ADVIA CENTAUR FT4, ADVIA CENTAUR T4

K132249 · Siemens Healthcare Diagnostics, Inc. · CEC · Sep 13, 2013 · Clinical Chemistry

Device Facts

Record IDK132249
Device NameADVIA CENTAUR FT4, ADVIA CENTAUR T4
ApplicantSiemens Healthcare Diagnostics, Inc.
Product CodeCEC · Clinical Chemistry
Decision DateSep 13, 2013
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 862.1695
Device ClassClass 2
AttributesReal-World Evidence, Pediatric

Real-World Evidence

SubmissionDeviceSponsorRWD SourcesRWE Use SummaryKey Tags
K132249 · Sep 13, 2013ADVIA CENTAUR FT4, ADVIA CENTAUR T4Siemens Healthcare Diagnostics, Inc.Clinical laboratory patient dataData from 454 pediatric patients were analyzed to establish and validate pediatric reference intervals for the ADVIA Centaur FT4 and T4 assays.pediatric reference intervals; clinical laboratory data

Clinical Evidence

Study DesignPopulationComparatorKey Endpoints
Retrospective analysis of clinical laboratory dataPediatric patients (72 infants, 190 children, 129 adolescents); Sample Size: 454Not applicable for this studyEstablishment of pediatric reference intervals for FT4 and T4

Indications for Use

ADVIA Centaur FT4 For in vitro diagnostic use in the quantitative determination of free thyroxine (FT4) in serum or plasma (heparinized or EDTA), using the ADVIA Centaur and ADVIA Centaur XP systems. Measurements of free thyroxine are used in the diagnosis and treatment of thyroid disease. ADVIA Centaur T4 For in vitro diagnostic use in the quantitative determination of thyroxine (T4) in serum using the ADVIA Centaur and ADVIA Centaur XP systems. Measurements of thyroxine are used in the diagnosis and treatment of thyroid disease.

Device Story

ADVIA Centaur FT4 and T4 are in vitro diagnostic assays for quantitative measurement of free thyroxine and total thyroxine in serum or plasma. Used on ADVIA Centaur and ADVIA Centaur XP systems in clinical laboratory settings. Assays utilize immunoassay technology to measure analyte concentrations; results assist clinicians in diagnosing and managing thyroid disease. This submission updates labeling to include pediatric reference intervals derived from 454 patients. No changes to device design, manufacturing, or analytical performance characteristics were made. Healthcare providers interpret results against established reference intervals to assess thyroid function.

Clinical Evidence

No new clinical or analytical performance testing was required. Pediatric reference intervals were established by analyzing data from 454 patients (72 infants, 190 children, 129 adolescents). Performance characteristics are cross-referenced to original 510(k) submissions (K905532, K080167) and system performance submission (K971418).

Technological Characteristics

Immunoassay-based diagnostic tests for use on ADVIA Centaur and ADVIA Centaur XP automated systems. Analytical measuring range: FT4 (0.1–12 ng/dL), T4 (0.3–30 µg/dL). No changes to materials, sensing principles, or software algorithms.

Indications for Use

Indicated for the quantitative determination of free thyroxine (FT4) and total thyroxine (T4) in serum or plasma to aid in the diagnosis and treatment of thyroid disease in patients, including pediatric populations (infants 01-23 months, children 02-12 years, and adolescents 13-21 years).

Regulatory Classification

Identification

A free thyroxine test system is a device intended to measure free (not protein bound) thyroxine (thyroid hormone) in serum or plasma. Levels of free thyroxine in plasma are thought to reflect the amount of thyroxine hormone available to the cells and may therefore determine the clinical metabolic status of thyroxine. Measurements obtained by this device are used in the diagnosis and treatment of thyroid diseases.

Submission Summary (Full Text)

{0} # 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION ## TRIAGE-QUICK REVIEW DECISION SUMMARY 510(k) #: k132249 This 510(k) was reviewed under OIR’s Triage-Quick Review Program. This program represents an internal workload management tool intended to reduce internal FDA review resources for 510(k) applications that are of good quality upon receipt by FDA. The information in the 510(k) is complete and supports a substantial equivalence (SE) determination. Please refer to the applicant’s 510(k) summary for a summary of the information that supports this SE determination.
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...