The Uric Acid assay is used for the quantitation of uric acid in serum, plasma, or urine on the ALCYON 300/300i Analyzer. Measurements of uric acid are used in the diagnosis and treatment of numerous renal and metabolic disorders, including renal failure, gout, leukemia, psoriasis, starvation or other wasting conditions, and of patients receiving cytotoxic drugs.
Device Story
In vitro diagnostic assay for quantitative determination of uric acid in human serum, plasma, or urine. Operates on ALCYON 300/300i Analyzer. Principle: uricase oxidizes uric acid to allantoin, producing hydrogen peroxide; peroxide reacts with 4-aminoantipyrine and TBHB via peroxidase to form quinoneimine dye. Change in absorbance at 550 nm measured; proportional to uric acid concentration. Used in clinical laboratories by trained personnel. Output provides uric acid concentration levels to assist clinicians in diagnosing and monitoring renal and metabolic disorders.
Clinical Evidence
Bench testing only. Method comparison studies performed on ALCYON Analyzer against predicate devices. Serum application: correlation coefficient 0.9918, slope 0.981, intercept 0.599 mg/dL. Urine application: correlation coefficient 0.9854, slope 0.951, intercept 0.435 mg/dL. Precision studies (within-run, between-run, between-day) showed total %CV ranging from 2.3% to 4.4%. Linearity up to 26 mg/dL; limit of quantitation 0.3 mg/dL.
Technological Characteristics
In vitro diagnostic clinical chemistry assay. Reagents: uricase, 4-aminoantipyrine, TBHB, peroxidase. Detection: spectrophotometric absorbance at 550 nm. System: ALCYON 300/300i Analyzer. Linear range up to 26 mg/dL; sensitivity 0.3 mg/dL.
Indications for Use
Indicated for the quantitative determination of uric acid in human serum, plasma, or urine for the diagnosis and treatment of renal and metabolic disorders, including renal failure, gout, leukemia, psoriasis, starvation, wasting conditions, and patients receiving cytotoxic drugs.
Regulatory Classification
Identification
A uric acid test system is a device intended to measure uric acid in serum, plasma, and urine. Measurements obtained by this device are used in the diagnosis and treatment of numerous renal and metabolic disorders, including renal failure, gout, leukemia, psoriasis, starvation or other wasting conditions, and of patients receiving cytotoxic drugs.
Predicate Devices
Roche Cobas Mira Plus Automated Chemistry Uric Acid assay (K922762)
{0}------------------------------------------------
# 510(k) Summary
Submitter's Name/Address Abbott Laboratories 1920 Hurd Drive Irving, Texas 75038
Contact Person Mark Littlefield Section Manager MS 1-8 Regulatory Affairs (972) 518-7861 Fax (972) 753-3367
| Date of Preparation of this Summary: | March 31, 1998 |
|--------------------------------------------------|----------------|
| Device Trade or Proprietary Name: | Uric Acid |
| Device Common/Usual Name or Classification Name: | Uric Acid |
| Classification Number/Class: | 75CDO/Class I |
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92. The assigned 510(k) number is:
## Test Description:
Uric Acid is an in vitro diagnostic assay for the quantitative determination of uric acid in human serum, plasma, or urine. The Uric Acid assay is a clinical chemistry assay in which uric acid is oxidized to allantoin by uricase with the production of hydrogen peroxide. The peroxide reacts with 4-aminoantipyrine and TBHB in the presence of peroxidase to yield a quinoneimine dye. The resulting change in absorbance at 550 nm is proportional to the uric acid concentration in the sample.
Uric Acid 510(k) April 29, 1998 Uric AcidE2. Iwp
Section II Page I
7
{1}------------------------------------------------
#### Substantial Equivalence:
The Uric Acid assay is substantially equivalent to the following devices:
- Roche® Cobas Mira® Plus Automated Chemistry Uric Acid assay (K922762) for . the serum application
- Boehringer Mannheim® Uric Acid assay (K873363) on the Hitachi® 717 Analyzer . for the urine application
These assays yield similar Performance Characteristics.
Similarities to Roche:
- Both assays are in vitro clinical chemistry methods. .
- Both assays can be used for the quantitative determination of uric acid. .
- . Both assays vield similar clinical results.
Differences to Roche:
- There is a minor difference between the assay ranges. .
Similarities to Boehringer Mannheim:
- Both assays are in vitro clinical chemistry methods. .
- . Both assays can be used for the quantitative determination of uric acid.
- . Both assays yield similar clinical results.
Differences to Boehringer Mannheim:
- . There is a minor difference between the assay ranges.
#### Intended Use:
The Uric Acid assay is used for the quantitation of uric acid in human serum, plasma, or urine on the ALCYON 300/300i Analyzer.
## Performance Characteristics:
Comparative performance studies were conducted using the ALCYON™ Analyzer. The Uric Acid assay method comparison yielded acceptable correlation with the Roche
Uric Acid 510(k) April 29, 1998 Uric AcidE2.1wp
Section II Page 2
{2}------------------------------------------------
Cobas Mira Plus Automated Chemistry System Uric Acid assay for the serum application and the Boehringer Mannheim Uric Acid assay on the Hitachi 717 Analyzer for the urine application. For the serum application, the correlation coefficient = 0.9918, slope = 0.981 and the Y-intercept = 0.599 mg/dL. For the urine application, the correlation coefficient = 0.9854, slope = 0.951 and the Y-intercept = 0.435 mg/dL. Precision studies were conducted using the Uric Acid assay. Within-run, between-run, and between-day studies were performed using two levels of control material. For the serum application, the total %CV for Level 1/Panel 111 is 2.8% and Level 2/Panel 112 is 2.3%. For the urine application, the total %CV for Level 1/Panel 131 is 4.4% and Level 2/Panel 132 is 3.2%. The Uric Acid assay is linear up to 26 mg/dL. The limit of quantitation (sensitivity) of the Uric Acid assay is 0.3 mg/dL. These data demonstrate that the performance of the Uric Acid assay is substantially equivalent to the performance of the Roche Cobas Mira Plus Automated Chemistry System Uric Acid assay for the serum application and the Boehringer Mannheim Uric Acid assay on the Hitachi 717 Analyzer for the urine application.
### Conclusion:
The Uric Acid assay is substantially equivalent to the Roche Cobas Mira Plus Automated Chemistry System Uric Acid assay for the serum application and the Boehringer Mannheim Uric Acid assay on the Hitachi 717Analyzer for the urine application as demonstrated by results obtained in the studies.
Uric Acid 510(k) April 29, 1998 Uric AcidE2. Iwp
Section II Page 3
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the Department of Health & Human Services USA. The logo features a stylized image of an eagle with three lines representing its body and wings. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" are arranged in a circular pattern around the eagle image.
Food and Drug Administration 2098 Gaither Road Rockville MD 20850
7 1998 MAY
Mark Littlefield Section Manager, Regulatory Affairs Abbott Laboratories 1920 Hurd Drive Irvinq, Texas 75038
K981189 Re : Uric Acid Requlatory Class: I Product Code: CDO March 31, 1998 Dated: Received: April 2, 1998
Dear Mr. Littlefield:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions The general controls provisions of the Act of the Act. include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major requlations affecting your device can be found in the Code of Federal Requlations, Title 21, Parts 800 to 895. ਜੋ substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory In addition, FDA may publish further announcements action. concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
{4}------------------------------------------------
Page 2
Under the Clinical Laboratory Improvement Amendments of 1988 (CLIA-88), this device may require a CLIA complexity categorization. To determine if it does, you should contact the Centers for Disease Control and Prevention (CDC) at (770) 488-7655.
This letter will allow you to begin marketing your device as described in your 510 (k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4588. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to
premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html".
Sincerely yours,
Steven Gutman
Steven I. Gutman, M.D., M.B.A. Director Division of Clinical Laboratory Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
510(k) Number (if known):
Device Name:
Indications For Use:
The Uric Acid assay is used for the quantitation of uric acid in serum, plasma, or urine on the ALCYON 300/300i Analyzer. Measurements of uric acid are used in the diagnosis and treatment of numerous renal and metabolic disorders, including renal failure, gout, leukemia, psoriasis, starvation or other wasting conditions, and of patients receiving cytotoxic drugs.
(Division Sign-Off)
Division of Clinical Laboratory Devices
510(k) Number k 981189
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE) Over-The-Counter Use _________________________________________________________________________________________________________________________________________________________ Prescription Use ﮯ OR (Per 21 CFR 801.109)
(Optional Format 1-2-96)
Uric Acid 510(k) April 29, 1998 Uric AcidE2.lvp
6
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.