ST AIA-PACK SHBG, ST AIA-PACK SHBG Calibrator Set
Device Facts
| Record ID | K143075 |
|---|---|
| Device Name | ST AIA-PACK SHBG, ST AIA-PACK SHBG Calibrator Set |
| Applicant | Tosoh Bioscience, Inc. |
| Product Code | CDZ · Clinical Chemistry |
| Decision Date | Jul 2, 2015 |
| Decision | SESE |
| Submission Type | Traditional |
| Regulation | 21 CFR 862.1680 |
| Device Class | Class 1 |
| Attributes | Real-World Evidence |
Real-World Evidence
| Submission | Device | Sponsor | RWD Sources | RWE Use Summary | Key Tags |
|---|---|---|---|---|---|
| K143075 · Jul 2, 2015 | ST AIA-PACK SHBG, ST AIA-PACK SHBG Calibrator Set | Tosoh Bioscience, Inc. | Clinical reference range study (488 apparently healthy individuals) | The sponsor used a cohort of 488 apparently healthy American and European individuals to establish and validate the reference intervals for the SHBG assay, consistent with CLSI C28-A3 guidelines. | Reference range; Clinical cohort; Healthy population |
Clinical Evidence
| Study Design | Population | Comparator | Key Endpoints |
|---|---|---|---|
| Retrospective/Observational cohort study for reference interval establishment | 488 apparently healthy American and European individuals; Sample Size: 488 | Not applicable for this study | Establishment of SHBG reference intervals by age and gender |
Indications for Use
ST AIA-PACK SHBG is designed for In Vitro Diagnostic Use Only for the quantitative measurement of sex hormone binding globulin (SHBG) in human serum or Na heparinized plasma on Tosoh AIA System Analyzers. The ST AIA-PACK SHBG assay is intended for use as an aid in the diagnosis of androgen disorders. The ST AIA-PACK SHBG Calibrator Set is intended for In Vitro Diagnostic Use Only for the calibration of the ST AIA-PACK SHBG assay.
Device Story
The ST AIA-PACK SHBG is a two-site immunoenzymometric assay performed in unit-dose test cups on Tosoh AIA System Analyzers. The device uses human serum or Na heparinized plasma as input. Inside the test cup, SHBG in the sample binds to monoclonal antibody immobilized on magnetic beads and enzyme-labeled monoclonal antibody. After washing to remove unbound components, a fluorogenic substrate (4-methylumbelliferyl phosphate) is added. The resulting fluorescence is directly proportional to the SHBG concentration. A standard curve is constructed using the ST AIA-PACK SHBG Calibrator Set to calculate unknown sample concentrations. The device is used in clinical laboratory settings by trained personnel. The output provides quantitative SHBG levels, which clinicians use as an aid in diagnosing androgen disorders, potentially guiding patient management and treatment decisions.
Clinical Evidence
No clinical diagnostic studies performed. Evidence consists of analytical performance data: precision (total CV 2.3-6.8%), linearity (R^2=0.9924-0.9997), LoQ (0.20 nmol/L), and method comparison against predicate (n=126, slope 0.949, R=0.991). Reference ranges established for pre/post-menopausal women and males (21-49 and >=50 years) using 122-123 subjects per group.
Technological Characteristics
Two-site immunoenzymometric assay. Components: unit dose test cups with lyophilized magnetic beads coated with anti-SHBG mouse monoclonal antibody and enzyme-labeled anti-SHBG mouse monoclonal antibody. Energy source: Tosoh AIA System Analyzer (fluorescence detection). Connectivity: System-integrated. Calibration: 6-point calibrator set traceable to NIBSC 08/266. Software: Embedded firmware for instrument control and data processing.
Indications for Use
Indicated for the quantitative measurement of sex hormone binding globulin (SHBG) in human serum or Na heparinized plasma to aid in the diagnosis of androgen disorders.
Regulatory Classification
Identification
A testosterone test system is a device intended to measure testosterone (a male sex hormone) in serum, plasma, and urine. Measurement of testosterone are used in the diagnosis and treatment of disorders involving the male sex hormones (androgens), including primary and secondary hypogonadism, delayed or precocious puberty, impotence in males and, in females hirsutism (excessive hair) and virilization (masculinization) due to tumors, polycystic ovaries, and adrenogenital syndromes.
Predicate Devices
- Architect SHBG Reagent Kit (k060818)
- Architect SHBG Calibrator Kit (k060818)