K111549 · Apnicure, Inc. · OZR · Mar 30, 2012 · Dental
Device Facts
Record ID
K111549
Device Name
ATTUNE SLEEP APNEA SYSTEM
Applicant
Apnicure, Inc.
Product Code
OZR · Dental
Decision Date
Mar 30, 2012
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 872.5570
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Attune Sleep Apnea System is indicated for home use in the treatment of obstructive sleep apnea (OSA) in adults.
Device Story
Attune Sleep Apnea System is an intraoral pressure gradient device for treating obstructive sleep apnea (OSA). The system comprises a bedside electronic console, a soft polymer mouthpiece, and connecting tubing. The patient wears the mouthpiece during sleep; the console generates gentle negative pressure delivered to the oral cavity to hold the tongue and soft palate forward, increasing airway patency and reducing obstruction. The console also collects saliva, monitors pressure, and records usage time. An optional laboratory console allows integration with polysomnography (PSG) systems for technician monitoring. The device is intended for home use by adult patients. Clinical benefits include reduction of the Apnea-Hypopnea Index (AHI).
Clinical Evidence
Clinical evidence includes two feasibility studies and one four-week, multi-center, prospective, open-label, randomized, single-arm pivotal trial (n=63 in primary endpoint cohort). Clinical success was defined as AHI reduction > 50% and treated AHI < 20. Success was observed in 41.3% of the primary endpoint cohort (mild: 4/15, moderate: 10/18, severe: 12/30).
Technological Characteristics
System consists of an electronic console, soft polymer mouthpiece, and flexible polymer tubing. Operates via negative pressure generation. Mouthpiece sizing determined by dental impression/template. Console includes power cord and optional PSG connectivity. Software validation and cleaning validation performed.
Indications for Use
Indicated for home use in the treatment of obstructive sleep apnea (OSA) in adults.
Regulatory Classification
Identification
Intraoral devices for snoring and intraoral devices for snoring and obstructive sleep apnea are devices that are worn during sleep to reduce the incidence of snoring and to treat obstructive sleep apnea. The devices are designed to increase the patency of the airway and to decrease air turbulence and airway obstruction. The classification includes palatal lifting devices, tongue retaining devices, and mandibular repositioning devices.
Special Controls
*Classification.* Class II (special controls). The special control for these devices is the FDA guidance document entitled “Class II Special Controls Guidance Document: Intraoral Devices for Snoring and/or Obstructive Sleep Apnea; Guidance for Industry and FDA.”
{0}------------------------------------------------
#### 2. 510(K) SUMMARY
MAR 3 0 2012
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92(a).
Date Summary Prepared: March 29, 2012
510(k) Number: K111549
#### 510(k) Owner Information:
ApniCure, Inc. 900 Chesapeake Drive Redwood City, CA 94063 Phone Number: (650) 361-9300 (650) 361-9399 Fax Number:
#### Contact Person:
Cindy Domecus, R.A.C. (US & EU) Principal, Domecus Consulting Services LLC (650) 343-4813
#### Device Information:
Trade Name: Attune Sleep Apnea System · Intraoral device for obstructive sleep apnea Common Name: Classification: Class II Intraoral devices for snoring and intraoral devices for snoring Classification Name: and obstructive sleep apnea (21 CFR 872.5570, Product Code OZR, Intraoral Pressure Gradient Device)
## Physical Description:
The Attune Sleep Apnea System consists of three (3) main components: a small electronic bedside console, a soft polymer mouthpiece, and a flexible polymer tube that connects the mouthpiece to the console. A mouthpiece holder is provided for mouthpiece storage and use during weekly system cleaning.
The mouthpiece is an intraoral device that is worn during sleep. The system is designed to increase airway patency and decrease airway obstruction.
#### Console
The console generates a gentle negative pressure, collects excess saliva, records patient use time, and monitors pressure. The console is provided with a power cord, which connects to a standard electrical outlet. An optional laboratory console is available for use in a sleep laboratory. The laboratory console includes wires that extend from the console. The wires connect to the sleep laboratory's polysomnography (PSG) system, allowing the sleep technicians to view the console's pressure on the same monitor as the other PSG channels.
{1}------------------------------------------------
# Mouthpiece
The mouthpiece is provided in ten (10) discrete sizes. Sleep technicians use a bite wax to obtain an impression of each patient's teeth and a sizing template to determine the best mouthpiece size.
# Tubing
The tubing connects to the console by screwing into the console base and to the mouthpiece with luer connectors.
# Use
The patient connects the system and places the mouthpiece in his or her mouth. The mouthpiece is worn during sleep. The console generates a gentle, negative pressure, which is delivered through the mouthpiece into the oral cavity and holds the tongue and soft palate out of the airway.
## Indications for Use:
The Attune Sleep Apnea System is indicated for home use in the treatment of obstructive sleep apnea (OSA) in adults.
## Predicate Devices:
The Attune Sleep Apnea System is substantially equivalent in intended use, indications for use, and technological characteristics to the following devices:
| Name | Manufacturer | 510(k) # |
|------------------------------------------------|----------------------|---------------------|
| SnoreSilencer Pro | Respironics | K033822 |
| Oasys Oral Airway System | Mark Abramson | K030440 |
| Provent Professional Sleep<br>Apnea Therapy | Ventus Medical | K071560 |
| Repose Bone Screw<br>System | Influence, Inc. | K981677 |
| Prelude III™ Tongue<br>Suspension System | Siesta Medical, Inc. | K110127 |
| Sleepstyle 200 Series<br>HC234 CPAP, Opus Mask | Fisher & Paykel | K040941,<br>K063036 |
| S8 Aspen CPAP | Resmed Ltd. | K091947 |
## Non-clinical Performance Data:
Non-clinical performance testing included functional testing, biocompatibility testing, software validation and cleaning validation. Results of non-clinical testing demonstrate that
{2}------------------------------------------------
the Attune Sleep Apnea System is safe and effective for its intended use and substantially equivalent to the predicates.
## Clinical Performance Data:
Clinical performance testing included data from two feasibility studies and one pivotal trial. The pivotal trial was a four-week, multi-center, prospective, open label, randomized firstnight order of control vs. treatment, single-arm trial of the Attune Sleep Apnea System for the treatment of obstructive sleep apnea. The objective of the study was to demonstrate safety and effectiveness of the Attune Sleep Apnea System. Clinical Success was determined per patient and was prospectively defined as AHI reduction of > 50% and treated AHI < 20. There were 63 subjects in the Primary Endpoint Cohort. Clinical Success was observed in 41.3% (4/15 mild, 10/18 moderate, 12/30 severe) of the Primary Endpoint Cohort comparing First Treatment Night with Control. Results of clinical testing demonstrate that the Attune Sleep Apnea System is safe and effective for its intended use and substantially equivalent to the predicates.
## Conclusion:
Based on the indications for use and non-clinical and clinical data provided in this 510(k) notification, the subject device has been shown to be safe, effective, and substantially equivalent to the currently marketed predicate devices.
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the Department of Health & Human Services USA. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" arranged around the perimeter. Inside the circle is an abstract symbol resembling an eagle or bird in flight, composed of three curved lines. The overall design is simple and conveys a sense of government authority and public service.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Mr. Chris Daniel Executive Vice President Apnicure, Incorporated 900 Chesapeake Drive Redwood City, California 94063
MAR 3 0 2012
Re: K111549 Trade/Device Name: Attune Sleep Apnea System Regulation Number: 21 CFR 872.5570 Regulation Name: Intraoral Devices for Snoring and Intraoral Devices for Snoring and Obstructive Sleep Apnea. Regulatory Class: II Product Code: OZR Dated: March 15, 2012 Received: March 26, 2012
Dear Mr. Daniel:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{4}------------------------------------------------
#### Page 2 - Mr. Daniel
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm 115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Anthony D. Watson, B.S., M.S., M.B.A. Director
Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
#### INDICATIONS FOR USE STATEMENT 1.
510(k) Number (if known):
Device Name: Attune Sleep Apnea System
Indications for Use:
The Attune Sleep Apnea System is indicated for home use in the treatment of obstructive sleep apnea (OSA) in adults. huller A
> (Division Sign-Off) Division of Anesthesiology, General Hospital :nfection Control, Dental Devices
1111549 510(k) Number: _
Prescription Use × (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
# (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
# Concurrence of CDRH, Office of Device Evaluation (ODE)
ApniCure, Inc.
CONFIDENTIAL
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.