CP k060205 |
| Measuring range | 2.0 to 20.0 mg/dL | 0.30 to 24.18 mg/dL
Automatic post-dilution: 96.72 mg/dL |
| LoD | 0.02 mg/dL | 0.28 mg/dL |
| LoQ | 1.00 mg/dL | |
| Calibration Frequency | 28 days | 34 days |
| On board stability | refrigerated area : 28 days | refrigerated area: 70 days |
| Precision | Within run
Level 2.37 mg/dL, CV=1.1%
Level 4.80 mg/dL, CV=1.5%
Level 9.55 mg/dL, CV=1.7%
Total
Level 2.37 mg/dL, CV=1.9%
Level 4.80 mg/dL, CV=1.7%
Level 9.55 mg/dL, CV=2.2% | Within run
Level 4.08 mg/dL, CV=1.25%
Level 6.34 mg/dL, CV=0.77%
Level 2.39 mg/dL, CV=2.48%
Level 3.48 mg/dL, CV=1.61%
Level 9.19 mg/dL, CV=1.38%
Total
Level 4.01 mg/dL, CV=2.50%
Level 6.35 mg/dL, CV=1.82%
Level 2.50 mg/dL, CV=3.56%
Level 11.44 mg/dL, V=1.38% |
| Method comparison | y=0.999x - 0.09 mg/dL
r2= 0.994
range: 2.02 to 20.08 mg/dL | y=1.04x + 0.15 mg/dL
r2= 0.998
range: 0.30 to 24.08 mg/dL |
| Limitations | Hemoglobin: No significant interference up to 50 mg/dL.
Triglycerides: No significant interference up to 732 mg/dL,
Unconjugated bilirubin: No significant interference up to 15 mg/dL
Conjugated bilirubin: No significant interference up to 1.5 mg/dL
Glucose: No significant interference up to 500 mg/dL. | Hemoglobin: No significant interference up to 125 mg/dL.
Triglycerides: No significant interference up to 262.5 mg/dL.
Total bilirubin: No significant interference up to 6 mg/dL.
Direct bilirubin: No significant interference up to 25 mg/dL. |
{5}
| Similarities | | |
| --- | --- | --- |
| Item | The ELITech Clinical Systems URIC ACID MONO SL | ABX Pentra Uric Acid CP k060205 |
| Intended use | For in vitro diagnostic use in the quantitative determination of uric acid. | Same |
| Assay protocol | Enzymatic determination using a chromogenic system in the presence of peroxidase and uricase (Trinder method). | Same |
| Appearance of reagents | Liquid form, ready to use | Same |
| Reagent storage | Store at 2-8 °C and protected from light. | Same |
| Expected values | Serum, plasma Women: 2.6 – 6.0 mg/dL Men: 3.5 – 7.2 mg/dL | Same |
| Differences | | |
| --- | --- | --- |
| Item | The ELITech Clinical Systems URIC ACID MONO SL | ABX Pentra Uric Acid CP k060205 |
| Composition | Reagent R: Phosphate buffer, pH 7.0 EHSPT Ferrocyanide Amino-4-antipyrine Uricase Peroxidase Sodium azide | Reagent 1: Phosphate buffer, pH 7.0 EHSPT Ascorbate oxidase Bovine albumin Sodium azide Reagent 2: 4-Aminoantipyrine Uricase Peroxidase Ferrocyanide Bovine albumin Sodium azide |
| Sample type | Serum Lithium heparin plasma | Serum Lithium heparin plasma Urine |
| Measuring range | 1.5 to 25.0 mg/dL Extended measuring range: 25.0-78.0 mg/dL | 0.18 to 25.00 mg/dL Automatic post-dilution: 75.00 mg/dL |
| LoD LoQ | 0.02 mg/dL 0.50 mg/dL | 0.19 mg/dL |
{6}
| Differences | | |
| --- | --- | --- |
| Item | The ELITech Clinical Systems URIC ACID MONO SL | ABX Pentra Uric Acid CP k060205 |
| Precision | Within run
Level 2.49 mg/dL, CV=0.8%
Level 5.19 mg/dL, CV=1.3%
Level 7.63 mg/dL, CV=1.1%
Total
Level 2.49 mg/dL, CV=2.6%
Level 5.19 mg/dL, CV=2.0%
Level 7.63 mg/dL, CV=2.1% | Within run
Level 4.62 mg/dL, CV=0.45%
Level 11.63 mg/dL, CV=0.34%
Level 2.53 mg/dL, CV=1.24%
Level 4.58 mg/dL, CV=0.91%
Level 7.19 mg/dL, CV=1.02%
Total
Level 4.64 mg/dL, CV=2.81%
Level 11.73 mg/dL, CV=1.39%
Level 4.67 mg/dL, CV=2.64%
Level 6.74 mg/dL, CV=2.51% |
| Method comparison | y=1.015 x + 0.03 mg/dL
r2= 0.999
range: 1.49 to 24.40 mg/dL | y=0.95 x + 0.09 mg/dL
r2= 0.996
range: 0.18 to 23.59 mg/dL |
| Limitations | Hemoglobin: No significant interference up to 50 mg/dL
Triglycerides: No significant interference up to 1070 mg/dL.
Unconjugated bilirubin: No significant interference up to 30 mg/dL.
Conjugated bilirubin: No significant interference up to 14.8 mg/dL
Glucose: No significant interference up to 500 mg/dL.
Ascorbic acid: Significant interference at all levels of ascorbic acid.
Methyldopa: No significant interference up to 1 mg/dL.
Calcium dobesilate: Induces falsely low results at therapeutic concentrations. | Hemoglobin: No significant interference up to 500 mg/dL.
Triglycerides: No significant interference up to 612.5 mg/dL.
Total bilirubin: No interference up to 36 mg/dL.
Direct bilirubin: No interference up to 30 mg/dL. |
| Calibration Frequency | 28 days | 15 days |
| On board stability | refrigerated area : 28 days | refrigerated area: 41 days |
| Measuring range | 1.5 to 25.0 mg/dL
Extended measuring range: 25.0-78.0 mg/dL | 0.18 to 25.00 mg/dL
Automatic post-dilution: 75.00 mg/dL |
{7}
| Similarities | | |
| --- | --- | --- |
| Item | The ELITech Clinical Systems UREA UV SL | ABX Pentra Urea CP k060205 |
| Intended use | For in vitro diagnostic use in the quantitative determination of urea nitrogen in serum and plasma. | Same |
| Assay protocol | Enzymatic UV method using Urease and Glutamate dehydrogenase | Same |
| Appearance of reagents | Liquid form, ready to use | Same |
| Reagent storage | Store at 2-8 °C and protected from light. | Same |
| Differences | | |
| --- | --- | --- |
| Item | The ELITech Clinical Systems UREA UV SL | ABX Pentra Urea CP k060205 |
| Expected values | Serum, plasma: Urea nitrogen (BUN) Adults (18-60 years)6-20 mg/dL Adults (60-90 years)8-23 mg/dL Adults (>90 years)10-31 mg/dL | Serum, plasma: Urea nitrogen (BUN) Adults: Global : 7.9-20.2 mg/dL Women < 50 years 7.3-18.8 mg/dL Women > 50 years 9.8-20.2 mg/dL Men < 50 years 9.0-20.5 mg/dL Men > 50 years 8.4-25.8 mg/dL Children: 1-3 years 5.1-16.8 mg/dL 4-13 years 7.0-16.8 mg/dL 14-19 years 8.1-21.1 mg/dL |
| Measuring range | 4.7 to 140.0 mg/dL Extended measuring range: 140.0-670.0 mg/dL | 1.03 to 140.3 mg/dL Automatic post-dilution: 701.5 mg/dL |
| LoD LoQ | 0.3 mg/dL 2.3 mg/dL | 0.9 mg/dL |
| Precision | Within run | Within run |
{8}
| Differences | | |
| --- | --- | --- |
| Item | The ELITech Clinical Systems UREA UV SL | ABX Pentra Urea CP k060205 |
| | Level 7.3 mg/dL CV=2.1% Level 29.2 mg/dLCV=0.8% Level 72.4 mg/dLCV=0.7% Total Level 7.3 mg/dL CV=2.8% Level 29.2 mg/dLCV=1.3% Level 72.4 mg/dLCV=1.6% | Level 18.7 mg/dL, CV=2.27% Level 72.8 mg/dL, CV=1.66% Level 6.0 mg/dL, CV=2.76% Level 20.9 mg/dL, CV=1.58% Level 85.5 mg/dL, CV=1.80% Total Level 18.5 mg/dL, CV=2.14% Level 71.7 mg/dL, CV=1.93% Level 19.2 mg/dL, CV=2.14% Level 70.1 mg/dL, CV=1.97% |
| Method comparison | y=0.991 x + 0.6 mg/dL r²= 0.999 range: 4.4 to 139.8 mg/dL | y=0.99 x - 0.06 mg/dL r²= 0.996 range: 1.03 to 138.89 mg/dL |
| Limitations | Hemoglobin: No significant interference up to 500 mg/dL. Turbidity: No significant interference up to 614 mg/dL triglyceride equivalent. Unconjugated bilirubin: No significant interference up to 36 mg/dL. Conjugated bilirubin: No significant interference up to 25 mg/dL. Ascorbic acid: No significant interference up to 20 mg/dL. Methyl-dopa: No significant interference up to 1 mg/dL. | Hemoglobin: No interference up to 460 mg/dL Triglycerides: No interference up to 612.5 mg/dL Total bilirubin: No interference up to 22.23 mg/dL Direct bilirubin: No interference up to 23.4 mg/dL |
| Calibration Frequency | 7 days | 8 days |
| On board stability | 14 days | 70 days |
| Similarities and Differences | | |
| --- | --- | --- |
| Item | ELITech Clinical Systems Elical 2 | Roche Calibrator (C.f.a.s) k033501 |
| Intended Use/Indications for Use | For in vitro diagnostic use in the calibration of quantitative methods | Same |
| Format | Lyophilized calibrator based on human serum with constituents added as required | Same |
{9}
10
| Similarities and Differences | | |
| --- | --- | --- |
| Item | ELITech Clinical Systems Elitrol I/Elitrol II | Roche Diagnostics Precinorm U and Precipath U k041227 |
| Intended Use/Indications for Use | For in vitro diagnostic use in accuracy control of quantitative methods | Same |
| Format | Lyophilized calibrator based on human serum with constituents added as required to obtain desired component levels | Same |
| Level | Two levels | Same |
| Stability | Lyophilized: store at 2-8° C and protect from light until the expiry date.
After reconstitution: < 12 hrs between 15-25° C, 5 days between 2-8° C, 4 weeks between -25° to -15° C (when frozen once). | Same |
| Instrument | Vital Scientific Selectra Junior Analyzer (also trademarked as the Vital Scientific Flexor Junior Analyzer) | Roche Analyzers |
{10}
11
# K. Standard/Guidance Document Referenced (if applicable):
CLSI EP7-A2: Interference Testing in Clinical Chemistry; Approved Guideline—2nd edition, Vol. 25, No. 27, Nov 2005.
CLSI EP9-A2: Method comparison and bias estimation using patient samples; Approved Guideline—2nd edition, Vol. 22, No. 19, Sept 2002.
CLSI EP6A: Evaluation of the linearity of the measurement of quantitative procedures: a statistical approach. Vol 23, No. 16, Apr 2003
CLSI EP5-A2: Evaluation of precision performance of quantitative measurement methods; Approved guideline—2nd edition, Vol. 24, No.25, Aug. 2004
Use of Symbols on Labels and in Labeling of in Vitro Diagnostic Devices Intended for Professional Use: Guidance for Industry and FDA Staff, Nov 2004.
Guidance for Industry and FDA Staff: Bundling Multiple Devices or Multiple Indications in a Single Submission, Nov 2003
Guidance for Industry and FDA Staff: Format for Traditional and Abbreviated 510(k)s, Aug. 2005.
# L. Test Principle:
1. The ELITech Clinical Systems PHOSPHORUS:
In the presence of sulfuric acid (acid medium), inorganic phosphate forms with ammonium molybdate an ammonium phosphomolybdate complex.
$$
\text{Ammonium molybdate} + \text{Sulfuric acid} \xrightarrow{\text{Phosphorus}} \text{Phosphomolybdate complex}
$$
The concentration of phosphomolybdate formed is directly proportional to the inorganic phosphate concentration in the sample and is measured at 340 nm.
2. The ELITech Clinical Systems URIC ACID MONO SL:
Uric acid is cleaved by uricase to form allantoine, CO₂ and hydrogen peroxide (H₂O₂). H₂O₂ reacts with 4-amino-antipyrine (4-AAP) and N-Ethyl-N-(2-Hydroxy-3-Sulfopropyl) m-Toluidine (EHSPT) under the catalytic action of peroxidase to form a colored quinoneimine. The absorbance of the quinoneimine at 546 nm is proportional to the concentration of uric acid in the sample.
$$
\begin{array}{l l}
\text{Uric acid} + 2\mathrm{H_2O} + \mathrm{O_2} & \xrightarrow{\text{Uricase}} \text{Allantoin} + \mathrm{CO_2} + \mathrm{H_2O_2} \\
2\mathrm{H_2O_2} + 4\text{-AAP} + \text{EHSPT} & \xrightarrow{\text{Peroxidase}} \text{Quinoneimine} + 4\mathrm{H_2O}
\end{array}
$$
3. The ELITech Clinical Systems UREA UV SL
{11}
Urea is hydrolyzed by urease to form ammonium and carbonate. The ammonium produce reacts with $\alpha$ -Ketoglutarate and NADH under the catalytic action of Glutamate Dehydrogenase (GlDH) to form L-Glutamate and $\mathrm{NAD^{+}}$ . The decrease in absorbance of NADH at $340~\mathrm{nm}$ is proportional to the concentration of urea in the sample.

# M. Performance Characteristics (if/when applicable):
# 1. Analytical performance:
# a. Precision/Reproducibility:
Within-run and Total precision results for the ELITech Clinical Systems Phosphorus, Uric Acid Mono SL, and Urea UV SL were obtained by spiking 3 serum pools to concentrations below, within, and above the expected value ranges for each analyte. The sponsor performed two runs per day, two measures per run, for twenty days according to CLSI protocol EP5-A2 using one Selectra Junior and one lot of reagent. The results are presented in the tables below:
1). The ELITech Clinical Systems PHOSPHORUS:
| | n | Mean (mg/dL) | Within-run SD | Total SD | Within-run CV% | Total CV% |
| --- | --- | --- | --- | --- | --- | --- |
| Low level | 80 | 2.37 | 0.03 | 0.04 | 1.1 | 1.9 |
| Medium level | 80 | 4.80 | 0.07 | 0.08 | 1.5 | 1.7 |
| High level | 80 | 9.55 | 0.16 | 0.21 | 1.7 | 2.2 |
The ELITech Clinical Systems URIC ACID MONO SL:
| | n | Mean (mg/dL) | Within-run SD | Total SD | Within-run CV% | Total CV% |
| --- | --- | --- | --- | --- | --- | --- |
| Low level | 80 | 2.49 | 0.02 | 0.06 | 0.8 | 2.6 |
| Medium level | 80 | 5.19 | 0.07 | 0.11 | 1.3 | 2.0 |
| High level | 80 | 7.63 | 0.08 | 0.16 | 1.1 | 2.1 |
The ELITech Clinical Systems UREA UV SL:
| | n | Mean (mg/dL) | Within-run SD | Total SD | Within-run CV% | Total CV% |
| --- | --- | --- | --- | --- | --- | --- |
| Low level | 80 | 7.3 | 0.16 | 0.20 | 2.1 | 2.8 |
| Medium level | 80 | 29.2 | 0.23 | 0.39 | 0.8 | 1.3 |
| High level | 80 | 72.4 | 0.53 | 1.12 | 0.7 | 1.6 |
{12}
# b. Linearity/assay reportable range:
The linearity of the Phosphorus, Uric Acid Mono SL and Urea UV SL was determined using the same protocol. Stock solutions for each device were gravimetrically prepared from commercially available material to make a high and low serum pools. High and low serum pools were mixed for each of the measurands to obtain 11 samples covering the proposed measuring range of each device. One Selectra Junior and one lot of reagent were used. Samples were tested in triplicate. The data analysis was performed and evaluated according to CLSI EP6-A using linear regression as well as second and third order non-linear fitted polynomial regressions. Linearity met the $1^{\text{st}}$ order polynomial for the three devices. The results are summarized below:
| Device | Linear Regression | R2 | Measuring Range |
| --- | --- | --- | --- |
| Phosphorus | y = 1.011x - 0.0947 | 0.9998 | 2.0-20.0 mg/dL |
| Uric Acid Mono SL | y = 0.9972x +0.1347 | 0.9994 | 1.5-25 mg/dL |
| Urea UV SL | y = 1.0177x - 0.8375 | 0.9985 | 4.7-140 mg/dL |
Recovery studies were performed for Uric Acid Mono SL and Urea to demonstrate that the analyzer can accurately measure manually or automated diluted samples. The study protocols for both measurands were the same. Stock spiking solutions were prepared for each analyte using commercially available dry reagents. 10 natural samples for each test, with known concentrations of uric acid or urea, were spiked with the stock solutions to obtain expected concentrations of $25.62 - 78.16\mathrm{mg / dL}$ of uric acid or $139.7 - 690.2\mathrm{mg / dL}$ of urea. All samples were then diluted 1:5 with 9 gm/L saline. Samples were analyzed in duplicate on one Spectra Junior using one lot of reagents and one calibration. The means were calculated for each sample, and multiplied by the dilution factor. The observed values were compared to the expected values to determine $\%$ recovery. The pre-determined acceptance criterion for recovery was within $10\%$ of the expected value. Results are summarized below.
Uric Acid Mono SL
| Expected Concentrations | Observed Results | Recovery % |
| --- | --- | --- |
| 25.62 | 23.49 | -8.3% |
| 29.91 | 28.02 | -6.3% |
| 36.39 | 33.35 | -8.4% |
| 42.51 | 40.69 | -4.3% |
| 47.03 | 44.21 | -6.0% |
| 54.29 | 53.05 | -2.3% |
| 59.95 | 59.31 | -1.1% |
| 65.46 | 64.67 | -1.2% |
| 70.62 | 69.27 | -1.9% |
| 78.16 | 77.41 | -1.0% |
{13}
Urea Nitrogen Mono SL
| Expected Concentrations | Observed Results | Recovery % |
| --- | --- | --- |
| 139.7 | 137.5 | -1.6% |
| 199.3 | 200.8 | 0.8% |
| 258.8 | 277.4 | 7.2% |
| 320.3 | 325.1 | 1.5% |
| 382.9 | 393.9 | 2.9% |
| 438.8 | 442.4 | 0.8% |
| 497.1 | 490.5 | -1.3% |
| 565.5 | 558.6 | -1.2% |
| 623.9 | 614.4 | -1.5% |
| 690.2 | 668.2 | -3.2% |
c. Traceability, Stability, Expected values (controls, calibrators, or methods):
ELICal 2 Calibrator for Phosphorus, Uric Acid Mono SL, and Urea UV SL are purchased from a commercially available source and are traceable to: a NIST traceable material, ID-MS, and SRM 909b, respectively.
Calibrator material is purchased from commercially available sources (previously cleared under k033501). Elical 2 calibrator is value assigned using multiple runs on two Vital Scientific Flexor Junior Analyzers. The target value of Elical 2 calibrator is the mean of the observed values range. After validation of the target value, a confidence range (high and low values) is calculated. Unreconstituted calibrators are stable for 11 months at $2 - 8^{\circ}\mathrm{C}$ . Reconstituted calibrator stability for all three analytes has been verified for the following temperatures and time limits: 12 hrs. at $15 - 25^{\circ}\mathrm{C}$ , 5 days at $2 - 8^{\circ}\mathrm{C}$ , 4 weeks at $-15^{\circ}\mathrm{C}$ and $-25^{\circ}\mathrm{C}$ .
Protocols and acceptance criteria were provided to support on-board calibration stability for 28 days for phosphorus and uric acid and 7 days for urea at $15 - 25^{\circ}\mathrm{C}$ .
Control material is purchased from a commercial vendor (previously cleared under k041227). The sponsor claimed the following for stability: before reconstitution, the shelf-life of the ELITech Clinical Systems Elitrol 1 and Elitrol II is 30 months at $2-8^{\circ}\mathrm{C}$ . After reconstitution the stability is 12 hours when stored at $15-25^{\circ}\mathrm{C}$ , 5 days when stored at $2-8^{\circ}\mathrm{C}$ or 4 weeks (when frozen once) at $-25^{\circ}$ and $-15^{\circ}\mathrm{C}$ . Elitrol I and II control solutions are value assigned using two Vital Scientific Flexor Junior Analyzers. Each sample is tested in triplicate over several days. The target value of Level I and II are the median of the observed values range. After validation of the target value, a confidence range (high and low values) is calculated.
Reagent stability for the Phosphorus, Uric Acid Mono SL, and Urea UV SL were evaluated on one Selectra Junior with one lot of reagent using three samples with low, medium and high concentrations of each measurand. These samples were aliquoted
{14}
and analyzed. On board stability are 28 days for phosphorus and uric acid and 14 days for urea.
# d. Detection limit:
The Limits of Blank (LoB), Detection (LoD) and Quantitation (LoQ) studies were conducted following CLSI EP17-A for Phosphorus, Uric Acid Mono SL, and Urea UV SL. One Selectra Junior and one lot of the respective reagents were used. The LoBs were determined by analyzing 60 blank samples. The LoD and LoQ studies each used 4 serum pools for the individual assays. The serum pools for the LoD had concentrations of the individual measurands approximately 4 times the LoB. LoQs were determined by analyzing 4 serum pools near the expected LoQ and evaluating the Total Error. Results are summarized below:
| Device | LoB | LoD | LoQ | Total Error |
| --- | --- | --- | --- | --- |
| Phosphorus | 0.01 mg/dL | 0.02 mg/dL | 1.00 mg/dL | 11.0% |
| Uric Acid Mono SL | 0.01 mg/dL | 0.02 mg/dL | 0.50 mg/dL | 4.8% |
| Urea UV SL | 0.1 mg/dL | 0.3 mg/dL | 2.3 mg/dL | 12.4% |
# e. Analytical specificity:
i. Endogenous interferences of turbidity (20% Intralipid solution), triglycerides, unconjugated bilirubin, conjugated bilirubin, hemoglobin and glucose were investigated for the ELITech Clinical Systems PHOSPHORUS, URIC ACID MONO SL, and UREA UV SL. Each device was tested in triplicate at two concentrations for each of the individual measurands. Recovery was compared to a sample with no interfering substance and significant interference was defined as recoveries $\geq 10\%$ . Urea was not tested for native triglycerides or glucose. The concentrations of interferents tested for the three devices are in the table below:
| Interferent | Range of interferent tested |
| --- | --- |
| Turbidity (Intralipid®) | up to 614 mg/dL |
| Triglycerides | up to 1000 mg/dL |
| Unconjugated Bilirubin | up to 30.0 mg/dL |
| Conjugated Bilirubin | up to 29.5 mg/dL |
| Hemoglobin | up to 500 mg/dL |
| Glucose | up to 550 mg/dL |
The concentrations of endogenous substances that demonstrated no significant interference for each device are below:
| | Phosphorus | Uric Acid Mono UV | Urea UV |
| --- | --- | --- | --- |
| Turbidity (Intralipid®) | See below* | See below* | up to 614 mg/dL |
| Triglycerides | up to 732 mg/dL | up to 1070 mg/dL | Not tested |
| Unconjugated Bilirubin | up to 15 mg/dL | up to 30 mg/dL | up to 30 mg/dL |
{15}
| Conjugated Bilirubin | up to 1.5 mg/dL | up to 14.8 mg/dL | up to 29.5 mg/dL |
| --- | --- | --- | --- |
| Hemoglobin | up to 50 mg/dL | up to 50 mg/dL | up to 500 mg/dL |
| Glucose | up to 500 mg/dL | up to 500 mg/dL | Not tested |
* Interference from turbidity was observed at both phosphorus concentrations and uric acid concentrations tested across all concentrations of Intralipid® (turbidity). Therefore, the sponsor recommends in the labeling that only clear (non-turbid) samples should be analyzed for phosphorus and uric acid.
Hemoglobin concentrations of 50 mg/dL also demonstrate interference for both phosphorus and uric acid. Therefore, the sponsor recommends in the labeling that no visibly hemolyzed samples should be analyzed for phosphorus and uric acid.
Conjugated bilirubin concentration of >1.5 mg/dL also demonstrated interference for phosphorus ≥10%, therefore, the sponsor put a limitation that icteric samples should not be used with the phosphorus assay.
Exogenous interferences were evaluated for PHOSPHORUS, URIC ACID MONO SL and UREA UV using the same concentrations of measurands as the endogenous interference studies. The substances were tested up to 1 mg/dL of methyl dopa, up to 20 mg/dL of ascorbic acid, and up to 50 mg/dL of calcium dobesilate. The concentrations of exogenous substances that demonstrated no significant interference for each device are below:
| | Phosphorus | Uric Acid Mono UV | Urea UV |
| --- | --- | --- | --- |
| Methyl dopa | Up to 1 mg/dL | up to 1 mg/dL | up to 1 mg/dL |
| Ascorbic Acid | Up to 20 mg/dL | See below** | up to 20 mg/dL |
| Calcium dobesilate | Not tested | See below** | Not tested |
**Ascorbic acid and calcium dobesilate demonstrated interference at all uric acid concentrations tested. Therefore, the sponsor has the following statement of limitation in their labeling:
Samples with any amount of ascorbic acid or Calcium dobesilate will have significant interference and should not be used for uric acid measurement."
f. Assay cut-off:
Not applicable.
2. Comparison studies:
a. Method comparison with predicate device:
1). ELITech Clinical Systems PHOSPHORUS: A combination of 100 human serum
{16}
and plasma samples were analyzed on the predicate and the Selectra Junior according to CLSI EP9-A2. Twenty samples were tested each day for 5 days. Phosphorus concentrations ranged from 2.0 mg/dL to 20.0 mg/dL.
| Slope (95%CI) | Intercept (95%CI) | r | Sy.x |
| --- | --- | --- | --- |
| 0.999 (0.994 to 1.004) | –0.09 (-0.12 to -0.06) | 0.999 | 0.12 |
2). ELITech Clinical Systems URIC ACID MONO SL: A combination of 100 human serum and plasma samples were analyzed on the predicate and the Selectra Junior according to CLSI EP9-A2. Twenty samples were tested each day for 5 days. Uric Acid concentrations ranged from 1.5 mg/dL to 24.0 mg/dL.
| Slope (95%CI) | Intercept (95%CI) | r | Sy.x |
| --- | --- | --- | --- |
| 1.015 (1.010 to 1.020) | 0.03 (-0.01 to 0.07) | 0.999 | 0.18 |
3). ELITech Clinical Systems UREA UV SL: ELITech Clinical Systems UREA UV SL: A combination of 100 human serum samples were analyzed on the predicate and the Selectra Junior according to CLSI EP9-A2 over 7 days. Urea concentrations ranged from 4.7 mg/dL to 140.0 mg/dL.
| Slope (95%CI) | Intercept (95%CI) | r | Sy.x |
| --- | --- | --- | --- |
| 0.991 (0.985 to 0.997) | 0.6 (0.4 to 0.9) | 0.999 | 1.3 mg/dL |
b. Matrix comparison:
Paired serum and lithium heparin samples were collected for each of the three devices. Analysis was performed on one Select Junior over a period of several days using one lot of reagents.
1). ELITech Clinical Systems PHOSPHORUS:
52 paired serum and lithium heparin samples were analyzed ranging from 2.0 mg/dL to 20.0 mg/dL on the Selectra Junior. The linear regression is :
| Slope (95%CI) | Intercept (95%CI) | r | Sy.x |
| --- | --- | --- | --- |
| 1.010 (1.003 to 1.016) | -0.2 (-0.27 to -0.14) | 0.999 | 0.21 |
2). ELITech Clinical Systems URIC ACID MONO SL:
50 paired serum and lithium heparin samples were analyzed ranging from 1.5 mg/dL to 25.0 mg/dL on the Selectra Junior. The linear regression is :
{17}
| Slope (95% CI) | Intercept (95% CI) | r | Sy.x |
| --- | --- | --- | --- |
| 0.996 (0.986 to 1.006) | -0.06 (-0.21 to 0.08) | 0.999 | 0.39 |
3). ELITech Clinical Systems UREA UV SL:
40 paired serum and lithium heparin samples were analyzed ranging from 4.7 mg/dL to 140.0 mg/dL on the Selectra Junior. The linear regression is :
| Slope (95% CI) | Intercept (95% CI) | r | Sy.x |
| --- | --- | --- | --- |
| 0.984 (0.970 to 0.999) | 0.5 (-0.5 to 1.6) | 0.998 | 2.6 |
3. Clinical studies:
a. Clinical Sensitivity: Not applicable
b. Clinical specificity: Not applicable
c. Other clinical supportive data (when a. and b. are not applicable): Not applicable
4. Clinical cut-off:
Not applicable.
5. Expected values/Reference range:
$^{1}$ Phosphorus: Serum, plasma: 2.7-4.5 mg/dL, 0.87-1.45 mmol/L
$^{2}$ Uric Acid: Serum, plasma: Women: 2.6 - 6.0 mg/dL, Men: 3.5 - 7.2 mg/dL
$^{3}$ Urea: Serum, plasma: Adults (18-60 years) 6-20 mg/dL, Adults (60-90 years) 8-23 mg/dL, Adults (>90 years) 10-31 mg/dL
1 Endres, D.B., Rude, R.K., Mineral and bone metabolism. Tietz Fundamentals of Clinical Chemistry, Burtis, C.A. and Ashwood, E.R. (W.B. Saunders eds. Philadelphia USA), (2001), 795.
2 Newman, D. J., Price, C. P., Non protein Nitrogen Metabolite. Tietz Fundamentals of Clinical Chemistry, 5th Ed., Burtis, C.A. & Ashwood, E.R. (W.B. Saunders eds. Philadelphia USA), (2001), 414.
$^{3}$ Tietz, N.W., Clinical guide to laboratory tests, 3rd Ed., (W.B. Saunders eds. Philadelphia USA), (1995), 622.
N. Other Supportive Instrument Performance Characteristics Data Not Covered In The "Performance Characteristics" Section above:
None
{18}
O. Proposed Labeling:
The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10.
P. Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
19
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.